510(k) Report

510(k) · Cleared

Recalled · Class I · 2026

Plum Duo™ Infusion System

K223607510(k) TraditionalCleared Aug 24, 2023265-day reviewClass II21 CFR 880.5725Recalls 6 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Icu Medical, Inc. · FRN Pump, Infusion

Cleared on Aug 24, 2023 as substantially equivalent to K161469 (Hospira, Inc., 2017) after a 265-day review. Its family tree reaches back 7 generations to 2001. 10 of the 29 devices in its family, counting itself, have a Class I recall. 2 later devices descend from it. It has been recalled 6 times, including a Class I recall.

Decided
Aug 24, 2023
Review
265 days
Class
Class II
Product code
FRN
Recalls
6 (1 Class I)
Predicates
1

Its longest line of descent runs back 7 generations to K011442 (2001). 26 ancestors in all, the oldest K780880 (1978). 2 later devices descend from it, 2 citing it directly. 10 of the 29 devices in this family, counting this one, have a Class I recall.

  1. Root · 7 generations backK011442 Abbott Plum a+Infusion Pump, Model 11971Abbott Laboratories · 2001 · Recalled (Class I)
  2. 6 generations backK031185 Abbott Plum a+ Infusion Pump with a HPL/RS Plug-and-Play ModuleAbbott Laboratories · 2003 · Recalled
  3. 5 generations backK042081 Plum a+ Infusion System with Hospira Mednet Software , and Plum a+3…Hospira, Inc. · 2004 · Recalled (Class I)
  4. 4 generations backK052052 Hospira Plum a+ Infusion Pump System V11.5, Hospira Plum a+ 3…Hospira, Inc. · 2005 · Recalled (Class I)
  5. 3 generations backK103224 Hospira Plum Infusion Set; Hospira Infusion Set with Yellow Stripe…Hospira, Inc. · 2011 · Recalled (Class I)
  6. 2 generations backK141789 Plum 360 Infusion System with Hospira Mednet, Smart Card Plug N Play…Hospira, Inc. · 2015 · Recalled (Class I)
  7. 1 generation backK161469 Plum 360(TM) Infusion System with MedNet/ Smart Card Plug and Play…Hospira, Inc. · 2017 · Recalled (Class I)
  8. This deviceK223607 Plum Duo™ Infusion SystemIcu Medical, Inc. · 2023
  9. Descendant · cites it directly · 0 of its ownK242114 Plum Solo™ Precision IV PumpIcu Medical, Inc. · 2025 · Recalled
  10. Descendant · cites it directly · 0 of its ownK242115 Plum Duo™ Precision IV PumpIcu Medical, Inc. · 2025 · Recalled

Class I recalls in this family

Most influential descendants

Showing 2 of 2 · See all
  • K242114Plum Solo™ Precision IV PumpIcu Medical, Inc. · 2025 · cites it directly0descendants
  • K242115Plum Duo™ Precision IV PumpIcu Medical, Inc. · 2025 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K011442Abbott Plum a+Infusion Pump, Model 11971Abbott Laboratories2001Class I
−6K031185Abbott Plum a+ Infusion Pump with a HPL/RS Plug-and-Play ModuleAbbott Laboratories20034
−5K042081Plum a+ Infusion System with Hospira Mednet Software , and Plum a+3 Infusion…Hospira, Inc.2004Class I
−4K052052Hospira Plum a+ Infusion Pump System V11.5, Hospira Plum a+ 3 Infusion Pump…Hospira, Inc.2005Class I
−3K103224Hospira Plum Infusion Set; Hospira Infusion Set with Yellow Stripe TubingHospira, Inc.2011Class I
−2K141789Plum 360 Infusion System with Hospira Mednet, Smart Card Plug N Play CE 3.0…Hospira, Inc.2015Class I
−1K161469Plum 360(TM) Infusion System with MedNet/ Smart Card Plug and Play ModuleHospira, Inc.2017Class I
0 (this device)K223607Plum Duo™ Infusion SystemIcu Medical, Inc.2023—
+1K242114Plum Solo™ Precision IV PumpIcu Medical, Inc.20253
+1K242115Plum Duo™ Precision IV PumpIcu Medical, Inc.20252

