GE Quasar Nuclear Medicine System

K022960510(k) TraditionalCleared Sep 19, 200213-day reviewClass II21 CFR 892.1200Recalls 24 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ge Medical Systems F.I. Haifa · KPS System, Tomography, Computed, Emission

Cleared on Sep 19, 2002 as substantially equivalent to K953801 (Elscint, Inc., 1995) and 2 other predicates after a 13-day review. Its family tree reaches back 3 generations to 1995. 9 of the 29 devices in its family, counting itself, have a Class I recall. 23 later devices descend from it. It has been recalled 24 times, including a Class I recall.

Decided
Sep 19, 2002
Review
13 days
Class
Class II
Product code
KPS
Recalls
24 (1 Class I)
Predicates
3

Its longest line of descent runs back 3 generations to K952674 (1995). 5 ancestors in all. 23 later devices descend from it, 4 citing it directly. 9 of the 29 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 1997: 1 other ancestor1995: 1 other ancestor, 1 with a Class I recall12004: 1 descendant cleared2010: 1 descendant cleared2009: 1 descendant cleared, 1 with a Class I recall12005: 1 descendant cleared2008: 2 descendants cleared2016: 2 descendants cleared, 1 with a Class I recall12025: 3 descendants cleared2018: 2 descendants cleared, 1 with a Class I recall12011: 3 descendants cleared, 2 with a Class I recall22021: 2 descendants cleared2012: 1 descendant cleared, 1 with a Class I recall12013: 1 descendant cleared2019: 1 descendant cleared2023: 1 descendant cleared2026: 1 descendant clearedK952674 · Transact: Attenuation Correction System for the Apex Cardial Gamma Camera · Elscint, Inc. · 1995 · recalledK980959 · Vtransact: Attenuation Correction System for Dual-Head Variable Angle Gamma Camera · Ge Medical Systems F.I. Haifa · 1998K991841 · Hawkeye Option for Dual-Head Variable Angle Gamma Camera · Ge Medical Systems F.I. Haifa · 1999 · Class I recall1995200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors, 1 with a Class I recall1 1 generation forward: 4 descendants, 1 with a Class I recall12 generations forward: 8 descendants, 3 with a Class I recall33 generations forward: 7 descendants, 2 with a Class I recall24 generations forward: 2 descendants5 generations forward: 1 descendant6 generations forward: 1 descendant K991841 · Hawkeye Option for Dual-Head Variable Angle Gamma Camera · Ge Medical Systems F.I. Haifa · 1999 · Class I recallK980959 · Vtransact: Attenuation Correction System for Dual-Head Variable Angle Gamma Camera · Ge Medical Systems F.I. Haifa · 1998K952674 · Transact: Attenuation Correction System for the Apex Cardial Gamma Camera · Elscint, Inc. · 1995 · recalled0+5
  1. Root · 3 generations backK952674 Transact: Attenuation Correction System for the Apex Cardial Gamma…Elscint, Inc. · 1995 · Recalled
  2. 2 generations backK980959 Vtransact: Attenuation Correction System for Dual-Head Variable…Ge Medical Systems F.I. Haifa · 1998
  3. 1 generation backK991841 Hawkeye Option for Dual-Head Variable Angle Gamma CameraGe Medical Systems F.I. Haifa · 1999 · Recalled (Class I)
  4. This deviceK022960 GE Quasar Nuclear Medicine SystemGe Medical Systems F.I. Haifa · 2002
  5. Descendant · cites it directly · 13 of its ownK093514 Discovery Model NM/CT 670Ge Medical Systems F.I. Haifa · 2009 · Recalled (Class I)
  6. Descendant · 2 generations on · 6 of its ownK153402 Discovery NM/CT 670 CZTGE Healthcare · 2016 · Recalled (Class I)
  7. Descendant · 3 generations on · 3 of its ownK182484 VERITON CT whole body SPECT/CT systemSpectrum Dynamics Medical, Ltd. · 2018
  8. Descendant · cites it directly · 3 of its ownK102231 Discovery NM 750BGe Medical Systems Israel, Functional Imaging · 2010
  9. Descendant · cites it directly · 3 of its ownK051855 VentriGe Medical Systems F.I. Haifa · 2005 · Recalled

