510(k) Report

510(k) · Cleared

Recalled · Class II · 2020

Blood Monitoring Unit, Model Bmu 40

K090147510(k) TraditionalCleared Apr 22, 200991-day reviewClass II21 CFR 870.4330Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Maquet Cardiopulmonary, AG · DRY Monitor, Blood-Gas, On-Line, Cardiopulmonary Bypass

Cleared on Apr 22, 2009 as substantially equivalent to K972962 (3M Company, 1997) and 1 other predicate after a 91-day review. Its family tree reaches back 4 generations to 1993. 2 of the 13 devices in its family, counting itself, have a Class I recall. 3 later devices descend from it. It has been recalled once.

Decided
Apr 22, 2009
Review
91 days
Class
Class II
Product code
DRY
Recalls
1
Predicates
2

Its longest line of descent runs back 4 generations to K931968 (1993). 9 ancestors in all, the oldest K854357 (1986). 3 later devices descend from it, 1 citing it directly. 2 of the 13 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 1997: 1 other ancestor1990: 1 other ancestor1989: 1 other ancestor1988: 1 other ancestor1986: 1 other ancestor2011: 1 descendant cleared, 1 with a Class I recall12013: 1 descendant cleared2014: 1 descendant cleared, 1 with a Class I recall1K931968 · Cdi 100 Extracorporeal Hematocrit/Oxygen Saturatio · 3M Health Care, Ltd. · 1993K062468 · Spectrum M2 Monitor · Spectrum Medical, Llp · 2006×2: 2 generations cleared close together, K071725 (2007), K072131 (2007) 1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 4 other ancestors 1 generation forward: 1 descendant, 1 with a Class I recall12 generations forward: 2 descendants, 1 with a Class I recall1 K072131 · M3 Monitor · Spectrum Medical, Ltd. · 2007K071725 · M2 Monitor · Spectrum Medical, Ltd. · 2007K062468 · Spectrum M2 Monitor · Spectrum Medical, Llp · 2006K931968 · Cdi 100 Extracorporeal Hematocrit/Oxygen Saturatio · 3M Health Care, Ltd. · 19930
  1. Root · 4 generations backK931968 Cdi 100 Extracorporeal Hematocrit/Oxygen Saturatio3M Health Care, Ltd. · 1993
  2. 3 generations backK062468 Spectrum M2 MonitorSpectrum Medical, Llp · 2006
  3. 2 generations backK071725 M2 MonitorSpectrum Medical, Ltd. · 2007
  4. 1 generation backK072131 M3 MonitorSpectrum Medical, Ltd. · 2007
  5. This deviceK090147 Blood Monitoring Unit, Model Bmu 40Maquet Cardiopulmonary, AG · 2009
  6. Descendant · cites it directly · 2 of its ownK102726 Cardiohelp SystemMaquet Cardiopulmonary, AG · 2011 · Recalled (Class I)
  7. Descendant · 2 generations on · 0 of its ownK133598 Cardiohelp SystemMaquet Cardiopulmonary, AG · 2014 · Recalled (Class I)
  8. Descendant · 2 generations on · 0 of its ownK133265 Quadrox-IR Adult and Small AdultMaquet Cardiopulmonary, AG · 2013

Class I recalls in this family

Most influential descendants

Showing 3 of 3 · See all
  • K102726Cardiohelp SystemMaquet Cardiopulmonary, AG · 2011 · cites it directly2descendants
  • K133598Cardiohelp SystemMaquet Cardiopulmonary, AG · 2014 · 2 generations on0descendants
  • K133265Quadrox-IR Adult and Small AdultMaquet Cardiopulmonary, AG · 2013 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K931968Cdi 100 Extracorporeal Hematocrit/Oxygen Saturatio3M Health Care, Ltd.1993—
−3K062468Spectrum M2 MonitorSpectrum Medical, Llp2006—
−2K071725M2 MonitorSpectrum Medical, Ltd.2007—
−1K072131M3 MonitorSpectrum Medical, Ltd.2007—
0 (this device)K090147Blood Monitoring Unit, Model Bmu 40Maquet Cardiopulmonary, AG2009—
+1K102726Cardiohelp SystemMaquet Cardiopulmonary, AG2011Class I
+2K133598Cardiohelp SystemMaquet Cardiopulmonary, AG2014Class I
+2K133265Quadrox-IR Adult and Small AdultMaquet Cardiopulmonary, AG2013—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198610
198810
198910
199010
199710