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
197810
198110
199110
199410
199910
200230
200320
200411
200520
200631
201021
201210

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202520
FIG. 1 — Lineage of K223607, 2001–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

10 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRN device

  1. #1 K253703 RemunityPRO+ Pump for Remodulin®…score 12.6 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K161469Plum 360(TM) Infusion System with MedNet/ Smart Card Plug and Play Module
Summary page 4
…eneration of the Plum™ family of devices that is based on the fundamental technology of the Plum 360™ Infusion System cleared under K161469. The pump design incorporates state-of-the-art features such as dual channel functionality,…
Hospira, Inc.2017Class I

Also cleared against K161469: 1 other device →

What descends from it? (2)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K242114Plum Solo™ Precision IV PumpIcu Medical, Inc.20250
K242115Plum Duo™ Precision IV PumpIcu Medical, Inc.20250

Has anything in its family been recalled? (10 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 10, 2026Class IIZ-2723-2026Under Investigation by firmIn a specific scenario, the pump will not decrement the Volume to be Infused (VTBI) displayed on the Delivery screen and may deliver up to twice the VTBI at the programmed rate.
Apr 29, 2026Class IIZ-2312-2026Software Design ChangeUnder certain conditions, the pump may not display a soft limit alert, and will not notify the clinician of a soft limit violation before confirming and starting delivery. This may potentially lead to under- or over-delivery.
Dec 15, 2025Class IIZ-1118-2026Software designPlum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy. The programmed flush volume is delivered from the primary line container at the piggyback therapy…
Dec 15, 2025Class IIZ-1094-2026Nonconforming Material/ComponentICU Medical received one lot of defective primary speakers from our supplier that may fail to produce an audible sound. If the primary speaker in your Plum Duo Infusion Pump fails, the pump may not produce audible alarms, alerts, or…
Jul 11, 2025Class IZ-2129-2025Under Investigation by firmICU Medical identified two sequences of programming events and alarm interactions that may cause the user interface to become unresponsive.
Apr 22, 2025Class IIZ-1990-2025Under Investigation by firmInternal testing found that Plum Duo pumps with software version 1.1.1 have a workflow that may result in bypassing the Maximum Dose Limit alert. The Maximum Dose Limit is set in the LifeShield Drug Library Manager (DLM) within the…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K161469Plum 360(TM) Infusion System with MedNet/ Smart Card Plug and Play ModuleHospira, Inc. · 2017 · Direct predicate · ◆ on the line of descent in FIG. 12 Class I recalls
  • K141789Plum 360 Infusion System with Hospira Mednet, Smart Card Plug N Play CE 3.0 Module for Plum a+ Infusion…Hospira, Inc. · 2015 · 2 generations back · ◆ on the line of descent in FIG. 12 Class I recalls
  • K103224Hospira Plum Infusion Set; Hospira Infusion Set with Yellow Stripe TubingHospira, Inc. · 2011 · 3 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K101677Hospira Infusion Blood SetsHospira, Inc. · 2010 · 3 generations back · ◆ in the 2010 ancestors bar of FIG. 18 Class I recalls
  • K042081Plum a+ Infusion System with Hospira Mednet Software , and Plum a+3 Infusion System with Hospira Mednet…Hospira, Inc. · 2004 · 3 generations back · ◆ on the line of descent in FIG. 14 Class I recalls
  • K060806Hospira Gemstar Infusion Pump System, 7 Therapy, Model 13087; 6 Therapy, Model 13086; Pain Management, Model…Hospira, Inc. · 2006 · 4 generations back · ◆ in the 2006 ancestors bar of FIG. 17 Class I recalls
  • K052052Hospira Plum a+ Infusion Pump System V11.5, Hospira Plum a+ 3 Infusion Pump System, V11.5Hospira, Inc. · 2005 · 4 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K041550Hospira Phoenix Infusion System with Hospira Mednet SoftwareAbbott Laboratories · 2004 · 4 generations back · ◆ in the 2004 ancestors bar of FIG. 12 Class I recalls
  • K011442Abbott Plum a+Infusion Pump, Model 11971Abbott Laboratories · 2001 · 5 generations back · ◆ on the line of descent in FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 22 years and passes through 3 companies; a new generation was cleared every 2.2 years on median.
  • 20 of the 26 ancestors have a recall of any class; 9 have a Class I recall.
  • Of its descendants, K242114 (Icu Medical, Inc., 2025) has gone furthest, with 0 descendants of its own.
  • 1 other device also cites its predicate K161469.

