510(k) Report

510(k) · Cleared

Recalled · Class II · 2023

3M Attest Steam Chemical Integrators (1243A, 1243B, 1243RE, 1243RES)

K220942510(k) SpecialCleared Jun 24, 202285-day reviewClass II21 CFR 880.2800Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

3M Company · JOJ Indicator, Physical/Chemical Sterilization Process

Cleared on Jun 24, 2022 as substantially equivalent to K193254 (3M Company, 2019) after an 85-day review. Its family tree reaches back 4 generations to 1977. 1 later device descends from it. It has been recalled 2 times.

Decided
Jun 24, 2022
Review
85 days
Class
Class II
Product code
JOJ
Recalls
2
Predicates
1

Its longest line of descent runs back 4 generations to K771080 (1977). 1 later device descends from it, 1 citing it directly.

Descendants, by year cleared 2026: 1 descendant clearedK771080 · Steam Sterilization Indicator · Info-Chem · 1977K101249 · Sterigage Model 1243RA, 1243RB, 1243RE · 3M Company · 2010×2: 2 generations cleared close together, K191236 (2019), K193254 (2019) 19801990200020102020
  1. Root · 4 generations backK771080 Steam Sterilization IndicatorInfo-Chem · 1977
  2. 3 generations backK101249 Sterigage Model 1243RA, 1243RB, 1243RE3M Company · 2010
  3. 2 generations backK191236 3M™ Attest™ Steam Chemical Integrators3M Company · 2019
  4. 1 generation backK193254 3M Attest Steam Chemical Integrator3M Company · 2019
  5. This deviceK220942 3M Attest Steam Chemical Integrators (1243A, 1243B, 1243RE, 1243RES)3M Company · 2022
  6. Descendant · cites it directly · 0 of its ownK261992 3M™ Attest™ Steam Chemical Integrator (1243A, 1243B, 1243RE, 1243RES)Solventum US, LLC · 2026

Most influential descendants

Showing 1 of 1 · See all
  • K2619923M™ Attest™ Steam Chemical Integrator (1243A, 1243B, 1243RE, 1243RES)Solventum US, LLC · 2026 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K771080Steam Sterilization IndicatorInfo-Chem1977—
−3K101249Sterigage Model 1243RA, 1243RB, 1243RE3M Company2010—
−2K1912363M™ Attest™ Steam Chemical Integrators3M Company2019—
−1K1932543M Attest Steam Chemical Integrator3M Company2019—
0 (this device)K2209423M Attest Steam Chemical Integrators (1243A, 1243B, 1243RE, 1243RES)3M Company2022—
+1K2619923M™ Attest™ Steam Chemical Integrator (1243A, 1243B, 1243RE, 1243RES)Solventum US, LLC2026—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202610
FIG. 1 — Lineage of K220942, 1977–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JOJ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JOJ device

  1. #1 K261992 3M™ Attest™ Steam Chemical Integrator (1243A…score 13.3 · 5 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K1932543M Attest Steam Chemical Integrator
Summary page 6
…ical/chemical sterilization process indicator Device Classification: Class II, 21 CFR § 880.2800(b) Product Code: JOJ 2. Predicate Device: K193254, 3M™ Attest™ Steam Chemical Integrators 3. Description of Device: 3M™ Attest™ Steam Chemical Integrators are chemical indica…
3M Company2019—

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K2619923M™ Attest™ Steam Chemical Integrator (1243A, 1243B, 1243RE, 1243RES)Solventum US, LLC20260

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 6, 2023Class IIZ-1610-2023Process control3M Attest Steam Chemical Integrators with dates of manufacture between November 6, 2022 and January 17, 2023 have an increased potential to leak ink during a sterilization cycle.
Apr 6, 2023Class IIZ-1609-2023Process control3M Attest Steam Chemical Integrators with dates of manufacture between November 6, 2022 and January 17, 2023 have an increased potential to leak ink during a sterilization cycle.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 45 years and passes through 2 companies; a new generation was cleared every 9.2 years on median.
  • None of the 4 ancestors has a recall.
  • Of its descendants, K261992 (Solventum US, LLC, 2026) has gone furthest, with 0 descendants of its own.

