510(k) · Cleared
Recalled · Class II · 2014Gambro Central Water Treatment System CWP 100 - Wro H
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Gambro Healthcare · FIP Subsystem, Water Purification
Cleared on Sep 23, 1998 as substantially equivalent to K931595 (Osmonics, Inc., 1994) after a 267-day review. Its family tree reaches back 1 generation to 1994. 14 later devices descend from it. It has been recalled 2 times.
- Decided
- Sep 23, 1998
- Review
- 267 days
- Class
- Class II
- Product code
- FIP
- Recalls
- 2
- Predicates
- 1
Its longest line of descent runs back 1 generation to K931595 (1994). 14 later devices descend from it, 5 citing it directly.
- Root · 1 generation backK931595 Reverse Osmosis Equipment for Water PurificationOsmonics, Inc. · 1994
- This deviceK974899 Gambro Central Water Treatment System CWP 100 - Wro HGambro Healthcare · 1998
- Descendant · cites it directly · 5 of its ownK124059 Aquaboss Ecoro Dia I (HT); Aquaboss Ecoro Dia II (HT)Lauer Membran Wassertechnik GmbH · 2013
- Descendant · cites it directly · 3 of its ownK133829 Aquabplus,Aquabplus & Aquabplus B2,Aquabplus&Aquabplus HF,Aquabplus&…Vivonic GmbH · 2014 · Recalled
- Descendant · cites it directly · 3 of its ownK093641 TANGO3 Water Storage Tank with Ozone Disinfection SystemTango3, LLC · 2010 · Recalled
- Descendant · cites it directly · 2 of its ownK131904 Mediqa Single Pass Reverse Osmosis System with Heat Sanitization…Ameriwater · 2013
- Descendant · 2 generations on · 1 of its ownK213507 AquaAFresenius Medical Care Rental Therapies Group, LLC · 2022
Showing 5 of 14 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 14 · See all- K124059Aquaboss Ecoro Dia I (HT); Aquaboss Ecoro Dia II (HT)Lauer Membran Wassertechnik GmbH · 2013 · cites it directly5descendants
- K133829Aquabplus,Aquabplus & Aquabplus B2,Aquabplus&Aquabplus HF,Aquabplus& Aquabplus B2&…Vivonic GmbH · 2014 · cites it directly3descendants
- K093641TANGO3 Water Storage Tank with Ozone Disinfection SystemTango3, LLC · 2010 · cites it directly3descendants
- K131904Mediqa Single Pass Reverse Osmosis System with Heat Sanitization, Mediqa Double Pass…Ameriwater · 2013 · cites it directly2descendants
- K213507AquaAFresenius Medical Care Rental Therapies Group, LLC · 2022 · 2 generations on1descendant
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K931595 | Reverse Osmosis Equipment for Water Purification | Osmonics, Inc. | 1994 | — |
| 0 (this device) | K974899 | Gambro Central Water Treatment System CWP 100 - Wro H | Gambro Healthcare | 1998 | — |
| +1 | K124059 | Aquaboss Ecoro Dia I (HT); Aquaboss Ecoro Dia II (HT) | Lauer Membran Wassertechnik GmbH | 2013 | — |
| +1 | K133829 | Aquabplus,Aquabplus & Aquabplus B2,Aquabplus&Aquabplus HF,Aquabplus& Aquabplus… | Vivonic GmbH | 2014 | 3 |
| +1 | K093641 | TANGO3 Water Storage Tank with Ozone Disinfection System | Tango3, LLC | 2010 | 1 |
| +1 | K131904 | Mediqa Single Pass Reverse Osmosis System with Heat Sanitization, Mediqa… | Ameriwater | 2013 | — |
| +2 | K213507 | AquaA | Fresenius Medical Care Rental Therapies Group, LLC | 2022 | — |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2000 | 1 | 0 |
| 2010 | 1 | 0 |
| 2013 | 2 | 0 |
| 2014 | 4 | 0 |
| 2015 | 1 | 0 |
| 2022 | 1 | 0 |
| 2023 | 2 | 0 |
| 2024 | 1 | 0 |
| 2025 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in FIP? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new FIP device
- #1 K251851 DIASAFEplusUS (F00013010)score 14.1 · 14 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K931595 | Reverse Osmosis Equipment for Water PurificationSummary page 4…n systems for hemodialysis in commercial distribution in the United States to include the Osmonics Osmo 23 G Series Reverse Osmosis Machines (FDA Document Control Number K931595 B. Both systems utilized reverse osmosis for purification of water to be used for hemodialysis. Both systems utilize a polyamide,… | 1994 |
What descends from it? (14)
See all 145 later devices cite it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K124059 | Aquaboss Ecoro Dia I (HT); Aquaboss Ecoro Dia II (HT) | 2013 | 5 |
| K133829 | Aquabplus,Aquabplus & Aquabplus B2,Aquabplus&Aquabplus HF,Aquabplus& Aquabplus B2&… | 2014 | 3 |
| K093641 | TANGO3 Water Storage Tank with Ozone Disinfection System | 2010 | 3 |
| K131904 | Mediqa Single Pass Reverse Osmosis System with Heat Sanitization, Mediqa Double Pass… | 2013 | 2 |
| K991916 | Virginia Water Systems' Water Purification System for Hemodialysis | 2000 | 0 |
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Apr 24, 2014 | Class II | Z-1563-2014 | Nonconforming Material/Component | Mar Cor Purification manufactured Central Water Plant (CWP) reverse osmosis system, models 102, 104 and 106. This correction has been initiated due to a potential situation with a specific lot of the water inlet solenoid valve. The… |
| Jun 27, 2012 | Class II | Z-1201-2014 | Component design/selection | There exists a potential situation with the crimp connection at the pump head assembly. The potential exists that the reverse osmosis pump head wiring could build resistance resulting in resistive heating of the surrounding components. If… |
No ancestor or descendant has a Class I recall.
What does this lineage say?
How does its category compare?
All FIP devicesFIP Subsystem, Water Purification: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · FIP
4in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · FIP
88in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · FIP
1in 2020
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · FIP
77in 2025
MAUDE reports naming product code FIP, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 1 | — | 0 | 1 |
| 2008 | 3 | 202 | 0 | 5 |
| 2009 | 3 | 167 | 0 | 0 |
| 2010 | 2 | — | 0 | 12 |
| 2011 | 4 | 116 | 0 | 4 |
| 2012 | 2 | — | 0 | 4 |
| 2013 | 4 | 125 | 0 | 41 |
| 2014 | 4 | 126 | 0 | 34 |
| 2015 | 2 | — | 0 | 12 |
| 2016 | 2 | — | 0 | 3 |
| 2017 | 1 | — | 0 | 6 |
| 2018 | 1 | — | 0 | 3 |
| 2019 | 2 | — | 0 | 13 |
| 2020 | 1 | — | 1 | 12 |
| 2021 | 1 | — | 0 | 35 |
| 2022 | 1 | — | 0 | 57 |
| 2023 | 3 | 69 | 0 | 140 |
| 2024 | 1 | — | 0 | 174 |
| 2025 | 4 | 88 | 0 | 77 |
| 2026 | 0 | — | 0 | — |
What is it indicated for?
The Gambro Central Water Treatment System CWP 100 — WRO H is designed to produce water of adquate quality for hemodialysis, both chemically and microbiologically with an adequate flow, provided that the feed water complies with the existing standards for drinking water and has been properly pre-treated.
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K974899 · printed Nov 4, 2026
Cite this page
510(k) Report. “K974899: Gambro Central Water Treatment System CWP 100 - Wro H.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K974899
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026