510(k) Report

510(k) · Cleared

Recalled · Class I · 2026

Ivenix Infusion System (IIS)

K183311510(k) TraditionalCleared Jun 7, 2019190-day reviewClass II21 CFR 880.5725Recalls 25 (13 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ivenix, Inc. · FRN Pump, Infusion

Cleared on Jun 7, 2019 as substantially equivalent to K173084 (Baxter Healthcare Corporation, 2018) and 1 other predicate after a 190-day review. Its family tree reaches back 7 generations to 2001. 15 of the 38 devices in its family, counting itself, have a Class I recall. 1 later device descends from it. It has been recalled 25 times, including 13 Class I recalls.

Decided
Jun 7, 2019
Review
190 days
Class
Class II
Product code
FRN
Recalls
25 (13 Class I)
Predicates
2

Its longest line of descent runs back 7 generations to K011442 (2001). 36 ancestors in all, the oldest K780880 (1978). 1 later device descends from it, 1 citing it directly. 15 of the 38 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2018: 1 other ancestor, 1 with a Class I recall12014: 1 other ancestor, 1 with a Class I recall12012: 1 other ancestor2010: 2 other ancestors, 1 with a Class I recall12006: 3 other ancestors, 1 with a Class I recall12005: 2 other ancestors2004: 2 other ancestors, 2 with a Class I recall21999: 1 other ancestor2003: 3 other ancestors2002: 3 other ancestors2001: 1 other ancestor, 1 with a Class I recall12000: 2 other ancestors, 1 with a Class I recall11995: 1 other ancestor1994: 2 other ancestors1991: 1 other ancestor1981: 1 other ancestor1978: 1 other ancestor1990: 1 other ancestor2021: 1 descendant clearedK011442 · Abbott Plum a+Infusion Pump, Model 11971 · Abbott Laboratories · 2001 · Class I recallK031185 · Abbott Plum a+ Infusion Pump with a HPL/RS Plug-and-Play Module · Abbott Laboratories · 2003 · recalledK042081 · Plum a+ Infusion System with Hospira Mednet Software , and Plum a+3 Infusion System with Hospira Mednet Software · Hospira, Inc. · 2004 · Class I recallK052052 · Hospira Plum a+ Infusion Pump System V11.5, Hospira Plum a+ 3 Infusion Pump System, V11.5 · Hospira, Inc. · 2005 · Class I recallK103224 · Hospira Plum Infusion Set; Hospira Infusion Set with Yellow Stripe Tubing · Hospira, Inc. · 2011 · Class I recallK141789 · Plum 360 Infusion System with Hospira Mednet, Smart Card Plug N Play CE 3.0 Module for Plum a+ Infusion System Upgrade · Hospira, Inc. · 2015 · Class I recallK161469 · Plum 360(TM) Infusion System with MedNet/ Smart Card Plug and Play Module · Hospira, Inc. · 2017 · Class I recall19801990200020102020
  1. Root · 7 generations backK011442 Abbott Plum a+Infusion Pump, Model 11971Abbott Laboratories · 2001 · Recalled (Class I)
  2. 6 generations backK031185 Abbott Plum a+ Infusion Pump with a HPL/RS Plug-and-Play ModuleAbbott Laboratories · 2003 · Recalled
  3. 5 generations backK042081 Plum a+ Infusion System with Hospira Mednet Software , and Plum a+3…Hospira, Inc. · 2004 · Recalled (Class I)
  4. 4 generations backK052052 Hospira Plum a+ Infusion Pump System V11.5, Hospira Plum a+ 3…Hospira, Inc. · 2005 · Recalled (Class I)
  5. 3 generations backK103224 Hospira Plum Infusion Set; Hospira Infusion Set with Yellow Stripe…Hospira, Inc. · 2011 · Recalled (Class I)
  6. 2 generations backK141789 Plum 360 Infusion System with Hospira Mednet, Smart Card Plug N Play…Hospira, Inc. · 2015 · Recalled (Class I)
  7. 1 generation backK161469 Plum 360(TM) Infusion System with MedNet/ Smart Card Plug and Play…Hospira, Inc. · 2017 · Recalled (Class I)
  8. This deviceK183311 Ivenix Infusion System (IIS)Ivenix, Inc. · 2019
  9. Descendant · cites it directly · 0 of its ownK213089 Ivenix Infusion System, LVP Epidural Administraion Set NRFit ConnectorIvenix, Inc. · 2021

