510(k) Report

510(k) · Cleared

Recalled · Class II · 2018

Wireless Portable Detector FD-W17

K090625510(k) TraditionalCleared Mar 24, 200915-day reviewClass II21 CFR 892.1680Recalls 28

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Philips Medical Systems North America Co. · MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)

Cleared on Mar 24, 2009 as substantially equivalent to K982795 (Philips Medical Systems, Inc., 1998) after a 15-day review. Its family tree reaches back 2 generations to 1994. 41 later devices descend from it. It has been recalled 28 times.

Decided
Mar 24, 2009
Review
15 days
Class
Class II
Product code
MQB
Recalls
28
Predicates
1

Its longest line of descent runs back 2 generations to K931071 (1994). 4 ancestors in all. 41 later devices descend from it, 7 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1997: 1 other ancestor1994: 1 other ancestor2011: 4 descendants cleared2013: 4 descendants cleared2009: 1 descendant cleared2014: 4 descendants cleared2010: 2 descendants cleared2017: 4 descendants cleared2019: 3 descendants cleared2016: 2 descendants cleared2023: 2 descendants cleared2021: 6 descendants cleared2020: 2 descendants cleared2024: 1 descendant cleared2018: 2 descendants cleared2022: 1 descendant cleared2025: 1 descendant cleared2026: 2 descendants clearedReference device: K051523 · Regius Console CS-2000 and Regius Console CS-3000, Medical Image Processing Workstation · Konica Minolta Medical & Graphic, Inc. · 2005 · recalledReference device: K063781 · Philips XD-S Direct Radiography Workstation/Package · Philips Medical Systems North America Co. · 2007 · recalledK931071 · Philips Thoravision · Philips Medical Systems, Inc. · 1994 · recalledK982795 · Philips Bucky Vision · Philips Medical Systems, Inc. · 1998 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 2 generations back: 2 other ancestors 1 generation forward: 7 descendants2 generations forward: 9 descendants3 generations forward: 11 descendants4 generations forward: 5 descendants5 generations forward: 3 descendants6 generations forward: 2 descendants7 generations forward: 3 descendants8 generations forward: 1 descendant K982795 · Philips Bucky Vision · Philips Medical Systems, Inc. · 1998 · recalledK931071 · Philips Thoravision · Philips Medical Systems, Inc. · 1994 · recalled0+5
  1. Root · 2 generations backK931071 Philips ThoravisionPhilips Medical Systems, Inc. · 1994 · Recalled
  2. 1 generation backK982795 Philips Bucky VisionPhilips Medical Systems, Inc. · 1998 · Recalled
  3. This deviceK090625 Wireless Portable Detector FD-W17Philips Medical Systems North America Co. · 2009
  4. Descendant · cites it directly · 14 of its ownK112132 Krystalrad 660 Digital Radiographic Portable Retrofit SystemMedicatech USA · 2011
  5. Descendant · 2 generations on · 11 of its ownK130883 Sedecal Digital Radiographic Upgrade Model Sdru-TSedecal., Sa. · 2013
  6. Descendant · 3 generations on · 7 of its ownK140825 Delworks DR SystemDel Medical, Inc. · 2014
  7. Descendant · 3 generations on · 6 of its ownK162529 DigiX FDXAllengers Medical Systems Limited · 2017
  8. Descendant · cites it directly · 6 of its ownK141736 Philips Prograde; Philips Eleva Workspot for Digital DiagnostPhilips Medical Systems Dmc GmbH · 2014 · Recalled

Showing 5 of 41 descendants. See all descendants

Descendants cleared per year

2009: 12010: 22011: 42013: 42014: 42016: 22017: 42018: 22019: 32020: 22021: 62022: 12023: 22024: 12025: 12026: 2
2009high 6 (2021) · low 0 (2012)2026

Most influential descendants

Showing 5 of 41 · See all
  • K112132Krystalrad 660 Digital Radiographic Portable Retrofit SystemMedicatech USA · 2011 · cites it directly14descendants
  • K130883Sedecal Digital Radiographic Upgrade Model Sdru-TSedecal., Sa. · 2013 · 2 generations on11descendants
  • K140825Delworks DR SystemDel Medical, Inc. · 2014 · 3 generations on7descendants
  • K162529DigiX FDXAllengers Medical Systems Limited · 2017 · 3 generations on6descendants
  • K141736Philips Prograde; Philips Eleva Workspot for Digital DiagnostPhilips Medical Systems Dmc GmbH · 2014 · cites it directly6descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K931071Philips ThoravisionPhilips Medical Systems, Inc.19941
−1K982795Philips Bucky VisionPhilips Medical Systems, Inc.199817
0 (this device)K090625Wireless Portable Detector FD-W17Philips Medical Systems North America Co.2009—
+1K112132Krystalrad 660 Digital Radiographic Portable Retrofit SystemMedicatech USA2011—
+2K130883Sedecal Digital Radiographic Upgrade Model Sdru-TSedecal., Sa.2013—
+3K140825Delworks DR SystemDel Medical, Inc.2014—
+3K162529DigiX FDXAllengers Medical Systems Limited2017—
+1K141736Philips Prograde; Philips Eleva Workspot for Digital DiagnostPhilips Medical Systems Dmc GmbH20146

