510(k) Report

510(k) · Cleared

Recalled · Class II · 2014

3M Cdi Blood Parameter Monitoring System 500

K972962510(k) TraditionalCleared Nov 6, 199787-day reviewClass II21 CFR 870.4330Recalls 26

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

3M Company · DRY Monitor, Blood-Gas, On-Line, Cardiopulmonary Bypass

Cleared on Nov 6, 1997 as substantially equivalent to K890113 (Cardiovascular Devices, Inc., 1989) and 2 other predicates after an 87-day review. Its family tree reaches back 1 generation to 1986. 2 of the 14 devices in its family, counting itself, have a Class I recall. 10 later devices descend from it. It has been recalled 26 times.

Decided
Nov 6, 1997
Review
87 days
Class
Class II
Product code
DRY
Recalls
26
Predicates
3

Its longest line of descent runs back 1 generation to K854357 (1986). 3 ancestors in all. 10 later devices descend from it, 3 citing it directly. 2 of the 14 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 1990: 1 other ancestor1989: 1 other ancestor2012: 1 descendant cleared2009: 1 descendant cleared2011: 2 descendants cleared, 1 with a Class I recall12018: 2 descendants cleared2013: 1 descendant cleared2014: 1 descendant cleared, 1 with a Class I recall12020: 1 descendant cleared2021: 1 descendant clearedK854357 · Gem-6, with Components · Diamond Electro-Tech, Inc. · 19861990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors 1 generation forward: 3 descendants2 generations forward: 2 descendants, 1 with a Class I recall13 generations forward: 3 descendants, 1 with a Class I recall14 generations forward: 1 descendant5 generations forward: 1 descendant K854357 · Gem-6, with Components · Diamond Electro-Tech, Inc. · 19860+5
  1. Root · 1 generation backK854357 Gem-6, with ComponentsDiamond Electro-Tech, Inc. · 1986
  2. This deviceK972962 3M Cdi Blood Parameter Monitoring System 5003M Company · 1997
  3. Descendant · cites it directly · 4 of its ownK110957 M4 MonitorSpectrum Medical, Llp · 2011
  4. Descendant · 2 generations on · 3 of its ownK173591 Quantum Diagnostic Module, Quantum Diagnostic Module - No GasSpectrum Medical, Ltd. · 2018
  5. Descendant · cites it directly · 3 of its ownK090147 Blood Monitoring Unit, Model Bmu 40Maquet Cardiopulmonary, AG · 2009 · Recalled
  6. Descendant · 3 generations on · 2 of its ownK181942 Quantum Ventilation ModuleSpectrum Medical, Ltd. · 2018
  7. Descendant · 2 generations on · 2 of its ownK102726 Cardiohelp SystemMaquet Cardiopulmonary, AG · 2011 · Recalled (Class I)

Showing 5 of 10 descendants. See all descendants

Descendants cleared per year

2009: 12011: 22012: 12013: 12014: 12018: 22020: 12021: 1
2009high 2 (2011) · low 0 (2010)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 10 · See all
  • K110957M4 MonitorSpectrum Medical, Llp · 2011 · cites it directly4descendants
  • K173591Quantum Diagnostic Module, Quantum Diagnostic Module - No GasSpectrum Medical, Ltd. · 2018 · 2 generations on3descendants
  • K090147Blood Monitoring Unit, Model Bmu 40Maquet Cardiopulmonary, AG · 2009 · cites it directly3descendants
  • K181942Quantum Ventilation ModuleSpectrum Medical, Ltd. · 2018 · 3 generations on2descendants
  • K102726Cardiohelp SystemMaquet Cardiopulmonary, AG · 2011 · 2 generations on2descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K854357Gem-6, with ComponentsDiamond Electro-Tech, Inc.1986—
0 (this device)K9729623M Cdi Blood Parameter Monitoring System 5003M Company1997—
+1K110957M4 MonitorSpectrum Medical, Llp2011—
+2K173591Quantum Diagnostic Module, Quantum Diagnostic Module - No GasSpectrum Medical, Ltd.2018—
+1K090147Blood Monitoring Unit, Model Bmu 40Maquet Cardiopulmonary, AG20091
+3K181942Quantum Ventilation ModuleSpectrum Medical, Ltd.2018—
+2K102726Cardiohelp SystemMaquet Cardiopulmonary, AG2011Class I

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198910
199010

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200910
201121
201210
201310
201411
201820
202010
202110
FIG. 1 — Lineage of K972962, 1986–2021Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DRY device

