Vcare Retractor/Elevator

K071907510(k) SpecialCleared Oct 5, 200787-day reviewClass II21 CFR 884.4530Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Conmed Corporation · LKF Cannula, Manipulator/Injector, Uterine

Cleared on Oct 5, 2007 as substantially equivalent to K955446 (Conmedcorp, 1996) after an 87-day review. Its family tree reaches back 1 generation to 1996. 7 later devices descend from it. It has been recalled 4 times.

Decided
Oct 5, 2007
Review
87 days
Class
Class II
Product code
LKF
Recalls
4
Predicates
1

Its longest line of descent runs back 1 generation to K955446 (1996). 7 later devices descend from it, 4 citing it directly.

Descendants, by year cleared 2015: 2 descendants cleared2014: 2 descendants cleared2017: 1 descendant cleared2018: 1 descendant cleared2022: 1 descendant clearedK955446 · Conmed Vcare Retractor/Elevator · Conmedcorp · 1996 · recalled1995200020052010201520202025
Descendants, by generation 1 generation forward: 4 descendants2 generations forward: 3 descendants K955446 · Conmed Vcare Retractor/Elevator · Conmedcorp · 1996 · recalled0
  1. Root · 1 generation backK955446 Conmed Vcare Retractor/ElevatorConmedcorp · 1996 · Recalled
  2. This deviceK071907 Vcare Retractor/ElevatorConmed Corporation · 2007
  3. Descendant · cites it directly · 1 of its ownK142716 VCAREConmed Corporation · 2015
  4. Descendant · cites it directly · 1 of its ownK142164 Manipulator; Manipulator ProGynetech Pty., Ltd. · 2014
  5. Descendant · cites it directly · 1 of its ownK131781 Clearview TotalClinical Innovations, LLC · 2014 · Recalled
  6. Descendant · 2 generations on · 0 of its ownK220202 Uterine ElevatOR PRO with OccludOR BalloonThe O R Company Pty, Ltd. · 2022
  7. Descendant · 2 generations on · 0 of its ownK172846 Uterine ManipulatorBeijing Hangtian Kadi Technology R&D Institute · 2018

Showing 5 of 7 descendants. See all descendants

Descendants cleared per year

2014: 22015: 22017: 12018: 12022: 1
2014high 2 (2014) · low 0 (2016)2026

Most influential descendants

Showing 5 of 7 · See all
  • K142716VCAREConmed Corporation · 2015 · cites it directly1descendant
  • K142164Manipulator; Manipulator ProGynetech Pty., Ltd. · 2014 · cites it directly1descendant
  • K131781Clearview TotalClinical Innovations, LLC · 2014 · cites it directly1descendant
  • K220202Uterine ElevatOR PRO with OccludOR BalloonThe O R Company Pty, Ltd. · 2022 · 2 generations on0descendants
  • K172846Uterine ManipulatorBeijing Hangtian Kadi Technology R&D Institute · 2018 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K955446Conmed Vcare Retractor/ElevatorConmedcorp19963
0 (this device)K071907Vcare Retractor/ElevatorConmed Corporation2007—
+1K142716VCAREConmed Corporation2015—
+1K142164Manipulator; Manipulator ProGynetech Pty., Ltd.2014—
+1K131781Clearview TotalClinical Innovations, LLC20141
+2K220202Uterine ElevatOR PRO with OccludOR BalloonThe O R Company Pty, Ltd.2022—
+2K172846Uterine ManipulatorBeijing Hangtian Kadi Technology R&D Institute2018—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201420
201520
201710
201810
202210
FIG. 1 — Lineage of K071907, 1996–2022Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LKF? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LKF device

  1. #1 K241693 FemChec Controlled Saline-Air Device (FCD-250)score 11.6 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K955446Conmed Vcare Retractor/Elevator
Summary page 1
…I PROCODE: LKF Classification Panel: Obstetrics/Gynecology Panel Regulation Number: Unknown Predicate Device: K955446 VCARE® ConMed Corp. (Vaginal-Cervical Ahluwalia's…
Conmedcorp19963

Also cleared against K955446: 1 other device →

What descends from it? (7)

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K142716VCAREConmed Corporation20151
K142164Manipulator; Manipulator ProGynetech Pty., Ltd.20141
K131781Clearview TotalClinical Innovations, LLC20141
K163214Unicare System, Unicare ManipulatorClearpath Surgical, Inc.20170

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 2, 2011Not classifiedZ-1520-2011Device DesignProduct manufactured after 4/11/2008, but before 8/01/2009, contains the earlier Instructions For Use, IFU (PN 17416, Rev. D) which does not contain the revisions on the addition of Warning Notes 10 and 11 and the expansion of Warning…
Sep 8, 2008Not classifiedZ-0175-2009Process change controlThe firm received several complaints sighting detachment of the balloon at the distal end of the shaft. It was also noted that the incidence of the forward cup slipping off the shaft and being retained in the vaginal canal had increased.
Sep 8, 2008Not classifiedZ-0176-2009Process change controlThe firm received several complaints sighting detachment of the balloon at the distal end of the shaft. It was also noted that the incidence of the forward cup slipping off the shaft and being retained in the vaginal canal had increased.
Sep 8, 2008Not classifiedZ-0177-2009Process change controlThe firm received several complaints sighting detachment of the balloon at the distal end of the shaft. It was also noted that the incidence of the forward cup slipping off the shaft and being retained in the vaginal canal had increased.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has a recall.
  • Of its descendants, K142716 (Conmed Corporation, 2015) has gone furthest, with 1 descendant of its own.
  • 1 other device also cites its predicate K955446.

How does its category compare?

All LKF devices

LKF Cannula, Manipulator/Injector, Uterine: FDA decisions, review times, recalls and adverse event reports over 20 years. 2007, this device’s year, is highlighted.

FDA decisions per year · LKF

2in 2025

2007: 32009: 22010: 62011: 32013: 32014: 32015: 52016: 12017: 12018: 42019: 12022: 42024: 32025: 2
2007high 6 (2010) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LKF

136in 2024

2007: 141 days median2010: 178 days median2011: 187 days median2013: 266 days median2014: 169 days median2015: 140 days median2018: 267 days median2022: 79 days median2024: 136 days median
2007high 267 (2018) · low 79 (2022)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LKF

3in 2022

2008: 22012: 22022: 3
2007high 3 (2022) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LKF

95in 2025

2007: 362008: 592009: 532010: 462011: 982012: 1042013: 532014: 452015: 532016: 622017: 642018: 232019: 762020: 1192021: 1892022: 1002023: 882024: 1252025: 95
2007high 189 (2021) · low 23 (2018)2025

MAUDE reports naming product code LKF, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20073141036
20080—259
20092—053
20106178046
20113187098
20120—2104
20133266053
20143169045
20155140053
20161—062
20171—064
20184267023
20191—076
20200—0119
20210—0189
20224793100
20230—088
202431360125
20252—095
20260—0—

What is it indicated for?

The ConMed VCARE® Retractor/Elevator is indicated for manipulation of the uterus and injection of fluids or gases during laparoscopic procedures such as Laparoscopic Assisted Vaginal Hysterectomy (LAVH), Total Laparoscopic Hysterectomy (TLH), minilap, laparoscopic tubal occlusion, or diagnostic laparoscopy and also maintains pneumoperitoneum by sealing the vagina once colpotomy is performed.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K071907: Vcare Retractor/Elevator.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K071907

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026