510(k) Report

510(k) · Cleared

Cleared

Discovery NM 750B

K102231510(k) TraditionalCleared Nov 3, 201086-day reviewClass I21 CFR 892.1100

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ge Medical Systems Israel, Functional Imaging · IYX Camera, Scintillation (Gamma)

Cleared on Nov 3, 2010 as substantially equivalent to K022960 (Ge Medical Systems F.I. Haifa, 2002) and 1 other predicate after an 86-day review. Its family tree reaches back 4 generations to 1995. 3 of the 14 devices in its family, counting itself, have a Class I recall. 3 later devices descend from it.

Decided
Nov 3, 2010
Review
86 days
Class
Class I
Product code
IYX
Recalls
None
Predicates
2

Its longest line of descent runs back 4 generations to K952674 (1995). 10 ancestors in all, the oldest K885054 (1989). 3 later devices descend from it, 2 citing it directly. 3 of the 14 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2000: 1 other ancestor1997: 2 other ancestors1995: 1 other ancestor, 1 with a Class I recall11992: 1 other ancestor1989: 1 other ancestor2016: 1 descendant cleared2011: 1 descendant cleared2013: 1 descendant clearedK952674 · Transact: Attenuation Correction System for the Apex Cardial Gamma Camera · Elscint, Inc. · 1995 · recalledK980959 · Vtransact: Attenuation Correction System for Dual-Head Variable Angle Gamma Camera · Ge Medical Systems F.I. Haifa · 1998K991841 · Hawkeye Option for Dual-Head Variable Angle Gamma Camera · Ge Medical Systems F.I. Haifa · 1999 · Class I recallK022960 · GE Quasar Nuclear Medicine System · Ge Medical Systems F.I. Haifa · 2002 · Class I recall1990200020102020
  1. Root · 4 generations backK952674 Transact: Attenuation Correction System for the Apex Cardial Gamma…Elscint, Inc. · 1995 · Recalled
  2. 3 generations backK980959 Vtransact: Attenuation Correction System for Dual-Head Variable…Ge Medical Systems F.I. Haifa · 1998
  3. 2 generations backK991841 Hawkeye Option for Dual-Head Variable Angle Gamma CameraGe Medical Systems F.I. Haifa · 1999 · Recalled (Class I)
  4. 1 generation backK022960 GE Quasar Nuclear Medicine SystemGe Medical Systems F.I. Haifa · 2002 · Recalled (Class I)
  5. This deviceK102231 Discovery NM 750BGe Medical Systems Israel, Functional Imaging · 2010
  6. Descendant · cites it directly · 1 of its ownK111791 Lumagem Molecular Breast Imaging SystemGamma Medica-Ideas Incorporated · 2011
  7. Descendant · cites it directly · 0 of its ownK160933 Discovery NM 750b BopsyGE Healthcare · 2016
  8. Descendant · 2 generations on · 0 of its ownK123408 Ergo Imaging SystemDigirad Corp. · 2013

Class I recalls in this family

Most influential descendants

Showing 3 of 3 · See all
  • K111791Lumagem Molecular Breast Imaging SystemGamma Medica-Ideas Incorporated · 2011 · cites it directly1descendant
  • K160933Discovery NM 750b BopsyGE Healthcare · 2016 · cites it directly0descendants
  • K123408Ergo Imaging SystemDigirad Corp. · 2013 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K952674Transact: Attenuation Correction System for the Apex Cardial Gamma CameraElscint, Inc.19951
−3K980959Vtransact: Attenuation Correction System for Dual-Head Variable Angle Gamma…Ge Medical Systems F.I. Haifa1998—
−2K991841Hawkeye Option for Dual-Head Variable Angle Gamma CameraGe Medical Systems F.I. Haifa1999Class I
−1K022960GE Quasar Nuclear Medicine SystemGe Medical Systems F.I. Haifa2002Class I
0 (this device)K102231Discovery NM 750BGe Medical Systems Israel, Functional Imaging2010—
+1K111791Lumagem Molecular Breast Imaging SystemGamma Medica-Ideas Incorporated2011—
+1K160933Discovery NM 750b BopsyGE Healthcare2016—
+2K123408Ergo Imaging SystemDigirad Corp.2013—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198910
199210
199511
199720
200010