Showing 5 of 23 descendants. See all descendants

Descendants cleared per year

2004: 12005: 12008: 22009: 12010: 12011: 32012: 12013: 12016: 22018: 22019: 12021: 22023: 12025: 32026: 1
2004high 3 (2011) · low 0 (2006)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 23 · See all
  • K093514Discovery Model NM/CT 670Ge Medical Systems F.I. Haifa · 2009 · cites it directly13descendants
  • K153402Discovery NM/CT 670 CZTGE Healthcare · 2016 · 2 generations on6descendants
  • K182484VERITON CT whole body SPECT/CT systemSpectrum Dynamics Medical, Ltd. · 2018 · 3 generations on3descendants
  • K102231Discovery NM 750BGe Medical Systems Israel, Functional Imaging · 2010 · cites it directly3descendants
  • K051855VentriGe Medical Systems F.I. Haifa · 2005 · cites it directly3descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K952674Transact: Attenuation Correction System for the Apex Cardial Gamma CameraElscint, Inc.19951
−2K980959Vtransact: Attenuation Correction System for Dual-Head Variable Angle Gamma…Ge Medical Systems F.I. Haifa1998—
−1K991841Hawkeye Option for Dual-Head Variable Angle Gamma CameraGe Medical Systems F.I. Haifa1999Class I
0 (this device)K022960GE Quasar Nuclear Medicine SystemGe Medical Systems F.I. Haifa2002—
+1K093514Discovery Model NM/CT 670Ge Medical Systems F.I. Haifa2009Class I
+2K153402Discovery NM/CT 670 CZTGE Healthcare2016Class I
+3K182484VERITON CT whole body SPECT/CT systemSpectrum Dynamics Medical, Ltd.2018—
+1K102231Discovery NM 750BGe Medical Systems Israel, Functional Imaging2010—
+1K051855VentriGe Medical Systems F.I. Haifa20053

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199511
199710

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200410
200510
200820
200911
201010
201132
201211
201310
201621
201821
201910
202120
202310
202530
202610
FIG. 1 — Lineage of K022960, 1995–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

9 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KPS device

  1. #1 K253564 uMI Panvivo (uMI Panvivo); uMI Panvivo (uMI…score 18.6 · 17 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K953801Varicam
Summary page 1
…Reason for 510(k) Submission Modification of legally marketed device. Identification of Legallv Marketed Equivalent Devices Elscint Apex VariCam - K953801 GE Hawkeye Option for Dual-Head Variable Angle Gamma Camera - K991841 GE CoDe Option; Volumetric Coinci…
Elscint, Inc.1995Class I
*K991841Hawkeye Option for Dual-Head Variable Angle Gamma Camera
Summary page 1
…ntification of Legallv Marketed Equivalent Devices Elscint Apex VariCam - K953801 GE Hawkeye Option for Dual-Head Variable Angle Gamma Camera - K991841 GE CoDe Option; Volumetric Coincidence Detection (CoDe) Imaging System for VariCam (Nuclear Medicine System)…
Ge Medical Systems F.I. Haifa1999Class I
*K962100Volmetric Coincidence Detection System
Summary page 1
…e Gamma Camera - K991841 GE CoDe Option; Volumetric Coincidence Detection (CoDe) Imaging System for VariCam (Nuclear Medicine System) - K962 100 24of51 DEPARTMENT OF HEALTH…
Elscint, Inc.1997—

Also cleared against K953801: 1 other device →

What descends from it? (23)

See all 23

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K093514Discovery Model NM/CT 670 Class IGe Medical Systems F.I. Haifa200913
K102231Discovery NM 750BGe Medical Systems Israel, Functional Imaging20103
K051855VentriGe Medical Systems F.I. Haifa20053
K043381Infinia LightspeedGe Medical Systems F.I. Haifa20041

