510(k) · Cleared
Recalled · Class II · 2013Blackstone SFS Parallel Rod Connectors
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Blackstone Medical, Inc. · MNI Orthosis, Spinal Pedicle Fixation
Cleared on Mar 13, 2008 as substantially equivalent to K994217 (Blackstone Medical, Inc., 2000) and 9 other predicates after a 28-day review. Its family tree reaches back 4 generations to 1998. 296 later devices descend from it. It has been recalled once.
- Decided
- Mar 13, 2008
- Review
- 28 days
- Class
- Class II
- Product code
- MNI
- Recalls
- 1
- Predicates
- 11
Its longest line of descent runs back 4 generations to K980447 (1998). 14 ancestors in all. 296 later devices descend from it, 8 citing it directly.
- Root · 4 generations backK980447 Depuy Motech Moss Miami Spinal SystemDePuy Orthopaedics, Inc. · 1998
- 3 generations backK013558 Blackstone Spinal Fixation System 4.5MM Mono-Axial ScrewBlackstone Medical, Inc. · 2002
- 2 generations backK022399 Blackstone Spinal Fixation System SpacerBlackstone Medical, Inc. · 2002
- 1 generation backK030241 Modification to Blackstone Spinal Fixation SystemBlackstone Medical, Inc. · 2003
- This deviceK080407 Blackstone SFS Parallel Rod ConnectorsBlackstone Medical, Inc. · 2008
- Descendant · cites it directly · 277 of its ownK081684 Blackstone Pedicle Screw SystemBlackstone Medical, Inc. · 2008 · Recalled
- Descendant · 2 generations on · 209 of its ownK082797 Blackstone Pedicle Screw SystemBlackstone Medical, Inc. · 2008
- Descendant · 2 generations on · 130 of its ownK092624 Firebird Spinal Fixation SystemBlackstone Medical, Inc. · 2009 · Recalled
- Descendant · cites it directly · 117 of its ownK100044 Firebird Spinal Fixation System, Firebird Side-Loading BodyBlackstone · 2010
- Descendant · 2 generations on · 116 of its ownK122901 Firebird Spinal Fixation System, Phoenix Mis Spinal Fixation SystemOrthofix · 2012
Showing 5 of 296 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 296 · See all- K081684Blackstone Pedicle Screw SystemBlackstone Medical, Inc. · 2008 · cites it directly277descendants
- K082797Blackstone Pedicle Screw SystemBlackstone Medical, Inc. · 2008 · 2 generations on209descendants
- K092624Firebird Spinal Fixation SystemBlackstone Medical, Inc. · 2009 · 2 generations on130descendants
- K100044Firebird Spinal Fixation System, Firebird Side-Loading BodyBlackstone · 2010 · cites it directly117descendants
- K122901Firebird Spinal Fixation System, Phoenix Mis Spinal Fixation SystemOrthofix · 2012 · 2 generations on116descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −4 | K980447 | Depuy Motech Moss Miami Spinal System | DePuy Orthopaedics, Inc. | 1998 | — |
| −3 | K013558 | Blackstone Spinal Fixation System 4.5MM Mono-Axial Screw | Blackstone Medical, Inc. | 2002 | — |
| −2 | K022399 | Blackstone Spinal Fixation System Spacer | Blackstone Medical, Inc. | 2002 | — |
| −1 | K030241 | Modification to Blackstone Spinal Fixation System | Blackstone Medical, Inc. | 2003 | — |
| 0 (this device) | K080407 | Blackstone SFS Parallel Rod Connectors | Blackstone Medical, Inc. | 2008 | — |
| +1 | K081684 | Blackstone Pedicle Screw System | Blackstone Medical, Inc. | 2008 | 3 |
| +2 | K082797 | Blackstone Pedicle Screw System | Blackstone Medical, Inc. | 2008 | — |
