510(k) · Cleared
Recalled · Class II · 20143M Attest 1491 Super Rapid Readout Biological Indicator; 3M Attest 490 Auto-Reader
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
3M Company-3M Health Care · FRC Indicator, Biological Sterilization Process
Cleared on Apr 19, 2011 as substantially equivalent to K900771 (3M Company, 1990) and 1 other predicate after a 165-day review. Its family tree reaches back 1 generation to 1990. 35 later devices descend from it. It has been recalled once.
- Decided
- Apr 19, 2011
- Review
- 165 days
- Class
- Class II
- Product code
- FRC
- Recalls
- 1
- Predicates
- 2
Its longest line of descent runs back 1 generation to K900771 (1990). 2 ancestors in all. 35 later devices descend from it, 5 citing it directly.
- Root · 1 generation backK900771 Attest Brand Rapid Readout Biological Indicator3M Company · 1990
- This deviceK103277 3M Attest 1491 Super Rapid Readout Biological Indicator; 3M Attest…3M Company-3M Health Care · 2011
- Descendant · cites it directly · 31 of its ownK121484 3M Attest (TM) Super Rapid Readout Biological Indicator, 3M Attest…3M Company · 2012
- Descendant · cites it directly · 12 of its ownK140392 3M Attest Rapid Readout Biological Indicator / 3M Attest Auto-Reader3M Company · 2014
- Descendant · 2 generations on · 11 of its ownK152060 3M Attest Rapid Readout Biological Indicator 12953M Company · 2015
- Descendant · 2 generations on · 11 of its ownK121593 3M Attest (TM) Super Radid Readout Steam Challenge Pack 3M Attest…3M Company · 2013
- Descendant · 3 generations on · 7 of its ownK173519 3M Attest Super Rapid Readout Steam Challenge Pack, 3M Attest Super…3M Company · 2018 · Recalled
Showing 5 of 35 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 35 · See all- K1214843M Attest (TM) Super Rapid Readout Biological Indicator, 3M Attest (TM) Auto-Reader3M Company · 2012 · cites it directly31descendants
- K1403923M Attest Rapid Readout Biological Indicator / 3M Attest Auto-Reader3M Company · 2014 · cites it directly12descendants
- K1520603M Attest Rapid Readout Biological Indicator 12953M Company · 2015 · 2 generations on11descendants
- K1215933M Attest (TM) Super Radid Readout Steam Challenge Pack 3M Attest (TM) Super Rapid 5…3M Company · 2013 · 2 generations on11descendants
- K1735193M Attest Super Rapid Readout Steam Challenge Pack, 3M Attest Super Rapid 5 Steam-Plus…3M Company · 2018 · 3 generations on7descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K900771 | Attest Brand Rapid Readout Biological Indicator | 3M Company | 1990 | — |
| 0 (this device) | K103277 | 3M Attest 1491 Super Rapid Readout Biological Indicator; 3M Attest 490… | 3M Company-3M Health Care | 2011 | — |
| +1 | K121484 | 3M Attest (TM) Super Rapid Readout Biological Indicator, 3M Attest (TM)… | 3M Company | 2012 | — |
| +1 | K140392 | 3M Attest Rapid Readout Biological Indicator / 3M Attest Auto-Reader | 3M Company | 2014 | — |
| +2 | K152060 | 3M Attest Rapid Readout Biological Indicator 1295 | 3M Company | 2015 | — |
| +2 | K121593 | 3M Attest (TM) Super Radid Readout Steam Challenge Pack 3M Attest (TM) Super… | 3M Company | 2013 | — |
| +3 | K173519 | 3M Attest Super Rapid Readout Steam Challenge Pack, 3M Attest Super Rapid 5… | 3M Company | 2018 | 1 |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2002 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2012 | 1 | 0 |
| 2013 | 1 | 0 |
| 2014 | 1 | 0 |
| 2015 | 1 | 0 |
| 2016 | 2 | 0 |
| 2017 | 3 | 0 |
| 2018 | 9 | 0 |
| 2020 | 8 | 0 |
| 2021 | 2 | 0 |
| 2022 | 2 | 0 |
| 2024 | 5 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in FRC? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new FRC device