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199410
199710

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200910
201020
201140
201340
201440
201620
201740
201820
201930
202020
202160
202210
202320
202410
202510
202620
FIG. 1 — Lineage of K090625, 1994–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MQB? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MQB device

  1. #1 K201528 Yushan X-Ray Flat Panel Detector with DROCscore 21.5 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K982795Philips Bucky Vision
Summary page 2
…considered substantially equivalent to the Pixium 4600 detector, which received FDA clearance on November 24, 1998, under the name Philips Bucky Vision in 510(k) Number K982795. The grid supression pre-processing is equivalent to the pre-processing used by the Konica Minolta Regius Console CS 2000/CS-300…
Philips Medical Systems, Inc.199817
†K051523Regius Console CS-2000 and Regius Console CS-3000, Medical Image Processing WorkstationKonica Minolta Medical & Graphic, Inc.20051
†K063781Philips XD-S Direct Radiography Workstation/PackagePhilips Medical Systems North America Co.20071

Also cleared against K982795: 10 other devices →

What descends from it? (41)

See all 41

7 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K112132Krystalrad 660 Digital Radiographic Portable Retrofit SystemMedicatech USA201114
K141736Philips Prograde; Philips Eleva Workspot for Digital DiagnostPhilips Medical Systems Dmc GmbH20146
K102615Wireless DR Imaging Option-WDR1Ge Medical Systems, LLC20106
K111725Mobilediagnost WDRSedecal., Sa.20114
K100400Prestodr PortableCmt Medical Technologies, Ltd.20103
K131483Philips Pixium 4343RCPhilips Medical Systems Dmc GmbH20131
K091752Viztek WLViztek, Inc.20090

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 27, 2018Class IIZ-2851-2018Device DesignThe hook does not securely hold the footplate in the vertical position. Therefore, the footplate can fall or swing down and injury operator or patient.
Jun 27, 2018Class IIZ-2867-2018Device DesignThe hook does not securely hold the footplate in the vertical position. Therefore, the footplate can fall or swing down and injury operator or patient.
Jun 27, 2018Class IIZ-2849-2018Device DesignThe hook does not securely hold the footplate in the vertical position. Therefore, the footplate can fall or swing down and injury operator or patient.
Jun 27, 2018Class IIZ-2843-2018Device DesignThe hook does not securely hold the footplate in the vertical position. Therefore, the footplate can fall or swing down and injury operator or patient.
Jun 27, 2018Class IIZ-2847-2018Device DesignThe hook does not securely hold the footplate in the vertical position. Therefore, the footplate can fall or swing down and injury operator or patient.
Jun 27, 2018Class IIZ-2863-2018Device DesignThe hook does not securely hold the footplate in the vertical position. Therefore, the footplate can fall or swing down and injury operator or patient.
Jun 27, 2018Class IIZ-2858-2018Device DesignThe hook does not securely hold the footplate in the vertical position. Therefore, the footplate can fall or swing down and injury operator or patient.
Jun 27, 2018Class IIZ-2861-2018Device DesignThe hook does not securely hold the footplate in the vertical position. Therefore, the footplate can fall or swing down and injury operator or patient.
Jun 27, 2018Class IIZ-2866-2018Device DesignThe hook does not securely hold the footplate in the vertical position. Therefore, the footplate can fall or swing down and injury operator or patient.
Jun 27, 2018Class IIZ-2846-2018Device DesignThe hook does not securely hold the footplate in the vertical position. Therefore, the footplate can fall or swing down and injury operator or patient.
Jun 27, 2018Class IIZ-2859-2018Device DesignThe hook does not securely hold the footplate in the vertical position. Therefore, the footplate can fall or swing down and injury operator or patient.
Jun 27, 2018Class IIZ-2860-2018Device DesignThe hook does not securely hold the footplate in the vertical position. Therefore, the footplate can fall or swing down and injury operator or patient.
Jun 27, 2018Class IIZ-2855-2018Device DesignThe hook does not securely hold the footplate in the vertical position. Therefore, the footplate can fall or swing down and injury operator or patient.
Jun 27, 2018Class IIZ-2848-2018Device DesignThe hook does not securely hold the footplate in the vertical position. Therefore, the footplate can fall or swing down and injury operator or patient.
Jun 27, 2018Class IIZ-2857-2018Device DesignThe hook does not securely hold the footplate in the vertical position. Therefore, the footplate can fall or swing down and injury operator or patient.
Jun 27, 2018Class IIZ-2868-2018Device DesignThe hook does not securely hold the footplate in the vertical position. Therefore, the footplate can fall or swing down and injury operator or patient.
Jun 27, 2018Class IIZ-2842-2018Device DesignThe hook does not securely hold the footplate in the vertical position. Therefore, the footplate can fall or swing down and injury operator or patient.
Jun 27, 2018Class IIZ-2852-2018Device DesignThe hook does not securely hold the footplate in the vertical position. Therefore, the footplate can fall or swing down and injury operator or patient.
Jun 27, 2018Class IIZ-2841-2018Device DesignThe hook does not securely hold the footplate in the vertical position. Therefore, the footplate can fall or swing down and injury operator or patient.
Jun 27, 2018Class IIZ-2856-2018Device DesignThe hook does not securely hold the footplate in the vertical position. Therefore, the footplate can fall or swing down and injury operator or patient.
Jun 27, 2018Class IIZ-2862-2018Device DesignThe hook does not securely hold the footplate in the vertical position. Therefore, the footplate can fall or swing down and injury operator or patient.
Jun 27, 2018Class IIZ-2865-2018Device DesignThe hook does not securely hold the footplate in the vertical position. Therefore, the footplate can fall or swing down and injury operator or patient.
Jun 27, 2018Class IIZ-2850-2018Device DesignThe hook does not securely hold the footplate in the vertical position. Therefore, the footplate can fall or swing down and injury operator or patient.
Jun 27, 2018Class IIZ-2844-2018Device DesignThe hook does not securely hold the footplate in the vertical position. Therefore, the footplate can fall or swing down and injury operator or patient.
Jun 27, 2018Class IIZ-2845-2018Device DesignThe hook does not securely hold the footplate in the vertical position. Therefore, the footplate can fall or swing down and injury operator or patient.
Jun 27, 2018Class IIZ-2853-2018Device DesignThe hook does not securely hold the footplate in the vertical position. Therefore, the footplate can fall or swing down and injury operator or patient.
Jun 27, 2018Class IIZ-2864-2018Device DesignThe hook does not securely hold the footplate in the vertical position. Therefore, the footplate can fall or swing down and injury operator or patient.
Jun 27, 2018Class IIZ-2854-2018Device DesignThe hook does not securely hold the footplate in the vertical position. Therefore, the footplate can fall or swing down and injury operator or patient.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 15 years and passes through 2 companies; a new generation was cleared every 10.3 years on median.
  • 3 of the 4 ancestors have a recall of any class.
  • Of its descendants, K112132 (Medicatech USA, 2011) has gone furthest, with 14 descendants of its own.
  • 10 other devices also cite its predicate K982795.

