510(k) Report

510(k) · Cleared

Cleared

Quantum Heater-Cooler

K212657510(k) TraditionalCleared May 6, 2022256-day reviewClass II21 CFR 870.4250

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Spectrum Medical, Ltd. · DWC Controller, Temperature, Cardiopulmonary Bypass

Cleared on May 6, 2022 as substantially equivalent to K052601 (Sorin Group Deutschland GmbH, 2006) after a 256-day review. Its family tree reaches back 3 generations to 1988.

Decided
May 6, 2022
Review
256 days
Class
Class II
Product code
DWC
Recalls
None
Predicates
1

Its longest line of descent runs back 3 generations to K883603 (1988). 5 ancestors in all, the oldest K811742 (1981). No later device cites it as a predicate yet.

  1. Root · 3 generations backK883603 Sarns Temp. Control and Monitor System (TCM II)3M Health Care, Sarns · 1988 · Recalled
  2. 2 generations backK031544 Heater Cooler Unit, Model Hcu 30Jostra AB · 2003 · Recalled
  3. 1 generation backK052601 Stockert Heater-Cooler System 3TSorin Group Deutschland GmbH · 2006 · Recalled
  4. This deviceK212657 Quantum Heater-CoolerSpectrum Medical, Ltd. · 2022
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K883603Sarns Temp. Control and Monitor System (TCM II)3M Health Care, Sarns19884
−2K031544Heater Cooler Unit, Model Hcu 30Jostra AB20034
−1K052601Stockert Heater-Cooler System 3TSorin Group Deutschland GmbH200618
0 (this device)K212657Quantum Heater-CoolerSpectrum Medical, Ltd.2022—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198110
200010
FIG. 1 — Lineage of K212657, 1988–2022Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DWC? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DWC device

  1. #1 K220635 Heater-Cooler System 3Tscore 5.3 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K052601Stockert Heater-Cooler System 3T
Summary page 4
…and Vascular Devices) III. PREDICATE DEVICE The predicate device for this submission is the Sorin/LivaNova Stöckert Heater-Cooler System 3T (K052601). IV. DEVICE DESCRIPTION The Quantum Heater-Cooler consists of the following components: • Quantum Heater-Cooler Unit (includ…
Sorin Group Deutschland GmbH200618

Also cleared against K052601: 2 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 34 years and passes through 4 companies; a new generation was cleared every 14.8 years on median.
  • 4 of the 5 ancestors have a recall of any class.
  • 2 other devices also cite its predicate K052601.

How does its category compare?

All DWC devices

DWC Controller, Temperature, Cardiopulmonary Bypass: FDA decisions, review times, recalls and adverse event reports over 20 years. 2022, this device’s year, is highlighted.

FDA decisions per year · DWC

2in 2022

2007: 12010: 12012: 12013: 12014: 12020: 12022: 2
2007high 2 (2022) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DWC

Too few DWC 510(k)s in any year since 2007 to give a median.

Recalls per year · DWC

5in 2023

2010: 12012: 22014: 12015: 12016: 12017: 12018: 32021: 52023: 5
2007high 5 (2021) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DWC

670in 2025

2007: 32009: 22010: 2432011: 2812012: 3512013: 2442014: 3782015: 5752016: 7872017: 7162018: 5642019: 6972020: 5482021: 6222022: 6052023: 9282024: 5552025: 670
2007high 928 (2023) · low 0 (2008)2025

MAUDE reports naming product code DWC, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—03
20080—00
20090—02
20101—1243
20110—0281
20121—2351
20131—0244
20141—1378
20150—1575
20160—1787
20170—1716
20180—3564
20190—0697
20201—0548
20210—5622
20222—0605
20230—5928
20240—0555
20250—0670
20260—0—

What is it indicated for?

The Spectrum Medical Quantum Heater-Cooler is intended to provide temperature-controlled fluid to compatible Quantum PureFlow heat exchanger devices (cardiopulmonary bypass heat exchangers and cardioplegia heat exchangers) to warm or cool a patient during cardiopulmonary bypass procedures lasting six (6) hours or less.

The Quantum Heater-Cooler is only intended to be used by trained Clinicians in a clinical environment.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K212657: Quantum Heater-Cooler.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K212657

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026