- #1 K250044 Celerity 5 HP Biological Indicator (LCB052)…score 14.5 · 10 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (2)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K900771 | Attest Brand Rapid Readout Biological Indicator | 1990 |
| * | K004009 | 3M Attest 290 Auto-ReaderSummary page 5…t to the predicate devices, the 3MTM AtteStTM 1291 Rapid Readout Biological Indicator for Steam, cleared under K90077 I, and 3MTM AttcStrM 290 Auto-reader, cleared under K004009, in terms of their intended use, physical properties and technological characteristics. There are no new questions of safety or e… | 2002 |
What descends from it? (35)
See all 355 later devices cite it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K121484 | 3M Attest (TM) Super Rapid Readout Biological Indicator, 3M Attest (TM) Auto-Reader | 2012 | 31 |
| K140392 | 3M Attest Rapid Readout Biological Indicator / 3M Attest Auto-Reader | 2014 | 12 |
| K202721 | Celerity 20 Steam Biological Indicator for IUSS | 2021 | 0 |
| K173584 | 3M Attest Super Rapid Readout Biological Indicator, 3M Attest Auto-reader, 3M Attest… | 2018 | 0 |
| K151482 | BCR-30-min BI™ and RBIP-ST30™ Readout Accessory | 2016 | 0 |
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Jun 9, 2014 | Class II | Z-2002-2014 | Labeling Change Control | Labeling on units shipped prior to May 16th, 2014 did not contain the statement "This product contains dry natural rubber" as required by the United States Food and Drug Administration (FDA) for medical devices. The non-slip pads at the… |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- None of the 2 ancestors has a recall.
- Of its descendants, K121484 (3M Company, 2012) has gone furthest, with 31 descendants of its own.
How does its category compare?
All FRC devicesFRC Indicator, Biological Sterilization Process: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.
FDA decisions per year · FRC
6in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · FRC
29in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · FRC
8so far in 2026, the most since 2007
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · FRC
9in 2025
MAUDE reports naming product code FRC, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 3 | 83 | 0 | 100 |
| 2008 | 1 | — | 0 | 60 |
| 2009 | 6 | 183 | 1 | 45 |
| 2010 | 4 | 165 | 0 | 57 |
| 2011 | 6 | 118 | 0 | 37 |
| 2012 | 9 | 178 | 2 | 46 |
| 2013 | 4 | 260 | 2 | 39 |
| 2014 | 13 | 174 | 0 | 51 |
| 2015 | 3 | 130 | 0 | 182 |
| 2016 | 2 | — | 0 | 482 |
| 2017 | 10 | 130 | 0 | 182 |
| 2018 | 18 | 97 | 0 | 22 |
| 2019 | 5 | 88 | 0 | 30 |
| 2020 | 10 | 150 | 0 | 16 |
| 2021 | 5 | 107 | 0 | 10 |
| 2022 | 5 | 193 | 4 | 9 |
| 2023 | 6 | 190 | 1 | 20 |
| 2024 | 8 | 105 | 0 | 22 |
| 2025 | 6 | 29 | 0 | 9 |
| 2026 | 1 | — | 8 | — |
What is it indicated for?
Use the 3M Attest 1491 Super Rapid Readout Biological Indicator in conjunction with the 3M Attest 490 Auto-reader to monitor the cycles below.
Sterilization Type: Gravity Displacement; Temperature: 270°F (132°C); Time: 3 minutes
Immediate Use Steam Sterilization Cycle (Flash); Temperature: 270°F (132°C); Time: 10 minutes
Temperature: 275°F (135°C); Time: 3 minutes
Temperature: 275°F (135°C); Time: 10 minutes
The 3M Attest 1491 Super Rapid Readout Biological Indicator provides a final fluorescent result in 30 minutes. An optional visual pH color change result is observed in 24 hours.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K103277 · printed Nov 6, 2026
Cite this page
510(k) Report. “K103277: 3M Attest 1491 Super Rapid Readout Biological Indicator; 3M Attest 490 Auto-Reader.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K103277
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026