510(k) Report

510(k) · Cleared

Recalled · Class II · 2026

Spacelabs Multi-Parameter Module

K103142510(k) SpecialCleared Mar 3, 2011129-day reviewClass II21 CFR 870.1025Recalls 14

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Spacelabs Healthcare, Inc. · DSI Detector And Alarm, Arrhythmia

Cleared on Mar 3, 2011 as substantially equivalent to K050605 (Spacelabs Medical, Inc., 2005) after a 129-day review. Its family tree reaches back 7 generations to 1997. It has been recalled 14 times.

Decided
Mar 3, 2011
Review
129 days
Class
Class II
Product code
DSI
Recalls
14
Predicates
1

Its longest line of descent runs back 7 generations to K962603 (1997). 26 ancestors in all, the oldest K882609 (1989). No later device cites it as a predicate yet.

  1. Root · 7 generations backK962603 Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & CablesMasimo Corp. · 1997
  2. 6 generations backK973887 Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and…Masimo Corp. · 1998
  3. 5 generations backK982255 Model 2000 Pulse OximeterIvy Biomedical Systems, Inc. · 1998
  4. 4 generations backK991823 N-395 Pulse Oximeter, Model N-395Nellcor Puritan Bennett, Inc. · 1999
  5. 3 generations backK012891 Oximax Pulse Oximetry System with N-595 Pulse Oximeter and Oximax…Nellcor Puritan Bennett, Inc. · 2002 · Recalled
  6. 2 generations backK050175 Spacelabs Medical Multiparameter Module 91496Spacelabs Medical, Inc. · 2005
  7. 1 generation backK050605 Spacelabs Medical Multiparameter Module 91496Spacelabs Medical, Inc. · 2005 · Recalled
  8. This deviceK103142 Spacelabs Multi-Parameter ModuleSpacelabs Healthcare, Inc. · 2011
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K962603Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & CablesMasimo Corp.1997—
−6K973887Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and CablesMasimo Corp.1998—
−5K982255Model 2000 Pulse OximeterIvy Biomedical Systems, Inc.1998—
−4K991823N-395 Pulse Oximeter, Model N-395Nellcor Puritan Bennett, Inc.1999—
−3K012891Oximax Pulse Oximetry System with N-595 Pulse Oximeter and Oximax Sensors and…Nellcor Puritan Bennett, Inc.20021
−2K050175Spacelabs Medical Multiparameter Module 91496Spacelabs Medical, Inc.2005—
−1K050605Spacelabs Medical Multiparameter Module 91496Spacelabs Medical, Inc.20051
0 (this device)K103142Spacelabs Multi-Parameter ModuleSpacelabs Healthcare, Inc.2011—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198910
199030
199110
199430
199530
199610
199720
199810
199910
200010
200310
200410
FIG. 1 — Lineage of K103142, 1997–2011Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DSI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DSI device

  1. #1 K243349 BodyGuardian Remote Monitoring System (BGRMS…score 13.1 · 9 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K050605Spacelabs Medical Multiparameter Module 91496
Summary page 1
…30, 21 CFR §880.2910, 21 CFR DSI, DPZ, DQA, DSK, DXG, DXN, FLL, LOS New Spacelabs Model Predicate 510(k) Predicate Number Manufacturer / Model Spacelabs Multi-parameter K050605 Spacelabs Medical Inc. / Spacelabs Medical Multi- parameter, Module 91496 Module Page 1 of 4 372 Device Descr…
Spacelabs Medical, Inc.20051

