510(k) Report

510(k) · Cleared

Cleared

Inspira ART100

K232788510(k) TraditionalCleared May 24, 2024256-day reviewClass II21 CFR 870.4380

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Inspira Technologies Oxy B.H.N., Ltd. · DWA Control, Pump Speed, Cardiopulmonary Bypass

Cleared on May 24, 2024 as substantially equivalent to K991864 (Jostra Medizintechnik AG, 2000) after a 256-day review. Its family tree reaches back 1 generation to 2000.

Decided
May 24, 2024
Review
256 days
Class
Class II
Product code
DWA
Recalls
None
Predicates
1

Its longest line of descent runs back 1 generation to K991864 (2000). No later device cites it as a predicate yet.

  1. Root · 1 generation backK991864 Rotaflow Centrifugal Pump SystemJostra Medizintechnik AG · 2000 · Recalled
  2. This deviceK232788 Inspira ART100Inspira Technologies Oxy B.H.N., Ltd. · 2024
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K991864Rotaflow Centrifugal Pump SystemJostra Medizintechnik AG20002
0 (this device)K232788Inspira ART100Inspira Technologies Oxy B.H.N., Ltd.2024—
FIG. 1 — Lineage of K232788, 2000–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DWA? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DWA device

  1. #1 K221491 Anivia SG1000 Pump Consolescore 9.3 · 6 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K991864Rotaflow Centrifugal Pump System
Summary page 4
…Classification Panel: Cardiovascular 3. Predicate Device: RotaFlow Centrifugal Pump System, cleared under K991864 4. Device Description: The INSPIRA ART100 system is a device for pumping blood without direct contact in an e…
Jostra Medizintechnik AG20002
†K221491Anivia SG1000 Pump ConsoleApmtd, Inc.2023—
†K230698Anivia SG1000 Pump ConsoleApmtd, Inc.2023—

Also cleared against K991864: 8 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has a recall.
  • 8 other devices also cite its predicate K991864.

How does its category compare?

All DWA devices

DWA Control, Pump Speed, Cardiopulmonary Bypass: FDA decisions, review times, recalls and adverse event reports over 20 years. 2024, this device’s year, is highlighted.

FDA decisions per year · DWA

1in 2024

2007: 12008: 32009: 22010: 22011: 22013: 22019: 12020: 22023: 22024: 1
2007high 3 (2008) · low 0 (2012)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DWA

44in 2008

2008: 44 days median
2007high 44 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DWA

1in 2022

2010: 12012: 32013: 12017: 12021: 12022: 1, 1 Class I
2007high 3 (2012) · low 0 (2007)2026

Class I (1 since 2007) Class II/III not classified

Adverse event reports per year · DWA

362in 2025

2007: 102008: 102009: 172010: 1102011: 1472012: 1022013: 1002014: 982015: 1032016: 1092017: 1402018: 2002019: 3762020: 1462021: 1552022: 2052023: 4382024: 3892025: 362
2007high 438 (2023) · low 10 (2007)2025

MAUDE reports naming product code DWA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—010
2008344010
20092—017
20102—1110
20112—0147
20120—3102
20132—1100
20140—098
20150—0103
20160—0109
20170—1140
20180—0200
20191—0376
20202—0146
20210—1155
20220—1 (1)205
20232—0438
20241—0389
20250—0362
20260—0—

What is it indicated for?

The INSPIRA ART100 system is intended for use in an extra corporeal perfusion circuit to pump blood during short duration cardiopulmonary bypass procedures lasting 6 hours or less.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K232788: Inspira ART100.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K232788

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026