510(k) Report

510(k) · Cleared

Recalled · Class II · 2016

Zimmer Periarticular Screws

K111447510(k) TraditionalCleared Oct 5, 2011134-day reviewClass II21 CFR 888.3040Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Zimmer, Inc. · HWC Screw, Fixation, Bone

Cleared on Oct 5, 2011 as substantially equivalent to K042598 (Zimmer, Inc., 2004) after a 134-day review. Its family tree reaches back 4 generations to 1996. 1 later device descends from it. It has been recalled 4 times.

Decided
Oct 5, 2011
Review
134 days
Class
Class II
Product code
HWC
Recalls
4
Predicates
1

Its longest line of descent runs back 4 generations to K962616 (1996). 9 ancestors in all. 1 later device descends from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2004: 1 other ancestor2000: 1 other ancestor2001: 1 other ancestor1998: 1 other ancestor1996: 1 other ancestor2014: 1 descendant clearedK962616 · Synthes Dorsal Distal Radius Plate (DDRP) System · Synthes (Usa) · 1996K982732 · Synthes (Usa) Distal Radius Plate System (DRPS) · Synthes · 1998K040593 · Periarticular Locking Plates and Screws, 2358, 2359 Series · Zimmer, Inc. · 2004 · recalledK042598 · Periarticular Locking Plates and Screws, 2357 and 2359 Series · Zimmer, Inc. · 2004 · recalled1995200020052010201520202025
Other ancestors, by generationDescendants, by generation 2 generations back: 2 other ancestors3 generations back: 2 other ancestors4 generations back: 1 other ancestor 1 generation forward: 1 descendant K042598 · Periarticular Locking Plates and Screws, 2357 and 2359 Series · Zimmer, Inc. · 2004 · recalledK040593 · Periarticular Locking Plates and Screws, 2358, 2359 Series · Zimmer, Inc. · 2004 · recalledK982732 · Synthes (Usa) Distal Radius Plate System (DRPS) · Synthes · 1998K962616 · Synthes Dorsal Distal Radius Plate (DDRP) System · Synthes (Usa) · 19960
  1. Root · 4 generations backK962616 Synthes Dorsal Distal Radius Plate (DDRP) SystemSynthes (Usa) · 1996
  2. 3 generations backK982732 Synthes (Usa) Distal Radius Plate System (DRPS)Synthes · 1998
  3. 2 generations backK040593 Periarticular Locking Plates and Screws, 2358, 2359 SeriesZimmer, Inc. · 2004 · Recalled
  4. 1 generation backK042598 Periarticular Locking Plates and Screws, 2357 and 2359 SeriesZimmer, Inc. · 2004 · Recalled
  5. This deviceK111447 Zimmer Periarticular ScrewsZimmer, Inc. · 2011
  6. Descendant · cites it directly · 0 of its ownK142579 Zimmer Periarticular Plating SystemZimmer, Inc. · 2014 · Recalled

Most influential descendants

Showing 1 of 1 · See all
  • K142579Zimmer Periarticular Plating SystemZimmer, Inc. · 2014 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K962616Synthes Dorsal Distal Radius Plate (DDRP) SystemSynthes (Usa)1996—
−3K982732Synthes (Usa) Distal Radius Plate System (DRPS)Synthes1998—
−2K040593Periarticular Locking Plates and Screws, 2358, 2359 SeriesZimmer, Inc.20042
−1K042598Periarticular Locking Plates and Screws, 2357 and 2359 SeriesZimmer, Inc.20041
0 (this device)K111447Zimmer Periarticular ScrewsZimmer, Inc.2011—
+1K142579Zimmer Periarticular Plating SystemZimmer, Inc.20142

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199610
199810
200010
200110
200410

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201410
FIG. 1 — Lineage of K111447, 1996–2014Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in HWC? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new HWC device

  1. #1 K203011 Monster Screw Systemscore 22.2 · 14 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K042598Periarticular Locking Plates and Screws, 2357 and 2359 Series
Summary page 1
…and performance characteristics to the predicate devices (Zimmer Periarticular Locking Plate System - K042598, cleared 10/29/2004). The proposed screws are provided sterile vs.…
Zimmer, Inc.20041

Also cleared against K042598: 7 other devices →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K142579Zimmer Periarticular Plating SystemZimmer, Inc.20140

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 4, 2016Class IIZ-1676-2016Packaging change controlA small hole may be present in the inner tray of specific lots of sterile Hip and Trauma Implant Screws.
Jan 11, 2016Class IIZ-1154-2016PackagingAs a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.
Jan 11, 2016Class IIZ-1105-2016PackagingAs a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.
Jul 22, 2014Class IIZ-2549-2014Package design/selectionZimmer is recalling sterile-packaged implants and instruments packaged in six different package configurations due to packaging design verification test failures. Specifically, multiple test samples from each of the six configurations…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 15 years and passes through 3 companies; a new generation was cleared every 5.5 years on median.
  • 3 of the 9 ancestors have a recall of any class.
  • Of its descendants, K142579 (Zimmer, Inc., 2014) has gone furthest, with 0 descendants of its own.
  • 7 other devices also cite its predicate K042598.

How does its category compare?

All HWC devices

HWC Screw, Fixation, Bone: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.

FDA decisions per year · HWC

15in 2025

2007: 352008: 332009: 272010: 232011: 352012: 202013: 252014: 312015: 412016: 262017: 402018: 202019: 302020: 282021: 262022: 262023: 242024: 172025: 152026 (partial year): 21
2007high 41 (2015) · low 15 (2025)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · HWC

111in 2025

2007: 77 days median2008: 76 days median2009: 113 days median2010: 110 days median2011: 88 days median2012: 99 days median2013: 90 days median2014: 88 days median2015: 85 days median2016: 131 days median2017: 127 days median2018: 106 days median2019: 108 days median2020: 75 days median2021: 84 days median2022: 167 days median2023: 104 days median2024: 90 days median2025: 111 days median2026 (partial year): 125 days median
2007high 167 (2022) · low 75 (2020)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · HWC

1in 2025

2007: 252008: 332009: 292010: 72011: 122012: 302013: 142014: 22015: 52016: 42017: 42018: 12019: 52020: 42021: 22022: 132025: 12026 (partial year): 2
2007high 33 (2008) · low 0 (2023)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · HWC

2,893in 2025

2007: 6692008: 6602009: 6512010: 7652011: 1,1042012: 3,3392013: 3,8502014: 2,5492015: 2,7112016: 2,2852017: 2,5612018: 3,6622019: 5,4632020: 4,8202021: 4,4462022: 3,9342023: 3,7182024: 3,5752025: 2,893
2007high 5,463 (2019) · low 651 (2009)2025

MAUDE reports naming product code HWC, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007357725669
2008337633660
20092711329651
2010231107765
20113588121,104
20122099303,339
20132590143,850
2014318822,549
2015418552,711
20162613142,285
20174012742,561
20182010613,662
20193010855,463
2020287544,820
2021268424,446
202226167133,934
20232410403,718
2024179003,575
20251511112,893
2026211252—

What is it indicated for?

Temporary internal fixation devices are designed to stabilize fractures during the normal healing process.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K111447: Zimmer Periarticular Screws.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K111447

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026