510(k) Report

510(k) · Decided

Recalled · Class II · 2021

Palindrome RT Chronic Catheter

K111817510(k) SpecialDecided Aug 30, 201164-day reviewClass II21 CFR 876.5540Recalls 13

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Covidien · MSD Catheter, Hemodialysis, Implanted

Cleared on Aug 30, 2011 as substantially equivalent to K092205 (Covidien Lp, Formerly Registered AS Kendall, 2009) and 1 other predicate after a 64-day review. Its family tree reaches back 4 generations to 2001. 1 of the 7 devices in its family, counting itself, has a Class I recall. It has been recalled 13 times.

Decided
Aug 30, 2011
Review
64 days
Class
Class II
Product code
MSD
Recalls
13
Predicates
2

Its longest line of descent runs back 4 generations to K010399 (2001). 6 ancestors in all, the oldest K002901 (2001). No later device cites it as a predicate yet. 1 of the 7 devices in this family, counting this one, has a Class I recall.

  1. Root · 4 generations backK010399 Diatek Cannon-Cath #CC2400, CC2800, CC3200, CC3600,CC5500. CC2400PC…Diatek, Inc. · 2001 · Recalled
  2. 3 generations backK040078 Cannon II PlusArrow International, Inc. · 2004 · Recalled
  3. 2 generations backK092205 Palindrome P Hemodialysis CatheterCovidien Lp, Formerly Registered AS Kendall · 2009
  4. 1 generation backK111372 Tal Pallindrome (TM) Symmetric Tip Dual Lumen Catheter, Mahurkar (R)…Covidien · 2011 · Recalled (Class I)
  5. This deviceK111817 Palindrome RT Chronic CatheterCovidien · 2011

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K010399Diatek Cannon-Cath #CC2400, CC2800, CC3200, CC3600,CC5500. CC2400PC, CC2800PC…Diatek, Inc.20012
−3K040078Cannon II PlusArrow International, Inc.200419
−2K092205Palindrome P Hemodialysis CatheterCovidien Lp, Formerly Registered AS Kendall2009—
−1K111372Tal Pallindrome (TM) Symmetric Tip Dual Lumen Catheter, Mahurkar (R) Maxid…Covidien2011Class I
0 (this device)K111817Palindrome RT Chronic CatheterCovidien2011—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200110
200410
FIG. 1 — Lineage of K111817, 2001–2011Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MSD device

  1. #1 K241581 Hemodialysis Catheterscore 12.3 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K092205Palindrome P Hemodialysis Catheter
Summary page 1
…theter Regulation Number: 21 CFR 876.5540 Product Code: MSD Classification: Class III 3. Predicate Devices: Palindrome"m RT Chronic Catheter (K092205 and Kil1372) 4. Device Description: The PalindroMeTM RT Chronic Catheter is supplied with a detached "proximal end"…
Covidien Lp, Formerly Registered AS Kendall2009—
*K111372Tal Pallindrome (TM) Symmetric Tip Dual Lumen Catheter, Mahurkar (R) Maxid (TM) Dual…
Cleared while this submission was under review
Summary page 2
…demonstrated that the modified Palindromem RT Chronic Catheter is substantially equivalent to the existing Palindromnem RT Chronic Catheter (K(092205 and KI111372). TMENT OFHEALTH &HUMAN SERVICESPulcHatSeve DDEPAR…
Covidien2011Class I

