A1CNOW+ (10 Test Kit, Professional Use) Model 3024, A1CNOW+(20 Test Kit, Professional Use) Model 3021, A1CNOW Self Chec

K090413510(k) SpecialCleared May 14, 200985-day reviewClass II21 CFR 864.7470Recalls 7

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Bayer Healthcare, LLC · LCP Assay, Glycosylated Hemoglobin

Cleared on May 14, 2009 as substantially equivalent to K051321 (Metrika, Inc., 2005) after an 85-day review. Its family tree reaches back 2 generations to 2003. 3 later devices descend from it. It has been recalled 7 times.

Decided
May 14, 2009
Review
85 days
Class
Class II
Product code
LCP
Recalls
7
Predicates
1

Its longest line of descent runs back 2 generations to K033847 (2003). 3 later devices descend from it, 3 citing it directly.

Descendants, by year cleared 2017: 1 descendant cleared2018: 1 descendant cleared2015: 1 descendant clearedK033847 · A1CNOW for Home and Professional Use · Metrika, Inc. · 2003K051321 · Inview · Metrika, Inc. · 2005 · recalled200020052010201520202025
Descendants, by generation 1 generation forward: 3 descendants K051321 · Inview · Metrika, Inc. · 2005 · recalledK033847 · A1CNOW for Home and Professional Use · Metrika, Inc. · 20030
  1. Root · 2 generations backK033847 A1CNOW for Home and Professional UseMetrika, Inc. · 2003
  2. 1 generation backK051321 InviewMetrika, Inc. · 2005 · Recalled
  3. This deviceK090413 A1CNOW+ (10 Test Kit, Professional Use) Model 3024, A1CNOW+(20 Test…Bayer Healthcare, LLC · 2009
  4. Descendant · cites it directly · 0 of its ownK182781 PTS Professional Chemistry KitPolymer Technology Systems, Inc. D/B/A Pts Diagnostics · 2018 · Recalled
  5. Descendant · cites it directly · 0 of its ownK161533 A1C EZ Glycohemoglobin Analysis SystemBiohermes Co., Ltd. · 2017
  6. Descendant · cites it directly · 0 of its ownK140827 SD A1cCare System, SD A1cCare Spoit Type Test Kit, SD HbA1c Control…Sd Biosensor, Inc. · 2015

Most influential descendants

Showing 3 of 3 · See all
  • K182781PTS Professional Chemistry KitPolymer Technology Systems, Inc. D/B/A Pts Diagnostics · 2018 · cites it directly0descendants
  • K161533A1C EZ Glycohemoglobin Analysis SystemBiohermes Co., Ltd. · 2017 · cites it directly0descendants
  • K140827SD A1cCare System, SD A1cCare Spoit Type Test Kit, SD HbA1c Control SetSd Biosensor, Inc. · 2015 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K033847A1CNOW for Home and Professional UseMetrika, Inc.2003—
−1K051321InviewMetrika, Inc.20051
0 (this device)K090413A1CNOW+ (10 Test Kit, Professional Use) Model 3024, A1CNOW+(20 Test Kit…Bayer Healthcare, LLC2009—
+1K182781PTS Professional Chemistry KitPolymer Technology Systems, Inc. D/B/A Pts Diagnostics20181
+1K161533A1C EZ Glycohemoglobin Analysis SystemBiohermes Co., Ltd.2017—
+1K140827SD A1cCare System, SD A1cCare Spoit Type Test Kit, SD HbA1c Control SetSd Biosensor, Inc.2015—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201510
201710
201810
FIG. 1 — Lineage of K090413, 2003–2018Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LCP? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LCP device

  1. #1 K221326 Nova Allegro HbA1c Assay, Nova Allegro Analyzerscore 9.6 · 5 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K051321Inview
Summary page 1
…AI cNow® Multi-Use for Home and Professional Use Special 510(k)- Device Modification for K051321 MAY 14 2009 510(K)…
Metrika, Inc.20051

