510(k) Report

510(k) · Cleared

Recalled · 2010

Aesculap- Meithke Progav Programmable Shunt System

K103003510(k) SpecialCleared Nov 9, 201028-day reviewClass II21 CFR 882.5550Recalls 5

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Aesculap, Inc. · JXG Shunt, Central Nervous System And Components

Cleared on Nov 9, 2010 as substantially equivalent to K062009 (Aesculap, Inc., 2007) after a 28-day review. Its family tree reaches back 7 generations to 1995. 4 later devices descend from it. It has been recalled 5 times.

Decided
Nov 9, 2010
Review
28 days
Class
Class II
Product code
JXG
Recalls
5
Predicates
1

Its longest line of descent runs back 7 generations to K944222 (1995). 17 ancestors in all, the oldest K861377 (1986). 4 later devices descend from it, 3 citing it directly.

  1. Root · 7 generations backK944222 Codman-Medos Valve Shunt SystemJohnson & Johnson Professionals, Inc. · 1995
  2. 6 generations backK973774 Codman Hakim Micro Precision ValveJohnson & Johnson Professionals, Inc. · 1997
  3. 5 generations backK980778 Hakim Micro Programmable Valve SystemJohnson & Johnson Professionals, Inc. · 1998
  4. 4 generations backK012052 Medtronic PS Medical Strata Valve, Models 42855, 42865, 92865, 92855Medtronic PS Medical · 2002
  5. 3 generations backK033850 Medtronic PS Medical Strata NSC Valve and Shunt Assemblies with and…Medtronic Neurosurgery · 2004 · Recalled
  6. 2 generations backK042465 PS Medical Strata II ValveMedtronic Neurosurgery · 2004 · Recalled
  7. 1 generation backK062009 Aesculap-Meithke Progav Programmable Valve, Models FV410T-FV451TAesculap, Inc. · 2007
  8. This deviceK103003 Aesculap- Meithke Progav Programmable Shunt SystemAesculap, Inc. · 2010
  9. Descendant · cites it directly · 2 of its ownK141687 Aesculap - Meithke Progav 2.0 Adjustable Shunt SystemAesculap, Inc. · 2015
  10. Descendant · cites it directly · 2 of its ownK120559 Aesculap-Meithke Prosa Adjustable Shunt SystemAesculap, Inc. · 2012 · Recalled
  11. Descendant · cites it directly · 1 of its ownK161853 proGAV Programmable Shunt System, proGAV 2.0 Adjustable Shunt…Aesculap, Inc. · 2017 · Recalled
  12. Descendant · 2 generations on · 0 of its ownK192266 M. Blue Adjustable Shunt SystemAesculap, Inc. · 2019

Most influential descendants

Showing 4 of 4 · See all
  • K141687Aesculap - Meithke Progav 2.0 Adjustable Shunt SystemAesculap, Inc. · 2015 · cites it directly2descendants
  • K120559Aesculap-Meithke Prosa Adjustable Shunt SystemAesculap, Inc. · 2012 · cites it directly2descendants
  • K161853proGAV Programmable Shunt System, proGAV 2.0 Adjustable Shunt System, proSA Adjustable…Aesculap, Inc. · 2017 · cites it directly1descendant
  • K192266M. Blue Adjustable Shunt SystemAesculap, Inc. · 2019 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K944222Codman-Medos Valve Shunt SystemJohnson & Johnson Professionals, Inc.1995—
−6K973774Codman Hakim Micro Precision ValveJohnson & Johnson Professionals, Inc.1997—
−5K980778Hakim Micro Programmable Valve SystemJohnson & Johnson Professionals, Inc.1998—
−4K012052Medtronic PS Medical Strata Valve, Models 42855, 42865, 92865, 92855Medtronic PS Medical2002—
−3K033850Medtronic PS Medical Strata NSC Valve and Shunt Assemblies with and Without…Medtronic Neurosurgery20042
−2K042465PS Medical Strata II ValveMedtronic Neurosurgery20044
−1K062009Aesculap-Meithke Progav Programmable Valve, Models FV410T-FV451TAesculap, Inc.2007—
0 (this device)K103003Aesculap- Meithke Progav Programmable Shunt SystemAesculap, Inc.2010—
+1K141687Aesculap - Meithke Progav 2.0 Adjustable Shunt SystemAesculap, Inc.2015—
+1K120559Aesculap-Meithke Prosa Adjustable Shunt SystemAesculap, Inc.20122
+1K161853proGAV Programmable Shunt System, proGAV 2.0 Adjustable Shunt System, proSA…Aesculap, Inc.20171
+2K192266M. Blue Adjustable Shunt SystemAesculap, Inc.2019—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198610
199010
199510
199810
200010
200220
200430