Also cleared against K092205: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 28, 2021Class IIZ-2368-2021Labeling designThis voluntary withdrawal is being conducted to implement point-of-use labeling on each device and enhancements to the Instructions for Use (IFU).
May 28, 2021Class IIZ-2369-2021Labeling designThis voluntary withdrawal is being conducted to implement point-of-use labeling on each device and enhancements to the Instructions for Use (IFU).
May 28, 2021Class IIZ-2364-2021Labeling designThis voluntary withdrawal is being conducted to implement point-of-use labeling on each device and enhancements to the Instructions for Use (IFU).
May 28, 2021Class IIZ-2366-2021Labeling designThis voluntary withdrawal is being conducted to implement point-of-use labeling on each device and enhancements to the Instructions for Use (IFU).
May 28, 2021Class IIZ-2358-2021Labeling designThis voluntary withdrawal is being conducted to implement point-of-use labeling on each device and enhancements to the Instructions for Use (IFU).
May 28, 2021Class IIZ-2367-2021Labeling designThis voluntary withdrawal is being conducted to implement point-of-use labeling on each device and enhancements to the Instructions for Use (IFU).
May 28, 2021Class IIZ-2363-2021Labeling designThis voluntary withdrawal is being conducted to implement point-of-use labeling on each device and enhancements to the Instructions for Use (IFU).
May 28, 2021Class IIZ-2362-2021Labeling designThis voluntary withdrawal is being conducted to implement point-of-use labeling on each device and enhancements to the Instructions for Use (IFU).
May 28, 2021Class IIZ-2370-2021Labeling designThis voluntary withdrawal is being conducted to implement point-of-use labeling on each device and enhancements to the Instructions for Use (IFU).
May 28, 2021Class IIZ-2361-2021Labeling designThis voluntary withdrawal is being conducted to implement point-of-use labeling on each device and enhancements to the Instructions for Use (IFU).
May 28, 2021Class IIZ-2359-2021Labeling designThis voluntary withdrawal is being conducted to implement point-of-use labeling on each device and enhancements to the Instructions for Use (IFU).
May 28, 2021Class IIZ-2360-2021Labeling designThis voluntary withdrawal is being conducted to implement point-of-use labeling on each device and enhancements to the Instructions for Use (IFU).
May 28, 2021Class IIZ-2365-2021Labeling designThis voluntary withdrawal is being conducted to implement point-of-use labeling on each device and enhancements to the Instructions for Use (IFU).

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K111372Tal Pallindrome (TM) Symmetric Tip Dual Lumen Catheter, Mahurkar (R) Maxid (TM) Dual Lumen Catheter…Covidien · 2011 · Direct predicate · ◆ on the line of descent in FIG. 12 Class I recalls

What does this lineage say?

  • The line of descent spans 10 years and passes through 4 companies; a new generation was cleared every 2.5 years on median.
  • 3 of the 6 ancestors have a recall of any class; 1 has a Class I recall.
  • 1 other device also cites its predicate K092205.

How does its category compare?

All MSD devices

MSD Catheter, Hemodialysis, Implanted: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.

FDA decisions per year · MSD

1in 2025

2007: 72008: 32009: 62010: 42011: 72012: 22013: 52014: 82015: 12016: 22017: 22018: 22019: 12020: 42021: 42025: 1
2007high 8 (2014) · low 0 (2022)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MSD

153in 2021

2007: 92 days median2008: 89 days median2009: 81 days median2010: 116 days median2011: 70 days median2013: 105 days median2014: 246 days median2020: 106 days median2021: 153 days median
2007high 246 (2014) · low 70 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MSD

1in 2013

2008: 472009: 492010: 112011: 242013: 12026 (partial year): 4, 4 Class I
2007high 49 (2009) · low 0 (2007)2026

Class I (4 since 2007) Class II/III not classified

Adverse event reports per year · MSD

1,054in 2025

2007: 1182008: 1742009: 1732010: 3062011: 2682012: 3672013: 3632014: 5062015: 5502016: 5322017: 4842018: 4632019: 6452020: 4012021: 5232022: 6102023: 5422024: 8412025: 1,054
2007high 1,054 (2025) · low 118 (2007)2025

MAUDE reports naming product code MSD, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20077920118
200838947174
200968149173
2010411611306
201177024268
20122—0367
201351051363
201482460506
20151—0550
20162—0532
20172—0484
20182—0463
20191—0645
202041060401
202141530523
20220—0610
20230—0542
20240—0841
20251—01,054
20260—4 (4)—

What is it indicated for?

The PalindromeTM RT Chronic Catheter is intended for acute and chronic hemodialysis, apheresis, and infusion, It may be inserted either percutaneously or by cut down. Catheters with greater than 40 cm implant length are indicated for femoral insertion.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K111817: Palindrome RT Chronic Catheter.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K111817

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026