Class I recalls in this family

Most influential descendants

Showing 1 of 1 · See all
  • K213089Ivenix Infusion System, LVP Epidural Administraion Set NRFit ConnectorIvenix, Inc. · 2021 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K011442Abbott Plum a+Infusion Pump, Model 11971Abbott Laboratories2001Class I
−6K031185Abbott Plum a+ Infusion Pump with a HPL/RS Plug-and-Play ModuleAbbott Laboratories20034
−5K042081Plum a+ Infusion System with Hospira Mednet Software , and Plum a+3 Infusion…Hospira, Inc.2004Class I
−4K052052Hospira Plum a+ Infusion Pump System V11.5, Hospira Plum a+ 3 Infusion Pump…Hospira, Inc.2005Class I
−3K103224Hospira Plum Infusion Set; Hospira Infusion Set with Yellow Stripe TubingHospira, Inc.2011Class I
−2K141789Plum 360 Infusion System with Hospira Mednet, Smart Card Plug N Play CE 3.0…Hospira, Inc.2015Class I
−1K161469Plum 360(TM) Infusion System with MedNet/ Smart Card Plug and Play ModuleHospira, Inc.2017Class I
0 (this device)K183311Ivenix Infusion System (IIS)Ivenix, Inc.2019—
+1K213089Ivenix Infusion System, LVP Epidural Administraion Set NRFit ConnectorIvenix, Inc.2021—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
197810
198110
199010
199110
199420
199510
199910
200021
200111
200230
200330
200422
200520
200631
201021
201210
201411
201811

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202110
FIG. 1 — Lineage of K183311, 2001–2021Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

15 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRN device

  1. #1 K253703 RemunityPRO+ Pump for Remodulin®…score 12.6 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K173084Spectrum IQ Infusion System, Dose IQ Safety Software
Summary page 4
…: 21 CFR 880.5725 Product Codes: FRN, PHC, FPA Device Class: Class II Predicate Device: K173084, Baxter SIGMA Spectra Infusion Pump K161469, Hospira Inc. Plum 360 Infusion System Device D…
Baxter Healthcare Corporation2018Class I
*K161469Plum 360(TM) Infusion System with MedNet/ Smart Card Plug and Play Module
Summary page 4
…FPA Device Class: Class II Predicate Device: K173084, Baxter SIGMA Spectra Infusion Pump K161469, Hospira Inc. Plum 360 Infusion System Device Description The Ivenix Infusion System (IIS) consists of three primary components:…
Hospira, Inc.2017Class I

Also cleared against K173084: 4 other devices →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K213089Ivenix Infusion System, LVP Epidural Administraion Set NRFit ConnectorIvenix, Inc.20210

