Syngo Imaging (Version V30)

K071114510(k) TraditionalCleared May 22, 200732-day reviewClass II21 CFR 892.2050Recalls 7

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Siemens AG Medical Solutions · LLZ System, Image Processing, Radiological

Cleared on May 22, 2007 after a 32-day review. The summary cites 2 numbers, none of which could be confirmed as its predicate (see the table below). 1 of the 182 devices in its family, counting itself, has a Class I recall. 181 later devices descend from it. It has been recalled 7 times.

Decided
May 22, 2007
Review
32 days
Class
Class II
Product code
LLZ
Recalls
7
Predicates
2

181 later devices descend from it, 4 citing it directly. 1 of the 182 devices in this family, counting this one, has a Class I recall.

Descendants, by year cleared 2008: 3 descendants cleared2010: 4 descendants cleared2009: 1 descendant cleared2013: 6 descendants cleared2011: 5 descendants cleared2012: 6 descendants cleared2015: 8 descendants cleared2018: 12 descendants cleared2016: 7 descendants cleared2014: 4 descendants cleared2019: 11 descendants cleared2020: 22 descendants cleared2017: 6 descendants cleared2023: 17 descendants cleared2021: 8 descendants cleared2025: 15 descendants cleared, 1 with a Class I recall12022: 16 descendants cleared2026: 15 descendants cleared2024: 15 descendants cleared20052010201520202025
  1. This deviceK071114 Syngo Imaging (Version V30)Siemens AG Medical Solutions · 2007
  2. Descendant · cites it directly · 177 of its ownK081734 Syngo Imaging (Version V31)Siemens AG Medical Solutions · 2008 · Recalled
  3. Descendant · 2 generations on · 176 of its ownK092519 Syngo.XSiemens AG Medical Solutions · 2009 · Recalled
  4. Descendant · 3 generations on · 99 of its ownK112020 Syngo.CT Vascular AnalysisSiemens Medi Cal Solutions, Inc. · 2011
  5. Descendant · 4 generations on · 63 of its ownK123920 Syngo.ViaSiemens AG Medical Solutions · 2013
  6. Descendant · 3 generations on · 54 of its ownK093621 Syngo.Pet&CT OncologySiemens Medical Solutions USA, Inc. · 2010

Showing 5 of 181 descendants. See all descendants

Descendants cleared per year

2008: 32009: 12010: 42011: 52012: 62013: 62014: 42015: 82016: 72017: 62018: 122019: 112020: 222021: 82022: 162023: 172024: 152025: 152026: 15
2008high 22 (2020) · low 1 (2009)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 181 · See all
  • K081734Syngo Imaging (Version V31)Siemens AG Medical Solutions · 2008 · cites it directly177descendants
  • K092519Syngo.XSiemens AG Medical Solutions · 2009 · 2 generations on176descendants
  • K112020Syngo.CT Vascular AnalysisSiemens Medi Cal Solutions, Inc. · 2011 · 3 generations on99descendants
  • K123920Syngo.ViaSiemens AG Medical Solutions · 2013 · 4 generations on63descendants
  • K093621Syngo.Pet&CT OncologySiemens Medical Solutions USA, Inc. · 2010 · 3 generations on54descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K071114Syngo Imaging (Version V30)Siemens AG Medical Solutions2007—
+1K081734Syngo Imaging (Version V31)Siemens AG Medical Solutions20085
+2K092519Syngo.XSiemens AG Medical Solutions20094
+3K112020Syngo.CT Vascular AnalysisSiemens Medi Cal Solutions, Inc.2011—
+4K123920Syngo.ViaSiemens AG Medical Solutions2013—
+3K093621Syngo.Pet&CT OncologySiemens Medical Solutions USA, Inc.2010—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200830
200910
201040
201150
201260
201360
201440
201580
201670
201760
2018120
2019110
2020220
202180
2022160
2023170
2024150
2025151
2026150
FIG. 1 — Lineage of K071114, 2007–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LLZ device

  1. #1 K230955 syngo Application Softwarescore 21.5 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary; numbers the summary cites that cannot be its predicate are marked – and kept off the family tree.

TypeNumberDeviceApplicantDecidedRecalls
Kept off the family treeK052409FDA’s K052409 is Silgimix Alginate Replacement Impression Material
Not matched in our reading of the summary (scans can blur digits), so it is kept off the family tree
Sultan Chemists, Inc.2005—
Kept off the family treeK021161FDA’s K021161 is TSO3 Ozone Chemical Indicator
Not matched in our reading of the summary (scans can blur digits), so it is kept off the family tree
Technologies of Sterilization With Oxone, Tso3, Inc.2003—

What descends from it? (181)

See all 181

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K081734Syngo Imaging (Version V31)Siemens AG Medical Solutions2008177
K082430Syngo Imaging XSSiemens AG Medical Solutions200811
K101666Syngo. Plaza VA20ASiemens AG Medical Solutions20109
K073418Syngo Imaging CCF VA10ASiemens AG Medical Solutions20080

