510(k) Report

510(k) · Decided

Recalled · 2008

Cannon II Plus

K040078510(k) SpecialDecided Feb 13, 200430-day reviewClass II21 CFR 876.5540Recalls 19

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Arrow International, Inc. · MSD Catheter, Hemodialysis, Implanted

Cleared on Feb 13, 2004 as substantially equivalent to K010399 (Diatek, Inc., 2001) after a 30-day review. Its family tree reaches back 1 generation to 2001. 3 of the 15 devices in its family, counting itself, have a Class I recall. 13 later devices descend from it. It has been recalled 19 times.

Decided
Feb 13, 2004
Review
30 days
Class
Class II
Product code
MSD
Recalls
19
Predicates
1

Its longest line of descent runs back 1 generation to K010399 (2001). 13 later devices descend from it, 2 citing it directly. 3 of the 15 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2009: 1 descendant cleared2004: 1 descendant cleared2011: 3 descendants cleared, 2 with a Class I recall22019: 1 descendant cleared2013: 1 descendant cleared, 1 with a Class I recall12014: 1 descendant cleared2021: 1 descendant cleared2017: 1 descendant cleared2025: 1 descendant cleared2020: 1 descendant cleared2018: 1 descendant clearedK010399 · Diatek Cannon-Cath #CC2400, CC2800, CC3200, CC3600,CC5500. CC2400PC, CC2800PC, CC3200PC (32 CM), CC3200PC (36 CM) · Diatek, Inc. · 2001 · recalled200020052010201520202025
Descendants, by generation 1 generation forward: 2 descendants2 generations forward: 2 descendants, 1 with a Class I recall13 generations forward: 3 descendants, 2 with a Class I recall24 generations forward: 2 descendants5 generations forward: 2 descendants6 generations forward: 2 descendants K010399 · Diatek Cannon-Cath #CC2400, CC2800, CC3200, CC3600,CC5500. CC2400PC, CC2800PC, CC3200PC (32 CM), CC3200PC (36 CM) · Diatek, Inc. · 2001 · recalled0+5
  1. Root · 1 generation backK010399 Diatek Cannon-Cath #CC2400, CC2800, CC3200, CC3600,CC5500. CC2400PC…Diatek, Inc. · 2001 · Recalled
  2. This deviceK040078 Cannon II PlusArrow International, Inc. · 2004
  3. Descendant · cites it directly · 11 of its ownK092205 Palindrome P Hemodialysis CatheterCovidien Lp, Formerly Registered AS Kendall · 2009
  4. Descendant · 2 generations on · 10 of its ownK111372 Tal Pallindrome (TM) Symmetric Tip Dual Lumen Catheter, Mahurkar (R)…Covidien · 2011 · Recalled (Class I)
  5. Descendant · 3 generations on · 5 of its ownK112477 Palindrome H, SI, Hsi Chronic CatheterCovidien · 2011 · Recalled (Class I)
  6. Descendant · 3 generations on · 4 of its ownK123196 Palindrome Precision H Chronic Catheter, Palindrome Precision SI…Covidien · 2013 · Recalled (Class I)
  7. Descendant · 4 generations on · 3 of its ownK140884 Symetrex Long Term Hemodialysis CatheterPhase One Medical · 2014

Showing 5 of 13 descendants. See all descendants

Descendants cleared per year

2004: 12009: 12011: 32013: 12014: 12017: 12018: 12019: 12020: 12021: 12025: 1
2004high 3 (2011) · low 0 (2005)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 13 · See all
  • K092205Palindrome P Hemodialysis CatheterCovidien Lp, Formerly Registered AS Kendall · 2009 · cites it directly11descendants
  • K111372Tal Pallindrome (TM) Symmetric Tip Dual Lumen Catheter, Mahurkar (R) Maxid (TM) Dual…Covidien · 2011 · 2 generations on10descendants
  • K112477Palindrome H, SI, Hsi Chronic CatheterCovidien · 2011 · 3 generations on5descendants
  • K123196Palindrome Precision H Chronic Catheter, Palindrome Precision SI Chronic Catheter…Covidien · 2013 · 3 generations on4descendants
  • K140884Symetrex Long Term Hemodialysis CatheterPhase One Medical · 2014 · 4 generations on3descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K010399Diatek Cannon-Cath #CC2400, CC2800, CC3200, CC3600,CC5500. CC2400PC, CC2800PC…Diatek, Inc.20012
0 (this device)K040078Cannon II PlusArrow International, Inc.2004—
+1K092205Palindrome P Hemodialysis CatheterCovidien Lp, Formerly Registered AS Kendall2009—
+2K111372Tal Pallindrome (TM) Symmetric Tip Dual Lumen Catheter, Mahurkar (R) Maxid…Covidien2011Class I
+3K112477Palindrome H, SI, Hsi Chronic CatheterCovidien2011Class I
+3K123196Palindrome Precision H Chronic Catheter, Palindrome Precision SI Chronic…Covidien2013Class I
+4K140884Symetrex Long Term Hemodialysis CatheterPhase One Medical2014—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200410
200910
201132
201311
201410
201710
201810
201910
202010
202110
202510
FIG. 1 — Lineage of K040078, 2001–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MSD device

  1. #1 K241581 Hemodialysis Catheterscore 12.3 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K010399Diatek Cannon-Cath #CC2400, CC2800, CC3200, CC3600,CC5500. CC2400PC, CC2800PC, CC3200PC…
Summary page 1
…e classification MSD, Class III, 21 CFR Part 876.5540, Catheter, Hemodialysis, name: Implanted Legally marketed Arrow Cannon Catheter TM (K010399). devices to which the device is substantially equivalent: Description of the The Arrow Cannon TM 11 Plus consists of a do…
Diatek, Inc.20012

