510(k) Report

510(k) · Cleared

Cleared

Sterigage Model 1243RA, 1243RB, 1243RE

K101249510(k) TraditionalCleared Sep 8, 2010127-day reviewClass II21 CFR 880.2800

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

3M Company · JOJ Indicator, Physical/Chemical Sterilization Process

Cleared on Sep 8, 2010 as substantially equivalent to K771080 (Info-Chem, 1977) after a 127-day review. Its family tree reaches back 1 generation to 1977. 4 later devices descend from it.

Decided
Sep 8, 2010
Review
127 days
Class
Class II
Product code
JOJ
Recalls
None
Predicates
1

Its longest line of descent runs back 1 generation to K771080 (1977). 4 later devices descend from it, 1 citing it directly.

Descendants, by year cleared 2019: 2 descendants cleared2022: 1 descendant cleared2026: 1 descendant clearedK771080 · Steam Sterilization Indicator · Info-Chem · 197719801990200020102020
Descendants, by generation 1 generation forward: 1 descendant2 generations forward: 1 descendant3 generations forward: 1 descendant4 generations forward: 1 descendant K771080 · Steam Sterilization Indicator · Info-Chem · 19770
  1. Root · 1 generation backK771080 Steam Sterilization IndicatorInfo-Chem · 1977
  2. This deviceK101249 Sterigage Model 1243RA, 1243RB, 1243RE3M Company · 2010
  3. Descendant · cites it directly · 3 of its ownK191236 3M™ Attest™ Steam Chemical Integrators3M Company · 2019
  4. Descendant · 2 generations on · 2 of its ownK193254 3M Attest Steam Chemical Integrator3M Company · 2019
  5. Descendant · 3 generations on · 1 of its ownK220942 3M Attest Steam Chemical Integrators (1243A, 1243B, 1243RE, 1243RES)3M Company · 2022 · Recalled
  6. Descendant · 4 generations on · 0 of its ownK261992 3M™ Attest™ Steam Chemical Integrator (1243A, 1243B, 1243RE, 1243RES)Solventum US, LLC · 2026

Most influential descendants

Showing 4 of 4 · See all
  • K1912363M™ Attest™ Steam Chemical Integrators3M Company · 2019 · cites it directly3descendants
  • K1932543M Attest Steam Chemical Integrator3M Company · 2019 · 2 generations on2descendants
  • K2209423M Attest Steam Chemical Integrators (1243A, 1243B, 1243RE, 1243RES)3M Company · 2022 · 3 generations on1descendant
  • K2619923M™ Attest™ Steam Chemical Integrator (1243A, 1243B, 1243RE, 1243RES)Solventum US, LLC · 2026 · 4 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K771080Steam Sterilization IndicatorInfo-Chem1977—
0 (this device)K101249Sterigage Model 1243RA, 1243RB, 1243RE3M Company2010—
+1K1912363M™ Attest™ Steam Chemical Integrators3M Company2019—
+2K1932543M Attest Steam Chemical Integrator3M Company2019—
+3K2209423M Attest Steam Chemical Integrators (1243A, 1243B, 1243RE, 1243RES)3M Company20222
+4K2619923M™ Attest™ Steam Chemical Integrator (1243A, 1243B, 1243RE, 1243RES)Solventum US, LLC2026—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201920
202210
202610
FIG. 1 — Lineage of K101249, 1977–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JOJ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JOJ device

  1. #1 K261992 3M™ Attest™ Steam Chemical Integrator (1243A…score 13.3 · 5 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K771080Steam Sterilization Indicator
Summary page 3
…SteriGageTm Chemical Integrator for Steam is substantially equival ent to the predicate device, the current 3M~m ComplyTm SteriGageTm Chemical Integrator, cleared under K771080 in terms of design, intended use, indications for use, composition, physical properties and technological characteristics. The on…
Info-Chem1977—

What descends from it? (4)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K1912363M™ Attest™ Steam Chemical Integrators3M Company20193

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has no recall.
  • Of its descendants, K191236 (3M Company, 2019) has gone furthest, with 3 descendants of its own.

How does its category compare?

All JOJ devices

JOJ Indicator, Physical/Chemical Sterilization Process: FDA decisions, review times, recalls and adverse event reports over 20 years. 2010, this device’s year, is highlighted.

FDA decisions per year · JOJ

6in 2025

2007: 82008: 92009: 72010: 62011: 132012: 72013: 62014: 62015: 82016: 42017: 72018: 52019: 122020: 72021: 42022: 82023: 42024: 122025: 62026 (partial year): 3
2007high 13 (2011) · low 4 (2016)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JOJ

90in 2025

2007: 140 days median2008: 57 days median2009: 188 days median2010: 115 days median2011: 151 days median2012: 224 days median2013: 212 days median2014: 198 days median2015: 188 days median2016: 150 days median2017: 172 days median2018: 150 days median2019: 38 days median2020: 108 days median2021: 87 days median2022: 90 days median2023: 53 days median2024: 71 days median2025: 90 days median2026 (partial year): 30 days median
2007high 224 (2012) · low 38 (2019)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JOJ

1in 2023

2008: 22009: 22011: 22021: 12023: 1
2007high 2 (2008) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JOJ

12in 2025

2007: 12008: 12009: 92010: 32011: 12012: 82013: 62014: 112015: 72016: 112017: 352018: 122019: 82020: 62021: 292022: 232023: 392024: 252025: 12
2007high 39 (2023) · low 1 (2007)2025

MAUDE reports naming product code JOJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007814001
200895721
2009718829
2010611503
20111315121
2012722408
2013621206
20146198011
2015818807
20164150011
20177172035
20185150012
2019123808
2020710806
2021487129
2022890023
2023453139
20241271025
2025690012
20263300—

What is it indicated for?

The 3M Comply SteriGage 1243RA, 1243RB, 1243RE Chemical Integrators for Steam are designed for pack control monitoring of the following cycles. Cycle Type: Gravity; Temperature: 250.0°F/121°C; Exposure Time: ≥30 minutes Cycle Type: Gravity; Temperature: 270°F/132°C; Exposure Time: ≥3 minutes Cycle Type: Vacuum-assisted (prevacuum); Temperature: 270°F/132°C; Exposure Time: >4 minutes (wrapped), >3 minutes (unwrapped) Cycle Type: Vacuum-assisted…

Read the full statement (112 words)

The 3M Comply SteriGage 1243RA, 1243RB, 1243RE Chemical Integrators for Steam are designed for pack control monitoring of the following cycles.

Cycle Type: Gravity; Temperature: 250.0°F/121°C; Exposure Time: ≥30 minutes

Cycle Type: Gravity; Temperature: 270°F/132°C; Exposure Time: ≥3 minutes

Cycle Type: Vacuum-assisted (prevacuum); Temperature: 270°F/132°C; Exposure Time: >4 minutes (wrapped), >3 minutes (unwrapped)

Cycle Type: Vacuum-assisted (prevacuum); Temperature: 273°F/134°C; Exposure Time: ≥4 minutes (wrapped), ≥3.5 minutes (unwrapped)

Cycle Type: Vacuum-assisted (prevacuum); Temperature: 275°F/135°C; Exposure Time: ≥3 minutes

The Minimum Stated Values for SteriGage as determined using a resistometer are shown below.

Minimum Stated Values for SteriGage

250°F (121°C): 16.5 minutes

270°F (132°C): 2.0 minutes

273°F (134°C): 1.4 minutes

275°F (135°C): 1.1 minute

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K101249: Sterigage Model 1243RA, 1243RB, 1243RE.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K101249

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026