510(k) Report

510(k) · Cleared

Recalled · Class II · 2024

Evis Exeraii 180 System

K100584510(k) TraditionalCleared Jul 2, 2010122-day reviewClass II21 CFR 876.1500Recalls 9

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Olympus Medical Systems Corporation · NWB Endoscope, Accessories, Narrow Band Spectrum

Cleared on Jul 2, 2010 as substantially equivalent to K051645 (Olympus Medical Systems Corporation, 2005) and 2 other predicates after a 122-day review. Its family tree reaches back 7 generations to 1993. 7 of the 160 devices in its family, counting itself, have a Class I recall. 132 later devices descend from it. It has been recalled 9 times.

Decided
Jul 2, 2010
Review
122 days
Class
Class II
Product code
NWB
Recalls
9
Predicates
3

Its longest line of descent runs back 7 generations to K931154 (1993). 27 ancestors in all, the oldest K843084 (1985). 132 later devices descend from it, 22 citing it directly. 7 of the 160 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2006: 1 other ancestor2005: 2 other ancestors2003: 1 other ancestor, 1 with a Class I recall12002: 1 other ancestor2001: 1 other ancestor2000: 3 other ancestors1996: 1 other ancestor1991: 2 other ancestors, 1 with a Class I recall11986: 1 other ancestor1995: 2 other ancestors1993: 1 other ancestor1992: 2 other ancestors1985: 2 other ancestors2012: 6 descendants cleared2018: 8 descendants cleared2016: 2 descendants cleared2013: 5 descendants cleared, 1 with a Class I recall12014: 3 descendants cleared2022: 7 descendants cleared2023: 11 descendants cleared2020: 16 descendants cleared2024: 15 descendants cleared2026: 14 descendants cleared2021: 9 descendants cleared, 1 with a Class I recall12015: 6 descendants cleared2019: 8 descendants cleared, 1 with a Class I recall12025: 19 descendants cleared, 1 with a Class I recall12017: 3 descendants clearedK931154 · Evis 200 System · Olympus Corp. · 1993K963033 · BF 240/P240/IT240 Bronchovideoscope & Accessories · Olympus America, Inc. · 1996 · recalledK981543 · Olympus LF-TP and LF-DP Tracheal Intubation Fiberscopes, Accessories and Ancillary Equipment · The Olympus Optical Co. · 1998 · recalled×2: 2 generations cleared close together, K013591 (2001), K021073 (2002) K031648 · Visera Rhino-Laryngovideoscope Olympus Enf Type V · Olympus Optical Co., Ltd. · 2003K061313 · Evis Exera 180 System · Olympus Medical Systems Corporation · 2006 · Class I recall1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors2 generations back: 7 other ancestors, 1 with a Class I recall13 generations back: 4 other ancestors, 1 with a Class I recall14 generations back: 7 other ancestors 1 generation forward: 22 descendants, 1 with a Class I recall12 generations forward: 39 descendants3 generations forward: 40 descendants, 2 with a Class I recall24 generations forward: 17 descendants, 1 with a Class I recall15 generations forward: 8 descendants6 generations forward: 3 descendants7 generations forward: 1 descendant8 generations forward: 1 descendant9 generations forward: 1 descendant K061313 · Evis Exera 180 System · Olympus Medical Systems Corporation · 2006 · Class I recallK031648 · Visera Rhino-Laryngovideoscope Olympus Enf Type V · Olympus Optical Co., Ltd. · 2003K021073 · Visera Rhinlaryngovideoscope Olympus Enf Type V · Olympus America, Inc. · 2002K013591 · Xenf-TP Rhino-Laryngofiberscope, Its Acessories and Ancillary Equipment · Olympus Optical Co., Ltd. · 2001K981543 · Olympus LF-TP and LF-DP Tracheal Intubation Fiberscopes, Accessories and Ancillary Equipment · The Olympus Optical Co. · 1998 · recalledK963033 · BF 240/P240/IT240 Bronchovideoscope & Accessories · Olympus America, Inc. · 1996 · recalledK931154 · Evis 200 System · Olympus Corp. · 1993−50+5
  1. Root · 7 generations backK931154 Evis 200 SystemOlympus Corp. · 1993
  2. 6 generations backK963033 BF 240/P240/IT240 Bronchovideoscope & AccessoriesOlympus America, Inc. · 1996 · Recalled
  3. 5 generations backK981543 Olympus LF-TP and LF-DP Tracheal Intubation Fiberscopes, Accessories…The Olympus Optical Co. · 1998 · Recalled
  4. 4 generations backK013591 Xenf-TP Rhino-Laryngofiberscope, Its Acessories and Ancillary…Olympus Optical Co., Ltd. · 2001
  5. 3 generations backK021073 Visera Rhinlaryngovideoscope Olympus Enf Type VOlympus America, Inc. · 2002
  6. 2 generations backK031648 Visera Rhino-Laryngovideoscope Olympus Enf Type VOlympus Optical Co., Ltd. · 2003
  7. 1 generation backK061313 Evis Exera 180 SystemOlympus Medical Systems Corporation · 2006 · Recalled (Class I)
  8. This deviceK100584 Evis Exeraii 180 SystemOlympus Medical Systems Corporation · 2010
  9. Descendant · cites it directly · 57 of its ownK112680 Evis Exera Iii Video SystemOlympus Medical Systems Corporation · 2012 · Recalled
  10. Descendant · cites it directly · 34 of its ownK111425 Olympus LTF-S190-10, Olumpus Otv-S190, Olympus CLV-S190Olympus Medical Systems Corp. · 2012 · Recalled
  11. Descendant · 2 generations on · 19 of its ownK123365 Olympus LTF-190-10-3D, Maj-Y0154, Olympus CV-190Olympus Medical Systems Corp. · 2013
  12. Descendant · 2 generations on · 19 of its ownK122831 Video System CenterOlympus Medical Systems Corp. · 2012
  13. Descendant · cites it directly · 16 of its ownK140149 Epx-4440HD Digital Video ProcessorFujifilm Medical Systems U.S.A, Inc. · 2014