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201210
201510
201710
201910
FIG. 1 — Lineage of K103003, 1995–2019Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JXG? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JXG device

  1. #1 K243531 Codman Libertís Shunt Catheter with Bactiseal…score 11.9 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K062009Aesculap-Meithke Progav Programmable Valve, Models FV410T-FV451T
Summary page 1
…TIAL EQUIVALENCE Aesculapa, Inc. believes that the modified proGAV Shunt is substantially equivalent to the proGAV Shunt of the Miethke proGAV Programmable Shunt System (K062009). The modified shunt is similar to the current proGAV Shunt. The only difference is the size of the internal magnets. The materia…
Aesculap, Inc.2007—

Also cleared against K062009: 4 other devices →

What descends from it? (4)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K141687Aesculap - Meithke Progav 2.0 Adjustable Shunt SystemAesculap, Inc.20152
K120559Aesculap-Meithke Prosa Adjustable Shunt SystemAesculap, Inc.20122
K161853proGAV Programmable Shunt System, proGAV 2.0 Adjustable Shunt System, proSA Adjustable…Aesculap, Inc.20171

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 16, 2010Not classifiedZ-0839-2011Device DesignReports of distal bond failure which has resulted in peeling of the bond and/or detachment of the distal end of the scoring element. May result in vessel injury or embolism.
Nov 16, 2010Not classifiedZ-0840-2011Device DesignReports of distal bond failure which has resulted in peeling of the bond and/or detachment of the distal end of the scoring element. May result in vessel injury or embolism.
Nov 16, 2010Not classifiedZ-0842-2011Device DesignReports of distal bond failure which has resulted in peeling of the bond and/or detachment of the distal end of the scoring element. May result in vessel injury or embolism.
Nov 16, 2010Not classifiedZ-0838-2011Device DesignReports of distal bond failure which has resulted in peeling of the bond and/or detachment of the distal end of the scoring element. May result in vessel injury or embolism.
Nov 16, 2010Not classifiedZ-0841-2011Device DesignReports of distal bond failure which has resulted in peeling of the bond and/or detachment of the distal end of the scoring element. May result in vessel injury or embolism.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 15 years and passes through 4 companies; a new generation was cleared every 2.2 years on median.
  • 5 of the 17 ancestors have a recall of any class.
  • Of its descendants, K141687 (Aesculap, Inc., 2015) has gone furthest, with 2 descendants of its own.
  • 4 other devices also cite its predicate K062009.

How does its category compare?

All JXG devices

JXG Shunt, Central Nervous System And Components: FDA decisions, review times, recalls and adverse event reports over 20 years. 2010, this device’s year, is highlighted.

FDA decisions per year · JXG

8in 2025

2007: 32008: 62009: 52010: 32011: 72012: 32013: 32015: 62016: 62017: 72018: 62019: 42020: 22021: 12022: 42023: 52024: 32025: 82026 (partial year): 2
2007high 8 (2025) · low 0 (2014)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JXG

255in 2025

2007: 184 days median2008: 60 days median2009: 106 days median2010: 30 days median2011: 136 days median2012: 78 days median2013: 108 days median2015: 276 days median2016: 143 days median2017: 90 days median2018: 102 days median2019: 101 days median2022: 101 days median2023: 134 days median2024: 90 days median2025: 255 days median
2007high 276 (2015) · low 30 (2010)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JXG

3in 2024

2008: 12009: 302011: 52012: 12013: 142014: 62017: 12020: 82023: 22024: 32026 (partial year): 1
2007high 30 (2009) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JXG

1,198in 2025

2007: 6492008: 6302009: 6292010: 8122011: 7872012: 9772013: 1,0042014: 1,2512015: 9842016: 1,0912017: 1,3352018: 1,4842019: 1,2952020: 1,2062021: 1,4672022: 1,3292023: 1,2212024: 1,2142025: 1,198
2007high 1,484 (2018) · low 629 (2009)2025

MAUDE reports naming product code JXG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200731840649
20086601630
2009510630629
20103300812
201171365787
20123781977
20133108141,004
20140—61,251
201562760984
2016614301,091
201779011,335
2018610201,484
2019410101,295
20202—81,206
20211—01,467
2022410101,329
2023513421,221
202439031,214
2025825501,198
20262—1—

What is it indicated for?

The Miethke proGAV Programmable Shunt System is intended to shunt cerebrospinal fluid (CSF) from the lateral ventricles of the brain into the peritoneum. Adjustments of the proGAV shunt can be verified by using the verification instrument but must be confirmed by radiograph (X-ray).

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K103003: Aesculap- Meithke Progav Programmable Shunt System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K103003

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026