510(k) Report

510(k) · Cleared

Cleared

STERRAD NX Sterilizer

K162007510(k) SpecialCleared Feb 16, 2017211-day reviewClass II21 CFR 880.6860

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Advanced Sterilization Products · MLR Sterilizer, Chemical

Cleared on Feb 16, 2017 as substantially equivalent to K042116 (Advanced Sterilization Products, 2005) and 2 other predicates after a 211-day review. Its family tree reaches back 5 generations to 2005.

Decided
Feb 16, 2017
Review
211 days
Class
Class II
Product code
MLR
Recalls
None
Predicates
3

Its longest line of descent runs back 5 generations to K042116 (2005). 10 ancestors in all, the oldest K981625 (1999). No later device cites it as a predicate yet.

  1. Root · 5 generations backK042116 Sterrad NX SterilizerAdvanced Sterilization Products · 2005 · Recalled
  2. 4 generations backK071385 Sterrad 100NX SterilizerAdvanced Sterilization Products · 2007 · Recalled
  3. 3 generations backK092622 Sterrad 100NX Sterilizer Express CycleAdvanced Sterilization Products · 2011
  4. 2 generations backK142454 STERRAD NX Sterilizer, STERRAD 100NX SterilizerAdvanced Sterilization Products · 2015 · Recalled
  5. 1 generation backK151725 STERRAD NX Sterilizer, STERRAD 100NX SterilizerAdvanced Sterilization Products · 2016 · Recalled
  6. This deviceK162007 STERRAD NX SterilizerAdvanced Sterilization Products · 2017
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K042116Sterrad NX SterilizerAdvanced Sterilization Products200510
−4K071385Sterrad 100NX SterilizerAdvanced Sterilization Products20075
−3K092622Sterrad 100NX Sterilizer Express CycleAdvanced Sterilization Products2011—
−2K142454STERRAD NX Sterilizer, STERRAD 100NX SterilizerAdvanced Sterilization Products20152
−1K151725STERRAD NX Sterilizer, STERRAD 100NX SterilizerAdvanced Sterilization Products20161
0 (this device)K162007STERRAD NX SterilizerAdvanced Sterilization Products2017—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199910
200010
200320
201210
FIG. 1 — Lineage of K162007, 2005–2017Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MLR? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MLR device

  1. #1 K250802 STERRAD® 100NX Sterilizer with ALLClear™…score 14.0 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K042116Sterrad NX Sterilizer
Summary page 5
…erilizer Device Class: Class II Product Code: MLR CFR Section: 21 CFR 880.6860 Predicate Device STERRAD NX Sterilizer originally cleared under K042116. Description The STERRAD NX Sterilizer is a self-contained, stand-alone system of hardware and software designed to sterilize m…
Advanced Sterilization Products200510
*K142454STERRAD NX Sterilizer, STERRAD 100NX Sterilizer
Summary page 9
…Predicate STERRAD NX Sterilizer Modified Characteristic K042116, K142454 and K151725 STERRAD NX Sterilizer Intended Use Designed for sterilization of both metal and…
Advanced Sterilization Products20152
*K151725STERRAD NX Sterilizer, STERRAD 100NX Sterilizer
Summary page 9
…Predicate STERRAD NX Sterilizer Modified Characteristic K042116, K142454 and K151725 STERRAD NX Sterilizer Intended Use Designed for sterilization of both metal and Sam…
Advanced Sterilization Products20161

Also cleared against K042116: 4 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 12 years and passes through 1 company; a new generation was cleared every 2.6 years on median.
  • 9 of the 10 ancestors have a recall of any class.
  • 4 other devices also cite its predicate K042116.

How does its category compare?

All MLR devices

MLR Sterilizer, Chemical: FDA decisions, review times, recalls and adverse event reports over 20 years. 2017, this device’s year, is highlighted.

FDA decisions per year · MLR

1in 2025

2007: 22009: 12011: 62012: 22013: 12014: 22015: 12016: 42017: 22018: 22019: 42021: 22022: 22023: 52024: 32025: 12026 (partial year): 1
2007high 6 (2011) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MLR

216in 2024

2011: 214 days median2016: 183 days median2019: 90 days median2023: 222 days median2024: 216 days median
2007high 222 (2023) · low 90 (2019)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MLR

1in 2017

2007: 52008: 82009: 22010: 12011: 22012: 32013: 92014: 32017: 1
2007high 9 (2013) · low 0 (2015)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MLR

46in 2025

2007: 2722008: 5362009: 4332010: 2112011: 342012: 1012013: 5412014: 3992015: 4122016: 2862017: 2792018: 2892019: 2222020: 1742021: 2022022: 1882023: 1402024: 442025: 46
2007high 541 (2013) · low 34 (2011)2025

MAUDE reports naming product code MLR, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—5272
20080—8536
20091—2433
20100—1211
20116214234
20122—3101
20131—9541
20142—3399
20151—0412
201641830286
20172—1279
20182—0289
20194900222
20200—0174
20212—0202
20222—0188
202352220140
20243216044
20251—046
20261—0—

What is it indicated for?

The STERRAD NX Sterilizer is designed for sterilization of both metal and nonmetal medical devices at low temperatures. The STERRAD sterilization process is a multiphase sterilization process that utilizes a combination of exposure to hydrogen peroxide vapor and plasma to affect sterilization. The STERRAD NX Sterilizer can sterilize instruments which have diffusion-restricted spaces, such as the hinged portion of forceps and scissors. Medical devices…

Read the full statement (281 words)

The STERRAD NX Sterilizer is designed for sterilization of both metal and nonmetal medical devices at low temperatures. The STERRAD sterilization process is a multiphase sterilization process that utilizes a combination of exposure to hydrogen peroxide vapor and plasma to affect sterilization. The STERRAD NX Sterilizer can sterilize instruments which have diffusion-restricted spaces, such as the hinged portion of forceps and scissors.

Medical devices with the following materials and dimensions can be processed in the STERRAD NX Sterilizer STANDARD cycle:

Single channel stainless steel lumens with

• An inside diameter of 1 mm or larger and a length of 150 mm or shorter†

• An inside diameter of 2 mm or larger and a length of 400 mm or shorter†

Medical devices, including most flexible endoscopes, with the following materials and dimensions can be processed in the STERRAD NX Sterilizer ADVANCED cycle:

Single channel stainless steel lumens with

• An inside diameter of 1 mm or larger and a length of 500 mm or shorter†

Single channel polyethylene and Teflon (polytetrafluoroethylene) flexible endoscope with

• An inside diameter of 1 mm or larger and length of 850 mm or shorter*

Note: With the exception of the 1 x 850 mm flexible endoscopes, validation studies were performed using a validation load consisting of one instrument tray weighing 10.7 lbs. The 1 x 850 mm flexible endoscope was validated without any additional load.

†The validation testing for this lumen size was conducted using a maximum of 10 lumens per load. Hospital loads should not exceed the maximum number of lumens validated by this testing.

*Only one flexible endoscope can be processed per sterilization cycle with or without a silicone mat. No additional load.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K162007: STERRAD NX Sterilizer.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K162007

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026