Wireless DR Imaging Option-WDR1

K102615510(k) TraditionalCleared Nov 18, 201070-day reviewClass II21 CFR 892.1680Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ge Medical Systems, LLC · MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)

Cleared on Nov 18, 2010 as substantially equivalent to K982196 (GE Medical Systems, 1998) and 1 other predicate after a 70-day review. Its family tree reaches back 3 generations to 1994. 6 later devices descend from it. It has been recalled once.

Decided
Nov 18, 2010
Review
70 days
Class
Class II
Product code
MQB
Recalls
1
Predicates
2

Its longest line of descent runs back 3 generations to K931071 (1994). 7 ancestors in all, the oldest K862120 (1986). 6 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1998: 1 other ancestor1986: 1 other ancestor1997: 1 other ancestor1994: 1 other ancestor2011: 1 descendant cleared2014: 1 descendant cleared2016: 1 descendant cleared2018: 2 descendants cleared2021: 1 descendant clearedK931071 · Philips Thoravision · Philips Medical Systems, Inc. · 1994 · recalledK982795 · Philips Bucky Vision · Philips Medical Systems, Inc. · 1998 · recalledK090625 · Wireless Portable Detector FD-W17 · Philips Medical Systems North America Co. · 2009 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 1 other ancestor3 generations back: 2 other ancestors 1 generation forward: 1 descendant2 generations forward: 1 descendant3 generations forward: 1 descendant4 generations forward: 1 descendant5 generations forward: 1 descendant6 generations forward: 1 descendant K090625 · Wireless Portable Detector FD-W17 · Philips Medical Systems North America Co. · 2009 · recalledK982795 · Philips Bucky Vision · Philips Medical Systems, Inc. · 1998 · recalledK931071 · Philips Thoravision · Philips Medical Systems, Inc. · 1994 · recalled0+5
  1. Root · 3 generations backK931071 Philips ThoravisionPhilips Medical Systems, Inc. · 1994 · Recalled
  2. 2 generations backK982795 Philips Bucky VisionPhilips Medical Systems, Inc. · 1998 · Recalled
  3. 1 generation backK090625 Wireless Portable Detector FD-W17Philips Medical Systems North America Co. · 2009 · Recalled
  4. This deviceK102615 Wireless DR Imaging Option-WDR1Ge Medical Systems, LLC · 2010
  5. Descendant · cites it directly · 5 of its ownK103476 BRIVOXR285AMX, Optima XR200AMX, OPTIMAXR220AMXGe Medical Systems, LLC · 2011 · Recalled
  6. Descendant · 2 generations on · 4 of its ownK142383 Optima XR200amx, Optima XR220amxGe Medical Systems, LLC · 2014 · Recalled
  7. Descendant · 3 generations on · 3 of its ownK162990 Optima XR240amxGe Medical Systems, LLC · 2016 · Recalled
  8. Descendant · 4 generations on · 2 of its ownK173602 Optima XR240amx, AutoGridGe Medical Systems, LLC · 2018 · Recalled
  9. Descendant · 5 generations on · 1 of its ownK182234 Optima XR240amxGe Medical Systems, LLC · 2018 · Recalled

Showing 5 of 6 descendants. See all descendants

Descendants cleared per year

2011: 12014: 12016: 12018: 22021: 1
2011high 2 (2018) · low 0 (2012)2026

Most influential descendants

Showing 5 of 6 · See all
  • K103476BRIVOXR285AMX, Optima XR200AMX, OPTIMAXR220AMXGe Medical Systems, LLC · 2011 · cites it directly5descendants
  • K142383Optima XR200amx, Optima XR220amxGe Medical Systems, LLC · 2014 · 2 generations on4descendants
  • K162990Optima XR240amxGe Medical Systems, LLC · 2016 · 3 generations on3descendants
  • K173602Optima XR240amx, AutoGridGe Medical Systems, LLC · 2018 · 4 generations on2descendants
  • K182234Optima XR240amxGe Medical Systems, LLC · 2018 · 5 generations on1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K931071Philips ThoravisionPhilips Medical Systems, Inc.19941
−2K982795Philips Bucky VisionPhilips Medical Systems, Inc.199817
−1K090625Wireless Portable Detector FD-W17Philips Medical Systems North America Co.200928
0 (this device)K102615Wireless DR Imaging Option-WDR1Ge Medical Systems, LLC2010—
+1K103476BRIVOXR285AMX, Optima XR200AMX, OPTIMAXR220AMXGe Medical Systems, LLC20118
+2K142383Optima XR200amx, Optima XR220amxGe Medical Systems, LLC20148
+3K162990Optima XR240amxGe Medical Systems, LLC20162
+4K173602Optima XR240amx, AutoGridGe Medical Systems, LLC20182
+5K182234Optima XR240amxGe Medical Systems, LLC20182