| +2 | K092624 | Firebird Spinal Fixation System | Blackstone Medical, Inc. | 2009 | 1 |
| +1 | K100044 | Firebird Spinal Fixation System, Firebird Side-Loading Body | Blackstone | 2010 | — |
| +2 | K122901 | Firebird Spinal Fixation System, Phoenix Mis Spinal Fixation System | Orthofix | 2012 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1998 | 1 | 0 |
| 1999 | 1 | 0 |
| 2000 | 1 | 0 |
| 2001 | 1 | 0 |
| 2002 | 4 | 0 |
| 2003 | 2 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2008 | 2 | 0 |
| 2009 | 1 | 0 |
| 2010 | 4 | 0 |
| 2011 | 2 | 0 |
| 2012 | 5 | 0 |
| 2013 | 9 | 0 |
| 2014 | 5 | 0 |
| 2015 | 9 | 0 |
| 2016 | 12 | 0 |
| 2017 | 23 | 0 |
| 2018 | 16 | 0 |
| 2019 | 17 | 0 |
| 2020 | 23 | 0 |
| 2021 | 16 | 0 |
| 2022 | 23 | 0 |
| 2023 | 36 | 0 |
| 2024 | 33 | 0 |
| 2025 | 34 | 0 |
| 2026 | 26 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in MNI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
- Candidate predicates for a device like this, rankedin the report
- Recall history across each candidate’s familyin the report
What was it cleared against? (11)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary; numbers the summary cites that cannot be its predicate are marked – and kept off the family tree.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K994217 | Blackstone Spinal Fixation SystemSummary page 1…System MNI - 888.3070 - Pedicle Screw Spinal System Substantially Equivalent Devices: BlackstoneTM SFS (K994217 SE 2-28-00) BlackstoneTM SFS 4.5mm Multi-Axial Screws (K020674 SE 4-3-02) BlackstoneTM SES 4.5mm Mono-Axial Scr… | 2000 |
| * | K020674 | Blackstone Spinal Fixation System 4.5MM Multi-Axial ScrewSummary page 1…3070 - Pedicle Screw Spinal System Substantially Equivalent Devices: BlackstoneTM SFS (K994217 SE 2-28-00) BlackstoneTM SFS 4.5mm Multi-Axial Screws (K020674 SE 4-3-02) BlackstoneTM SES 4.5mm Mono-Axial Screws (K0I13558 SE 1-23-02) BlackstoneTM SFS 2nd Gen. Cross-Conne… | 2002 |
| Kept off the family tree | K011358 | Blackstone SES 4.5mm Mono-Axial Screws Not in FDA’s 510(k), De Novo or PMA lists as we hold them (a misreading, a typo, or a number FDA does not list), so it is kept off the family tree | |
| * | K003735 | Blackstone Spinal Fixation System, Second-Generation Cross-ConnectorSummary page 1…4.5mm Multi-Axial Screws (K020674 SE 4-3-02) BlackstoneTM SES 4.5mm Mono-Axial Screws (K0I13558 SE 1-23-02) BlackstoneTM SFS 2nd Gen. Cross-Connector (K003735 SE 5-8-0 1) BlackstoneTM SFS Modified Multi-Axial Screws (K023498 SE 11-13-02) BlackstoneTM SF5 Hooks (K013885… | 2001 |
| * | K023498 | Modification to Blackstone Spinal Fixation SystemSummary page 1…mm Mono-Axial Screws (K0I13558 SE 1-23-02) BlackstoneTM SFS 2nd Gen. Cross-Connector (K003735 SE 5-8-0 1) BlackstoneTM SFS Modified Multi-Axial Screws (K023498 SE 11-13-02) BlackstoneTM SF5 Hooks (K013885 SE 2-1-02) BlackstoneTM SFS Spacers (K022399 SE 8-6-02) B… | 2002 |
| * | K013885 | Blackstone Spinal Fixation System Posterior HooksSummary page 1…ackstoneTM SFS 2nd Gen. Cross-Connector (K003735 SE 5-8-0 1) BlackstoneTM SFS Modified Multi-Axial Screws (K023498 SE 11-13-02) BlackstoneTM SF5 Hooks (K013885 SE 2-1-02) BlackstoneTM SFS Spacers (K022399 SE 8-6-02) BlackstoneTM SFS Staple & Washer (K022605 SE 8-21-02)… | 2002 |