What descends from it? (3)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K182781PTS Professional Chemistry KitPolymer Technology Systems, Inc. D/B/A Pts Diagnostics20180
K161533A1C EZ Glycohemoglobin Analysis SystemBiohermes Co., Ltd.20170
K140827SD A1cCare System, SD A1cCare Spoit Type Test Kit, SD HbA1c Control SetSd Biosensor, Inc.20150

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 18, 2024Class IIZ-0223-2025Under Investigation by firmThe reason for this recall is PTS Diagnostics, Inc. received complaints of non-matching cartridge and analyzer codes within the product test system box. Use of cartridges and analyzers with non-matching codes may result in inaccurate test…
Dec 2, 2016Class IIZ-0683-2020Software designA numerical value less than 4% or greater than 13% may be displayed instead of the expected <4% or >13% result, due to a software bug.
Dec 2, 2016Class IIZ-0684-2020Software designA numerical value less than 4% or greater than 13% may be displayed instead of the expected <4% or >13% result, due to a software bug.
Dec 2, 2016Class IIZ-0686-2020Software designA numerical value less than 4% or greater than 13% may be displayed instead of the expected <4% or >13% result, due to a software bug.
Dec 2, 2016Class IIZ-0682-2020Software designA numerical value less than 4% or greater than 13% may be displayed instead of the expected <4% or >13% result, due to a software bug.
Dec 2, 2016Class IIZ-0685-2020Software designA numerical value less than 4% or greater than 13% may be displayed instead of the expected <4% or >13% result, due to a software bug.
Aug 26, 2013Class IIZ-0136-2014Nonconforming Material/ComponentTest results from these lots may produce a high-bias reading (by up to 20%) that is outside of limits defined for stability and product accuracy when product is stored at room temperature four (4) to five (5) months after release.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 6 years and passes through 2 companies; a new generation was cleared every 3.9 years on median.
  • 1 of the 2 ancestors has a recall of any class.
  • Of its descendants, K182781 (Polymer Technology Systems, Inc. D/B/A Pts Diagnostics, 2018) has gone furthest, with 0 descendants of its own.

How does its category compare?

All LCP devices

LCP Assay, Glycosylated Hemoglobin: FDA decisions, review times, recalls and adverse event reports over 20 years. 2009, this device’s year, is highlighted.

FDA decisions per year · LCP

1in 2024

2007: 72008: 52009: 62010: 22011: 102012: 92013: 22014: 82015: 32016: 22017: 52018: 22019: 42020: 12023: 22024: 1
2007high 10 (2011) · low 0 (2021)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LCP

269in 2019

2007: 133 days median2008: 206 days median2009: 158 days median2011: 176 days median2012: 143 days median2014: 127 days median2015: 239 days median2017: 218 days median2019: 269 days median
2007high 269 (2019) · low 127 (2014)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LCP

1in 2025

2007: 12008: 32009: 62010: 52011: 512013: 42014: 22015: 22016: 62017: 12018: 12022: 12024: 12025: 1
2007high 51 (2011) · low 0 (2012)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LCP

22in 2025

2007: 52008: 112009: 152010: 42011: 272012: 512013: 1932014: 1822015: 642016: 292017: 1492018: 1,9112019: 1142020: 132021: 72022: 82023: 272024: 392025: 22
2007high 1,911 (2018) · low 4 (2010)2025

MAUDE reports naming product code LCP, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007713315
20085206311
20096158615
20102—54
2011101765127
20129143051
20132—4193
201481272182
20153239264
20162—629
201752181149
20182—11,911
201942690114
20201—013
20210—07
20220—18
20232—027
20241—139
20250—122
20260—0—

What is it indicated for?

The A1CNow multi-use test provides quantitative measurement of the percent of glycated hemoglobin (%HbAlc, %A1C) levels in whole blood samples. The test is for home use and professional use for monitoring glycemic control in people with diabetes.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K090413: A1CNOW+ (10 Test Kit, Professional Use) Model 3024, A1CNOW+(20 Test Kit, Professional Use) Model 3021, A1CNOW Self Chec.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K090413

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026