Showing 5 of 132 descendants. See all descendants

Descendants cleared per year

2012: 62013: 52014: 32015: 62016: 22017: 32018: 82019: 82020: 162021: 92022: 72023: 112024: 152025: 192026: 14
2012high 19 (2025) · low 2 (2016)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 132 · See all
  • K112680Evis Exera Iii Video SystemOlympus Medical Systems Corporation · 2012 · cites it directly57descendants
  • K111425Olympus LTF-S190-10, Olumpus Otv-S190, Olympus CLV-S190Olympus Medical Systems Corp. · 2012 · cites it directly34descendants
  • K123365Olympus LTF-190-10-3D, Maj-Y0154, Olympus CV-190Olympus Medical Systems Corp. · 2013 · 2 generations on19descendants
  • K122831Video System CenterOlympus Medical Systems Corp. · 2012 · 2 generations on19descendants
  • K140149Epx-4440HD Digital Video ProcessorFujifilm Medical Systems U.S.A, Inc. · 2014 · cites it directly16descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K931154Evis 200 SystemOlympus Corp.1993—
−6K963033BF 240/P240/IT240 Bronchovideoscope & AccessoriesOlympus America, Inc.19966
−5K981543Olympus LF-TP and LF-DP Tracheal Intubation Fiberscopes, Accessories and…The Olympus Optical Co.19985
−4K013591Xenf-TP Rhino-Laryngofiberscope, Its Acessories and Ancillary EquipmentOlympus Optical Co., Ltd.2001—
−3K021073Visera Rhinlaryngovideoscope Olympus Enf Type VOlympus America, Inc.2002—
−2K031648Visera Rhino-Laryngovideoscope Olympus Enf Type VOlympus Optical Co., Ltd.2003—
−1K061313Evis Exera 180 SystemOlympus Medical Systems Corporation2006Class I
0 (this device)K100584Evis Exeraii 180 SystemOlympus Medical Systems Corporation2010—
+1K112680Evis Exera Iii Video SystemOlympus Medical Systems Corporation20126
+1K111425Olympus LTF-S190-10, Olumpus Otv-S190, Olympus CLV-S190Olympus Medical Systems Corp.20121
+2K123365Olympus LTF-190-10-3D, Maj-Y0154, Olympus CV-190Olympus Medical Systems Corp.2013—
+2K122831Video System CenterOlympus Medical Systems Corp.2012—
+1K140149Epx-4440HD Digital Video ProcessorFujifilm Medical Systems U.S.A, Inc.2014—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198520
198610
199121
199220
199310
199520
199610
200030
200110
200210
200311
200520
200610