Has anything in its family been recalled? (9 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 20, 2025Class IIZ-2097-2025Software designUnintended radial detector motion may occur during patient setup or during patient scan if system does not have correct version of gantry software installed. Unintended detector motion may result in life-threatening injury.
Dec 18, 2024Class IIZ-0888-2025Package design/selectionGE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been…
Dec 18, 2024Class IIZ-0897-2025Package design/selectionGE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been…
Dec 18, 2024Class IIZ-0898-2025Package design/selectionGE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been…
Dec 18, 2024Class IIZ-0899-2025Package design/selectionGE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been…
Oct 14, 2014Class IIZ-0185-2015Nonconforming Material/ComponentGE is issuing this recall due to a potential safety issue related to a portion of the system falling onto the patient during a scan due to fasteners being loose that secured the camera to the gantry. GE is updating the Preventative…
Oct 14, 2014Class IIZ-0186-2015Nonconforming Material/ComponentGE is issuing this recall due to a potential safety issue related to a portion of the system falling onto the patient during a scan due to fasteners being loose that secured the camera to the gantry. GE is updating the Preventative…
Oct 14, 2014Class IIZ-0188-2015Nonconforming Material/ComponentGE is issuing this recall due to a potential safety issue related to a portion of the system falling onto the patient during a scan due to fasteners being loose that secured the camera to the gantry. GE is updating the Preventative…
Oct 14, 2014Class IIZ-0189-2015Nonconforming Material/ComponentGE is issuing this recall due to a potential safety issue related to a portion of the system falling onto the patient during a scan due to fasteners being loose that secured the camera to the gantry. GE is updating the Preventative…
Oct 14, 2014Class IIZ-0187-2015Nonconforming Material/ComponentGE is updating the Preventative Maintenance procedure and schedule, and added an additional Preventative Maintenance check for loose fasteners on the Nuclear Medicaine systems. This check is being added to prevent a portion of the system…
Oct 14, 2014Class IIZ-0184-2015Nonconforming Material/ComponentGE is issuing this recall due to a potential safety issue related to a portion of the system falling onto the patient during a scan due to fasteners being loose that secured the camera to the gantry. GE is updating the Preventative…
Feb 7, 2014Class IIZ-1198-2014Device DesignPotential safety issue related to unintended radial detector motion, which may occur during patient setup or at the end of a patient scan. This failure mode has been observed on a single system, in which the detector moved to the home…
Feb 7, 2014Class IIZ-1200-2014Device DesignPotential safety issue related to unintended radial detector motion, which may occur during patient setup or at the end of a patient scan. This failure mode has been observed on a single system, in which the detector moved to the home…
Feb 7, 2014Class IIZ-1199-2014Device DesignPotential safety issue related to unintended radial detector motion, which may occur during patient setup or at the end of a patient scan. This failure mode has been observed on a single system, in which the detector moved to the home…
Jun 13, 2013Class IZ-1787-2013Device DesignGE became aware of an incident at a VA Medical Center facility in the US in which a patient died due to injuries sustained while being scanned on an Infinia Hawkeye 4 Nuclear Medicine System.
Mar 6, 2012Class IIZ-1078-2013Process controlAccelerated fatigue of the lateral axis motion subsystem of the Infinia Computed Emission Tomography System could result in mechanical failure causing the detector to slide; may impact patient and operators.
Sep 8, 2008Not classifiedZ-1095-2009Software designThere is a possibility for misregistration when brain scans are performed with Infinia with the Hawkeye option. This misregistration could present the potential for misdiagnosis when the patient headrest sags under the weight of a head.
Sep 8, 2008Not classifiedZ-1094-2009Software designThere is a possibility for misregistration when brain scans are performed with Infinia with the Hawkeye option. This misregistration could present the potential for misdiagnosis when the patient headrest sags under the weight of a head.
Jun 2, 2008Not classifiedZ-1001-2009Device DesignGE Healthcare has recently became aware of excessive pressure applied by the collimator pressure sensitive cover of Infinia (Infinia, Infinia Hawkeye, Infinia Hawkeye 4) that may impact patient safety. It was reported at a customer site…
Jun 2, 2008Not classifiedZ-1003-2009Device DesignGE Healthcare has recently became aware of excessive pressure applied by the collimator pressure sensitive cover of Infinia (Infinia, Infinia Hawkeye, Infinia Hawkeye 4) that may impact patient safety. It was reported at a customer site…
Jun 2, 2008Not classifiedZ-1002-2009Device DesignGE Healthcare has recently became aware of excessive pressure applied by the collimator pressure sensitive cover of Infinia (Infinia, Infinia Hawkeye, Infinia Hawkeye 4) that may impact patient safety. It was reported at a customer site…
Jun 2, 2008Not classifiedZ-1121-2009OtherThe metal pin in the adjustable head holder, which is used to acquire Brain SPECT data on Infinia systems (Infinia, Infinia Hawkeye, and Infinia Hawkeye 4), creates an artifact on the patient's cerebellum, which may lead to the potential…
Jun 2, 2008Not classifiedZ-1120-2009OtherThe metal pin in the adjustable head holder, which is used to acquire Brain SPECT data on Infinia systems (Infinia, Infinia Hawkeye, and Infinia Hawkeye 4), creates an artifact on the patient's cerebellum, which may lead to the potential…
Jun 2, 2008Not classifiedZ-1119-2009OtherThe metal pin in the adjustable head holder, which is used to acquire Brain SPECT data on Infinia systems (Infinia, Infinia Hawkeye, and Infinia Hawkeye 4), creates an artifact on the patient's cerebellum, which may lead to the potential…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K991841Hawkeye Option for Dual-Head Variable Angle Gamma CameraGe Medical Systems F.I. Haifa · 1999 · Direct predicate · ◆ on the line of descent in FIG. 11 Class I recall
  • K953801VaricamElscint, Inc. · 1995 · Direct predicate · ◆ in the 1995 ancestors bar of FIG. 11 Class I recall
  • K093514Discovery Model NM/CT 670Ge Medical Systems F.I. Haifa · 2009 · Cites it directly · ◆ in the 2009 descendants bar of FIG. 12 Class I recalls
  • K153402Discovery NM/CT 670 CZTGE Healthcare · 2016 · 2 generations on · ◆ in the 2016 descendants bar of FIG. 11 Class I recall
  • K173816NM/CT 850, NM/CT 860GE Healthcare · 2018 · 2 generations on · ◆ in the 2018 descendants bar of FIG. 11 Class I recall
  • K111445Discovery NM 630GE Healthcare · 2011 · 2 generations on · ◆ in the 2011 descendants bar of FIG. 12 Class I recalls
  • K113350Brivo NM 615GE Healthcare · 2011 · 3 generations on · ◆ in the 2011 descendants bar of FIG. 12 Class I recalls
  • K121019Optima NM/CT 640GE Healthcare · 2012 · 3 generations on · ◆ in the 2012 descendants bar of FIG. 12 Class I recalls