Also cleared against K010399: 2 other devices →

What descends from it? (13)

See all 13

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K092205Palindrome P Hemodialysis CatheterCovidien Lp, Formerly Registered AS Kendall200911
K040666Step-Tip Chronic Hemodialysis CatheterArrow International, Inc.20040

Has anything in its family been recalled? (3 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 10, 2008Not classifiedZ-2267-2008Device DesignThe tips may not have been adequately welded to the catheter body.
Jun 10, 2008Not classifiedZ-2264-2008Device DesignThe tips may not have been adequately welded to the catheter body.
Jun 10, 2008Not classifiedZ-2253-2008Device DesignThe tips may not have been adequately welded to the catheter body.
Jun 10, 2008Not classifiedZ-2257-2008Device DesignThe tips may not have been adequately welded to the catheter body.
Jun 10, 2008Not classifiedZ-2258-2008Device DesignThe tips may not have been adequately welded to the catheter body.
Jun 10, 2008Not classifiedZ-2263-2008Device DesignThe tips may not have been adequately welded to the catheter body.
Jun 10, 2008Not classifiedZ-2262-2008Device DesignThe tips may not have been adequately welded to the catheter body.
Jun 10, 2008Not classifiedZ-2256-2008Device DesignThe tips may not have been adequately welded to the catheter body.
Jun 10, 2008Not classifiedZ-2261-2008Device DesignThe tips may not have been adequately welded to the catheter body.
Jun 10, 2008Not classifiedZ-2255-2008Device DesignThe tips may not have been adequately welded to the catheter body.
Jun 10, 2008Not classifiedZ-2268-2008Device DesignThe tips may not have been adequately welded to the catheter body.
Jun 10, 2008Not classifiedZ-2254-2008Device DesignThe tips may not have been adequately welded to the catheter body.
Jun 10, 2008Not classifiedZ-2269-2008Device DesignThe tips may not have been adequately welded to the catheter body.
Jun 10, 2008Not classifiedZ-2259-2008Device DesignThe tips may not have been adequately welded to the catheter body.
Jun 10, 2008Not classifiedZ-2270-2008Device DesignThe tips may not have been adequately welded to the catheter body.
Jun 10, 2008Not classifiedZ-2266-2008Device DesignThe tips may not have been adequately welded to the catheter body.
Jun 10, 2008Not classifiedZ-2252-2008Device DesignThe tips may not have been adequately welded to the catheter body.
Jun 10, 2008Not classifiedZ-2265-2008Device DesignThe tips may not have been adequately welded to the catheter body.
Jun 10, 2008Not classifiedZ-2260-2008Device DesignThe tips may not have been adequately welded to the catheter body.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K111372Tal Pallindrome (TM) Symmetric Tip Dual Lumen Catheter, Mahurkar (R) Maxid (TM) Dual Lumen Catheter…Covidien · 2011 · 2 generations on · ◆ in the 2011 descendants bar of FIG. 12 Class I recalls
  • K112477Palindrome H, SI, Hsi Chronic CatheterCovidien · 2011 · 3 generations on · ◆ in the 2011 descendants bar of FIG. 13 Class I recalls
  • K123196Palindrome Precision H Chronic Catheter, Palindrome Precision SI Chronic Catheter, Palindrome Precision Hsi…Covidien · 2013 · 3 generations on · ◆ in the 2013 descendants bar of FIG. 14 Class I recalls

What does this lineage say?

  • Its one ancestor has a recall.
  • Of its descendants, K092205 (Covidien Lp, Formerly Registered AS Kendall, 2009) has gone furthest, with 11 descendants of its own.
  • 2 other devices also cite its predicate K010399.

How does its category compare?

All MSD devices

MSD Catheter, Hemodialysis, Implanted: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · MSD

1in 2025

2007: 72008: 32009: 62010: 42011: 72012: 22013: 52014: 82015: 12016: 22017: 22018: 22019: 12020: 42021: 42025: 1
2007high 8 (2014) · low 0 (2022)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MSD

153in 2021

2007: 92 days median2008: 89 days median2009: 81 days median2010: 116 days median2011: 70 days median2013: 105 days median2014: 246 days median2020: 106 days median2021: 153 days median
2007high 246 (2014) · low 70 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MSD

1in 2013

2008: 472009: 492010: 112011: 242013: 12026 (partial year): 4, 4 Class I
2007high 49 (2009) · low 0 (2007)2026

Class I (4 since 2007) Class II/III not classified

Adverse event reports per year · MSD

1,054in 2025

2007: 1182008: 1742009: 1732010: 3062011: 2682012: 3672013: 3632014: 5062015: 5502016: 5322017: 4842018: 4632019: 6452020: 4012021: 5232022: 6102023: 5422024: 8412025: 1,054
2007high 1,054 (2025) · low 118 (2007)2025

MAUDE reports naming product code MSD, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20077920118
200838947174
200968149173
2010411611306
201177024268
20122—0367
201351051363
201482460506
20151—0550
20162—0532
20172—0484
20182—0463
20191—0645
202041060401
202141530523
20220—0610
20230—0542
20240—0841
20251—01,054
20260—4 (4)—

What is it indicated for?

The Arrow CannonTM IIPlus is indicated for use in attaining long-term vascular access for hemodialysis and apheresis. The Cannon TM II Plus is inserted percutaneously and is preferentially placed into the internal jugular (IJ) vein. Alternately, this catheter may be inserted into the subclavian vein although the jugular vein is the preferred site. Catheters greater than 40 cm are intended for femoral vein insertion. The CannonTM II Plus is intended for use in adult patients.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K040078: Cannon II Plus.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K040078

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026