510(k) Report

510(k) · Cleared

Recalled · Class II · 2015

Syngo Imaging XS

K082430510(k) TraditionalCleared Nov 14, 200884-day reviewClass II21 CFR 892.2050Recalls 12

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Siemens AG Medical Solutions · LLZ System, Image Processing, Radiological

Cleared on Nov 14, 2008 as substantially equivalent to K071114 (Siemens AG Medical Solutions, 2007) and 1 other predicate after an 84-day review. Its family tree reaches back 5 generations to 1995. 11 later devices descend from it. It has been recalled 12 times.

Decided
Nov 14, 2008
Review
84 days
Class
Class II
Product code
LLZ
Recalls
12
Predicates
2

Its longest line of descent runs back 5 generations to K945188 (1995). 7 ancestors in all. 11 later devices descend from it, 1 citing it directly.

Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor3 generations back: 1 other ancestor 1 generation forward: 1 descendant2 generations forward: 1 descendant3 generations forward: 1 descendant4 generations forward: 2 descendants5 generations forward: 2 descendants6 generations forward: 1 descendant7 generations forward: 1 descendant8 generations forward: 1 descendant9 generations forward: 1 descendant K072205 · Visage Pacs/CS, Release Version 5.0 · Mercury Computer Systems, Inc. · 2007K012086 · Aquariusnet Server · Terarecon,Inc. · 2001K994329 · Iivs Integrated Image Viewing Station · Terarecon,Inc. · 2000K961023 · Aidp (Accuimage Image Display Processor) · Accuimage, Inc. · 1996K945188 · Cemax Vipstation and/or Software · Cemax, Inc. · 1995−50+5
  1. Root · 5 generations backK945188 Cemax Vipstation and/or SoftwareCemax, Inc. · 1995
  2. 4 generations backK961023 Aidp (Accuimage Image Display Processor)Accuimage, Inc. · 1996
  3. 3 generations backK994329 Iivs Integrated Image Viewing StationTerarecon,Inc. · 2000
  4. 2 generations backK012086 Aquariusnet ServerTerarecon,Inc. · 2001
  5. 1 generation backK072205 Visage Pacs/CS, Release Version 5.0Mercury Computer Systems, Inc. · 2007
  6. This deviceK082430 Syngo Imaging XSSiemens AG Medical Solutions · 2008
  7. Descendant · cites it directly · 10 of its ownK093612 Syngo.PlazaSiemens AG Medical Solutions · 2010 · Recalled
  8. Descendant · 2 generations on · 9 of its ownK101666 Syngo. Plaza VA20ASiemens AG Medical Solutions · 2010 · Recalled
  9. Descendant · 3 generations on · 8 of its ownK132532 Syngo.Plaza VB10ASiemens AG Healthcare · 2013 · Recalled
  10. Descendant · 4 generations on · 6 of its ownK142750 JiveXVisus Technology Transfer GmbH · 2015
  11. Descendant · 5 generations on · 4 of its ownK162008 JiveXVisus Technology Transfer GmbH · 2016

Showing 5 of 11 descendants. See all descendants

Descendants cleared per year

2010: 22013: 12015: 12016: 12018: 22020: 22021: 12023: 1
2010high 2 (2010) · low 0 (2011)2026

Most influential descendants

Showing 5 of 11 · See all
  • K093612Syngo.PlazaSiemens AG Medical Solutions · 2010 · cites it directly10descendants
  • K101666Syngo. Plaza VA20ASiemens AG Medical Solutions · 2010 · 2 generations on9descendants
  • K132532Syngo.Plaza VB10ASiemens AG Healthcare · 2013 · 3 generations on8descendants
  • K142750JiveXVisus Technology Transfer GmbH · 2015 · 4 generations on6descendants
  • K162008JiveXVisus Technology Transfer GmbH · 2016 · 5 generations on4descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K945188Cemax Vipstation and/or SoftwareCemax, Inc.1995—
−4K961023Aidp (Accuimage Image Display Processor)Accuimage, Inc.1996—
−3K994329Iivs Integrated Image Viewing StationTerarecon,Inc.2000—
−2K012086Aquariusnet ServerTerarecon,Inc.2001—
−1K072205Visage Pacs/CS, Release Version 5.0Mercury Computer Systems, Inc.2007—
0 (this device)K082430Syngo Imaging XSSiemens AG Medical Solutions2008—
+1K093612Syngo.PlazaSiemens AG Medical Solutions201016
+2K101666Syngo. Plaza VA20ASiemens AG Medical Solutions20103
+3K132532Syngo.Plaza VB10ASiemens AG Healthcare201311
+4K142750JiveXVisus Technology Transfer GmbH2015—
+5K162008JiveXVisus Technology Transfer GmbH2016—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199710
200710