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201111
201310
201411
FIG. 1 — Lineage of K090147, 1993–2014Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DRY device

  1. #1 K234065 CDI OneView Monitoring Systemscore 5.5 · 2 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K9729623M Cdi Blood Parameter Monitoring System 500
Summary page 1
…Cardiopulmonary bypass rn-line blood gas sensor Predicate Devices: CD[ Blood Parameter Monitoring System 500, K972962 M3 Monitor, K072 131 Device Description: The Blood Monitoring Unit BMU 40 monitors blood parameters d…
3M Company199726
*K072131M3 Monitor
Summary page 1
…n-line blood gas sensor Predicate Devices: CD[ Blood Parameter Monitoring System 500, K972962 M3 Monitor, K072 131 Device Description: The Blood Monitoring Unit BMU 40 monitors blood parameters during cardiopulmonary bypass or similar procedur…
Spectrum Medical, Ltd.2007—

Also cleared against K972962: 2 other devices →

What descends from it? (3)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K102726Cardiohelp System Class IMaquet Cardiopulmonary, AG20112

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 30, 2020Class IIZ-0443-2021Under Investigation by firmThe Printed Circuit Board Assembly (PCBA) are not in compliance with the IEC standard (60601 1, third edition). The BMU 40 units affected by this recall were designed with a short creepage distance of 1.8 mm between the two distinct and…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K102726Cardiohelp SystemMaquet Cardiopulmonary, AG · 2011 · Cites it directly · ◆ in the 2011 descendants bar of FIG. 11 Class I recall
  • K133598Cardiohelp SystemMaquet Cardiopulmonary, AG · 2014 · 2 generations on · ◆ in the 2014 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 16 years and passes through 4 companies; a new generation was cleared every 1.7 years on median.
  • 3 of the 9 ancestors have a recall of any class.
  • Of its descendants, K102726 (Maquet Cardiopulmonary, AG, 2011) has gone furthest, with 2 descendants of its own.
  • 2 other devices also cite its predicate K972962.

How does its category compare?

All DRY devices

DRY Monitor, Blood-Gas, On-Line, Cardiopulmonary Bypass: FDA decisions, review times, recalls and adverse event reports over 20 years. 2009, this device’s year, is highlighted.

FDA decisions per year · DRY

1in 2024

2007: 22009: 22011: 22012: 12014: 22017: 12018: 32020: 12021: 12024: 1
2007high 3 (2018) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DRY

88in 2018

2018: 88 days median
2007high 88 (2018)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DRY

1in 2020

2008: 12010: 62011: 142012: 52015: 12020: 1
2007high 14 (2011) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DRY

134in 2025

2008: 12009: 22010: 1,0702011: 2,2682012: 1672013: 1572014: 1562015: 2312016: 2402017: 612018: 1652019: 632020: 352021: 422022: 372023: 412024: 722025: 134
2007high 2,268 (2011) · low 0 (2007)2025

MAUDE reports naming product code DRY, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—00
20080—11
20092—02
20100—61,070
20112—142,268
20121—5167
20130—0157
20142—0156
20150—1231
20160—0240
20171—061
20183880165
20190—063
20201—135
20211—042
20220—037
20230—041
20241—072
20250—0134
20260—0—

What is it indicated for?

The intended purpose of the MAQUET Blood Monitoring Unit BMU 40 is to monitor blood parameters during cardiopulmonary bypass (CPB) or similar procedures with extracorporeal circulation, which require continuous monitoring of the arterial and/or venous blood parameters: partial pressure of oxygen (pO2), temperature (Ta and Tv), oxygen saturation (SO2), hemoglobin (Hb) and hematocrit (Hct). Oxygen consumption (VO2) can also be calculated. Blood flow…

Read the full statement (106 words)

The intended purpose of the MAQUET Blood Monitoring Unit BMU 40 is to monitor blood parameters during cardiopulmonary bypass (CPB) or similar procedures with extracorporeal circulation, which require continuous monitoring of the arterial and/or venous blood parameters: partial pressure of oxygen (pO2), temperature (Ta and Tv), oxygen saturation (SO2), hemoglobin (Hb) and hematocrit (Hct). Oxygen consumption (VO2) can also be calculated. Blood flow (QBlood) can be entered manually or values can be received from a connected heart-lung machine.

The duration of application of the disposable products (arterial BMU Sensor and venous BMU Cell) is limited to six hours.

The BMU 40 is designed for continuous operation.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K090147: Blood Monitoring Unit, Model Bmu 40.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K090147

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026