510(k) Report

510(k) · Cleared

Cleared

Neutral Luer Activated Device and Extension Sets with Neutral Luer Activated Device

K120443510(k) TraditionalCleared May 22, 201298-day reviewClass II21 CFR 880.5440

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Baxter Healthcare Corporation - Renal Division · FPA Set, Administration, Intravascular

Cleared on May 22, 2012 as substantially equivalent to K003225 (Baxter Healthcare Corp, 2000) and 2 other predicates after a 98-day review. Its family tree reaches back 4 generations to 1995. 28 later devices descend from it.

Decided
May 22, 2012
Review
98 days
Class
Class II
Product code
FPA
Recalls
None
Predicates
3

Its longest line of descent runs back 4 generations to K942988 (1995). 7 ancestors in all. 28 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2008: 1 other ancestor2000: 1 other ancestor1997: 1 other ancestor2013: 2 descendants cleared2021: 6 descendants cleared2017: 4 descendants cleared2014: 2 descendants cleared2020: 2 descendants cleared2016: 1 descendant cleared2023: 3 descendants cleared2019: 2 descendants cleared2026: 2 descendants cleared2025: 2 descendants cleared2024: 1 descendant cleared2022: 1 descendant clearedK942988 · Filtered Extension Sets · B. Braun of America, Inc. · 1995 · recalledK955585 · V2 Injection Site · B.Braun Medical, Inc. · 1996 · recalledK973167 · Positive Pressure Connector · Icu Medical, Inc. · 1997K991653 · Invision-Plus Injection Port Connector System · Rymed Technologies, Inc. · 1999 · recalled1995200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors3 generations back: 1 other ancestor 1 generation forward: 2 descendants2 generations forward: 6 descendants3 generations forward: 8 descendants4 generations forward: 8 descendants5 generations forward: 2 descendants6 generations forward: 2 descendants K991653 · Invision-Plus Injection Port Connector System · Rymed Technologies, Inc. · 1999 · recalledK973167 · Positive Pressure Connector · Icu Medical, Inc. · 1997K955585 · V2 Injection Site · B.Braun Medical, Inc. · 1996 · recalledK942988 · Filtered Extension Sets · B. Braun of America, Inc. · 1995 · recalled0+5
  1. Root · 4 generations backK942988 Filtered Extension SetsB. Braun of America, Inc. · 1995 · Recalled
  2. 3 generations backK955585 V2 Injection SiteB.Braun Medical, Inc. · 1996 · Recalled
  3. 2 generations backK973167 Positive Pressure ConnectorIcu Medical, Inc. · 1997
  4. 1 generation backK991653 Invision-Plus Injection Port Connector SystemRymed Technologies, Inc. · 1999 · Recalled
  5. This deviceK120443 Neutral Luer Activated Device and Extension Sets with Neutral Luer…Baxter Healthcare Corporation - Renal Division · 2012
  6. Descendant · cites it directly · 26 of its ownK132413 MZ1000Care Fusion · 2013
  7. Descendant · 2 generations on · 13 of its ownK140831 Maxzero Extension Set with Needless ConnectorCare Fusion · 2014
  8. Descendant · 2 generations on · 6 of its ownK170336 BD Nexiva Closed IV Catheter System (Single Port) with MaxZero…Becton, Dickinson and Company · 2017
  9. Descendant · 3 generations on · 5 of its ownK183399 BD Nexiva Closed IV Catheter SystemBecton Dickinson Infusion Therapy Systems, Inc. · 2019 · Recalled
  10. Descendant · 3 generations on · 4 of its ownK162804 Q2 Low Pressure Power Injection Extension SetQuest Medical, Inc. · 2017

Showing 5 of 28 descendants. See all descendants

Descendants cleared per year

2013: 22014: 22016: 12017: 42019: 22020: 22021: 62022: 12023: 32024: 12025: 22026: 2
2013high 6 (2021) · low 0 (2015)2026

Most influential descendants

Showing 5 of 28 · See all
  • K132413MZ1000Care Fusion · 2013 · cites it directly26descendants
  • K140831Maxzero Extension Set with Needless ConnectorCare Fusion · 2014 · 2 generations on13descendants
  • K170336BD Nexiva Closed IV Catheter System (Single Port) with MaxZero Needleless ConnectorBecton, Dickinson and Company · 2017 · 2 generations on6descendants
  • K183399BD Nexiva Closed IV Catheter SystemBecton Dickinson Infusion Therapy Systems, Inc. · 2019 · 3 generations on5descendants
  • K162804Q2 Low Pressure Power Injection Extension SetQuest Medical, Inc. · 2017 · 3 generations on4descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K942988Filtered Extension SetsB. Braun of America, Inc.199532
−3K955585V2 Injection SiteB.Braun Medical, Inc.199627
−2K973167Positive Pressure ConnectorIcu Medical, Inc.1997—
−1K991653Invision-Plus Injection Port Connector SystemRymed Technologies, Inc.19994
0 (this device)K120443Neutral Luer Activated Device and Extension Sets with Neutral Luer Activated…Baxter Healthcare Corporation - Renal Division2012—
+1K132413MZ1000Care Fusion2013—
+2K140831Maxzero Extension Set with Needless ConnectorCare Fusion2014—
+2K170336BD Nexiva Closed IV Catheter System (Single Port) with MaxZero Needleless…Becton, Dickinson and Company2017—
+3K183399BD Nexiva Closed IV Catheter SystemBecton Dickinson Infusion Therapy Systems, Inc.20193
+3K162804Q2 Low Pressure Power Injection Extension SetQuest Medical, Inc.2017—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199710
200010
200810

