510(k) Report

510(k) · Decided

Recalled · Class II · 2015

Split Cath Iii

K111651510(k) TraditionalDecided Dec 8, 2011178-day reviewClass II21 CFR 876.5540Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Medical Components, Inc. · MSD Catheter, Hemodialysis, Implanted

Cleared on Dec 8, 2011 as substantially equivalent to K040318 (Medcomp, 2005) and 4 other predicates after a 178-day review. Its family tree reaches back 9 generations to 2000. 2 of the 36 devices in its family, counting itself, have a Class I recall. It has been recalled once.

Decided
Dec 8, 2011
Review
178 days
Class
Class II
Product code
MSD
Recalls
1
Predicates
5

Its longest line of descent runs back 9 generations to K001901 (2000). 35 ancestors in all, the oldest K830256 (1983). No later device cites it as a predicate yet. 2 of the 36 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year cleared 2005: 5 other ancestors2002: 2 other ancestors1999: 2 other ancestors2008: 1 other ancestor2006: 2 other ancestors2003: 2 other ancestors1998: 3 other ancestors1989: 1 other ancestor2004: 1 other ancestor2001: 2 other ancestors2000: 2 other ancestors, 1 with a Class I recall11996: 1 other ancestor1983: 1 other ancestor1986: 1 other ancestor, 1 with a Class I recall1K001901 · Poly Per-Q-Cath Picc Catheter,Model 3236100 and Poly Per-Q-Cath Midline Catheter, Model 3134135 · C.R. Bard, Inc. · 2000 · recalledK012902 · Poly Per-Q-Cath Single Lumen Picc Catheter, Model 3143100; Poly Per-Q-Cath Dual Lumen Picc Catheter, Model 3246100 · C.R. Bard, Inc. · 2001K031129 · Poly Per-Q-Cath Picc Catheters; Poly Radpicc · C.R. Bard, Inc. · 2003K033389 · Powerpicc Catheters, Models 3175155, 3175135 · C.R. Bard, Inc. · 2004 · recalled×3: 3 generations cleared close together, K050185 (2005), K051417 (2005), K051991 (2005) K070996 · 4 FR Single Lumen (SL) Powerpicc Catheter · C.R. Bard, Inc. · 2007K091953 · Pro-Picc · Medcomp · 20091980198519901995200020052010
  1. Root · 9 generations backK001901 Poly Per-Q-Cath Picc Catheter,Model 3236100 and Poly Per-Q-Cath…C.R. Bard, Inc. · 2000 · Recalled
  2. 5 generations between
  3. 3 generations backK051991 Powerpicc, Poly Per-Q-Cath, 6 FR TL Poly Per-Q-Cath, Power Hohn and…C.R. Bard, Inc. · 2005
  4. 2 generations backK070996 4 FR Single Lumen (SL) Powerpicc CatheterC.R. Bard, Inc. · 2007
  5. 1 generation backK091953 Pro-PiccMedcomp · 2009
  6. This deviceK111651 Split Cath IiiMedical Components, Inc. · 2011
Show all 9 generations
  1. Root · 9 generations backK001901 Poly Per-Q-Cath Picc Catheter,Model 3236100 and Poly Per-Q-Cath…C.R. Bard, Inc. · 2000 · Recalled
  2. 8 generations backK012902 Poly Per-Q-Cath Single Lumen Picc Catheter, Model 3143100; Poly…C.R. Bard, Inc. · 2001
  3. 7 generations backK031129 Poly Per-Q-Cath Picc Catheters; Poly RadpiccC.R. Bard, Inc. · 2003
  4. 6 generations backK033389 Powerpicc Catheters, Models 3175155, 3175135C.R. Bard, Inc. · 2004 · Recalled
  5. 5 generations backK050185 Powerhohn and Powerline CathetersC.R. Bard, Inc. · 2005
  6. 4 generations backK051417 6 FR FL Powerhohn and Powerline CatheterC.R. Bard, Inc. · 2005 · Recalled
  7. 3 generations backK051991 Powerpicc, Poly Per-Q-Cath, 6 FR TL Poly Per-Q-Cath, Power Hohn and…C.R. Bard, Inc. · 2005
  8. 2 generations backK070996 4 FR Single Lumen (SL) Powerpicc CatheterC.R. Bard, Inc. · 2007
  9. 1 generation backK091953 Pro-PiccMedcomp · 2009
  10. This deviceK111651 Split Cath IiiMedical Components, Inc. · 2011

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−9K001901Poly Per-Q-Cath Picc Catheter,Model 3236100 and Poly Per-Q-Cath Midline…C.R. Bard, Inc.20002
−8K012902Poly Per-Q-Cath Single Lumen Picc Catheter, Model 3143100; Poly Per-Q-Cath…C.R. Bard, Inc.2001—
−7K031129Poly Per-Q-Cath Picc Catheters; Poly RadpiccC.R. Bard, Inc.2003—
−6K033389Powerpicc Catheters, Models 3175155, 3175135C.R. Bard, Inc.20041
−5K050185Powerhohn and Powerline CathetersC.R. Bard, Inc.2005—
−4K0514176 FR FL Powerhohn and Powerline CatheterC.R. Bard, Inc.20051
−3K051991Powerpicc, Poly Per-Q-Cath, 6 FR TL Poly Per-Q-Cath, Power Hohn and Power LineC.R. Bard, Inc.2005—
−2K0709964 FR Single Lumen (SL) Powerpicc CatheterC.R. Bard, Inc.2007—
−1K091953Pro-PiccMedcomp2009—
0 (this device)K111651Split Cath IiiMedical Components, Inc.2011—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198310
198611
198910
199610
199830
199920
200021
200120
200220
200320
200410
200550
200620
200810
FIG. 1 — Lineage of K111651, 2000–2011Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MSD device

