510(k) Report

510(k) · Cleared

Cleared

Single use stone retrieval balloons

K253132510(k) TraditionalCleared Mar 6, 2026162-day reviewClass II21 CFR 876.5010

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Zhejiang Chuangxiang Medical Technology Co., Ltd. · FGE Stents, Drains And Dilators For The Biliary Ducts

Cleared on Mar 6, 2026 as substantially equivalent to K211021 (Micro-Tech (Nanjing) Co., Ltd., 2022) after a 162-day review. Its family tree reaches back 3 generations to 1993.

Decided
Mar 6, 2026
Review
162 days
Class
Class II
Product code
FGE
Recalls
None
Predicates
1

Its longest line of descent runs back 3 generations to K931619 (1993). 4 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 3 generations backK931619 Microvasive Extractor, Retrieval Balloon CatheterBoston Scientific Corp · 1993
  2. 2 generations backK102082 Extractor Pro RX Retrieval Ballons, Extractor Pro XL Retrieval…Boston Scientific Corp · 2010 · Recalled
  3. 1 generation backK211021 Sterile Biliary Stone Retrieval Balloon Catheter, Retrieval Balloon…Micro-Tech (Nanjing) Co., Ltd. · 2022
  4. This deviceK253132 Single use stone retrieval balloonsZhejiang Chuangxiang Medical Technology Co., Ltd. · 2026
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K931619Microvasive Extractor, Retrieval Balloon CatheterBoston Scientific Corp1993—
−2K102082Extractor Pro RX Retrieval Ballons, Extractor Pro XL Retrieval Ballons…Boston Scientific Corp20105
−1K211021Sterile Biliary Stone Retrieval Balloon Catheter, Retrieval Balloon /…Micro-Tech (Nanjing) Co., Ltd.2022—
0 (this device)K253132Single use stone retrieval balloonsZhejiang Chuangxiang Medical Technology Co., Ltd.2026—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200410
FIG. 1 — Lineage of K253132, 1993–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FGE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FGE device

  1. #1 K240464 WallFlex Biliary RX Stent System; Epic Biliary…score 14.5 · 13 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K211021Sterile Biliary Stone Retrieval Balloon Catheter, Retrieval Balloon / short-wire…
Summary page 5
…lation class: II Classification name: Biliary catheter and accessories Regulation number:876.5010 Product code: FGE 4. Predicative device identification 510(k) Number: K211021 Device Name: Sterile Biliary Stone Retrieval Balloon Catheter, Retrieval Balloon / short-wire compatible Product Code: FGE 5. De…
Micro-Tech (Nanjing) Co., Ltd.2022—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 33 years and passes through 3 companies; a new generation was cleared every 11.7 years on median.
  • 1 of the 4 ancestors has a recall of any class.

How does its category compare?

All FGE devices

FGE Stents, Drains And Dilators For The Biliary Ducts: FDA decisions, review times, recalls and adverse event reports over 20 years. 2026, this device’s year, is highlighted.

FDA decisions per year · FGE

7in 2025

2007: 42008: 112009: 72010: 92011: 92012: 82013: 52014: 82015: 82016: 42017: 92018: 72019: 22020: 42021: 12022: 62023: 62024: 62025: 72026 (partial year): 8
2007high 11 (2008) · low 1 (2021)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FGE

257in 2025

2007: 316 days median2008: 214 days median2009: 160 days median2010: 81 days median2011: 114 days median2012: 96 days median2013: 161 days median2014: 156 days median2015: 79 days median2016: 49 days median2017: 59 days median2018: 215 days median2020: 121 days median2022: 101 days median2023: 199 days median2024: 188 days median2025: 257 days median2026 (partial year): 153 days median
2007high 316 (2007) · low 49 (2016)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FGE

1in 2024

2007: 542008: 932009: 182010: 322011: 22012: 42013: 32017: 12018: 432019: 132020: 12021: 272022: 212023: 192024: 12026 (partial year): 11
2007high 93 (2008) · low 0 (2014)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FGE

1,923in 2025

2007: 1,0062008: 1,4652009: 1,6552010: 1,4342011: 1,3212012: 1,1832013: 1,0602014: 1,2702015: 1,5432016: 1,4532017: 1,5152018: 1,6812019: 1,8962020: 2,1822021: 2,2972022: 1,7492023: 1,6152024: 1,9482025: 1,923
2007high 2,297 (2021) · low 1,006 (2007)2025

MAUDE reports naming product code FGE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20074316541,006
200811214931,465
20097160181,655
2010981321,434
2011911421,321
201289641,183
2013516131,060
2014815601,270
201587901,543
201644901,453
201795911,515
20187215431,681
20192—131,896
2020412112,182
20211—272,297
20226101211,749
20236199191,615
2024618811,948
2025725701,923
2026815311—

What is it indicated for?

The Retrieval Balloon is indicated for use endoscopically to remove stones from biliary system, or to

facilitate injection of contrast medium while occluding the duct with the balloon. The device is for adult use

only.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K253132: Single use stone retrieval balloons.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K253132

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026