510(k) Report

510(k) · Cleared

Recalled · Class II · 2022

Luminos Agile

K111292510(k) SpecialCleared Jun 3, 201128-day reviewClass II21 CFR 892.1650Recalls 6

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Siemens Medical Solutions USA, Inc. · JAA System, X-Ray, Fluoroscopic, Image-Intensified

Cleared on Jun 3, 2011 as substantially equivalent to K051602 (Siemens Medical Systems, Inc., 2005) and 1 other predicate after a 28-day review. Its family tree reaches back 8 generations to 1980. 2 later devices descend from it. It has been recalled 6 times.

Decided
Jun 3, 2011
Review
28 days
Class
Class II
Product code
JAA
Recalls
6
Predicates
2

Its longest line of descent runs back 8 generations to K801623 (1980). 57 ancestors in all. 2 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2005: 1 other ancestor2002: 4 other ancestors1999: 3 other ancestors1995: 12 other ancestors2001: 1 other ancestor2000: 3 other ancestors1997: 4 other ancestors2004: 2 other ancestors1996: 4 other ancestors1993: 3 other ancestors1989: 1 other ancestor1986: 1 other ancestor1992: 2 other ancestors1991: 4 other ancestors1988: 2 other ancestors1981: 1 other ancestor1994: 1 other ancestor2018: 1 descendant cleared2025: 1 descendant clearedK801623 · Cavitron Ultrasonic Surg. Aspir. Cusa · Cavitron Corp. · 1980K962952 · Olympus Ultrasonic Surgical System (Usu) · Olympus Optical, Co. · 1996×2: 2 generations cleared close together, K972114 (1998), K981993 (1998) K994231 · Siemens Integrated Operating System (Sios) · Siemens Medical Corp. · 2000×2: 2 generations cleared close together, K052202 (2006), K061054 (2006) K062623 · Axiom Luminos DRF · Siemens Medical Solutions USA, Inc. · 2007 · recalled19801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 3 other ancestors3 generations back: 4 other ancestors4 generations back: 8 other ancestors5 generations back: 7 other ancestors6 generations back: 13 other ancestors7 generations back: 9 other ancestors8 generations back: 4 other ancestors 1 generation forward: 1 descendant2 generations forward: 1 descendant K062623 · Axiom Luminos DRF · Siemens Medical Solutions USA, Inc. · 2007 · recalledK061054 · Axiom Aristos FX Plus · Siemens Medical Solutions USA, Inc. · 2006 · recalledK052202 · Axiom Artis - Modular Angiographic System · Siemens Medical Solutions USA, Inc. · 2006 · recalledK994231 · Siemens Integrated Operating System (Sios) · Siemens Medical Corp. · 2000K981993 · Olympus Integrated Endosurgery System Endoalpha · The Olympus Optical Co. · 1998K972114 · Olympus Sonosurg System · The Olympus Optical Co. · 1998K962952 · Olympus Ultrasonic Surgical System (Usu) · Olympus Optical, Co. · 1996K801623 · Cavitron Ultrasonic Surg. Aspir. Cusa · Cavitron Corp. · 1980−50
  1. Root · 8 generations backK801623 Cavitron Ultrasonic Surg. Aspir. CusaCavitron Corp. · 1980
  2. 7 generations backK962952 Olympus Ultrasonic Surgical System (Usu)Olympus Optical, Co. · 1996
  3. 6 generations backK972114 Olympus Sonosurg SystemThe Olympus Optical Co. · 1998
  4. 5 generations backK981993 Olympus Integrated Endosurgery System EndoalphaThe Olympus Optical Co. · 1998
  5. 4 generations backK994231 Siemens Integrated Operating System (Sios)Siemens Medical Corp. · 2000
  6. 3 generations backK052202 Axiom Artis - Modular Angiographic SystemSiemens Medical Solutions USA, Inc. · 2006 · Recalled
  7. 2 generations backK061054 Axiom Aristos FX PlusSiemens Medical Solutions USA, Inc. · 2006 · Recalled
  8. 1 generation backK062623 Axiom Luminos DRFSiemens Medical Solutions USA, Inc. · 2007 · Recalled
  9. This deviceK111292 Luminos AgileSiemens Medical Solutions USA, Inc. · 2011
  10. Descendant · cites it directly · 1 of its ownK173639 Luminos dRF Max, Luminos Agile Max, Uroskop Omnia Max, Multitom RaxSiemens Medi Cal Solutions, Inc. · 2018 · Recalled
  11. Descendant · 2 generations on · 0 of its ownK250660 LUMINOS Q.namix T; LUMINOS Q.namix RSiemens Medical Solutions · 2025 · Recalled

Most influential descendants

Showing 2 of 2 · See all
  • K173639Luminos dRF Max, Luminos Agile Max, Uroskop Omnia Max, Multitom RaxSiemens Medi Cal Solutions, Inc. · 2018 · cites it directly1descendant
  • K250660LUMINOS Q.namix T; LUMINOS Q.namix RSiemens Medical Solutions · 2025 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−8K801623Cavitron Ultrasonic Surg. Aspir. CusaCavitron Corp.1980—
−7K962952Olympus Ultrasonic Surgical System (Usu)Olympus Optical, Co.1996—
−6K972114Olympus Sonosurg SystemThe Olympus Optical Co.1998—
−5K981993Olympus Integrated Endosurgery System EndoalphaThe Olympus Optical Co.1998—
−4K994231Siemens Integrated Operating System (Sios)Siemens Medical Corp.2000—
−3K052202Axiom Artis - Modular Angiographic SystemSiemens Medical Solutions USA, Inc.200629
−2K061054Axiom Aristos FX PlusSiemens Medical Solutions USA, Inc.20063
−1K062623Axiom Luminos DRFSiemens Medical Solutions USA, Inc.200720
0 (this device)K111292Luminos AgileSiemens Medical Solutions USA, Inc.2011—
+1K173639Luminos dRF Max, Luminos Agile Max, Uroskop Omnia Max, Multitom RaxSiemens Medi Cal Solutions, Inc.201823
+2K250660LUMINOS Q.namix T; LUMINOS Q.namix RSiemens Medical Solutions20252

