510(k) Report

510(k) · Cleared

Recalled · Class II · 2023

Cardioquip Modular Cooler-Heater Model MCH-1000

K102147510(k) TraditionalCleared Nov 19, 2010112-day reviewClass II21 CFR 870.4250Recalls 9

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Cardioquip Llp · DWC Controller, Temperature, Cardiopulmonary Bypass

Cleared on Nov 19, 2010 after a 112-day review. The FDA summary names no predicate by number. It has been recalled 9 times.

Decided
Nov 19, 2010
Review
112 days
Class
Class II
Product code
DWC
Recalls
9
Predicates
0

The FDA summary names no predicate by number.

FIG. 1 — Lineage of K102147Source: FDA 510(k) database · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DWC? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DWC device

  1. #1 K220635 Heater-Cooler System 3Tscore 5.3 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

The FDA summary names no predicate by number.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 24, 2023Class IIZ-0565-2024Process controlThere is the potential that cooler-heater devices may contain bacterial contamination which could result in patient infection.
Oct 24, 2023Class IIZ-0566-2024Process controlThere is the potential that cooler-heater devices may contain bacterial contamination which could result in patient infection.
Oct 19, 2023Class IIZ-0438-2024Device DesignThere is a potential for the heating-chamber to malfunction resulting in melting/burning of the heater chamber components.
Oct 19, 2023Class IIZ-0439-2024Device DesignThere is a potential for the heating-chamber to malfunction resulting in melting/burning of the heater chamber components.
Jun 24, 2021Class IIZ-0069-2022Under Investigation by firmThere is the potential bacterial including Nontuberculous mycobacterial (NTM) to grow in the water system of heater-cooler devices which could result in contaminated water potentially aerosolized into the operating room during surgery…
May 12, 2021Class IIZ-0036-2022Error in labelingLabeling does not include guidance for proper water-quality maintenance and device inspection for heater cooler devices.
May 12, 2021Class IIZ-0035-2022Error in labelingLabeling does not include guidance for proper water-quality maintenance and device inspection for heater cooler devices.
May 12, 2021Class IIZ-0037-2022Error in labelingLabeling does not include guidance for proper water-quality maintenance and device inspection for heater cooler devices.
May 12, 2021Class IIZ-0038-2022Error in labelingLabeling does not include guidance for proper water-quality maintenance and device inspection for heater cooler devices.

No ancestor or descendant has a Class I recall.

How does its category compare?

All DWC devices

DWC Controller, Temperature, Cardiopulmonary Bypass: FDA decisions, review times, recalls and adverse event reports over 20 years. 2010, this device’s year, is highlighted.

FDA decisions per year · DWC

2in 2022

2007: 12010: 12012: 12013: 12014: 12020: 12022: 2
2007high 2 (2022) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DWC

Too few DWC 510(k)s in any year since 2007 to give a median.

Recalls per year · DWC

5in 2023

2010: 12012: 22014: 12015: 12016: 12017: 12018: 32021: 52023: 5
2007high 5 (2021) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DWC

670in 2025

2007: 32009: 22010: 2432011: 2812012: 3512013: 2442014: 3782015: 5752016: 7872017: 7162018: 5642019: 6972020: 5482021: 6222022: 6052023: 9282024: 5552025: 670
2007high 928 (2023) · low 0 (2008)2025

MAUDE reports naming product code DWC, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—03
20080—00
20090—02
20101—1243
20110—0281
20121—2351
20131—0244
20141—1378
20150—1575
20160—1787
20170—1716
20180—3564
20190—0697
20201—0548
20210—5622
20222—0605
20230—5928
20240—0555
20250—0670
20260—0—

What is it indicated for?

The CardioQuip Modular Cooler-Heater (MCH) is indicated to supply temperature-controlled water to heat exchange devices (e.g.; CPB Heat Exchangers, Cardioplegia Heat Exchangers and thermal regulating system blankets) to help control a patient's temperature during cardiopulmonary bypass procedures lasting not longer than six hours.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K102147: Cardioquip Modular Cooler-Heater Model MCH-1000.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K102147

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Page last updated Nov 3, 2026