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201260
201351
201430
201560
201620
201730
201880
201981
2020160
202191
202270
2023110
2024150
2025191
2026140
FIG. 1 — Lineage of K100584, 1993–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

7 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new NWB device

  1. #1 K211882 HD-550 Video Endoscope Systemscore 8.5 · 9 generations · no Class I recall
  2. #2 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K051645Evis Exera 160A System
Summary page 2
…0AY2L Subject Devnce Predicate D v 5':2 i (Part of this Submlsswn) e lca e_ svice ' “Né ) EVIS EXERA Il XENON LIGHT SOURCE OLYMPUS CLV-180 EVIS EXERA Xenon Light Source K051645 (Hereinafter referred to as CLV-180) Olympus CLV-160A (K061313, K062049) EVIS EXERA Il VIDEO SYSTEM CENTER OLYMPUS CV-180 EVIS EX…
Olympus Medical Systems Corporation20053
*K061313Evis Exera 180 System
Summary page 2
…swn) e lca e_ svice ' “Né ) EVIS EXERA Il XENON LIGHT SOURCE OLYMPUS CLV-180 EVIS EXERA Xenon Light Source K051645 (Hereinafter referred to as CLV-180) Olympus CLV-160A (K061313, K062049) EVIS EXERA Il VIDEO SYSTEM CENTER OLYMPUS CV-180 EVIS EXERA Video System Center K051645 (Hereinafter referred to as CV-…
Olympus Medical Systems Corporation2006Class I
*K062049Evis Exera II 180 System
Summary page 2
…a e_ svice ' “Né ) EVIS EXERA Il XENON LIGHT SOURCE OLYMPUS CLV-180 EVIS EXERA Xenon Light Source K051645 (Hereinafter referred to as CLV-180) Olympus CLV-160A (K061313, K062049) EVIS EXERA Il VIDEO SYSTEM CENTER OLYMPUS CV-180 EVIS EXERA Video System Center K051645 (Hereinafter referred to as CV-180) .Oly…
Olympus Medical Systems Corporation2006—

Also cleared against K051645: 8 other devices →

What descends from it? (132)

See all 132

22 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K112680Evis Exera Iii Video SystemOlympus Medical Systems Corporation201257
K111425Olympus LTF-S190-10, Olumpus Otv-S190, Olympus CLV-S190Olympus Medical Systems Corp.201234
K140149Epx-4440HD Digital Video ProcessorFujifilm Medical Systems U.S.A, Inc.201416
K121959Evis Exera Iii Video System Class IOlympus Medical Systems Corp.201314
K131855Pentax Video Colonoscopes (EC Family)Pentax Medical Company201413
K131902Pentax Video Upper G.I. Scopes (EG Family)Pentax Medical Company201410
K230332Ambu® aScope™ Colon; Ambu® aBox™ 2Ambu A/S20236
K122470Pentax Epk-I5010 Video ProcessorPentax Medical20136
K212382Ambu® aScope™ Gastro, Ambu® aBox™2Ambu A/S20222
K201098Ambu Duodeno SystemAmbu Innovation GmbH20202