| * | K022399 | Blackstone Spinal Fixation System SpacerSummary page 1…8-0 1) BlackstoneTM SFS Modified Multi-Axial Screws (K023498 SE 11-13-02) BlackstoneTM SF5 Hooks (K013885 SE 2-1-02) BlackstoneTM SFS Spacers (K022399 SE 8-6-02) BlackstoneTM SFS Staple & Washer (K022605 SE 8-21-02) Blackstone TMN SFS Axial Domino (K030241 SE 2-… | 2002 |
| * | K022605 | Blackstone Spinal Fixation System Staple & WasherSummary page 1…K023498 SE 11-13-02) BlackstoneTM SF5 Hooks (K013885 SE 2-1-02) BlackstoneTM SFS Spacers (K022399 SE 8-6-02) BlackstoneTM SFS Staple & Washer (K022605 SE 8-21-02) Blackstone TMN SFS Axial Domino (K030241 SE 2-21-03) BlackstoneTM SES Rigid Cross Connector (K03086… | 2002 |
| * | K030241 | Modification to Blackstone Spinal Fixation SystemSummary page 1…SE 2-1-02) BlackstoneTM SFS Spacers (K022399 SE 8-6-02) BlackstoneTM SFS Staple & Washer (K022605 SE 8-21-02) Blackstone TMN SFS Axial Domino (K030241 SE 2-21-03) BlackstoneTM SES Rigid Cross Connector (K030862 SE 4-17-03) BlackstoneTM SES Lateral Offset (K03058… | 2003 |
| * | K030862 | Blackstone Spinal Fixation System Rigid Cross ConnectorSummary page 1…BlackstoneTM SFS Staple & Washer (K022605 SE 8-21-02) Blackstone TMN SFS Axial Domino (K030241 SE 2-21-03) BlackstoneTM SES Rigid Cross Connector (K030862 SE 4-17-03) BlackstoneTM SES Lateral Offset (K030581 SE 6-26-03) Device Description: The Blackstone Spinal Fixation Sy… | 2003 |
| * | K030581 | Blackstone Spinal Fixation System Lateral Offset (System Modification)Summary page 1…Blackstone TMN SFS Axial Domino (K030241 SE 2-21-03) BlackstoneTM SES Rigid Cross Connector (K030862 SE 4-17-03) BlackstoneTM SES Lateral Offset (K030581 SE 6-26-03) Device Description: The Blackstone Spinal Fixation System (SFS) is comprised of titanium alloy (Ti-6AL-4V ELI per A… | 2003 |
What descends from it? (296)
See all 2968 later devices cite it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K081684 | Blackstone Pedicle Screw System | 2008 | 277 |
| K100044 | Firebird Spinal Fixation System, Firebird Side-Loading Body | 2010 | 117 |
| K093926 | Firebird Spinal Fixation System | 2010 | 88 |
| K171082 | Firebird Spinal Fixation System, Phoenix MIS Spinal Fixation System, JANUS Midline… | 2017 | 61 |
| K170647 | Connector System | 2017 | 18 |
| K140260 | Skyhawk Lateral Plate System | 2014 | 14 |
| K150184 | Dymaxeon Spine System | 2015 | 4 |
| K211710 | Ascent POCT System, Centurion POCT System, Spinal Fixation System (SFS), Firebird Spinal… | 2021 | 2 |
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Feb 1, 2013 | Class II | Z-0896-2013 | Device Design | There is a possibility that the Set Screw (p/n 90-2923) sub-component of the Parallel Rod Connector Assembly (p/n 52-6800) & Axial Rod Connector (p/n 52-6700) may crack during final tightening. Orthofix Inc received six (6) customer… |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 10 years and passes through 2 companies; a new generation was cleared every 3.8 years on median.
- None of the 14 ancestors has a recall.
- Of its descendants, K081684 (Blackstone Medical, Inc., 2008) has gone furthest, with 277 descendants of its own.
- 9 other devices also cite its predicate K994217.
