510(k) Report

510(k) · Cleared

Recalled · Class II · 2013

Sterrad Cyclesure Biological Indicator

K103222510(k) TraditionalCleared Feb 25, 2011116-day reviewClass II21 CFR 880.2800Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Advanced Sterilization Products · FRC Indicator, Biological Sterilization Process

Cleared on Feb 25, 2011 as substantially equivalent to K071014 (Advanced Sterilization Products, 2007) after a 116-day review. Its family tree reaches back 3 generations to 2002. 20 later devices descend from it. It has been recalled 4 times.

Decided
Feb 25, 2011
Review
116 days
Class
Class II
Product code
FRC
Recalls
4
Predicates
1

Its longest line of descent runs back 3 generations to K994055 (2002). 20 later devices descend from it, 2 citing it directly.

Descendants, by year cleared 2012: 3 descendants cleared2014: 2 descendants cleared2015: 2 descendants cleared2017: 2 descendants cleared2021: 2 descendants cleared2018: 3 descendants cleared2016: 1 descendant cleared2020: 2 descendants cleared2024: 2 descendants cleared2022: 1 descendant clearedK994055 · Sterrad Cyclesure Biological Indicator · Advanced Sterilization Products · 2002 · recalledK031226 · Sterrad Cyclesure Biological Indicator · Advanced Sterilization Products · 2003 · recalledK071014 · Modification to Sterrad Cyclesure Biological Indicator · Advanced Sterilization Products · 2007200020052010201520202025
Descendants, by generation 1 generation forward: 2 descendants2 generations forward: 1 descendant3 generations forward: 3 descendants4 generations forward: 5 descendants5 generations forward: 3 descendants6 generations forward: 2 descendants7 generations forward: 2 descendants8 generations forward: 1 descendant9 generations forward: 1 descendant K071014 · Modification to Sterrad Cyclesure Biological Indicator · Advanced Sterilization Products · 2007K031226 · Sterrad Cyclesure Biological Indicator · Advanced Sterilization Products · 2003 · recalledK994055 · Sterrad Cyclesure Biological Indicator · Advanced Sterilization Products · 2002 · recalled0+5
  1. Root · 3 generations backK994055 Sterrad Cyclesure Biological IndicatorAdvanced Sterilization Products · 2002 · Recalled
  2. 2 generations backK031226 Sterrad Cyclesure Biological IndicatorAdvanced Sterilization Products · 2003 · Recalled
  3. 1 generation backK071014 Modification to Sterrad Cyclesure Biological IndicatorAdvanced Sterilization Products · 2007
  4. This deviceK103222 Sterrad Cyclesure Biological IndicatorAdvanced Sterilization Products · 2011
  5. Descendant · cites it directly · 18 of its ownK111375 Sterrad (R) Cyclesure (R) 24 Biological IndicatorAdvanced Sterilization Products · 2012
  6. Descendant · 2 generations on · 17 of its ownK123017 Sterrad Cyclesure 24 Biological Indicator Model 14324Advanced Sterilization Products · 2012
  7. Descendant · 3 generations on · 12 of its ownK140392 3M Attest Rapid Readout Biological Indicator / 3M Attest Auto-Reader3M Company · 2014
  8. Descendant · 4 generations on · 11 of its ownK152060 3M Attest Rapid Readout Biological Indicator 12953M Company · 2015
  9. Descendant · 5 generations on · 6 of its ownK160546 3M Attest Rapid Readout Biological Indicator3M Company · 2016

Showing 5 of 20 descendants. See all descendants

Descendants cleared per year

2012: 32014: 22015: 22016: 12017: 22018: 32020: 22021: 22022: 12024: 2
2012high 3 (2012) · low 0 (2013)2026

Most influential descendants

Showing 5 of 20 · See all
  • K111375Sterrad (R) Cyclesure (R) 24 Biological IndicatorAdvanced Sterilization Products · 2012 · cites it directly18descendants
  • K123017Sterrad Cyclesure 24 Biological Indicator Model 14324Advanced Sterilization Products · 2012 · 2 generations on17descendants
  • K1403923M Attest Rapid Readout Biological Indicator / 3M Attest Auto-Reader3M Company · 2014 · 3 generations on12descendants
  • K1520603M Attest Rapid Readout Biological Indicator 12953M Company · 2015 · 4 generations on11descendants
  • K1605463M Attest Rapid Readout Biological Indicator3M Company · 2016 · 5 generations on6descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K994055Sterrad Cyclesure Biological IndicatorAdvanced Sterilization Products20021
−2K031226Sterrad Cyclesure Biological IndicatorAdvanced Sterilization Products20031
−1K071014Modification to Sterrad Cyclesure Biological IndicatorAdvanced Sterilization Products2007—
0 (this device)K103222Sterrad Cyclesure Biological IndicatorAdvanced Sterilization Products2011—
+1K111375Sterrad (R) Cyclesure (R) 24 Biological IndicatorAdvanced Sterilization Products2012—
+2K123017Sterrad Cyclesure 24 Biological Indicator Model 14324Advanced Sterilization Products2012—
+3K1403923M Attest Rapid Readout Biological Indicator / 3M Attest Auto-Reader3M Company2014—
+4K1520603M Attest Rapid Readout Biological Indicator 12953M Company2015—
+5K1605463M Attest Rapid Readout Biological Indicator3M Company2016—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201230
201420
201520
201610
201720
201830
202020
202120
202210
202420
FIG. 1 — Lineage of K103222, 2002–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FRC? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRC device

