510(k) Report

510(k) · Cleared

Recalled · Class II · 2023

Tracheal Intubation Fibervideoscope, Models LF-Y0004 and LF-Y0005

K082720510(k) TraditionalCleared Nov 5, 200849-day reviewClass II21 CFR 874.4680Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Olympus Medical Systems Corporation · EOQ Bronchoscope (Flexible Or Rigid)

Cleared on Nov 5, 2008 as substantially equivalent to K981543 (The Olympus Optical Co., 1998) and 2 other predicates after a 49-day review. Its family tree reaches back 4 generations to 1993. 1 of the 11 devices in its family, counting itself, has a Class I recall. 1 later device descends from it. It has been recalled 3 times.

Decided
Nov 5, 2008
Review
49 days
Class
Class II
Product code
EOQ
Recalls
3
Predicates
3

Its longest line of descent runs back 4 generations to K931154 (1993). 9 ancestors in all, the oldest K850978 (1985). 1 later device descends from it, 1 citing it directly. 1 of the 11 devices in this family, counting this one, has a Class I recall.

  1. Root · 4 generations backK931154 Evis 200 SystemOlympus Corp. · 1993
  2. 3 generations backK963033 BF 240/P240/IT240 Bronchovideoscope & AccessoriesOlympus America, Inc. · 1996 · Recalled
  3. 2 generations backK023984 Modification to Olympus Bronchoscopes BF-40 Series, BF-240 Series…The Olympus Optical Co. · 2003 · Recalled (Class I)
  4. 1 generation backK033225 Evis Exera Bronchofibervideoscope Olympus BF Type XP160F, Its…Olympus Optical Co., Ltd. · 2003 · Recalled
  5. This deviceK082720 Tracheal Intubation Fibervideoscope, Models LF-Y0004 and LF-Y0005Olympus Medical Systems Corporation · 2008
  6. Descendant · cites it directly · 0 of its ownK201300 Airway MobilescopeOlympus Medical Systems Corp. · 2021 · Recalled

Class I recalls in this family

Most influential descendants

Showing 1 of 1 · See all
  • K201300Airway MobilescopeOlympus Medical Systems Corp. · 2021 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K931154Evis 200 SystemOlympus Corp.1993—
−3K963033BF 240/P240/IT240 Bronchovideoscope & AccessoriesOlympus America, Inc.19966
−2K023984Modification to Olympus Bronchoscopes BF-40 Series, BF-240 Series, and BF-160…The Olympus Optical Co.2003Class I
−1K033225Evis Exera Bronchofibervideoscope Olympus BF Type XP160F, Its Accessories and…Olympus Optical Co., Ltd.20035
0 (this device)K082720Tracheal Intubation Fibervideoscope, Models LF-Y0004 and LF-Y0005Olympus Medical Systems Corporation2008—
+1K201300Airway MobilescopeOlympus Medical Systems Corp.20212

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198510
199110
199310
199510
199810

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202110
FIG. 1 — Lineage of K082720, 1993–2021Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new EOQ device

  1. #1 K223299 Ambu® aScope™ 5 Broncho HD 5.0/2.2, Ambu®…score 18.4 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K981543Olympus LF-TP and LF-DP Tracheal Intubation Fiberscopes, Accessories and Ancillary…
Summary page 2
…redicate Devices Subject Dqyftes rec ei"PD'S TRACHEAL INTUBATION Olympus LF-DP Tracheal Intubation K981543 FIBERVIDEOSCOPE LF-Y0004 Fiberscope EVIS EXERA BRONCHOFIBERVIDEO…
The Olympus Optical Co.19985
*K033225Evis Exera Bronchofibervideoscope Olympus BF Type XP160F, Its Accessories and Ancillary…
Summary page 2
…tubation K981543 FIBERVIDEOSCOPE LF-Y0004 Fiberscope EVIS EXERA BRONCHOFIBERVIDEO K033225 TRACHEAL INTUBATION SCOPE OLYPUS BF TYPE XP160F FIBERVIDEOSCOPE LF-Y0005 Olympus Bronchoscope BF-X…
Olympus Optical Co., Ltd.20035
*K023984Modification to Olympus Bronchoscopes BF-40 Series, BF-240 Series, and BF-160 Series
Summary page 2
…K033225 TRACHEAL INTUBATION SCOPE OLYPUS BF TYPE XP160F FIBERVIDEOSCOPE LF-Y0005 Olympus Bronchoscope BF-XT40 K023984 4 Device Description The TRACHEAL INTUBATION FIBERVIDEOSCOPE LF-Y0004 and LF-Y0005 are all-in-one cordless endoscope…
The Olympus Optical Co.2003Class I