How does its category compare?

All JOJ devices

JOJ Indicator, Physical/Chemical Sterilization Process: FDA decisions, review times, recalls and adverse event reports over 20 years. 2022, this device’s year, is highlighted.

FDA decisions per year · JOJ

6in 2025

2007: 82008: 92009: 72010: 62011: 132012: 72013: 62014: 62015: 82016: 42017: 72018: 52019: 122020: 72021: 42022: 82023: 42024: 122025: 62026 (partial year): 3
2007high 13 (2011) · low 4 (2016)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JOJ

90in 2025

2007: 140 days median2008: 57 days median2009: 188 days median2010: 115 days median2011: 151 days median2012: 224 days median2013: 212 days median2014: 198 days median2015: 188 days median2016: 150 days median2017: 172 days median2018: 150 days median2019: 38 days median2020: 108 days median2021: 87 days median2022: 90 days median2023: 53 days median2024: 71 days median2025: 90 days median2026 (partial year): 30 days median
2007high 224 (2012) · low 38 (2019)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JOJ

1in 2023

2008: 22009: 22011: 22021: 12023: 1
2007high 2 (2008) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JOJ

12in 2025

2007: 12008: 12009: 92010: 32011: 12012: 82013: 62014: 112015: 72016: 112017: 352018: 122019: 82020: 62021: 292022: 232023: 392024: 252025: 12
2007high 39 (2023) · low 1 (2007)2025

MAUDE reports naming product code JOJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007814001
200895721
2009718829
2010611503
20111315121
2012722408
2013621206
20146198011
2015818807
20164150011
20177172035
20185150012
2019123808
2020710806
2021487129
2022890023
2023453139
20241271025
2025690012
20263300—

What is it indicated for?

The 3M™ Attest™ Steam Chemical Integrators are designed to respond to all critical parameters over a specified range of steam sterilization cycles. The integrating indicator is intended to be placed in each pack, pouch, container, tray or other containment device to function as an independent monitor of critical parameters for the following sterilization cycles: Cycle Type Temperature Exposure Time Gravity 250°F/121°C 30 minutes Gravity 270°F/132°C 3…

Read the full statement (183 words)

The 3M™ Attest™ Steam Chemical Integrators are designed to respond to all critical parameters over a specified range of steam sterilization cycles. The integrating indicator is intended to be placed in each pack, pouch, container, tray or other containment device to function as an independent monitor of critical parameters for the following sterilization cycles:

Cycle Type Temperature Exposure Time

Gravity 250°F/121°C 30 minutes

Gravity 270°F/132°C 3 minutes

Gravity 270°F/132°C 4 minutes

Gravity 270°F/132°C 10 minutes

Gravity 270°F/132°C 15 minutes

Gravity 270°F/132°C 25 minutes

Gravity 275°F/135°C 3 minutes

Gravity 275°F/135°C 10 minutes

Dynamic Air Removal 250°F/121°C 30 minutes

Dynamic Air Removal 270°F/132°C 4 minutes

Dynamic Air Removal 270°F/132°C 5 minutes

Dynamic Air Removal 270°F/132°C 6 minutes

Dynamic Air Removal 270°F/132°C 7 minutes

Dynamic Air Removal 270°F/132°C 8 minutes

Dynamic Air Removal 270°F/132°C 9 minutes

Dynamic Air Removal 270°F/132°C 10 minutes

Dynamic Air Removal 273°F/134°C 3 minutes

Dynamic Air Removal 273°F/134°C 4 minutes

Dynamic Air Removal 275°F/135°C 3 minutes

Minimum Stated Values for 3M™ Attest™ Steam Chemical Integrators as determined in a resistometer:

250°F/121°C 270°F/132°C 273°F/134°C 275°F/135°C

16.5 Minutes 2.0 Minutes 1.4 Minutes 1.2 Minutes

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K220942: 3M Attest Steam Chemical Integrators (1243A, 1243B, 1243RE, 1243RES).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K220942

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026