510(k) Report

510(k) · Cleared

Recalled · Class III · 2019

Architect Sirolimus Assay

K070822510(k) TraditionalCleared Sep 28, 2007186-day reviewClass II21 CFR 862.3840Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Fujirebio Diagnostics,Inc. · NRP Sirolimus Test System

Cleared on Sep 28, 2007 as substantially equivalent to K042411 (Axis-Shield, Ltd., 2005) after a 186-day review. Its family tree reaches back 2 generations to 2004. It has been recalled 2 times.

Decided
Sep 28, 2007
Review
186 days
Class
Class II
Product code
NRP
Recalls
2
Predicates
1

Its longest line of descent runs back 2 generations to DEN040008 (2004). No later device cites it as a predicate yet.

  1. Root · 2 generations backDEN040008 Cedia Sirolimus AssayMicrogenics Corp. · 2004
  2. 1 generation backK042411 Abbott Imx Sirolimus Microparticle Enzyme ImmunoassayAxis-Shield, Ltd. · 2005 · Recalled
  3. This deviceK070822 Architect Sirolimus AssayFujirebio Diagnostics,Inc. · 2007
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2DEN040008Cedia Sirolimus AssayMicrogenics Corp.2004—
−1K042411Abbott Imx Sirolimus Microparticle Enzyme ImmunoassayAxis-Shield, Ltd.20052
0 (this device)K070822Architect Sirolimus AssayFujirebio Diagnostics,Inc.2007—
FIG. 1 — Lineage of K070822, 2004–2007Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in NRP? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K042411Abbott Imx Sirolimus Microparticle Enzyme Immunoassay
Summary page 1
…ry Class: Class 11 Product Code: NRP DLJ) Predicate Device ABBOTT IMx® Sirolimus Microparticie Enzyme Immunoassay (K04241 1) Device Description The ARCHITECT Sirolimus assay is a delayed one-step immunoassay for the quantitative determination of sirol…
Axis-Shield, Ltd.20052

Also cleared against K042411: 2 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 9, 2019Class IIIZ-1355-2019Under Investigation by firmDevices were delivered without the required dry ice.
May 27, 2010Not classifiedZ-2172-2010Equipment maintenanceThe Architect Sirolimus assay lot 80162M100 is exhibiting higher than normal reports of barcode read errors.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 3 years and passes through 3 companies; a new generation was cleared every 2.5 years on median.
  • 1 of the 2 ancestors has a recall of any class.
  • 2 other devices also cite its predicate K042411.

How does its category compare?

All NRP devices

NRP Sirolimus Test System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2007, this device’s year, is highlighted.

FDA decisions per year · NRP

1in 2009

2007: 12008: 12009: 1
2007high 1 (2007) · low 0 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · NRP

Too few NRP 510(k)s in any year since 2007 to give a median.

Recalls per year · NRP

2in 2010

2010: 2
2007high 2 (2010) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · NRP

1in 2025

2010: 122011: 12014: 12015: 22016: 72017: 12018: 12019: 12022: 12025: 1
2007high 12 (2010) · low 0 (2007)2025

MAUDE reports naming product code NRP, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—00
20081—00
20091—00
20100—212
20110—01
20120—00
20130—00
20140—01
20150—02
20160—07
20170—01
20180—01
20190—01
20200—00
20210—00
20220—01
20230—00
20240—00
20250—01
20260—0—

What is it indicated for?

Reagent Kit The ARCHITECT Sirolimus assay is a chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of sirolimus in human whole blood on the ARCHITECT i System. The ARCHITECT Sirolimus assay is to be used as an aid in the management of renal transplant patients receiving sirolimus therapy. Calibrator Kit The ARCHITECT Sirolimus Calibrators are for the calibration of the ARCHITECT i System when used for the quantitative…

Read the full statement (112 words)

Reagent Kit

The ARCHITECT Sirolimus assay is a chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of sirolimus in human whole blood on the ARCHITECT i System. The ARCHITECT Sirolimus assay is to be used as an aid in the management of renal transplant patients receiving sirolimus therapy.

Calibrator Kit

The ARCHITECT Sirolimus Calibrators are for the calibration of the ARCHITECT i System when used for the quantitative determination of sirolimus in human whole blood.

Whole Blood Precipitation Reagent

The ARCHITECT Sirolimus Whole Blood Precipitation Reagent is for the extraction of sirolimus from samples (human whole blood patient specimens, controls, and ARCHITECT Sirolimus Calibrators) to be tested on the ARCHITECT i System.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K070822: Architect Sirolimus Assay.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K070822

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026