How does its category compare?
All MNI devicesMNI Orthosis, Spinal Pedicle Fixation: FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.
FDA decisions per year · MNI
1in 2017
2026 to Oct 25, outlined.
Median 510(k) review, days · MNI
168in 2016
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · MNI
1in 2018
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · MNI
80in 2025
MAUDE reports naming product code MNI, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 20 | 82 | 26 | 83 |
| 2008 | 23 | 64 | 8 | 108 |
| 2009 | 18 | 86 | 3 | 138 |
| 2010 | 16 | 74 | 1 | 126 |
| 2011 | 17 | 111 | 0 | 146 |
| 2012 | 19 | 75 | 1 | 200 |
| 2013 | 14 | 77 | 4 | 824 |
| 2014 | 6 | 104 | 0 | 301 |
| 2015 | 12 | 54 | 1 | 260 |
| 2016 | 7 | 168 | 0 | 117 |
| 2017 | 1 | — | 0 | 247 |
| 2018 | 0 | — | 1 | 134 |
| 2019 | 0 | — | 0 | 188 |
| 2020 | 0 | — | 0 | 115 |
| 2021 | 0 | — | 0 | 270 |
| 2022 | 0 | — | 0 | 214 |
| 2023 | 0 | — | 0 | 122 |
| 2024 | 0 | — | 0 | 114 |
| 2025 | 0 | — | 0 | 80 |
| 2026 | 0 | — | 0 | — |
What is it indicated for?
The Blackstone Spinal Fixation System is intended for non-cervical use in the spine. The Blackstone Spinal Fixation System, when used for pedicle screw fixation, is intended only for patients: a) Having severe spondylolisthesis (Grades 3 and 4) at the LS-S1 joint; b) Who are receiving fusion using autogenous bone graft only; c) Who are having the device fixed or attached to the lumbar and sacral spine (L3 and below); and d) Who are having the device…
Read the full statement (301 words)
The Blackstone Spinal Fixation System is intended for non-cervical use in the spine. The Blackstone Spinal Fixation System, when used for pedicle screw fixation, is intended only for patients:
a) Having severe spondylolisthesis (Grades 3 and 4) at the LS-S1 joint;
b) Who are receiving fusion using autogenous bone graft only;
c) Who are having the device fixed or attached to the lumbar and sacral spine (L3 and below); and
d) Who are having the device removed after the development of a solid fusion mass.
The Blackstone Spinal Fixation System, when used as a pedicle screw system in skeletally mature patients, is intended to provide immobilization and stabilization of spinal segments, as an adjunct to fusion, in treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine:
a) Degenerative spondylolistheses with objective evidence of neurologic impainnent;
b) Fracture;
c) Dislocation;
d) Scoliosis;
e) Kyphosis;
f) Spinal tumor; and
g) Failed previous fusion (pseudarthrosis).
The Blackstone Spinal Fixation System, when used for anterolateral non-pedicle screw fixation to the non-cervical spine, is intended for the following indications:
a) Degenerative disc disease (as defined as back pain of discogenic origin with degenerative disc confirmed by history and radiographic studies);
b) spondylolistheses;
c) spinal stenosis;
d) spinal deformities (i.e., scoliosis, kyphosis, lordosis);
e) tumor;
f) pseudoarthrosis;
g) previous failed fusion; and
h) trauma (i.e., fracture or dislocation).
The Blackstone Spinal Fixation System, when used for posterior non-pedicle screw fixation system of the non-cervical spine, is intended for the following indications:
a) Degenerative disc disease (as defined as back pain of discogenic origin with degenerative disc confirmed by history and radiographic studies);
b) spondylolistheses;
c) spinal stenosis;
d) spinal deformities (i.e., scoliosis, kyphosis, lordosis);
e) tumor;
f) pseudoarthrosis;
g) previous failed fusion; and
h) trauma (i.e., fracture or dislocation)
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K080407 · printed Nov 6, 2026
Cite this page
510(k) Report. “K080407: Blackstone SFS Parallel Rod Connectors.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K080407
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026