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201320
201420
201610
201740
201920
202020
202160
202210
202330
202410
202520
202620
FIG. 1 — Lineage of K120443, 1995–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FPA? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FPA device

  1. #1 K243841 Sparta Infusion Set for Insulinscore 13.4 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K003225Solution Administration Sets with Capless Luer Activated Valve
Summary page 2
…evices are cited in support of the substantial equivalence determination: * Baxter Healthcare Solution Administration Sets with Capped Luer Activated Valve, K003225 cleared on October 19, 2000 " Pressure Rated Extension Sets, K083472 cleared on November 18, 2008 9 InVision Plus Injec…
Baxter Healthcare Corp200018
*K083472Medegen Pressure Rated Extension Sets
Summary page 2
…* Baxter Healthcare Solution Administration Sets with Capped Luer Activated Valve, K003225 cleared on October 19, 2000 " Pressure Rated Extension Sets, K083472 cleared on November 18, 2008 9 InVision Plus Injection Port Systems, K99 1653, cleared June 24, 1999 DESCRIPTION OF THE DEV…
Medegen, Inc.20082
*K991653Invision-Plus Injection Port Connector System
Summary page 2
…Valve, K003225 cleared on October 19, 2000 " Pressure Rated Extension Sets, K083472 cleared on November 18, 2008 9 InVision Plus Injection Port Systems, K99 1653, cleared June 24, 1999 DESCRIPTION OF THE DEVICE: The Neutral Luer Activated Device (LAD) and Extension Sets with Neutral Luer…
Rymed Technologies, Inc.19994

Also cleared against K003225: 4 other devices →

What descends from it? (28)

See all 28

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K132413MZ1000Care Fusion201326
K132734Neutral Luer Activated Device (One-Link Needle-Free IV Connector) and Extension SetsBaxter Healthcare Corp20130

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 17 years and passes through 5 companies; a new generation was cleared every 1.6 years on median.
  • 6 of the 7 ancestors have a recall of any class.
  • Of its descendants, K132413 (Care Fusion, 2013) has gone furthest, with 26 descendants of its own.
  • 4 other devices also cite its predicate K003225.

How does its category compare?

All FPA devices

FPA Set, Administration, Intravascular: FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.

FDA decisions per year · FPA

11in 2025

2007: 162008: 182009: 102010: 232011: 122012: 212013: 132014: 222015: 172016: 122017: 172018: 42019: 72020: 102021: 132022: 142023: 142024: 72025: 112026 (partial year): 5
2007high 23 (2010) · low 4 (2018)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FPA

225in 2025

2007: 75 days median2008: 89 days median2009: 166 days median2010: 120 days median2011: 87 days median2012: 96 days median2013: 90 days median2014: 89 days median2015: 123 days median2016: 195 days median2017: 240 days median2018: 177 days median2019: 161 days median2020: 327 days median2021: 186 days median2022: 213 days median2023: 340 days median2024: 244 days median2025: 225 days median2026 (partial year): 266 days median
2007high 340 (2023) · low 75 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FPA

98in 2025

2007: 172008: 152009: 1662010: 582011: 282012: 16, 4 Class I2013: 642014: 242015: 62016: 102017: 132018: 42019: 132020: 1072021: 12022: 9, 5 Class I2023: 12, 4 Class I2024: 13, 1 Class I2025: 98, 1 Class I2026 (partial year): 10, 2 Class I
2007high 166 (2009) · low 1 (2021)2026

Class I (17 since 2007) Class II/III not classified

Adverse event reports per year · FPA

34,506in 2025

2007: 1,4992008: 1,9412009: 2,5122010: 2,5372011: 4,4072012: 4,9302013: 4,6512014: 4,6952015: 6,6252016: 8,4702017: 6,4622018: 6,5742019: 11,2272020: 9,5632021: 10,4142022: 13,3932023: 10,9162024: 47,2102025: 34,506
2007high 47,210 (2024) · low 1,499 (2007)2025

MAUDE reports naming product code FPA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071675171,499
20081889151,941
2009101661662,512
201023120582,537
20111287284,407
2012219616 (4)4,930
20131390644,651
20142289244,695
20151712366,625
201612195108,470
201717240136,462
2018417746,574
201971611311,227
2020103271079,563
202113186110,414
2022142139 (5)13,393
20231434012 (4)10,916
2024724413 (1)47,210
20251122598 (1)34,506
2026526610 (2)—

What is it indicated for?

The Baxter Neutral Luer Activated Device is intended for single patient use with a vascular access device for the administration of drugs and solutions without needles, thus eliminating the potential for needle-stick injuries during use. This device is an in-line injection site which can be connected to standard male Luer adapters (e.g., syringes or sets) for the continuous or intermittent fluid administration or the withdrawal of fluids. This device may be used with low pressure power injectors having a maximum pressure of 325 psi (2241 kPa) and a maximum flow rate of 10 mL/s.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K120443: Neutral Luer Activated Device and Extension Sets with Neutral Luer Activated Device.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K120443

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.