510(k) Report

510(k) · Cleared

Recalled · Class I · 2013

Hospira Infusion Blood Sets

K101677510(k) SpecialCleared Aug 11, 201057-day reviewClass II21 CFR 880.5440Recalls 13 (8 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Hospira, Inc. · FPA Set, Administration, Intravascular

Cleared on Aug 11, 2010 as substantially equivalent to K063239 (Hospira, Inc., 2006) and 2 other predicates after a 57-day review. Its family tree reaches back 2 generations to 2003. 5 of the 35 devices in its family, counting itself, have a Class I recall. 30 later devices descend from it. It has been recalled 13 times, including 8 Class I recalls.

Decided
Aug 11, 2010
Review
57 days
Class
Class II
Product code
FPA
Recalls
13 (8 Class I)
Predicates
3

Its longest line of descent runs back 2 generations to K030467 (2003). 4 ancestors in all, the oldest K780880 (1978). 30 later devices descend from it, 8 citing it directly. 5 of the 35 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2003: 1 other ancestor1978: 1 other ancestor2014: 3 descendants cleared2015: 6 descendants cleared, 1 with a Class I recall12012: 1 descendant cleared2016: 4 descendants cleared2017: 3 descendants cleared, 1 with a Class I recall12022: 3 descendants cleared2026: 1 descendant cleared2020: 3 descendants cleared2023: 1 descendant cleared, 1 with a Class I recall12019: 1 descendant cleared, 1 with a Class I recall12021: 2 descendants cleared2025: 2 descendants clearedK030467 · Gravitech TM Flow Controller Primary Set/Gravitech Flow Controller Extension Set · Iv Medical, Inc. · 2003K063239 · Lifeshield Latex-Free Gravitech Flow Controller I.V. Sets, Models 17528, 17527, 17525 & 17526 · Hospira, Inc. · 2006 · recalled19801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors 1 generation forward: 8 descendants, 1 with a Class I recall12 generations forward: 10 descendants, 1 with a Class I recall13 generations forward: 9 descendants, 2 with a Class I recall24 generations forward: 3 descendants K063239 · Lifeshield Latex-Free Gravitech Flow Controller I.V. Sets, Models 17528, 17527, 17525 & 17526 · Hospira, Inc. · 2006 · recalledK030467 · Gravitech TM Flow Controller Primary Set/Gravitech Flow Controller Extension Set · Iv Medical, Inc. · 20030
  1. Root · 2 generations backK030467 Gravitech TM Flow Controller Primary Set/Gravitech Flow Controller…Iv Medical, Inc. · 2003
  2. 1 generation backK063239 Lifeshield Latex-Free Gravitech Flow Controller I.V. Sets, Models…Hospira, Inc. · 2006 · Recalled
  3. This deviceK101677 Hospira Infusion Blood SetsHospira, Inc. · 2010
  4. Descendant · cites it directly · 19 of its ownK113683 Lifeshield Vision Infusion Set with Pre-Pierced ResealHospira, Inc. · 2012
  5. Descendant · cites it directly · 9 of its ownK141789 Plum 360 Infusion System with Hospira Mednet, Smart Card Plug N Play…Hospira, Inc. · 2015 · Recalled (Class I)
  6. Descendant · cites it directly · 6 of its ownK140838 IV Administration Set with Hand PumpB.Braun Medical, Inc. · 2014
  7. Descendant · 2 generations on · 5 of its ownK161469 Plum 360(TM) Infusion System with MedNet/ Smart Card Plug and Play…Hospira, Inc. · 2017 · Recalled (Class I)
  8. Descendant · cites it directly · 4 of its ownK142974 Extension SetsHospira, Inc. · 2015

Showing 5 of 30 descendants. See all descendants

Descendants cleared per year

2012: 12014: 32015: 62016: 42017: 32019: 12020: 32021: 22022: 32023: 12025: 22026: 1
2012high 6 (2015) · low 0 (2013)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 30 · See all
  • K113683Lifeshield Vision Infusion Set with Pre-Pierced ResealHospira, Inc. · 2012 · cites it directly19descendants
  • K141789Plum 360 Infusion System with Hospira Mednet, Smart Card Plug N Play CE 3.0 Module for…Hospira, Inc. · 2015 · cites it directly9descendants
  • K140838IV Administration Set with Hand PumpB.Braun Medical, Inc. · 2014 · cites it directly6descendants
  • K161469Plum 360(TM) Infusion System with MedNet/ Smart Card Plug and Play ModuleHospira, Inc. · 2017 · 2 generations on5descendants
  • K142974Extension SetsHospira, Inc. · 2015 · cites it directly4descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K030467Gravitech TM Flow Controller Primary Set/Gravitech Flow Controller Extension SetIv Medical, Inc.2003—
−1K063239Lifeshield Latex-Free Gravitech Flow Controller I.V. Sets, Models 17528…Hospira, Inc.20067
0 (this device)K101677Hospira Infusion Blood SetsHospira, Inc.2010—
+1K113683Lifeshield Vision Infusion Set with Pre-Pierced ResealHospira, Inc.2012—
+1K141789Plum 360 Infusion System with Hospira Mednet, Smart Card Plug N Play CE 3.0…Hospira, Inc.2015Class I
+1K140838IV Administration Set with Hand PumpB.Braun Medical, Inc.2014—
+2K161469Plum 360(TM) Infusion System with MedNet/ Smart Card Plug and Play ModuleHospira, Inc.2017Class I
+1K142974Extension SetsHospira, Inc.2015—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
197810
200310