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201020
201310
201510
201610
201820
202020
202110
202310
FIG. 1 — Lineage of K082430, 1995–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LLZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LLZ device

  1. #1 K230955 syngo Application Softwarescore 21.5 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K071114Syngo Imaging (Version V30)
Summary page 3
…tion, is substantially equivalent to the following commercially available devices: Siemens syngo Imaging V30 K071114 Mercury VISAGE Pacs/CS K072205 The syngo Imaging XS describe…
Siemens AG Medical Solutions20077
*K072205Visage Pacs/CS, Release Version 5.0
Summary page 3
…mens syngo Imaging V30 K071114 Mercury VISAGE Pacs/CS K072205 The syngo Imaging XS described in this 510(k) has the same intended use and similar technical characteristics as the dev…
Mercury Computer Systems, Inc.2007—
†K042221Coronis 5MP Medical Flat Panel Display SystemBarco NV Barcoview2004—
†K043122SMD 21500 or DSB-2103-D-5MP-5M Pixel Grayscale Flat Panel DisplaySiemens AG2005—
†K032202Dome CX Digital Flat-Panel Display System, Models C2 and C5IPlanar Systems, Inc.2003—

Also cleared against K071114: 3 other devices →

What descends from it? (11)

See all 11

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K093612Syngo.PlazaSiemens AG Medical Solutions201010

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 17, 2015Class IIZ-0550-2016Software in the Use EnvironmentFor the Syngo Workflow SLR System with Software Ver: VB10C: Printouts may be printed in incorrect anatomical size when using syngo Imaging XS filming application in conjunction with a printer not released for anatomical print usage. This…
Nov 11, 2014Class IIZ-0576-2015Component design/selectionPrintouts may be printed in incorrect anatomical size when using syngo.plaza or syngo Imaging XS filming application in conjunction with a printer not released for anatomical print usage. This may happen due to an invalid combination of…
Jun 30, 2014Class IIZ-2111-2014Software designSiemens has become aware of the potential malfunction when using syngo Imaging XS, version VA70A or higher. In some cases a miscalculation may occur for the grey scale values in the functions Region of Interest, Pixel Lens, Edge…
Jul 18, 2012Class IIZ-2264-2012Software designThere is a potential malfunction when using syngo Imaging XS, version VA70A or higher. In some situations it can happen that a miscalculation is done for the grey scale values in the functions "Regions of Interest", "Pixel Lens", "Edge…
Jan 26, 2012Not classifiedZ-1064-2012Software designIn September 2011 firm issued two Customer Safety Advisory Notices for syngo Imaging XS. Firm became aware of an unintended behavior when using syngo Imaging XS, version VA70A and VA70B with Client/Server Installations. Loading additional…
Sep 20, 2011Not classifiedZ-0514-2012Labeling False and MisleadingUnintended behavior when using syngo Imaging XS, all versions. The "Rename" functionality can cause an unintended patient merge if the renamed patient is sent to another system.
Sep 19, 2011Not classifiedZ-0515-2012Software designUnintended behavior when using syngo Imaging XS, versions VA70A and VA70B Client/Serve installations. Loading additional images/series from Patient Browser using the thumbnails can cause an unintended deletion of images/series.
Apr 6, 2010Not classifiedZ-1892-2010Device DesignSiemens has become aware of a potential malfunction and hazard to patients when using Syngo Imaging version lines VB20, VB30, VB35A in combination with a RIS that is violating DICOM Standard by creating non-unique study instance UIDs.
Feb 16, 2010Not classifiedZ-1423-2010Software designimages can be overwritten
Sep 30, 2009Not classifiedZ-0174-2010Software designIncomplete amount of image data; specifically, The image count in the series table of the online and archive database will display the full amount of images received at the Syngo Imaging XS as archived. However, the zip file on the NAS or…
Sep 29, 2009Not classifiedZ-0270-2010Software designDistorted image. The "Distorted" image cannot be used for diagnostic findings, and, depending on the customer scenario, a repeated exam could be necessary if the image cannot be reported on any other workplace or recovered from another…
Sep 29, 2009Not classifiedZ-0271-2010Software designDistorted image. The "Distorted" image cannot be used for diagnostic findings, and, depending on the customer scenario, a repeated exam could be necessary if the image cannot be reported on any other workplace or recovered from another…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 13 years and passes through 5 companies; a new generation was cleared every 1.3 years on median.
  • 1 of the 7 ancestors has a recall of any class.
  • Of its descendants, K093612 (Siemens AG Medical Solutions, 2010) has gone furthest, with 10 descendants of its own.
  • 3 other devices also cite its predicate K071114.

