510(k) Report

510(k) · Cleared

Recalled · Class II · 2025

Extractor Pro RX Retrieval Ballons, Extractor Pro XL Retrieval Ballons, Extractor Pro DL Retrieval Ballons Model M005470

K102082510(k) SpecialCleared Aug 23, 201028-day reviewClass II21 CFR 876.5010Recalls 5

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Boston Scientific Corp · FGE Stents, Drains And Dilators For The Biliary Ducts

Cleared on Aug 23, 2010 as substantially equivalent to K041606 (Boston Scientific Corp, 2004) and 1 other predicate after a 28-day review. Its family tree reaches back 1 generation to 1993. 3 later devices descend from it. It has been recalled 5 times.

Decided
Aug 23, 2010
Review
28 days
Class
Class II
Product code
FGE
Recalls
5
Predicates
2

Its longest line of descent runs back 1 generation to K931619 (1993). 2 ancestors in all. 3 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2004: 1 other ancestor2022: 1 descendant cleared2013: 1 descendant cleared2026: 1 descendant clearedK931619 · Microvasive Extractor, Retrieval Balloon Catheter · Boston Scientific Corp · 19931990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor 1 generation forward: 2 descendants2 generations forward: 1 descendant K931619 · Microvasive Extractor, Retrieval Balloon Catheter · Boston Scientific Corp · 19930
  1. Root · 1 generation backK931619 Microvasive Extractor, Retrieval Balloon CatheterBoston Scientific Corp · 1993
  2. This deviceK102082 Extractor Pro RX Retrieval Ballons, Extractor Pro XL Retrieval…Boston Scientific Corp · 2010
  3. Descendant · cites it directly · 1 of its ownK211021 Sterile Biliary Stone Retrieval Balloon Catheter, Retrieval Balloon…Micro-Tech (Nanjing) Co., Ltd. · 2022
  4. Descendant · 2 generations on · 0 of its ownK253132 Single use stone retrieval balloonsZhejiang Chuangxiang Medical Technology Co., Ltd. · 2026
  5. Descendant · cites it directly · 0 of its ownK131642 RSH Biliary Stone Removal DeviceRsh, LLC · 2013

Most influential descendants

Showing 3 of 3 · See all
  • K211021Sterile Biliary Stone Retrieval Balloon Catheter, Retrieval Balloon / short-wire…Micro-Tech (Nanjing) Co., Ltd. · 2022 · cites it directly1descendant
  • K253132Single use stone retrieval balloonsZhejiang Chuangxiang Medical Technology Co., Ltd. · 2026 · 2 generations on0descendants
  • K131642RSH Biliary Stone Removal DeviceRsh, LLC · 2013 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K931619Microvasive Extractor, Retrieval Balloon CatheterBoston Scientific Corp1993—
0 (this device)K102082Extractor Pro RX Retrieval Ballons, Extractor Pro XL Retrieval Ballons…Boston Scientific Corp2010—
+1K211021Sterile Biliary Stone Retrieval Balloon Catheter, Retrieval Balloon /…Micro-Tech (Nanjing) Co., Ltd.2022—
+2K253132Single use stone retrieval balloonsZhejiang Chuangxiang Medical Technology Co., Ltd.2026—
+1K131642RSH Biliary Stone Removal DeviceRsh, LLC2013—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200410

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201310
202210
202610
FIG. 1 — Lineage of K102082, 1993–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FGE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FGE device

  1. #1 K240464 WallFlex Biliary RX Stent System; Epic Biliary…score 14.5 · 13 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K041606Extractor RX Retrieval Balloon
Summary page 1
…ssification: Class H1 3. Predicate Device: Trade Name: ExtractorTm 1RX Retrieval Balloon Manufacturer and Clearance Number: Boston Scientific Corporation, K04 1606 Classification Name: Endoscopic Biliary Stone Retrieval Balloon Catheter Regulation Number: 876.50 10 Produc…
Boston Scientific Corp2004—
*K931619Microvasive Extractor, Retrieval Balloon Catheter
Summary page 1
…E Classification: Class H Trade Name: Endoscopic Biliary Stone Retrieval Catheter Manufacturer and Clearance Number: Boston Scientific Corporation, K93 1619 Classification Name: Catheter, Biliary, Surgical Regulation Number: 876.5010 Product Code: GCA Classi…
Boston Scientific Corp1993—

