Sorin Centrifugal Pump 5 (CP5)

K112225510(k) SpecialCleared Sep 20, 201148-day reviewClass II21 CFR 870.4380Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Sorin Group Deutschland GmbH · DWA Control, Pump Speed, Cardiopulmonary Bypass

Cleared on Sep 20, 2011 as substantially equivalent to K103468 (Sorin Group Deutschland GmbH, 2011) and 1 other predicate after a 48-day review. Its family tree reaches back 3 generations to 1997. 2 later devices descend from it. It has been recalled 4 times.

Decided
Sep 20, 2011
Review
48 days
Class
Class II
Product code
DWA
Recalls
4
Predicates
2

Its longest line of descent runs back 3 generations to K971520 (1997). 7 ancestors in all, the oldest K950990 (1996). 2 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2011: 1 other ancestor2009: 1 other ancestor2006: 1 other ancestor1996: 1 other ancestor2023: 2 descendants clearedK971520 · S3 Cyclic RPM Control · Stoeckert Instrumente · 1997K042374 · Stockert SCP Rhythm · Stockert Instrumente GmbH · 2004K071318 · Modification to: Stockert S5 System · Sorin Group Deutschland GmbH · 2007 · recalled1995200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 3 other ancestors 1 generation forward: 1 descendant2 generations forward: 1 descendant K071318 · Modification to: Stockert S5 System · Sorin Group Deutschland GmbH · 2007 · recalledK042374 · Stockert SCP Rhythm · Stockert Instrumente GmbH · 2004K971520 · S3 Cyclic RPM Control · Stoeckert Instrumente · 19970
  1. Root · 3 generations backK971520 S3 Cyclic RPM ControlStoeckert Instrumente · 1997
  2. 2 generations backK042374 Stockert SCP RhythmStockert Instrumente GmbH · 2004
  3. 1 generation backK071318 Modification to: Stockert S5 SystemSorin Group Deutschland GmbH · 2007 · Recalled
  4. This deviceK112225 Sorin Centrifugal Pump 5 (CP5)Sorin Group Deutschland GmbH · 2011
  5. Descendant · cites it directly · 1 of its ownK221373 Essenz HLMLivanova Deutschland, GmbH · 2023
  6. Descendant · 2 generations on · 0 of its ownK232291 Essenz HLM, Essenz ILBMLivanova Deutschland, GmbH · 2023 · Recalled

Most influential descendants

Showing 2 of 2 · See all
  • K221373Essenz HLMLivanova Deutschland, GmbH · 2023 · cites it directly1descendant
  • K232291Essenz HLM, Essenz ILBMLivanova Deutschland, GmbH · 2023 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K971520S3 Cyclic RPM ControlStoeckert Instrumente1997—
−2K042374Stockert SCP RhythmStockert Instrumente GmbH2004—
−1K071318Modification to: Stockert S5 SystemSorin Group Deutschland GmbH200713
0 (this device)K112225Sorin Centrifugal Pump 5 (CP5)Sorin Group Deutschland GmbH2011—
+1K221373Essenz HLMLivanova Deutschland, GmbH2023—
+2K232291Essenz HLM, Essenz ILBMLivanova Deutschland, GmbH20231

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199610
200610
200910
201110

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202320
FIG. 1 — Lineage of K112225, 1997–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DWA? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DWA device

  1. #1 K221491 Anivia SG1000 Pump Consolescore 9.3 · 6 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K103468Stockert Centrifugal Pump (SCP Plus) (Complete) or (for S5/C5 Systems)
Summary page 1
…speed control device 870.4380; Product Code: DWA) 3. PREDICATE DEVICE/s * Stackert Centrifugal Pump System Plus for the S5/C5 System (K103468) * Pulse Mode Control for the S5 System (K07 1318) 4. DEVICE DESCRIPTION *Physical description…
Sorin Group Deutschland GmbH2011—
*K071318Modification to: Stockert S5 System
Summary page 1
…; Product Code: DWA) 3. PREDICATE DEVICE/s * Stackert Centrifugal Pump System Plus for the S5/C5 System (K103468) * Pulse Mode Control for the S5 System (K07 1318) 4. DEVICE DESCRIPTION *Physical description…
Sorin Group Deutschland GmbH200713

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K221373Essenz HLMLivanova Deutschland, GmbH20231

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 17, 2021Class IIIZ-0787-2022Process controlOne ferrite ring expected to be installed on centrifugal pump system power cable may not be present.
Oct 30, 2017Class IIZ-0401-2018Device DesignPossibility that the automatic line closure by the Electronic Remote-Controlled (ERC) clamp after detection of an air bubble could take longer than is intended by design in the CP5 system.
Dec 4, 2013Class IIZ-0857-2014Device DesignSorin is initiating a field correction on S5/C5 Heart-lung machine and CP5 centrifugal pump system due to reports of pump speed control knob failures resulting in no response.
Feb 23, 2012Not classifiedZ-1329-2012Nonconforming Material/ComponentThe pump speed control knob is attached to the shaft of an encoder that is used to adjust the speed. Insufficient lubrication within the encoder increases friction over time making it more difficult to turn the pump speed control knob.…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 14 years and passes through 3 companies; a new generation was cleared every 4.2 years on median.
  • 2 of the 7 ancestors have a recall of any class.
  • Of its descendants, K221373 (Livanova Deutschland, GmbH, 2023) has gone furthest, with 1 descendant of its own.

How does its category compare?

All DWA devices

DWA Control, Pump Speed, Cardiopulmonary Bypass: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.

FDA decisions per year · DWA

1in 2024

2007: 12008: 32009: 22010: 22011: 22013: 22019: 12020: 22023: 22024: 1
2007high 3 (2008) · low 0 (2012)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DWA

44in 2008

2008: 44 days median
2007high 44 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DWA

1in 2022

2010: 12012: 32013: 12017: 12021: 12022: 1, 1 Class I
2007high 3 (2012) · low 0 (2007)2026

Class I (1 since 2007) Class II/III not classified

Adverse event reports per year · DWA

362in 2025

2007: 102008: 102009: 172010: 1102011: 1472012: 1022013: 1002014: 982015: 1032016: 1092017: 1402018: 2002019: 3762020: 1462021: 1552022: 2052023: 4382024: 3892025: 362
2007high 438 (2023) · low 10 (2007)2025

MAUDE reports naming product code DWA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—010
2008344010
20092—017
20102—1110
20112—0147
20120—3102
20132—1100
20140—098
20150—0103
20160—0109
20170—1140
20180—0200
20191—0376
20202—0146
20210—1155
20220—1 (1)205
20232—0438
20241—0389
20250—0362
20260—0—

What is it indicated for?

The Sorin Centrifugal Pump (CP5) is a cardiopulmonary bypass speed control device indicated for use exclusively with the Sorin Revolution®5, for speed-controlled pumping through the cardiopulmonary bypass circuit for typical durations of six hours or less.

The 0P5 has been qualified only for typical durations of six hours or less, appropriate to cardiopulmonary bypass procedures and has not been qualified through in vitro, in vivo, or clinical studies, for long term use as a bridge to transplant, pending recovery of the natural heart, or extracorporeal membrane oxygenation (ECMVO) procedures.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K112225: Sorin Centrifugal Pump 5 (CP5).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K112225

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026