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198610
199410
199710
199810

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201110
201410
201610
201820
202110
FIG. 1 — Lineage of K102615, 1994–2021Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MQB? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MQB device

  1. #1 K201528 Yushan X-Ray Flat Panel Detector with DROCscore 21.5 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K982196Digital Radiographic Imaging System
Summary page 1
…al, tethered cable or direct connection (docked). The Wireless detector is very similar to the digital detector currently in use as the GE Digital Radiographic Detector (K982196), and the Philips 5-2 KIOZG (S P LD GE Medical Systems, LLC 510(k) Premarket Notification Wireless DR Imaging O…
GE Medical Systems1998—
*K090625Wireless Portable Detector FD-W17
Summary page 2
…KIOZG (S P LD GE Medical Systems, LLC 510(k) Premarket Notification Wireless DR Imaging Option WDRI Medical Systems Wireless Portable Detector FD-W17 (K090625) comprised of amorphous silicon with a cesium iodide scintillator. It is used on radiographic X-ray systems as a substitute for…
Philips Medical Systems North America Co.200928

Also cleared against K982196: 3 other devices →

What descends from it? (6)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K103476BRIVOXR285AMX, Optima XR200AMX, OPTIMAXR220AMXGe Medical Systems, LLC20115

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 20, 2014Class IIZ-1078-2014Software designGE Healthcare has recently become aware of a software issue associated with patient selection from the worklist on your WDR1(Precision 500D and Proteus XR/a equipped with Wireless Digital Radiography Option) that may impact patient…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 16 years and passes through 3 companies; a new generation was cleared every 4.2 years on median.
  • 5 of the 7 ancestors have a recall of any class.
  • Of its descendants, K103476 (Ge Medical Systems, LLC, 2011) has gone furthest, with 5 descendants of its own.
  • 3 other devices also cite its predicate K982196.

How does its category compare?

All MQB devices

MQB Solid State X-Ray Imager (Flat Panel/Digital Imager): FDA decisions, review times, recalls and adverse event reports over 20 years. 2010, this device’s year, is highlighted.

FDA decisions per year · MQB

8in 2025

2007: 132008: 82009: 152010: 112011: 242012: 222013: 212014: 322015: 212016: 222017: 312018: 162019: 222020: 192021: 212022: 312023: 102024: 82025: 82026 (partial year): 3
2007high 32 (2014) · low 8 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MQB

127in 2025

2007: 22 days median2008: 15 days median2009: 57 days median2010: 116 days median2011: 102 days median2012: 80 days median2013: 105 days median2014: 102 days median2015: 81 days median2016: 92 days median2017: 30 days median2018: 30 days median2019: 30 days median2020: 28 days median2021: 48 days median2022: 48 days median2023: 30 days median2024: 187 days median2025: 127 days median2026 (partial year): 245 days median
2007high 245 (2026) · low 15 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MQB

1in 2025

2007: 152008: 362009: 192010: 142011: 102012: 82013: 62017: 12018: 42019: 12020: 42021: 12025: 12026 (partial year): 1
2007high 36 (2008) · low 0 (2014)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MQB

141in 2025

2007: 252008: 302009: 342010: 392011: 1012012: 2672013: 1592014: 292015: 212016: 262017: 202018: 842019: 452020: 92021: 262022: 132023: 512024: 912025: 141
2007high 267 (2012) · low 9 (2020)2025

MAUDE reports naming product code MQB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200713221525
20088153630
200915571934
2010111161439
20112410210101
201222808267
2013211056159
201432102029
20152181021
20162292026
20173130120
20181630484
20192230145
2020192849
20212148126
20223148013
20231030051
20248187091
202581271141
202632451—

What is it indicated for?

The Wireless DR Imaging Option -WDRI, when used with a radiographic imaging system, is indicated for use in generating radiographic images of human anatomy. It is intended for use in all routine radiography exams, and specialized areas including pediatric work, intensive care and trauma, wherever conventional screen-film systems or CR systems maybe used.

The device is not intended for mammography or fluoroscopy applications.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K102615: Wireless DR Imaging Option-WDR1.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K102615

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.