  1. #1 K250044 Celerity 5 HP Biological Indicator (LCB052)…score 14.5 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K071014Modification to Sterrad Cyclesure Biological Indicator
Summary page 1
…assification Regulation: 21CFR880.2200 Proprietary Name: STERRAD® CYCLESURE® 24 Biological indicator 2. PREDICATE DEVICES STERRAD® CycleSure® Biological indicator, K071014, May 24, 2007 3. INDICATIONS FOR USE The STERRAD® CYCLESURE® Biological Indicator is intended to be used as a standard meth…
Advanced Sterilization Products2007—

Also cleared against K071014: 3 other devices →

What descends from it? (20)

See all 20

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K111375Sterrad (R) Cyclesure (R) 24 Biological IndicatorAdvanced Sterilization Products201218
K122044Sterrad (R) Cyclesure (R) 24 Biological IndicatorAdvanced Sterilization Products20120

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 26, 2013Class IIZ-0604-2014Nonconforming Material/ComponentAdvanced Sterilization Products (ASP) is recalling the STERRAD Cyclesure Biological Indicator (BI) because ASP has identified a microscopic intermittent and low frequency flaw in the outer vial of the STERRAD Cyclesure 24 Biological…
Oct 3, 2012Class IIZ-0520-2013Nonconforming Material/ComponentAdvanced Sterilization Products (ASP) is recalling all lots of Sterrad Cyclesure 24 Biological Indicator (BI) manufactured between February 2008 and December 2011 because it may not have adequate data to support the entire duration of the…
Jun 28, 2012Class IIZ-2100-2012Incorrect or no expiration dateAdvanced Sterilization Products (ASP) is recalling certain lots of the STERRAD CYCLESURE 24 Biological Indicator (BI) product because it does not have adequate data to support the entire duration of the product's labeled shelf-life.
Feb 17, 2012Not classifiedZ-1773-2012Process controlThe recall was initiated by Advanced Sterilization Products because they recently determined that certain lots of raw materials that were rejected during incoming inspection acceptance test were used for the manufacture of a specific lot…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 9 years and passes through 1 company; a new generation was cleared every 3.8 years on median.
  • 2 of the 3 ancestors have a recall of any class.
  • Of its descendants, K111375 (Advanced Sterilization Products, 2012) has gone furthest, with 18 descendants of its own.
  • 3 other devices also cite its predicate K071014.

How does its category compare?

All FRC devices

FRC Indicator, Biological Sterilization Process: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.

FDA decisions per year · FRC

6in 2025

2007: 32008: 12009: 62010: 42011: 62012: 92013: 42014: 132015: 32016: 22017: 102018: 182019: 52020: 102021: 52022: 52023: 62024: 82025: 62026 (partial year): 1
2007high 18 (2018) · low 1 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRC

29in 2025

2007: 83 days median2009: 183 days median2010: 165 days median2011: 118 days median2012: 178 days median2013: 260 days median2014: 174 days median2015: 130 days median2017: 130 days median2018: 97 days median2019: 88 days median2020: 150 days median2021: 107 days median2022: 193 days median2023: 190 days median2024: 105 days median2025: 29 days median
2007high 260 (2013) · low 29 (2025)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRC

8so far in 2026, the most since 2007

2009: 12012: 22013: 22022: 42023: 12026 (partial year): 8
2007high 8 (2026) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FRC

9in 2025

2007: 1002008: 602009: 452010: 572011: 372012: 462013: 392014: 512015: 1822016: 4822017: 1822018: 222019: 302020: 162021: 102022: 92023: 202024: 222025: 9
2007high 482 (2016) · low 9 (2022)2025

MAUDE reports naming product code FRC, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20073830100
20081—060
20096183145
20104165057
20116118037
20129178246
20134260239
201413174051
201531300182
20162—0482
2017101300182
20181897022
2019588030
202010150016
20215107010
2022519349
20236190120
20248105022
202562909
20261—8—

What is it indicated for?

The STERRAD® CYCLESURE® 24 Biological Indicator is intended to be used as a standard method for frequent monitoring of the following STERRAD® Sterilization Systems and Cycles:

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K103222: Sterrad Cyclesure Biological Indicator.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K103222

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026