Also cleared against K981543: 6 other devices →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K201300Airway MobilescopeOlympus Medical Systems Corp.20210

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 25, 2023Class IIZ-0340-2024Under Investigation by firmChannel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead…
Jul 27, 2023Class IIZ-2495-2023Labeling Change ControlComplaint of endoscope model becoming lodged in the endotracheal tube connector due to diameter being too large.
Aug 8, 2018Class IIZ-3311-2018OtherThe attachment of non-Olympus accessories to the bronchoscope s instrument channel port may have resulted in more applied force than expected and lead to loosening of the instrument channel port.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K023984Modification to Olympus Bronchoscopes BF-40 Series, BF-240 Series, and BF-160 SeriesThe Olympus Optical Co. · 2003 · Direct predicate · ◆ on the line of descent in FIG. 114 Class I recalls

What does this lineage say?

  • The line of descent spans 15 years and passes through 5 companies; a new generation was cleared every 5.1 years on median.
  • 6 of the 9 ancestors have a recall of any class; 1 has a Class I recall.
  • Of its descendants, K201300 (Olympus Medical Systems Corp., 2021) has gone furthest, with 0 descendants of its own.
  • 6 other devices also cite its predicate K981543.

How does its category compare?

All EOQ devices

EOQ Bronchoscope (Flexible Or Rigid): FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.

FDA decisions per year · EOQ

20in 2025

2007: 22008: 22009: 32012: 22013: 42014: 32015: 62016: 32017: 42018: 72019: 112020: 102021: 152022: 112023: 122024: 132025: 202026 (partial year): 9
2007high 20 (2025) · low 0 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · EOQ

147in 2025

2009: 24 days median2013: 178 days median2014: 302 days median2015: 206 days median2016: 262 days median2017: 136 days median2018: 212 days median2019: 147 days median2020: 103 days median2021: 169 days median2022: 134 days median2023: 160 days median2024: 130 days median2025: 147 days median2026 (partial year): 206 days median
2007high 302 (2014) · low 24 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · EOQ

25in 2025

2011: 12012: 42016: 12019: 12020: 12021: 92022: 52023: 18, 4 Class I2024: 42025: 25, 24 Class I
2007high 25 (2025) · low 0 (2007)2026

Class I (28 since 2007) Class II/III not classified

Adverse event reports per year · EOQ

6,546in 2025

2007: 392008: 872009: 1142010: 1002011: 962012: 932013: 932014: 982015: 1322016: 2972017: 3372018: 2792019: 6542020: 5152021: 1,3582022: 2,5412023: 5,0642024: 6,0512025: 6,546
2007high 6,546 (2025) · low 39 (2007)2025

MAUDE reports naming product code EOQ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—039
20082—087
20093240114
20100—0100
20110—196
20122—493
20134178093
20143302098
201562060132
201632621297
201741360337
201872120279
2019111471654
2020101031515
20211516991,358
20221113452,541
20231216018 (4)5,064
20241313046,051
20252014725 (24)6,546
202692060—

What is it indicated for?

These instruments have been designed to be used with a Suction Pump and EndoTherapy accessories and other ancillary equipment for airway management, which includes endoscopic treatment, diagnosis, and observation to access airway anatomy, endotracheal/endobronchial intubation and management.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K082720: Tracheal Intubation Fibervideoscope, Models LF-Y0004 and LF-Y0005.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K082720

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026