510(k) Report

510(k) · Cleared

Cleared

AmeriWater Heatsan, 3-phase, 208v, 60hz, AmeriWater Heatsan, 3-phase, 460V, 60hz, Ameriwater Heatsan, 3-phase, 575v, 60hz

K142285510(k) TraditionalCleared Dec 1, 2014105-day reviewClass II21 CFR 876.5665

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ameriwater · FIP Subsystem, Water Purification

Cleared on Dec 1, 2014 as substantially equivalent to K093641 (Tango3, LLC, 2010) after a 105-day review. Its family tree reaches back 3 generations to 1989.

Decided
Dec 1, 2014
Review
105 days
Class
Class II
Product code
FIP
Recalls
None
Predicates
1

Its longest line of descent runs back 3 generations to K882564 (1989). 8 ancestors in all, the oldest K830575 (1983). No later device cites it as a predicate yet.

  1. Root · 3 generations backK882564 Renalin Cold SterilantMinntech Corp. · 1989
  2. 2 generations backK962959 Peracidin Dialyzer Reprocessing ConcentrateHdc Medical, Inc. · 1998
  3. 1 generation backK093641 TANGO3 Water Storage Tank with Ozone Disinfection SystemTango3, LLC · 2010 · Recalled
  4. This deviceK142285 AmeriWater Heatsan, 3-phase, 208v, 60hz, AmeriWater Heatsan…Ameriwater · 2014
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K882564Renalin Cold SterilantMinntech Corp.1989—
−2K962959Peracidin Dialyzer Reprocessing ConcentrateHdc Medical, Inc.1998—
−1K093641TANGO3 Water Storage Tank with Ozone Disinfection SystemTango3, LLC20101
0 (this device)K142285AmeriWater Heatsan, 3-phase, 208v, 60hz, AmeriWater Heatsan, 3-phase, 460V…Ameriwater2014—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198310
198910
199410
199810
200510
FIG. 1 — Lineage of K142285, 1989–2014Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FIP? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FIP device

  1. #1 K251851 DIASAFEplusUS (F00013010)score 14.1 · 14 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K093641TANGO3 Water Storage Tank with Ozone Disinfection System
Summary page 4
…r §876.5665 Panel: Gastroenterology Product Code: FIP Equivalent Device: K093641, TANGO3 Ozone System Device Description: The AmeriWater Heatsan is intended for use in hospitals, clinics, or dialysis cen…
Tango3, LLC20101
†K974899Gambro Central Water Treatment System CWP 100 - Wro HGambro Healthcare19982
†K924695Aseptech Portable RO+ SystemDayton Water Systems1995—
†K991519Ameriwater Purification System for HemodialysisAmeriwater2000—
†K111740Ameriwater Mro Portable Reverse Osmosis System - Single Patient, 1 to 3 StationAmeriwater20111
†K051031Ameriwater Solution Mix and Distribution System for HemodialysisAmeriwater2006—

Also cleared against K093641: 2 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 25 years and passes through 4 companies; a new generation was cleared every 9.4 years on median.
  • 2 of the 8 ancestors have a recall of any class.
  • 2 other devices also cite its predicate K093641.

How does its category compare?

All FIP devices

FIP Subsystem, Water Purification: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.

FDA decisions per year · FIP

4in 2025

2007: 12008: 32009: 32010: 22011: 42012: 22013: 42014: 42015: 22016: 22017: 12018: 12019: 22020: 12021: 12022: 12023: 32024: 12025: 4
2007high 4 (2011) · low 1 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FIP

88in 2025

2008: 202 days median2009: 167 days median2011: 116 days median2013: 125 days median2014: 126 days median2023: 69 days median2025: 88 days median
2007high 202 (2008) · low 69 (2023)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FIP

1in 2020

2020: 1
2007high 1 (2020) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FIP

77in 2025

2007: 12008: 52010: 122011: 42012: 42013: 412014: 342015: 122016: 32017: 62018: 32019: 132020: 122021: 352022: 572023: 1402024: 1742025: 77
2007high 174 (2024) · low 0 (2009)2025

MAUDE reports naming product code FIP, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—01
2008320205
2009316700
20102—012
2011411604
20122—04
20134125041
20144126034
20152—012
20162—03
20171—06
20181—03
20192—013
20201—112
20211—035
20221—057
20233690140
20241—0174
2025488077
20260—0—

What is it indicated for?

The AmeriWater Heatsan is an optional accessory for water treatment systems intended for use in hemodialysis applications. The Heatsan is designed to heat water received from the water purification system, and circulate the heated water around the distribution loop at a controlled sanitizing temperature (minimum temperature of 185°F) for a programmed length of time (minimum hold time of 2 hours) to provide heat disinfection of the water distribution loop. The AmeriWater Heatsan is intended for use in hospitals, clinics, or dialysis centers to provide heat disinfection resulting in total viable microbial counts below 50 CFU/ml.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K142285: AmeriWater Heatsan, 3-phase, 208v, 60hz, AmeriWater Heatsan, 3-phase, 460V, 60hz, Ameriwater Heatsan, 3-phase, 575v, 60hz.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K142285

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026