How does its category compare?

All FRN devices

FRN Pump, Infusion: FDA decisions, review times, recalls and adverse event reports over 20 years. 2023, this device’s year, is highlighted.

FDA decisions per year · FRN

8in 2025

2007: 122008: 82009: 92010: 32011: 32012: 42013: 62014: 62015: 32016: 32017: 32018: 32019: 22020: 12021: 22022: 62023: 72024: 52025: 82026 (partial year): 2
2007high 12 (2007) · low 1 (2020)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRN

172in 2025

2007: 102 days median2008: 89 days median2009: 66 days median2010: 93 days median2011: 103 days median2012: 217 days median2013: 312 days median2014: 172 days median2015: 232 days median2016: 301 days median2017: 305 days median2018: 267 days median2022: 268 days median2023: 265 days median2024: 103 days median2025: 172 days median
2007high 312 (2013) · low 66 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRN

7in 2025

2007: 152008: 242009: 1022010: 372011: 692012: 42, 2 Class I2013: 22, 3 Class I2014: 4, 3 Class I2015: 52016: 7, 1 Class I2017: 132018: 12019: 18, 1 Class I2020: 16, 5 Class I2021: 3, 2 Class I2022: 22023: 16, 10 Class I2024: 9, 4 Class I2025: 7, 4 Class I2026 (partial year): 11, 2 Class I
2007high 102 (2009) · low 1 (2018)2026

Class I (37 since 2007) Class II/III not classified

Adverse event reports per year · FRN

52,792in 2025

2007: 1,0862008: 8,2612009: 1,9522010: 6,1332011: 41,3672012: 15,4822013: 40,0822014: 94,1802015: 26,7422016: 16,5602017: 12,0602018: 16,5912019: 15,8502020: 94,1862021: 199,3422022: 888,0202023: 165,8382024: 57,7302025: 52,792
2007high 888,020 (2022) · low 1,086 (2007)2025

MAUDE reports naming product code FRN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200712102151,086
2008889248,261
20099661021,952
2010393376,133
201131036941,367
2012421742 (2)15,482
2013631222 (3)40,082
201461724 (3)94,180
20153232526,742
201633017 (1)16,560
201733051312,060
20183267116,591
20192—18 (1)15,850
20201—16 (5)94,186
20212—3 (2)199,342
202262682888,020
2023726516 (10)165,838
202451039 (4)57,730
202581727 (4)52,792
20262—11 (2)—

What is it indicated for?

The Plum Duo Infusion System is intended for parenteral fluids and medications through clinically acceptable routes (limited to intravenous, intra-arterial, and subcutaneous therapies).

The Plum Duo Infusion System is intended for use in clinical environments in the hospital environment and other outpatient healthcare facilities by licensed healthcare professionals. These healthcare professionals are trained in the use of the infusion pump and the administration of therapies consistent with the intended use.

The Plum Duo Infusion System is intended for adult, pediatric (including infants and children), and neonatal patient populations.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K223607: Plum Duo™ Infusion System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K223607

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.