510(k) Report

510(k) · Cleared

Recalled · Class II · 2012

TANGO3 Water Storage Tank with Ozone Disinfection System

K093641510(k) TraditionalCleared Dec 8, 2010379-day reviewClass II21 CFR 876.5820Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Tango3, LLC · FIN Tank, Holding, Dialysis

Cleared on Dec 8, 2010 as substantially equivalent to K830575 (Renal Systems, Inc., 1983) and 3 other predicates after a 379-day review. Its family tree reaches back 2 generations to 1989. 3 later devices descend from it. It has been recalled once.

Decided
Dec 8, 2010
Review
379 days
Class
Class II
Product code
FIN
Recalls
1
Predicates
4

Its longest line of descent runs back 2 generations to K882564 (1989). 7 ancestors in all, the oldest K830575 (1983). 3 later devices descend from it, 3 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2005: 1 other ancestor1998: 1 other ancestor1983: 1 other ancestor1994: 1 other ancestor1989: 1 other ancestor2015: 1 descendant cleared2014: 2 descendants clearedK882564 · Renalin Cold Sterilant · Minntech Corp. · 1989K962959 · Peracidin Dialyzer Reprocessing Concentrate · Hdc Medical, Inc. · 199819801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 3 other ancestors2 generations back: 2 other ancestors 1 generation forward: 3 descendants K962959 · Peracidin Dialyzer Reprocessing Concentrate · Hdc Medical, Inc. · 1998K882564 · Renalin Cold Sterilant · Minntech Corp. · 19890
  1. Root · 2 generations backK882564 Renalin Cold SterilantMinntech Corp. · 1989
  2. 1 generation backK962959 Peracidin Dialyzer Reprocessing ConcentrateHdc Medical, Inc. · 1998
  3. This deviceK093641 TANGO3 Water Storage Tank with Ozone Disinfection SystemTango3, LLC · 2010
  4. Descendant · cites it directly · 0 of its ownK140984 TANGO3 Water Storage Tank with Ozone Disinfection SystemTango3, LLC · 2015
  5. Descendant · cites it directly · 0 of its ownK141213 Ameriwater Ozone Disinfection SystemAmeriwater · 2014
  6. Descendant · cites it directly · 0 of its ownK142285 AmeriWater Heatsan, 3-phase, 208v, 60hz, AmeriWater Heatsan…Ameriwater · 2014

Most influential descendants

Showing 3 of 3 · See all
  • K140984TANGO3 Water Storage Tank with Ozone Disinfection SystemTango3, LLC · 2015 · cites it directly0descendants
  • K141213Ameriwater Ozone Disinfection SystemAmeriwater · 2014 · cites it directly0descendants
  • K142285AmeriWater Heatsan, 3-phase, 208v, 60hz, AmeriWater Heatsan, 3-phase, 460V, 60hz…Ameriwater · 2014 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K882564Renalin Cold SterilantMinntech Corp.1989—
−1K962959Peracidin Dialyzer Reprocessing ConcentrateHdc Medical, Inc.1998—
0 (this device)K093641TANGO3 Water Storage Tank with Ozone Disinfection SystemTango3, LLC2010—
+1K140984TANGO3 Water Storage Tank with Ozone Disinfection SystemTango3, LLC2015—
+1K141213Ameriwater Ozone Disinfection SystemAmeriwater2014—
+1K142285AmeriWater Heatsan, 3-phase, 208v, 60hz, AmeriWater Heatsan, 3-phase, 460V…Ameriwater2014—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198310
198910
199410
199810
200510

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201420
201510
FIG. 1 — Lineage of K093641, 1989–2015Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FIN? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (6)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K830575Renalin (Tentative
Summary page 2
…inntech Corporation) and other Peracetic Acid products such as Renalin® (K(830575) (Minintech Corporation) and Peracidin® (K962959) (AlcavisHDC) which are commonly used in hemodialysis clinics to disinfe…
Renal Systems, Inc.1983—
*K962959Peracidin Dialyzer Reprocessing Concentrate
Summary page 2
…oducts such as Renalin® (K(830575) (Minintech Corporation) and Peracidin® (K962959) (AlcavisHDC) which are commonly used in hemodialysis clinics to disinfect the reverse osmosis water purification…
Hdc Medical, Inc.1998—
*K043207Ozone Generator
Summary page 2
…with soak times of 30 minutes. Devices with equivalent intended use, technological characteristics and operational characteristics: 0 3Z (K(043207) - The O3Z Ozone System is an optional accessory for the Solution Delivery Systems (SDS) manufactured by GE Infrastructur…
Ge Infrastructure Water & Process Technologies2005—
*K974899Gambro Central Water Treatment System CWP 100 - Wro H
Summary page 2
…tion must be performed during non-operational hours just as TANGO 3. Devices with equivalent intended use and operational characteristics: *CWP 100 - WRO H (K974899) (heat disinfection unit for the distribution. loop only) - The CWP 100 - WRO H has a heat disinfection unit that produces…
Gambro Healthcare19982
*—Minnecare® (Minntech Corporation)
Named without a submission number
—
*—Sodium hypochlorite at 500 ppm dilution
Named without a submission number
—

Also cleared against K830575: 1 other device →

What descends from it? (3)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K140984TANGO3 Water Storage Tank with Ozone Disinfection SystemTango3, LLC20150
K141213Ameriwater Ozone Disinfection SystemAmeriwater20140
K142285AmeriWater Heatsan, 3-phase, 208v, 60hz, AmeriWater Heatsan, 3-phase, 460V, 60hz…Ameriwater20140

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 26, 2012Class IIZ-2138-2012Process controlDuring an FDA inspection, the firm was notified of Quality System deficiencies, which included lack of procedures for design control and customized modifications that were either not reviewed or received proper clearance from FDA. There…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 21 years and passes through 3 companies; a new generation was cleared every 12.5 years on median.
  • 1 of the 7 ancestors has a recall of any class.
  • Of its descendants, K140984 (Tango3, LLC, 2015) has gone furthest, with 0 descendants of its own.
  • 1 other device also cites its predicate K830575.

How does its category compare?

All FIN devices

FIN Tank, Holding, Dialysis: FDA decisions, review times, recalls and adverse event reports over 20 years. 2010, this device’s year, is highlighted.

FDA decisions per year · FIN

1in 2015

2010: 12012: 12014: 12015: 1
2007high 1 (2010) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FIN

Too few FIN 510(k)s in any year since 2007 to give a median.

Recalls per year · FIN

1in 2019

2019: 1
2007high 1 (2019) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FIN

1in 2018

2015: 12017: 12018: 1
2007high 1 (2015) · low 0 (2007)2025

MAUDE reports naming product code FIN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20101—00
20110—00
20121—00
20130—00
20141—00
20151—01
20160—00
20170—01
20180—01
20190—10
20200—00
20210—00
20220—00
20230—00
20240—00
20250—00
20260—0—

What is it indicated for?

The TANGO 3 Water Storage Tank with Ozone Disinfection System is intended to be used for disinfection of the water distribution system of a dialysis facility. The tank of the TANGO is also used as the water holding tank of the distribution system. The disinfection process is completely automated. Ozone concentration during disinfection is between 0.2 ppmn and 0.3 ppm. The distribution system will be exposed to ozone for one (1) period of 45 minutes and three (3) subsequent periods of 30 minutes with adequate water flushes between them and at the end, leaving the distribution loop without ozone.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K093641: TANGO3 Water Storage Tank with Ozone Disinfection System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K093641

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026