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198110
198610
198820
198910
199140
199220
199330
199410
1995120
199640
199740
199930
200030
200110
200240
200420
200510

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201810
202510
FIG. 1 — Lineage of K111292, 1980–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JAA? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JAA device

  1. #1 K253103 Nexus DRF Digital X-ray Imaging Systemscore 13.2 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K051602Axiom Sireskop SD
Summary page 2
…commercially available Siemens systems, the AXIOM Luminos TF (AXIOM Sireskop SD) and the AXIOM Luminos dRF. The AXIOM Sireskop SD was described in premarket notification K051602 which received FDA Clearance on 07/07/2005. The AXIOM Sireskop SD was renamed later to AXIOM Luminos TF for marketing reasons. Th…
Siemens Medical Systems, Inc.20055
*K062623Axiom Luminos DRF
Summary page 2
…spot COMPACT digital imaging system. The table design remains unchanged while the new imaging chain is based on the AXIOM Luminos dRF described in premarket notification K062623 which received FDA Clearance on August 22. 2007. 5. Intended Use: The Luminos Agile is intended to be used as a universal diagno…
Siemens Medical Solutions USA, Inc.200720

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K173639Luminos dRF Max, Luminos Agile Max, Uroskop Omnia Max, Multitom RaxSiemens Medi Cal Solutions, Inc.20181

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 9, 2022Class IIZ-1001-2023Process controlThere is potential for the footrest to detach from the patient table during use.
Apr 13, 2015Class IIZ-1669-2015Software designStudy and all acquired images deleted when using systems with software version VE10E. Acquisitions have to be repeated. Connection with portable detectors (MAX wi-D, MAX mini) is sporadic in image recovery. Pressing undo button on the RAD…
Dec 15, 2014Class IIZ-0994-2015Software designPotential for composed images to be flipped before being sent to PACS on systems with software version VE10 and SmartOrtho license. Flipped images may be reversed so the associated annotations, e.g. labels (R/L), may be displayed…
Sep 2, 2014Class IIZ-0005-2015Software designTwo malfunctions posing potential risk to patients were identified with Siemens Ysio Max, AXIOM Luminos dRF Max, Uroskop Omnia Max, and Luminos Agile Max systems with a specific set of serial numbers that will cause lost images due to an…
Sep 3, 2013Class IIZ-0374-2014Software designA potential malfunction and hazard to patients exists when using the Ysio, Luminos Agile, Luminos dRF, and Uroskop Omnia systems with software version VC10, which may result in radiation continuing longer than required and overexposed…
Feb 15, 2012Not classifiedZ-1159-2012Software designFirm has become aware of unintended behavior when using Ysio, AXIOM dRF, AXIOM Luminos Agile/TF and Uroskop Omnia with RAD Fluoro Uro function using software version VC10A. Due to an error in the image pipeline, images may display dark…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 31 years and passes through 5 companies; a new generation was cleared every 2.0 years on median.
  • 19 of the 57 ancestors have a recall of any class.
  • Of its descendants, K173639 (Siemens Medi Cal Solutions, Inc., 2018) has gone furthest, with 1 descendant of its own.

How does its category compare?

All JAA devices

JAA System, X-Ray, Fluoroscopic, Image-Intensified: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.

FDA decisions per year · JAA

2in 2025

2007: 12008: 62010: 12011: 42012: 12013: 62014: 62015: 32016: 22017: 12018: 32019: 62020: 42021: 62022: 12023: 22024: 32025: 22026 (partial year): 1
2007high 6 (2008) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JAA

89in 2024

2008: 18 days median2011: 48 days median2013: 129 days median2014: 276 days median2015: 197 days median2018: 30 days median2019: 49 days median2020: 39 days median2021: 104 days median2024: 89 days median
2007high 276 (2014) · low 18 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JAA

4in 2025

2007: 352008: 212009: 182010: 62011: 72012: 22013: 32014: 12015: 12016: 72017: 32018: 12019: 22020: 22021: 22022: 52023: 92024: 22025: 4
2007high 35 (2007) · low 1 (2014)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JAA

33in 2025

2007: 9,5172008: 9,4842009: 12,0402010: 5,8032011: 14,0002012: 11,6842013: 13,7062014: 10,2652015: 2,9222016: 352017: 562018: 362019: 492020: 392021: 352022: 342023: 412024: 252025: 33
2007high 14,000 (2011) · low 25 (2024)2025

MAUDE reports naming product code JAA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—359,517
2008618219,484
20090—1812,040
20101—65,803
2011448714,000
20121—211,684
20136129313,706
20146276110,265
2015319712,922
20162—735
20171—356
2018330136
2019649249
2020439239
20216104235
20221—534
20232—941
2024389225
20252—433
20261—0—

What is it indicated for?

The Luminos Agile is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic studies. Using either film cassettes or a digital mobile flat detector, it can perform a range of applications including general R/F, angiography and pediatric examinations.

Luminos Agile may be used for emergency treatment on an outpatient basis, as well as for bedside examinations.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K111292: Luminos Agile.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K111292

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026