Has anything in its family been recalled? (7 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 11, 2024Class IIZ-1208-2024Nonconforming Material/ComponentSpecific repaired colonovideoscopes were assembled without an adhesive which protects them from certain physical conditions such as high levels of vibration, fluctuation in temperature or severe shock loads.
Jan 11, 2024Class IIZ-1209-2024Nonconforming Material/ComponentSpecific repaired colonovideoscopes were assembled without an adhesive which protects them from certain physical conditions such as high levels of vibration, fluctuation in temperature or severe shock loads.
Jan 11, 2024Class IIZ-1214-2024Nonconforming Material/ComponentSpecific repaired colonovideoscopes were assembled without an adhesive which protects them from certain physical conditions such as high levels of vibration, fluctuation in temperature or severe shock loads.
Jan 11, 2024Class IIZ-1205-2024Nonconforming Material/ComponentSpecific repaired colonovideoscopes were assembled without an adhesive which protects them from certain physical conditions such as high levels of vibration, fluctuation in temperature or severe shock loads.
Sep 25, 2023Class IIZ-0329-2024Under Investigation by firmChannel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead…
Sep 25, 2023Class IIZ-0334-2024Under Investigation by firmChannel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead…
Dec 21, 2022Class IIZ-1132-2023Process designA single CF-Q180AL colonovideoscope was utilized in a veterinary endoscopy procedure in advance of being assigned to a medical facility as a service loaner in error, potential for microbial contamination
Apr 22, 2022Class IIZ-1274-2022Nonconforming Material/ComponentA (Camera unit) CCD intended for a different model of colonoscope may have been installed in the CF-H180AL, viewing angle would be narrower than a normal CF-H180AL and the use of EndoTherapy devices may stick out (approximately 1.4 mm)…
Dec 17, 2018Class IIZ-1009-2019Under Investigation by firmThe adhesive used in the repair of the endoscope was incorrectly prepared. It is unknown what affect the nonconforming adhesive mixture would have on the durability or effectiveness of the gluing operation for the endoscope repair.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K061313Evis Exera 180 SystemOlympus Medical Systems Corporation · 2006 · Direct predicate · ◆ on the line of descent in FIG. 16 Class I recalls
  • K023984Modification to Olympus Bronchoscopes BF-40 Series, BF-240 Series, and BF-160 SeriesThe Olympus Optical Co. · 2003 · 2 generations back · ◆ in the 2003 ancestors bar of FIG. 114 Class I recalls
  • K912120Urf-P2 Ureterorenofiberscope/CholedochofiberscopeOlympus Corp. · 1991 · 3 generations back · ◆ in the 1991 ancestors bar of FIG. 11 Class I recall
  • K121959Evis Exera Iii Video SystemOlympus Medical Systems Corp. · 2013 · Cites it directly · ◆ in the 2013 descendants bar of FIG. 18 Class I recalls
  • K183419EVIS EXERA III Bronchovideoscope Olympus BF-XT190Olympus Medical Systems Corp. · 2019 · 3 generations on · ◆ in the 2019 descendants bar of FIG. 12 Class I recalls
  • K250862Evis Exera Iii Bronchovideoscope (Olympus BF-XP190); Evis Exera Iii Bronchovideoscope (Olympus BF-P190)…Olympus Medical Systems Corp. · 2025 · 3 generations on · ◆ in the 2025 descendants bar of FIG. 14 Class I recalls
  • K201758EVIS EXERA III Bronchovideoscope Olympus BF-XP190, EVIS EXERA III Bronchovideoscope Olympus BF-P190Olympus Medical Systems Corp. · 2021 · 4 generations on · ◆ in the 2021 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 17 years and passes through 5 companies; a new generation was cleared every 2.9 years on median.
  • 15 of the 27 ancestors have a recall of any class; 3 have a Class I recall.
  • Of its descendants, K112680 (Olympus Medical Systems Corporation, 2012) has gone furthest, with 57 descendants of its own.
  • 8 other devices also cite its predicate K051645.