  1. #1 K234065 CDI OneView Monitoring Systemscore 5.5 · 2 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K890113The Cdi (R) System 400
Summary page 1
…Device Classification Name: Cardiopulmonary bypass on-line blood gas monitor and cardiopulmonary bypass in- line blood gas sensor (21 CFR 870.4330) Predicate Devices: K890113 System 400 CDI Extracorporeal Blood Gas Monitoring System K902654 System 100 CDI Hematocrit/Oxygen Saturation Monitoring System K…
Cardiovascular Devices, Inc.1989—
*K902654Cdi (TM) System 100 Monitoring System
Summary page 1
…gas monitor and cardiopulmonary bypass in- line blood gas sensor (21 CFR 870.4330) Predicate Devices: K890113 System 400 CDI Extracorporeal Blood Gas Monitoring System K902654 System 100 CDI Hematocrit/Oxygen Saturation Monitoring System K854357 Mallinckrodt Gem 6 Blood Gas/Electrolyte Monitor Device De…
3M Health Care, Ltd.19905
*K854357Gem-6, with Components
Summary page 1
…FR 870.4330) Predicate Devices: K890113 System 400 CDI Extracorporeal Blood Gas Monitoring System K902654 System 100 CDI Hematocrit/Oxygen Saturation Monitoring System K854357 Mallinckrodt Gem 6 Blood Gas/Electrolyte Monitor Device Description: The 3M CDI Blood Parameter Monitoring System 500 is an AC-…
Diamond Electro-Tech, Inc.1986—

Also cleared against K890113: 1 other device →

What descends from it? (10)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K110957M4 MonitorSpectrum Medical, Llp20114
K090147Blood Monitoring Unit, Model Bmu 40Maquet Cardiopulmonary, AG20093
K123039Cdi Blood Parameter Monitoring System 500Terumo Cardiovascular Systems Corp.20120

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 9, 2014Class IIZ-1634-2014Release of Material/Component prior to receiving test resultsThere is a potential for inaccuracies in blood gas readings/measurements. There is a potential risk of respiratory alkalosis or cerebral vasoconstriction with low paCO2. If the Hemoglobin and Hematocrit measurements are not accurate…
Feb 27, 2013Class IIZ-1731-2013Software designTerumo discovered through its own testing that CDI 500 Blood Parameter Monitoring System software contains inaccuracies associated with a software algorithm, and a software update has been developed to address this issue and a premarket…
Feb 27, 2013Class IIZ-1732-2013Software designTerumo discovered through its own testing that CDI 500 Blood Parameter Monitoring System software contains unused (turned off) accuracy correction factors for slope (gain) and offset (bias) for Hematocrit/Saturation calculations.
Feb 27, 2013Class IIZ-1730-2013Software designTerumo discovered through its own testing that CDI 500 Blood Parameter Monitoring System software contains inaccuracies associated with a software algorithm, and a software update has been developed to address this issue and a premarket…
Feb 27, 2013Class IIZ-1734-2013Software designTerumo discovered through its own testing that CDI 500 Blood Parameter Monitoring System software contains inaccuracies associated with a software algorithm, and a software update has been developed to address this issue and a premarket…
Feb 27, 2013Class IIZ-1733-2013Software designTerumo discovered through its own testing that CDI 500 Blood Parameter Monitoring System software contains inaccuracies associated with a software algorithm, and a software update has been developed to address this issue and a premarket…
Feb 3, 2012Not classifiedZ-1379-2012Labeling designReplacement of the SBC batteries (system batteries) should be managed as "routine maintenance" instead of as an in-house Service Procedure because some systems may be relying on batteries beyond their useful life.
Feb 3, 2012Not classifiedZ-1380-2012Labeling designReplacement of the SBC batteries (system batteries) should be managed as "routine maintenance" instead of as an in-house Service Procedure because some systems may be relying on batteries beyond their useful life.
Feb 3, 2012Not classifiedZ-1381-2012Labeling designReplacement of the SBC batteries (system batteries) should be managed as "routine maintenance" instead of as an in-house Service Procedure because some systems may be relying on batteries beyond their useful life.
Feb 3, 2012Not classifiedZ-1378-2012Labeling designReplacement of the SBC batteries (system batteries) should be managed as "routine maintenance" instead of as an in-house Service Procedure because some systems may be relying on batteries beyond their useful life.
Feb 3, 2012Not classifiedZ-1382-2012Labeling designReplacement of the SBC batteries (system batteries) should be managed as "routine maintenance" instead of as an in-house Service Procedure because some systems may be relying on batteries beyond their useful life.
Dec 6, 2011Not classifiedZ-0672-2012Component change controlIn a recent review of in-house Service Procedures for the CDI Blood Parameter Monitoring System 500, Terumo Cardiovascular Systems (Terumo CVS) found that one of the procedures it had been performing on older systems should have been…
Dec 6, 2011Not classifiedZ-0673-2012Component change controlIn a recent review of in-house Service Procedures for the CDI Blood Parameter Monitoring System 500, Terumo Cardiovascular Systems (Terumo CVS) found that one of the procedures it had been performing on older systems should have been…
Dec 6, 2011Not classifiedZ-0674-2012Component change controlIn a recent review of in-house Service Procedures for the CDI Blood Parameter Monitoring System 500, Terumo Cardiovascular Systems (Terumo CVS) found that one of the procedures it had been performing on older systems should have been…
Dec 6, 2011Not classifiedZ-0675-2012Component change controlIn a recent review of in-house Service Procedures for the CDI Blood Parameter Monitoring System 500, Terumo Cardiovascular Systems (Terumo CVS) found that one of the procedures it had been performing on older systems should have been…
Sep 23, 2011Not classifiedZ-0142-2012Component design/selectionTerumo CVS identified a defective circuit board, which could potentially cause the CDI monitor to not function after a power loss. Once power is lost, the monitor will not power up again, even if AC or battery power are restored.
Apr 21, 2011Not classifiedZ-2679-2011Device DesignInaccurate readings may result after Methylene Blue, an intravascular dye, is administered to the patient. The presence of Methylene Blue may interfere with the readings from the shunt sensor for the CDI 500 system in a way that would…
Apr 21, 2011Not classifiedZ-2678-2011Device DesignInaccurate readings may result after Methylene Blue, an intravascular dye, is administered to the patient. The presence of Methylene Blue may interfere with the readings from the shunt sensor for the CDI 500 system in a way that would…
Apr 21, 2011Not classifiedZ-2682-2011Device DesignInaccurate readings may result after Methylene Blue, an intravascular dye, is administered to the patient. The presence of Methylene Blue may interfere with the readings from the shunt sensor for the CDI 500 system in a way that would…
Apr 21, 2011Not classifiedZ-2681-2011Device DesignInaccurate readings may result after Methylene Blue, an intravascular dye, is administered to the patient. The presence of Methylene Blue may interfere with the readings from the shunt sensor for the CDI 500 system in a way that would…
Apr 21, 2011Not classifiedZ-2680-2011Device DesignInaccurate readings may result after Methylene Blue, an intravascular dye, is administered to the patient. The presence of Methylene Blue may interfere with the readings from the shunt sensor for the CDI 500 system in a way that would…
Nov 1, 2010Not classifiedZ-2188-2011Device DesignThe CDI 500 monitors may display inaccurate values. These errors include inaccurate potassium, CO2, hematocrit, O2 saturation, pH, hemoglobin and other miscellaneous inaccurate values.
Nov 1, 2010Not classifiedZ-2189-2011Device DesignThe CDI 500 monitor may display inaccurate values. These errors include inaccurate potassium, CO2, hematocrit, O2 saturation, pH, hemoglobin and other miscellaneous inaccuracate values.
Nov 1, 2010Not classifiedZ-2185-2011Device DesignThe firm has received 234 complaints of inaccuracy of displayed values on the CDI 101 and 500 monitors. In some cases the customer reported inaccurate potassium values. Other inaccuracies include CO2, hematocrit, O2 saturation, pH…
Nov 1, 2010Not classifiedZ-2186-2011Device DesignThe firm has received 234 complaints of inaccuracy of displayed values on the CDI 101 and 500 monitors. In some cases the customer reported inaccurate potassium values. Other inaccuracies include CO2, hematocrit, O2 saturation, pH…
Nov 1, 2010Not classifiedZ-2187-2011Device DesignThe CDI 500 monitor may display inaccurate values. These errors include inaccurate potassium, CO2, hematocrit, O2 saturation, pH, hemoglobin and other miscellaneous inaccuracate values.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K102726Cardiohelp SystemMaquet Cardiopulmonary, AG · 2011 · 2 generations on · ◆ in the 2011 descendants bar of FIG. 11 Class I recall
  • K133598Cardiohelp SystemMaquet Cardiopulmonary, AG · 2014 · 3 generations on · ◆ in the 2014 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • 1 of the 3 ancestors has a recall of any class.
  • Of its descendants, K110957 (Spectrum Medical, Llp, 2011) has gone furthest, with 4 descendants of its own.
  • 1 other device also cites its predicate K890113.