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 24, 2026Class IIZ-2243-2026Software designGE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the…
Apr 24, 2026Class IIZ-2246-2026Software designGE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the…
Apr 24, 2026Class IIZ-2247-2026Software designGE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the…
Apr 24, 2026Class IIZ-2245-2026Software designGE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the…
Apr 24, 2026Class IIZ-2244-2026Software designGE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the…
Mar 26, 2026Class IIZ-2052-2026Process controlDue to a manufacturing issue (i.e. malfunctioning of electrical Leakage tester).
Feb 27, 2026Class IIZ-1793-2026Component change controlMulti-parameter command module has a circuit board issue and when in Auto mode, Cardiac Output (CO) measurement may activate early, resulting in Bad Curve error with no CO measurement or Irregular Curve error with CO measurement not…
Apr 12, 2023Class IIZ-1503-2023Component change controlCommand modules will display the following when parameter processing is suspended, depending on software version: 1) Older versions will have waveforms present with no alarms, 2) Newer versions have waveforms absent with no alarms. This…
Jul 1, 2016Class IIZ-2343-2016Vendor change controlThe firm has received multiple reports of the Non-Invasive Blood Pressure (NIBP) parameter becoming non-functional with an associated message (No Reading). In addition, there is one of the following three error messages: Inflate Error, HW…
Apr 21, 2015Class IIZ-1577-2015Software designModules resetting. This failure mode happens when the Spacelabs Command Module is configured with the Masimo SpO2 option (-M), the Spacelabs Respiration option (-R) and the Masimo SpO2 PCBA, PN: 010-1136-02. With these two options and the…
Apr 17, 2015Class IIZ-1542-2015Software designUltraview SL Command Modules which were upgraded with the Masimo SpO2 Option, Model 91496-M was affected. Customer reported that during the time that the module is resetting (~10 seconds), monitoring of all parameters will be suspended.…
May 8, 2014Class IIZ-1705-2014Software Manufacturing/Software DeploymentA software anomaly causes temperature alarm limits to change to limits which are not physiologically possible and result in nuisance false positive alarms.
Nov 3, 2011Not classifiedZ-0690-2012Process change controlNIBP pump inside the Ultraview SL Command Module, Model 91496, became loose. The pump is suspended by the air hoses and electrical connections, but there can be enough movement to damage internal components and cause the module to fail.
Jul 13, 2011Not classifiedZ-0776-2012Software designOne Spacelabs Medical Ultraview SL Command Module, Model 91496, was removed from customer due to software error.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 14 years and passes through 5 companies; a new generation was cleared every 11 months on median.
  • 7 of the 26 ancestors have a recall of any class.

How does its category compare?

All DSI devices

DSI Detector And Alarm, Arrhythmia: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.

FDA decisions per year · DSI

1in 2025

2007: 72008: 52009: 62010: 62011: 62012: 32013: 12014: 42015: 92016: 62018: 22019: 32020: 22023: 22024: 12025: 12026 (partial year): 1
2007high 9 (2015) · low 0 (2017)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DSI

243in 2019

2007: 151 days median2008: 70 days median2009: 73 days median2010: 111 days median2011: 114 days median2012: 113 days median2014: 200 days median2015: 127 days median2016: 224 days median2019: 243 days median
2007high 243 (2019) · low 70 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DSI

18in 2025

2007: 42008: 82009: 32010: 52011: 32012: 12018: 12019: 42020: 22022: 12023: 22025: 182026 (partial year): 1
2007high 18 (2025) · low 0 (2013)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DSI

712in 2025

2007: 642008: 1562009: 2792010: 8232011: 7932012: 7662013: 7952014: 1,0002015: 1,6452016: 4,3522017: 3,5942018: 3,6482019: 3,4852020: 3,0842021: 3,0922022: 1,6382023: 1,6162024: 1,3202025: 712
2007high 4,352 (2016) · low 64 (2007)2025

MAUDE reports naming product code DSI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20077151464
20085708156
20096733279
201061115823
201161143793
201231131766
20131—0795
2014420001,000
2015912701,645
2016622404,352
20170—03,594
20182—13,648
2019324343,485
20202—23,084
20210—03,092
20220—11,638
20232—21,616
20241—01,320
20251—18712
20261—1—

What is it indicated for?

The Spacelabs Multi-parameter Module is intended for use with the Patient Care Management System (PCMS) to acquire, monitor, and process various clinical parameters from an adult or neonate/infant populations in any type of clinical environment other than home use.

Physiological parameters that may be monitored include ECG with arrhythmia detection, respiration, invasive and noninvasive blood pressure, temperature, oxygen saturation (SpO2) and cardiac output. Acquired data may then be communicated to an information network for display, recording, editing and analysis.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K103142: Spacelabs Multi-Parameter Module.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K103142

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026