Also cleared against K041606: 1 other device →

What descends from it? (3)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K211021Sterile Biliary Stone Retrieval Balloon Catheter, Retrieval Balloon / short-wire…Micro-Tech (Nanjing) Co., Ltd.20221
K131642RSH Biliary Stone Removal DeviceRsh, LLC20130

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 28, 2025Class IIZ-2618-2025Process controlThe product in incorrectly labeled. The label indicates that the skive hole should be above the balloon, while it is actually positioned below, and vice versa.
Apr 27, 2015Class IIZ-1941-2015Employee errorOne lot of the Extractor" Pro RX Retrieval Balloon Catheter (9-12mm balloon) contains an incorrect syringe size (12-15mm), which is larger than the size identified on the label.
Dec 18, 2013Class IIZ-0710-2014Process controlExtractor Pro XL Retrieval Balloon Catheters may have the incorrect skive hole location or balloon size
Dec 18, 2013Class IIZ-0711-2014Process controlExtractor Pro XL Retrieval Balloon Catheters may have the incorrect skive hole location or balloon size
Dec 18, 2013Class IIZ-0712-2014Process controlExtractor Pro XL Retrieval Balloon Catheters may have the incorrect skive hole location or balloon size

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • None of the 2 ancestors has a recall.
  • Of its descendants, K211021 (Micro-Tech (Nanjing) Co., Ltd., 2022) has gone furthest, with 1 descendant of its own.
  • 1 other device also cites its predicate K041606.

How does its category compare?

All FGE devices

FGE Stents, Drains And Dilators For The Biliary Ducts: FDA decisions, review times, recalls and adverse event reports over 20 years. 2010, this device’s year, is highlighted.

FDA decisions per year · FGE

7in 2025

2007: 42008: 112009: 72010: 92011: 92012: 82013: 52014: 82015: 82016: 42017: 92018: 72019: 22020: 42021: 12022: 62023: 62024: 62025: 72026 (partial year): 8
2007high 11 (2008) · low 1 (2021)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FGE

257in 2025

2007: 316 days median2008: 214 days median2009: 160 days median2010: 81 days median2011: 114 days median2012: 96 days median2013: 161 days median2014: 156 days median2015: 79 days median2016: 49 days median2017: 59 days median2018: 215 days median2020: 121 days median2022: 101 days median2023: 199 days median2024: 188 days median2025: 257 days median2026 (partial year): 153 days median
2007high 316 (2007) · low 49 (2016)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FGE

1in 2024

2007: 542008: 932009: 182010: 322011: 22012: 42013: 32017: 12018: 432019: 132020: 12021: 272022: 212023: 192024: 12026 (partial year): 11
2007high 93 (2008) · low 0 (2014)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FGE

1,923in 2025

2007: 1,0062008: 1,4652009: 1,6552010: 1,4342011: 1,3212012: 1,1832013: 1,0602014: 1,2702015: 1,5432016: 1,4532017: 1,5152018: 1,6812019: 1,8962020: 2,1822021: 2,2972022: 1,7492023: 1,6152024: 1,9482025: 1,923
2007high 2,297 (2021) · low 1,006 (2007)2025

MAUDE reports naming product code FGE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20074316541,006
200811214931,465
20097160181,655
2010981321,434
2011911421,321
201289641,183
2013516131,060
2014815601,270
201587901,543
201644901,453
201795911,515
20187215431,681
20192—131,896
2020412112,182
20211—272,297
20226101211,749
20236199191,615
2024618811,948
2025725701,923
2026815311—

What is it indicated for?

The Extractor Pro Retrieval Balloon Catheters are indicated for use endoscopically to remove stones from the biliary system, or to facilitate injection of contrast medium while occluding the duct with the balloon.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K102082: Extractor Pro RX Retrieval Ballons, Extractor Pro XL Retrieval Ballons, Extractor Pro DL Retrieval Ballons Model M005470.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K102082

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026