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201210
201430
201561
201640
201731
201911
202030
202120
202230
202311
202520
202610
FIG. 1 — Lineage of K101677, 2003–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

5 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FPA device

  1. #1 K243841 Sparta Infusion Set for Insulinscore 13.4 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K063239Lifeshield Latex-Free Gravitech Flow Controller I.V. Sets, Models 17528, 17527, 17525 &…
Summary page 1
…[21 Claimed CFR Lifeshield Latex-Free Veniloop Connector with Prepierced Injection 807.92(a)(3)] Site, Non-DEHP (K063239) Lifeshield Latex-Free Plum Non-Vented Blood set (K030002) HEMA Y…
Hospira, Inc.20067
*K030002Lifeshield Primary IV Set Conv. Pin. Clave; LF LC 5000 Plumset-Dual W/Conv Pin & Cap…
Summary page 1
…ector with Prepierced Injection 807.92(a)(3)] Site, Non-DEHP (K063239) Lifeshield Latex-Free Plum Non-Vented Blood set (K030002) HEMA Y-Type Transfusion Blood Filter Pump Set 100 Inch-SL with P…
Abbott Laboratories200329
*K780880Y-Type Blood Set
Summary page 1
…(K030002) HEMA Y-Type Transfusion Blood Filter Pump Set 100 Inch-SL with Prepierced Reseals (Latex-Free) (K780880) Reason for, 510(k) ~ To offer a non-DEHP PVC alternative Infusion set with a manual submission b…
Abbott Laboratories19784
*—Hospira Administration Blood Sets with Manual Blood Pump
Named without a submission number
—

Also cleared against K063239: 1 other device →

What descends from it? (30)

See all 30

8 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K113683Lifeshield Vision Infusion Set with Pre-Pierced ResealHospira, Inc.201219
K141789Plum 360 Infusion System with Hospira Mednet, Smart Card Plug N Play CE 3.0 Module for… Class IHospira, Inc.20159
K140838IV Administration Set with Hand PumpB.Braun Medical, Inc.20146
K142974Extension SetsHospira, Inc.20154
K142367Hospira Primary SetsHospira, Inc.20144
K142433Hospira Extension SetHospira, Inc.20143
K143087Hospira Blood SetHospira, Inc.20152
K143082IV Administration Sets with 200(micro)m Blood FilterB.Braun Medical, Inc.20151