How does its category compare?

All DRY devices

DRY Monitor, Blood-Gas, On-Line, Cardiopulmonary Bypass: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · DRY

1in 2024

2007: 22009: 22011: 22012: 12014: 22017: 12018: 32020: 12021: 12024: 1
2007high 3 (2018) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DRY

88in 2018

2018: 88 days median
2007high 88 (2018)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DRY

1in 2020

2008: 12010: 62011: 142012: 52015: 12020: 1
2007high 14 (2011) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DRY

134in 2025

2008: 12009: 22010: 1,0702011: 2,2682012: 1672013: 1572014: 1562015: 2312016: 2402017: 612018: 1652019: 632020: 352021: 422022: 372023: 412024: 722025: 134
2007high 2,268 (2011) · low 0 (2007)2025

MAUDE reports naming product code DRY, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—00
20080—11
20092—02
20100—61,070
20112—142,268
20121—5167
20130—0157
20142—0156
20150—1231
20160—0240
20171—061
20183880165
20190—063
20201—135
20211—042
20220—037
20230—041
20241—072
20250—0134
20260—0—

What is it indicated for?

The 3M CDI System 500 provides continuous, on-line monitoring of the extracorporeal partial pressure of oxygen and carbon dioxide, pH, potassium, oxygen saturation, hematocrit, hemoglobin and temperature. In addition, calculated values of base excess, bicarbonate, oxygen saturation, and oxygen consumption may also be provided. These parameters are displayed at either actual temperature or adjusted to 37" C. For documentation purposes, the system 500’s integral printer provides a hard copy of displayed parameters.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K972962: 3M Cdi Blood Parameter Monitoring System 500.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K972962

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.