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 18, 2016Class IIZ-1349-2016Software designTo inform users about the possible incorrect values for Distance Measurements when using certain modalities in combination with syngo Imaging.
Oct 28, 2015Class IIZ-0319-2016Software change controlTo provide supplementary information regarding the release of the syngo Imaging Software Version VB36D_HF02. The software provides improvements for all syngo Imaging installations running on the software Version VB36D.
Jun 11, 2015Class IIZ-1972-2015Software designDuring certain clinical workflows safety issues may occur involving monitor assignment order and incorrect display and measurements for images with non square pixel size.
Jan 12, 2009Not classifiedZ-1235-2009Software designIncorrect and/or incomplete data: Several potential safety issues while working with the syngo Image product including, but not limited to, data loss, errors in patient data due to different patient IDs, delays in image availability…
Aug 22, 2008Not classifiedZ-0102-2009Software designImage may not be visible after merge.
Mar 10, 2008Not classifiedZ-1643-2008Software designThe firm discovered a problem where the selected patient images on the device may display an additional image from another patient or study.
Nov 27, 2007Not classifiedZ-0732-2008Software designInconsistent data - A buffer overflow may occur between information system and picture system if the system experiences a high load. (e.g. RIS/PACS communication is unusally high)

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K250443MAGNETOM Avanto Fit; MAGNETOM Skyra Fit; MAGNETOM Sola Fit; MAGNETOM Viato.MobileSiemens Healthcare GmbH · 2025 · 7 generations on · ◆ in the 2025 descendants bar of FIG. 12 Class I recalls

What does this lineage say?

  • Of its descendants, K081734 (Siemens AG Medical Solutions, 2008) has gone furthest, with 177 descendants of its own.

How does its category compare?

All LLZ devices

LLZ System, Image Processing, Radiological: FDA decisions, review times, recalls and adverse event reports over 20 years. 2007, this device’s year, is highlighted.

FDA decisions per year · LLZ

58in 2025

2007: 762008: 942009: 702010: 512011: 1062012: 822013: 952014: 782015: 642016: 822017: 782018: 1042019: 742020: 812021: 832022: 652023: 872024: 622025: 582026 (partial year): 45
2007high 106 (2011) · low 51 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LLZ

173in 2025

2007: 36 days median2008: 52 days median2009: 62 days median2010: 98 days median2011: 79 days median2012: 87 days median2013: 84 days median2014: 89 days median2015: 90 days median2016: 90 days median2017: 107 days median2018: 105 days median2019: 107 days median2020: 115 days median2021: 116 days median2022: 137 days median2023: 157 days median2024: 131 days median2025: 173 days median2026 (partial year): 110 days median
2007high 173 (2025) · low 36 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LLZ

11in 2025

2007: 202008: 322009: 212010: 172011: 302012: 122013: 642014: 62015: 162016: 122017: 52018: 72019: 22020: 62021: 32022: 42023: 82024: 62025: 112026 (partial year): 14
2007high 64 (2013) · low 2 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LLZ

116in 2025

2007: 1522008: 1542009: 1452010: 1452011: 4542012: 2662013: 1,1302014: 15,6312015: 2112016: 2902017: 1042018: 722019: 882020: 822021: 702022: 282023: 262024: 422025: 116
2007high 15,631 (2014) · low 26 (2023)2025

MAUDE reports naming product code LLZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007763620152
2008945232154
2009706221145
2010519817145
20111067930454
2012828712266
20139584641,130
20147889615,631
2015649016211
2016829012290
2017781075104
2018104105772
201974107288
202081115682
202183116370
202265137428
202387157826
202462131642
20255817311116
20264511014—

What is it indicated for?

syngo® Imaging is a Picture Archiving and Communication System (PACS) intended to display, process, read, report, communicate, distribute, store and archive digital radiological images, including digital mammography images. It supports the physician in diagnosis and treatment planning. For primary image diagnosis in mammography only uncompressed or non-lossy compressed images and only pre-processed DICOM "For Presentation" images must be used. Also only…

Read the full statement (109 words)

syngo® Imaging is a Picture Archiving and Communication System (PACS) intended to display, process, read, report, communicate, distribute, store and archive digital radiological images, including digital mammography images.

It supports the physician in diagnosis and treatment planning.

For primary image diagnosis in mammography only uncompressed or non-lossy compressed images and only pre-processed DICOM "For Presentation" images must be used.

Also only monitors which received FDA approval for mammography must be used.

,syngo®Imaging also supports storage and archiving of DJCOM Structured Reports.

In a comprehensive imaging suite syngo® Imaging integrates Radiology Information Systems (RIS) to enable customer specific workflows.

Enhanced syngo® Imaging workplaces use a variety of advanced postprocessing applications.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K071114: Syngo Imaging (Version V30).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K071114

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026