Has anything in its family been recalled? (15 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 6, 2026Class IZ-2224-2026Device DesignFresenius Kabi has been made aware via customer complaints that dropping or severe jarring of the Ivenix LVP may cause damage to internal components that may not be apparent or readily visible and the devices therefore remained in…
May 6, 2026Class IZ-2223-2026Software design (manufacturing process)Potential software anomaly that can cause the pump to report a false battery-health value (typically 69%) and trigger a battery-depletion alarm that will shut down the pump unless it is reconnected to AC mains power.
Feb 4, 2026Class IIZ-2274-2026Device DesignPotential for the Cassette Loading Lever to break.
Nov 21, 2025Class IIZ-1048-2026Use errorEmphasizing instructions for LVP duration programming located in the IFU.
Nov 14, 2025Class IZ-0885-2026Software designSoftware version 5.10.1 and earlier contain anomalies that have the potential to cause serious patient harm or death.
Nov 3, 2025Class IZ-0586-2026Process controlIncorrect assembly of specific lot of LVP Primary Administration Set that could result in serious patient injury.
Nov 3, 2025Class IIZ-0848-2026Labeling designDownstream Occlusion alarms may occur during procedures using low flow infusion rates when a check valve* is added to the Ivenix Administration Set.
Sep 3, 2025Class IIZ-0378-2026Process change controlPotential issue that can cause the device to register "phantom" touches in the lower-left corner of the touchscreen during operation.
Jun 24, 2025Class IIZ-2232-2025Process designPotential for external cassette leaks
Jun 24, 2025Class IIZ-2235-2025Process designPotential for external cassette leaks
Jun 24, 2025Class IIZ-2234-2025Process designPotential for external cassette leaks
Jun 24, 2025Class IIZ-2233-2025Process designPotential for external cassette leaks
May 12, 2025Class IZ-1869-2025Under Investigation by firmMis-assembly error of Blood Products Administrations Sets where the 200-micron drip chamber filter may be incorrectly positioned and unable to filter out large blood particulates.
Jan 10, 2025Class IZ-1019-2025Software Design ChangeLarge Volume Pump Software, version 5.9.2 and earlier has potential for the following anomalies: 1. The pump may become nonfunctional if during an alarm condition the Pause Audio option is repeated 70 time or more. 2. If a secondary…
Dec 6, 2024Class IZ-0763-2025Process controlPotential pneumatic valve failure that may cause the LVP to stop an active infusion and/or could prevent the pump from being used.
Aug 30, 2024Class IZ-3153-2024Software designThe software has anomalies that have the potential to cause alarms, nonfunctioning pump, or unresponsive screen while continuing therapy. These could cause serious patient harm or death.
Aug 1, 2024Class IZ-2721-2024Process controlA defect may cause an uncontrolled medication flow condition which may result in a dose greater than prescribed for the patient (i.e., medication overdose).
Mar 7, 2024Class IZ-1484-2024Software Design ChangeSoftware has anomalies that have the potential to cause underdose, overdose, or delay in therapy which could lead to serious patient harm or death.
Nov 29, 2023Class IZ-0549-2024Process change controlThe device may experience mechanical interference on the Fluid Valve pins, which will trigger a Pump Problem alarm. The issue may lead to delay or interruption of therapy. Depending on the therapy and duration of delay/interruption, the…
May 10, 2023Class IIZ-1283-2024Software designRetroactively reported corrections from 2023: 1) A software defect may cause an incorrect (Fail-Stop) alarm when an administration set is loaded or coupled while the pump is executing the power-up sequence. May lead to delay in therapy.…
Mar 10, 2023Class IZ-1313-2023Process controlFluid ingress that can cause a loss of electrical funtion and failure of the Set ID Sensory, resulting in the prevention of infusion or halt of infusion.
Sep 16, 2022Class IIZ-1282-2024Software designRetroactively reported correction from 2022: A software defect may cause an incorrect (Fail-Stop) Alarm when an administration set is loaded with an upstream occlusion present, resulting in a non-recoverable (Fail-Stop) Pump Problem…
Sep 8, 2022Class IZ-0002-2023Software designThe display screen may become frozen and unresponsive to user input, triggering an alarm that is evidenced by a buzzer sound and flashing red LEDs as a secondary notification. However, as the screen goes blank it does not indicate an…
Apr 22, 2022Class IIZ-2635-2023Device DesignStart-up time for the flow rate range greater than 50 ml/hr, and less than 200 ml/hr may exceed the specified 6 minutes, depending on the characteristics of air bubble entrapment within the Intermediate Pumping Chamber (IPC) of the…
Apr 22, 2022Class IIZ-1381-2022Software designDownstream occlusion alarm is sometimes immediately followed by a pump problem alarm.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K173084Spectrum IQ Infusion System, Dose IQ Safety SoftwareBaxter Healthcare Corporation · 2018 · Direct predicate · ◆ in the 2018 ancestors bar of FIG. 14 Class I recalls
  • K161469Plum 360(TM) Infusion System with MedNet/ Smart Card Plug and Play ModuleHospira, Inc. · 2017 · Direct predicate · ◆ on the line of descent in FIG. 12 Class I recalls
  • K141789Plum 360 Infusion System with Hospira Mednet, Smart Card Plug N Play CE 3.0 Module for Plum a+ Infusion…Hospira, Inc. · 2015 · 2 generations back · ◆ on the line of descent in FIG. 12 Class I recalls
  • K133801Baxter Sigma Spectrum Infusion Pump with Master Drug LibraryBaxter Healthcare Corporation · 2014 · 2 generations back · ◆ in the 2014 ancestors bar of FIG. 15 Class I recalls
  • K103224Hospira Plum Infusion Set; Hospira Infusion Set with Yellow Stripe TubingHospira, Inc. · 2011 · 3 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K101677Hospira Infusion Blood SetsHospira, Inc. · 2010 · 3 generations back · ◆ in the 2010 ancestors bar of FIG. 18 Class I recalls
  • K042121Spectrum and Spectrum with Master Drug LibrarySigma Intl. · 2004 · 3 generations back · ◆ in the 2004 ancestors bar of FIG. 13 Class I recalls
  • K042081Plum a+ Infusion System with Hospira Mednet Software , and Plum a+3 Infusion System with Hospira Mednet…Hospira, Inc. · 2004 · 3 generations back · ◆ on the line of descent in FIG. 14 Class I recalls
  • K060806Hospira Gemstar Infusion Pump System, 7 Therapy, Model 13087; 6 Therapy, Model 13086; Pain Management, Model…Hospira, Inc. · 2006 · 4 generations back · ◆ in the 2006 ancestors bar of FIG. 17 Class I recalls
  • K052052Hospira Plum a+ Infusion Pump System V11.5, Hospira Plum a+ 3 Infusion Pump System, V11.5Hospira, Inc. · 2005 · 4 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K041550Hospira Phoenix Infusion System with Hospira Mednet SoftwareAbbott Laboratories · 2004 · 4 generations back · ◆ in the 2004 ancestors bar of FIG. 12 Class I recalls
  • K011975Horizon Outlook with DosecomB.Braun Medical, Inc. · 2001 · 4 generations back · ◆ in the 2001 ancestors bar of FIG. 11 Class I recall

Nearest 12 of 14 shown.