How does its category compare?

All MQB devices

MQB Solid State X-Ray Imager (Flat Panel/Digital Imager): FDA decisions, review times, recalls and adverse event reports over 20 years. 2009, this device’s year, is highlighted.

FDA decisions per year · MQB

8in 2025

2007: 132008: 82009: 152010: 112011: 242012: 222013: 212014: 322015: 212016: 222017: 312018: 162019: 222020: 192021: 212022: 312023: 102024: 82025: 82026 (partial year): 3
2007high 32 (2014) · low 8 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MQB

127in 2025

2007: 22 days median2008: 15 days median2009: 57 days median2010: 116 days median2011: 102 days median2012: 80 days median2013: 105 days median2014: 102 days median2015: 81 days median2016: 92 days median2017: 30 days median2018: 30 days median2019: 30 days median2020: 28 days median2021: 48 days median2022: 48 days median2023: 30 days median2024: 187 days median2025: 127 days median2026 (partial year): 245 days median
2007high 245 (2026) · low 15 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MQB

1in 2025

2007: 152008: 362009: 192010: 142011: 102012: 82013: 62017: 12018: 42019: 12020: 42021: 12025: 12026 (partial year): 1
2007high 36 (2008) · low 0 (2014)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MQB

141in 2025

2007: 252008: 302009: 342010: 392011: 1012012: 2672013: 1592014: 292015: 212016: 262017: 202018: 842019: 452020: 92021: 262022: 132023: 512024: 912025: 141
2007high 267 (2012) · low 9 (2020)2025

MAUDE reports naming product code MQB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200713221525
20088153630
200915571934
2010111161439
20112410210101
201222808267
2013211056159
201432102029
20152181021
20162292026
20173130120
20181630484
20192230145
2020192849
20212148126
20223148013
20231030051
20248187091
202581271141
202632451—

What is it indicated for?

As a part of a radiographic system, the Wireless Portable Detector FD-W17 is intended to acquire digital radiographic images. The Wireless Portable Detector FD-W17 is suitable for all routine radiography exams, including specialist areas like intensive care, trauma, or pediatric work, excluding fluoroscopy, angiography and mammography.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K090625: Wireless Portable Detector FD-W17.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K090625

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026