How does its category compare?

All LLZ devices

LLZ System, Image Processing, Radiological: FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.

FDA decisions per year · LLZ

58in 2025

2007: 762008: 942009: 702010: 512011: 1062012: 822013: 952014: 782015: 642016: 822017: 782018: 1042019: 742020: 812021: 832022: 652023: 872024: 622025: 582026 (partial year): 45
2007high 106 (2011) · low 51 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LLZ

173in 2025

2007: 36 days median2008: 52 days median2009: 62 days median2010: 98 days median2011: 79 days median2012: 87 days median2013: 84 days median2014: 89 days median2015: 90 days median2016: 90 days median2017: 107 days median2018: 105 days median2019: 107 days median2020: 115 days median2021: 116 days median2022: 137 days median2023: 157 days median2024: 131 days median2025: 173 days median2026 (partial year): 110 days median
2007high 173 (2025) · low 36 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LLZ

11in 2025

2007: 202008: 322009: 212010: 172011: 302012: 122013: 642014: 62015: 162016: 122017: 52018: 72019: 22020: 62021: 32022: 42023: 82024: 62025: 112026 (partial year): 14
2007high 64 (2013) · low 2 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LLZ

116in 2025

2007: 1522008: 1542009: 1452010: 1452011: 4542012: 2662013: 1,1302014: 15,6312015: 2112016: 2902017: 1042018: 722019: 882020: 822021: 702022: 282023: 262024: 422025: 116
2007high 15,631 (2014) · low 26 (2023)2025

MAUDE reports naming product code LLZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007763620152
2008945232154
2009706221145
2010519817145
20111067930454
2012828712266
20139584641,130
20147889615,631
2015649016211
2016829012290
2017781075104
2018104105772
201974107288
202081115682
202183116370
202265137428
202387157826
202462131642
20255817311116
20264511014—

What is it indicated for?

syngo Imaging XS is a Picture Archiving and Communication System (PACS) intended to display, process, read, report, communicate, distribute, store and archive digital medical images. It supports the physician in diagnosis and treatment planning. syngo Imaging XS also supports storage and archiving of Structured DICOM Reports. In a comprehensive imaging suite syngo Imaging XS integrates Hospital/Radiology Information Systems (HIS/RIS) to enable customer…

Read the full statement (98 words)

syngo Imaging XS is a Picture Archiving and Communication System (PACS) intended to display, process, read, report, communicate, distribute, store and archive digital medical images.

It supports the physician in diagnosis and treatment planning.

syngo Imaging XS also supports storage and archiving of Structured DICOM Reports.

In a comprehensive imaging suite syngo Imaging XS integrates Hospital/Radiology Information Systems (HIS/RIS) to enable customer specific workflows.

syngo Imaging XS optionally uses a variety of advanced postprocessing applications.

Note:

Web based image distribution is not intended for reporting.

syngo Imaging XS does not support the display of mammography images for diagnosis.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K082430: Syngo Imaging XS.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K082430

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026