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201110
201310
201610
FIG. 1 — Lineage of K102231, 1995–2016Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new IYX device

  1. #1 K230221 QDOSE® Multi-purpose Voxel Dosimetry…score 6.2 · 2 generations · no Class I recall
  2. #2 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K022960GE Quasar Nuclear Medicine System
Summary page 1
…Scintillation (gamma) camera Classification Names: 2I1CFR 892.1 100 Product Code: 90 IYX Predicate Device(s): K022960- INFINIA; K993813 LumaGEM scintillation Gamma camera Device Description:…
Ge Medical Systems F.I. Haifa2002Class I
*K993813Lumagem Scintillation Camera
Summary page 1
…ma) camera Classification Names: 2I1CFR 892.1 100 Product Code: 90 IYX Predicate Device(s): K022960- INFINIA; K993813 LumaGEM scintillation Gamma camera Device Description: The Discovery NM…
Gamma Medica2000—

Also cleared against K022960: 3 other devices →

What descends from it? (3)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K111791Lumagem Molecular Breast Imaging SystemGamma Medica-Ideas Incorporated20111
K160933Discovery NM 750b BopsyGE Healthcare20160

Has anything in its family been recalled? (3 Class I)

No FDA recall names this device.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K022960GE Quasar Nuclear Medicine SystemGe Medical Systems F.I. Haifa · 2002 · Direct predicate · ◆ on the line of descent in FIG. 11 Class I recall
  • K991841Hawkeye Option for Dual-Head Variable Angle Gamma CameraGe Medical Systems F.I. Haifa · 1999 · 2 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K953801VaricamElscint, Inc. · 1995 · 2 generations back · ◆ in the 1995 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 15 years and passes through 3 companies; a new generation was cleared every 3.1 years on median.
  • 4 of the 10 ancestors have a recall of any class; 3 have a Class I recall.
  • Of its descendants, K111791 (Gamma Medica-Ideas Incorporated, 2011) has gone furthest, with 1 descendant of its own.
  • 3 other devices also cite its predicate K022960.

How does its category compare?

All IYX devices

IYX Camera, Scintillation (Gamma): FDA decisions, review times, recalls and adverse event reports over 20 years. 2010, this device’s year, is highlighted.

FDA decisions per year · IYX

2in 2023

2007: 22008: 22009: 22010: 42011: 32013: 12014: 12016: 22017: 12023: 2
2007high 4 (2010) · low 0 (2012)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · IYX

82in 2011

2010: 82 days median2011: 82 days median
2007high 82 (2010)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · IYX

2in 2025

2025: 2
2007high 2 (2025) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · IYX

1in 2022

2007: 112008: 102009: 82010: 52011: 62012: 12013: 12014: 32015: 42016: 22018: 12022: 1
2007high 11 (2007) · low 0 (2017)2025

MAUDE reports naming product code IYX, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—011
20082—010
20092—08
201048205
201138206
20120—01
20131—01
20141—03
20150—04
20162—02
20171—00
20180—01
20190—00
20200—00
20210—00
20220—01
20232—00
20240—00
20250—20
20260—0—

What is it indicated for?

The Discovery NM 750b Gamma Camera is intended to measure and image the distribution of selected single photon emission radioisotopes in the human body to aid in the evaluation of lesions. The resultant images are intended to be reviewed by qualified medical professionals. The Discovery NM 750b Gamma Camera is intended for diagnostic imaging of the breast and other small body parts. The Discovery NM 750b Gamma Camera when used for breast imaging is…

Read the full statement (128 words)

The Discovery NM 750b Gamma Camera is intended to measure and image the distribution of selected single photon emission radioisotopes in the human body to aid in the evaluation of lesions. The resultant images are intended to be reviewed by qualified medical professionals. The Discovery NM 750b Gamma Camera is intended for diagnostic imaging of the breast and other small body parts. The Discovery NM 750b Gamma Camera when used for breast imaging is intended as an adjunct to mammography or other breast imaging modalities (it is not intended for primary screening of the population). The Discovery NM 750b Gamma Camera is indicated for planar and dynamic planar scintigraphy in the energy range 80-200keV for the detection and display of radioisotope tracer uptake in patients of all ages.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K102231: Discovery NM 750B.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K102231

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.