Has anything in its family been recalled? (5 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 25, 2013Class IIZ-1126-2014Process designSeveral lots of the LifeShield Latex-Free HEMA Y-Type Blood Set IV Administration Set were assembled incorrectly.
Sep 9, 2013Class IIZ-2060-2016Device DesignIntravenous fluid may leak at the CLAVE secondary port of Hospira PlumSet IV administration sets.
Apr 1, 2013Class IZ-1850-2013Device DesignIt was discovered that there was a potential for the piercing pin on certain Hospira blood sets to puncture the outer wall of blood bags during insertion of the pin into the blood bag.
Apr 1, 2013Class IZ-1853-2013Device DesignIt was discovered that there was a potential for the piercing pin on certain Hospira blood sets to puncture the outer wall of blood bags during insertion of the pin into the blood bag.
Apr 1, 2013Class IZ-1854-2013Device DesignIt was discovered that there was a potential for the piercing pin on certain Hospira blood sets to puncture the outer wall of blood bags during insertion of the pin into the blood bag.
Apr 1, 2013Class IZ-1857-2013Device DesignIt was discovered that there was a potential for the piercing pin on certain Hospira blood sets to puncture the outer wall of blood bags during insertion of the pin into the blood bag.
Apr 1, 2013Class IZ-1852-2013Device DesignIt was discovered that there was a potential for the piercing pin on certain Hospira blood sets to puncture the outer wall of blood bags during insertion of the pin into the blood bag.
Apr 1, 2013Class IZ-1858-2013Device DesignIt was discovered that there was a potential for the piercing pin on certain Hospira blood sets to puncture the outer wall of blood bags during insertion of the pin into the blood bag.
Apr 1, 2013Class IZ-1851-2013Device DesignIt was discovered that there was a potential for the piercing pin on certain Hospira blood sets to puncture the outer wall of blood bags during insertion of the pin into the blood bag.
Apr 1, 2013Class IZ-1855-2013Device DesignIt was discovered that there was a potential for the piercing pin on certain Hospira blood sets to puncture the outer wall of blood bags during insertion of the pin into the blood bag.
Jul 13, 2012Class IIZ-2075-2012Employee errorThe float valve in the burette sticks to the burette wall and does not open or close properly.
Jan 12, 2012Not classifiedZ-0817-2012Component design/selectionSilicone applied to the piercing pin of the blood set is reducing retention force of the piercing pin while in use. This could cause the piercing pin to slip out of the administration container.
Nov 10, 2011Not classifiedZ-0292-2012Employee errorThe float valve in the burette sticks to the burette wall and does not open or close properly.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K141789Plum 360 Infusion System with Hospira Mednet, Smart Card Plug N Play CE 3.0 Module for Plum a+ Infusion…Hospira, Inc. · 2015 · Cites it directly · ◆ in the 2015 descendants bar of FIG. 12 Class I recalls
  • K161469Plum 360(TM) Infusion System with MedNet/ Smart Card Plug and Play ModuleHospira, Inc. · 2017 · 2 generations on · ◆ in the 2017 descendants bar of FIG. 12 Class I recalls
  • K223607Plum Duo™ Infusion SystemIcu Medical, Inc. · 2023 · 3 generations on · ◆ in the 2023 descendants bar of FIG. 11 Class I recall
  • K183311Ivenix Infusion System (IIS)Ivenix, Inc. · 2019 · 3 generations on · ◆ in the 2019 descendants bar of FIG. 113 Class I recalls

What does this lineage say?

  • The line of descent spans 7 years and passes through 2 companies; a new generation was cleared every 3.7 years on median.
  • 3 of the 4 ancestors have a recall of any class.
  • Of its descendants, K113683 (Hospira, Inc., 2012) has gone furthest, with 19 descendants of its own.
  • 1 other device also cites its predicate K063239.

How does its category compare?

All FPA devices

FPA Set, Administration, Intravascular: FDA decisions, review times, recalls and adverse event reports over 20 years. 2010, this device’s year, is highlighted.

FDA decisions per year · FPA

11in 2025

2007: 162008: 182009: 102010: 232011: 122012: 212013: 132014: 222015: 172016: 122017: 172018: 42019: 72020: 102021: 132022: 142023: 142024: 72025: 112026 (partial year): 5
2007high 23 (2010) · low 4 (2018)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FPA

225in 2025

2007: 75 days median2008: 89 days median2009: 166 days median2010: 120 days median2011: 87 days median2012: 96 days median2013: 90 days median2014: 89 days median2015: 123 days median2016: 195 days median2017: 240 days median2018: 177 days median2019: 161 days median2020: 327 days median2021: 186 days median2022: 213 days median2023: 340 days median2024: 244 days median2025: 225 days median2026 (partial year): 266 days median
2007high 340 (2023) · low 75 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FPA

98in 2025

2007: 172008: 152009: 1662010: 582011: 282012: 16, 4 Class I2013: 642014: 242015: 62016: 102017: 132018: 42019: 132020: 1072021: 12022: 9, 5 Class I2023: 12, 4 Class I2024: 13, 1 Class I2025: 98, 1 Class I2026 (partial year): 10, 2 Class I
2007high 166 (2009) · low 1 (2021)2026

Class I (17 since 2007) Class II/III not classified

Adverse event reports per year · FPA

34,506in 2025

2007: 1,4992008: 1,9412009: 2,5122010: 2,5372011: 4,4072012: 4,9302013: 4,6512014: 4,6952015: 6,6252016: 8,4702017: 6,4622018: 6,5742019: 11,2272020: 9,5632021: 10,4142022: 13,3932023: 10,9162024: 47,2102025: 34,506
2007high 47,210 (2024) · low 1,499 (2007)2025

MAUDE reports naming product code FPA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071675171,499
20081889151,941
2009101661662,512
201023120582,537
20111287284,407
2012219616 (4)4,930
20131390644,651
20142289244,695
20151712366,625
201612195108,470
201717240136,462
2018417746,574
201971611311,227
2020103271079,563
202113186110,414
2022142139 (5)13,393
20231434012 (4)10,916
2024724413 (1)47,210
20251122598 (1)34,506
2026526610 (2)—

What is it indicated for?

Hospira Administration sets are intended for the delivery of fluids including but not limited to blood and blood products from a container into a patients vascular system.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K101677: Hospira Infusion Blood Sets.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K101677

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026