What does this lineage say?

  • The line of descent spans 18 years and passes through 3 companies; a new generation was cleared every 2.2 years on median.
  • 27 of the 36 ancestors have a recall of any class; 14 have a Class I recall.
  • Of its descendants, K213089 (Ivenix, Inc., 2021) has gone furthest, with 0 descendants of its own.
  • 4 other devices also cite its predicate K173084.

How does its category compare?

All FRN devices

FRN Pump, Infusion: FDA decisions, review times, recalls and adverse event reports over 20 years. 2019, this device’s year, is highlighted.

FDA decisions per year · FRN

8in 2025

2007: 122008: 82009: 92010: 32011: 32012: 42013: 62014: 62015: 32016: 32017: 32018: 32019: 22020: 12021: 22022: 62023: 72024: 52025: 82026 (partial year): 2
2007high 12 (2007) · low 1 (2020)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRN

172in 2025

2007: 102 days median2008: 89 days median2009: 66 days median2010: 93 days median2011: 103 days median2012: 217 days median2013: 312 days median2014: 172 days median2015: 232 days median2016: 301 days median2017: 305 days median2018: 267 days median2022: 268 days median2023: 265 days median2024: 103 days median2025: 172 days median
2007high 312 (2013) · low 66 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRN

7in 2025

2007: 152008: 242009: 1022010: 372011: 692012: 42, 2 Class I2013: 22, 3 Class I2014: 4, 3 Class I2015: 52016: 7, 1 Class I2017: 132018: 12019: 18, 1 Class I2020: 16, 5 Class I2021: 3, 2 Class I2022: 22023: 16, 10 Class I2024: 9, 4 Class I2025: 7, 4 Class I2026 (partial year): 11, 2 Class I
2007high 102 (2009) · low 1 (2018)2026

Class I (37 since 2007) Class II/III not classified

Adverse event reports per year · FRN

52,792in 2025

2007: 1,0862008: 8,2612009: 1,9522010: 6,1332011: 41,3672012: 15,4822013: 40,0822014: 94,1802015: 26,7422016: 16,5602017: 12,0602018: 16,5912019: 15,8502020: 94,1862021: 199,3422022: 888,0202023: 165,8382024: 57,7302025: 52,792
2007high 888,020 (2022) · low 1,086 (2007)2025

MAUDE reports naming product code FRN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200712102151,086
2008889248,261
20099661021,952
2010393376,133
201131036941,367
2012421742 (2)15,482
2013631222 (3)40,082
201461724 (3)94,180
20153232526,742
201633017 (1)16,560
201733051312,060
20183267116,591
20192—18 (1)15,850
20201—16 (5)94,186
20212—3 (2)199,342
202262682888,020
2023726516 (10)165,838
202451039 (4)57,730
202581727 (4)52,792
20262—11 (2)—

What is it indicated for?

The Ivenix Infusion System is indicated for use in a hospital and in outpatient care environments for the controlled administration of fluids through clinically accepted routes of administration: intravenous, intra-arterial, epidural, and subcutaneous, to adults, pediatric and neonate patients. Administered fluids may be pharmaceutical drugs, red blood cells, platelets, plasma, and other mixtures required for patient therapy. The Ivenix Large Volume Pump…

Read the full statement (176 words)

The Ivenix Infusion System is indicated for use in a hospital and in outpatient care environments for the controlled administration of fluids through clinically accepted routes of administration: intravenous, intra-arterial, epidural, and subcutaneous, to adults, pediatric and neonate patients. Administered fluids may be pharmaceutical drugs, red blood cells, platelets, plasma, and other mixtures required for patient therapy.

The Ivenix Large Volume Pump (LVP) is indicated for use only with the Ivenix sterile, single use, disposable administration sets, including:

Primary dual inlet, single outlet, 1 Y-site

Primary single inlet, single outlet

Blood Set, dual inlet, single outlet, 1 Y-site, for administration of red blood cells, platelets and plasma

Microbore single inlet, single outlet, for administration of epidural fluids

Microbore dual inlet, single outlet, 1 Y-site

Infusion Management System (IMS):

The IMS provides information to the clinician regarding the use of the Ivenix Large Volume Pump (LVP) by way of a drug library, LVP configurations, and by providing remote information regarding LVP status. It also provides information on pump usage data reports to various functions within an institution.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K183311: Ivenix Infusion System (IIS).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K183311

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026