  1. #1 K241581 Hemodialysis Catheterscore 12.3 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (5)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K040318Medcomp Split-Cath II
Summary page 1
…s CFR Reference: 21 CFR 876.5540, Class IIl Classification Panel: Gastroenterology / Urology Premarket Notification: K040318, concurrence date February 3, 2011 K020465, concurrence date July 22, 2002…
Medcomp2005—
*K020465Medcomp Ash Split-Cath II
Summary page 1
…fication Panel: Gastroenterology / Urology Premarket Notification: K040318, concurrence date February 3, 2011 K020465, concurrence date July 22, 2002 K091 953, concurrence date September 16, 2009…
Medcomp2002—
*K091953Pro-Picc
Summary page 1
…ion: K040318, concurrence date February 3, 2011 K020465, concurrence date July 22, 2002 K091 953, concurrence date September 16, 2009 K051280, concurrence date November 30, 2005…
Medcomp2009—
*K051280Medcomp Split Cath II
Summary page 1
…K020465, concurrence date July 22, 2002 K091 953, concurrence date September 16, 2009 K051280, concurrence date November 30, 2005 K981 125, concurrence date February 26, 1999 Performa…
Medcomp2005—
*K981125Medcomp Tesio Catheter, Models MBR-6, MBL-6, MBS-6
Summary page 1
…K091 953, concurrence date September 16, 2009 K051280, concurrence date November 30, 2005 K981 125, concurrence date February 26, 1999 Performance Standards: Performance standards have not been established by FDA under section…
Medical Components, Inc.19991

Also cleared against K040318: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 5, 2015Class IIZ-1351-2015Process controlMedcomp has initiated the recall of Duo-Flow 400XL Catheter, Hemo-Flow Catheter, SLX Catheter, Triple Lumen Infusion Catheter, CT rated PICC, Split Cath III, PC Split Cath III because the product shipped was out of specification. One BI…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K993691Arrowg +Ard Blue Plus Antimicrobial Multi-Lumen Central Venous CatheterArrow Intl., Inc. · 2000 · 5 generations back · ◆ in the 2000 ancestors bar of FIG. 11 Class I recall
  • K862056[Central Venous][Multi-Lumen] Catheter W/Flex TipArrow Intl., Inc. · 1986 · 6 generations back · ◆ in the 1986 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 11 years and passes through 3 companies; a new generation was cleared every 1.2 years on median.
  • 14 of the 35 ancestors have a recall of any class; 2 have a Class I recall.
  • 1 other device also cites its predicate K040318.

How does its category compare?

All MSD devices

MSD Catheter, Hemodialysis, Implanted: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.

FDA decisions per year · MSD

1in 2025

2007: 72008: 32009: 62010: 42011: 72012: 22013: 52014: 82015: 12016: 22017: 22018: 22019: 12020: 42021: 42025: 1
2007high 8 (2014) · low 0 (2022)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MSD

153in 2021

2007: 92 days median2008: 89 days median2009: 81 days median2010: 116 days median2011: 70 days median2013: 105 days median2014: 246 days median2020: 106 days median2021: 153 days median
2007high 246 (2014) · low 70 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MSD

1in 2013

2008: 472009: 492010: 112011: 242013: 12026 (partial year): 4, 4 Class I
2007high 49 (2009) · low 0 (2007)2026

Class I (4 since 2007) Class II/III not classified

Adverse event reports per year · MSD

1,054in 2025

2007: 1182008: 1742009: 1732010: 3062011: 2682012: 3672013: 3632014: 5062015: 5502016: 5322017: 4842018: 4632019: 6452020: 4012021: 5232022: 6102023: 5422024: 8412025: 1,054
2007high 1,054 (2025) · low 118 (2007)2025

MAUDE reports naming product code MSD, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20077920118
200838947174
200968149173
2010411611306
201177024268
20122—0367
201351051363
201482460506
20151—0550
20162—0532
20172—0484
20182—0463
20191—0645
202041060401
202141530523
20220—0610
20230—0542
20240—0841
20251—01,054
20260—4 (4)—

What is it indicated for?

The Medcomp® Split Cath® III is indicated for use in attaining long term vascular access for Hemodialysis and Apheresis in the adult patient. It may be inserted percutaneously and is primarily placed in the internal jugular vein. Alternate insertion sites include the subclavian vein and inferior vena cava as required. Catheters greater than 40cm are intended for femoral vein or inferior vena cava insertion. Translumbar insertion via inferior vena cava is…

Read the full statement (131 words)

The Medcomp® Split Cath® III is indicated for use in attaining long term vascular access for Hemodialysis and Apheresis in the adult patient.

It may be inserted percutaneously and is primarily placed in the internal jugular vein.

Alternate insertion sites include the subclavian vein and inferior vena cava as required.

Catheters greater than 40cm are intended for femoral vein or inferior vena cava insertion.

Translumbar insertion via inferior vena cava is indicated when all other access sites are identified as non-viable.

The Medcomp® Split Cath® III is indicated for use in attaining long term vascular access for Hemodialysis and Apheresis.

It may be inserted percutaneously and is primarily placed in the internal jugular vein.

Alternate insertion sites include the subclavian vein.

Catheters greater than 40cm are intended for femoral vein insertion.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K111651: Split Cath Iii.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K111651

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026