What does this lineage say?

  • The line of descent spans 7 years and passes through 2 companies; a new generation was cleared every 2.7 years on median.
  • 3 of the 5 ancestors have a recall of any class; 2 have a Class I recall.
  • Of its descendants, K093514 (Ge Medical Systems F.I. Haifa, 2009) has gone furthest, with 13 descendants of its own.
  • 1 other device also cites its predicate K953801.

How does its category compare?

All KPS devices

KPS System, Tomography, Computed, Emission: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · KPS

12in 2025

2007: 82008: 172009: 92010: 32011: 72012: 62013: 112014: 112015: 32016: 112017: 112018: 132019: 32020: 122021: 152022: 52023: 122024: 112025: 122026 (partial year): 6
2007high 17 (2008) · low 3 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KPS

65in 2025

2007: 38 days median2008: 25 days median2009: 42 days median2010: 91 days median2011: 28 days median2012: 94 days median2013: 84 days median2014: 92 days median2015: 112 days median2016: 65 days median2017: 104 days median2018: 59 days median2019: 60 days median2020: 72 days median2021: 66 days median2022: 60 days median2023: 63 days median2024: 59 days median2025: 65 days median2026 (partial year): 67 days median
2007high 112 (2015) · low 25 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KPS

16in 2025

2007: 222008: 162009: 142010: 42011: 342012: 132013: 132014: 182015: 52016: 162017: 112018: 12, 3 Class I2019: 5, 5 Class I2020: 82021: 42022: 8, 6 Class I2023: 6, 3 Class I2024: 152025: 162026 (partial year): 6
2007high 34 (2011) · low 4 (2010)2026

Class I (17 since 2007) Class II/III not classified

Adverse event reports per year · KPS

36in 2025

2007: 382008: 112009: 152010: 442011: 562012: 342013: 2402014: 332015: 242016: 102017: 312018: 262019: 1052020: 222021: 102022: 152023: 222024: 232025: 36
2007high 240 (2013) · low 10 (2016)2025

MAUDE reports naming product code KPS, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20078382238
200817251611
20099421415
2010391444
20117283456
20126941334
2013118413240
201411921833
20153112524
201611651610
2017111041131
2018135912 (3)26
20193605 (5)105
20201272822
20211566410
20225608 (6)15
202312636 (3)22
202411591523
202512651636
20266676—

What is it indicated for?

The intended use of the Quasar system is to perform general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, using a variety of scanning modes supported by various acquisition types and optional imaging features designed to enhance image quality in Oncology, Cardiology, Neurology and other clinical diagnostic imaging applications. The scanning modes include planar (Static, Multi-gated, Dynamic, Whole…

Read the full statement (121 words)

The intended use of the Quasar system is to perform general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, using a variety of scanning modes supported by various acquisition types and optional imaging features designed to enhance image quality in Oncology, Cardiology, Neurology and other clinical diagnostic imaging applications.

The scanning modes include planar (Static, Multi-gated, Dynamic, Whole body scanning) and tomographic (SPECT , Gated SPECT , Whole body SPECT , Camera based PET - also known as Coincidence Detection). Acquisition types include single and multi-isotope/multi-peak frame/list mode single-photon and positron imaging. Optional imaging-enhancement features include assortment of collimators, gating by physiological signals, real-time automatic body contouring, and CT-based attenuation correction and functional anatomic mapping.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K022960: GE Quasar Nuclear Medicine System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K022960

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026