How does its category compare?

All NWB devices

NWB Endoscope, Accessories, Narrow Band Spectrum: FDA decisions, review times, recalls and adverse event reports over 20 years. 2010, this device’s year, is highlighted.

FDA decisions per year · NWB

2in 2022

2008: 12009: 12010: 12014: 12018: 12022: 2
2007high 2 (2022) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · NWB

Too few NWB 510(k)s in any year since 2007 to give a median.

Recalls per year · NWB

5in 2023

2018: 32023: 5, 2 Class I
2007high 5 (2023) · low 0 (2007)2026

Class I (2 since 2007) Class II/III not classified

Adverse event reports per year · NWB

2,847in 2025

2007: 22008: 32009: 12010: 22011: 12012: 12013: 92014: 222015: 322016: 252017: 172018: 42019: 32020: 8232021: 1,4532022: 1,8282023: 2,1802024: 2,6482025: 2,847
2007high 2,847 (2025) · low 1 (2009)2025

MAUDE reports naming product code NWB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—02
20081—03
20091—01
20101—02
20110—01
20120—01
20130—09
20141—022
20150—032
20160—025
20170—017
20181—34
20190—03
20200—0823
20210—01,453
20222—01,828
20230—5 (2)2,180
20240—02,648
20250—02,847
20260—0—

What is it indicated for?

EVIS EXERA II XENON LIGHT SOURCE OLYMPUS CLV-180. This light source has been designed to be used with Olympus endoscopes, video system center, and other ancillary equipment for endoscopic diagnosis, treatment and video observation. EVIS EXERA II VIDEO SYSTEM CENTER OLYMPUS CV-180 This video system center has been designed to be used with OLYMPUS camera heads, endoscopes, light sources, monitors, endo-therapy accessories and other ancillary equipment for…

Read the full statement (284 words)

EVIS EXERA II XENON LIGHT SOURCE OLYMPUS CLV-180.

This light source has been designed to be used with Olympus endoscopes, video system center, and other ancillary equipment for endoscopic diagnosis, treatment and video observation.

EVIS EXERA II VIDEO SYSTEM CENTER OLYMPUS CV-180

This video system center has been designed to be used with OLYMPUS camera heads, endoscopes, light sources, monitors, endo-therapy accessories and other ancillary equipment for endoscopic diagnosis, treatment and video observation.

EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF TYPE N180, OLYMPUS GIF TYPE XP180N

These instruments have been designed to be used with an Olympus video system center, light source, documentation equipment, monitor, EndloTherapy accessories (such as a biopsy forceps), and other ancillary equipment for transoral or transnasal endloscopy and endoscopic surgery within the upper digestive tract (including the esophagus, stomach, and duodenum).

EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF TYPE Q180. OLYMPUS GIF TYPE H180

These instruments have been designed to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endloscopy and endloscopic surgery within the upper digestive tract (including the esophagus, stomach, and duodenum).

EVIS EXERA II COLONOVIDEOSCOPE OLYMPUS CF TYPE Q180AL. OLYMPUS CF TYPE Q180AI, OLYMPUS CF TYPE H180AL, OLYMPUS CF TYPE H180AI. OLYMPUS PCF TYPE Q180AL (Pediatric). OLYMPUS PCF TYPE Q180AI (Pediatric), OLYMPUS PCF TYPE H180AL (Pediatric). OLYMPUS PCF TYPE H180AI (Pediatric),

These instruments have been designed to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endloscopic surgery within the lower digestive tract (including the anus, rectum, sigmoid colon, colon, and ileocecal valve).

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K100584: Evis Exeraii 180 System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K100584

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026