510(k) Report

510(k) · Cleared

Recalled · Class II · 2016

Flashpak

K113776510(k) TraditionalCleared Aug 8, 2012230-day reviewClass II21 CFR 880.6850Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Symmetry Medical · FRG Wrap, Sterilization

Cleared on Aug 8, 2012 as substantially equivalent to K871202 (Performance Instruments Corp., 1987) after a 230-day review. Its family tree reaches back 1 generation to 1987. 1 later device descends from it. It has been recalled 3 times.

Decided
Aug 8, 2012
Review
230 days
Class
Class II
Product code
FRG
Recalls
3
Predicates
1

Its longest line of descent runs back 1 generation to K871202 (1987). 1 later device descends from it, 1 citing it directly.

Descendants, by year cleared 2015: 1 descendant clearedK871202 · Flashpak Container System · Performance Instruments Corp. · 19871990200020102020
Descendants, by generation 1 generation forward: 1 descendant K871202 · Flashpak Container System · Performance Instruments Corp. · 19870
  1. Root · 1 generation backK871202 Flashpak Container SystemPerformance Instruments Corp. · 1987
  2. This deviceK113776 FlashpakSymmetry Medical · 2012
  3. Descendant · cites it directly · 0 of its ownK141826 Flashpak Sterilization Container SystemSymmetry Medical Inc- Manchester · 2015

Most influential descendants

Showing 1 of 1 · See all
  • K141826Flashpak Sterilization Container SystemSymmetry Medical Inc- Manchester · 2015 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K871202Flashpak Container SystemPerformance Instruments Corp.1987—
0 (this device)K113776FlashpakSymmetry Medical2012—
+1K141826Flashpak Sterilization Container SystemSymmetry Medical Inc- Manchester2015—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201510
FIG. 1 — Lineage of K113776, 1987–2015Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FRG? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRG device

  1. #1 K234050 HALYARD® ONE-STEP® Sterilization Wrap…score 15.8 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K871202Flashpak Container System
Summary page 2
…igid Intended Use statement was included in the container system to be used statement is K87 1202 submission. The during flash sterilization by eqfuivalent albeit Intended Use following is t…
Performance Instruments Corp.1987—

Also cleared against K871202: 4 other devices →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K141826Flashpak Sterilization Container SystemSymmetry Medical Inc- Manchester20150

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 29, 2016Class IIZ-1011-2017Nonconforming Material/ComponentSealant on the base trim of sterilizer container may be deteriorated or absent which could provide a path for contaminants to enter the container under certain conditions post-sterilization.
Nov 29, 2016Class IIZ-1012-2017Nonconforming Material/ComponentSealant on the base trim of sterilizer container may be deteriorated or absent which could provide a path for contaminants to enter the container under certain conditions post-sterilization.
Nov 29, 2016Class IIZ-1010-2017Nonconforming Material/ComponentSealant on the base trim of sterilizer container may be deteriorated or absent which could provide a path for contaminants to enter the container under certain conditions post-sterilization.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has no recall.
  • Of its descendants, K141826 (Symmetry Medical Inc- Manchester, 2015) has gone furthest, with 0 descendants of its own.
  • 4 other devices also cite its predicate K871202.

How does its category compare?

All FRG devices

FRG Wrap, Sterilization: FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.

FDA decisions per year · FRG

9in 2025

2007: 32008: 22009: 42010: 42011: 62012: 102013: 62014: 82015: 122016: 92017: 92018: 52019: 92020: 52021: 42022: 92023: 92024: 112025: 92026 (partial year): 4
2007high 12 (2015) · low 2 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRG

151in 2025

2007: 182 days median2009: 171 days median2010: 213 days median2011: 110 days median2012: 133 days median2013: 146 days median2014: 177 days median2015: 196 days median2016: 131 days median2017: 260 days median2018: 150 days median2019: 258 days median2020: 122 days median2021: 170 days median2022: 217 days median2023: 230 days median2024: 182 days median2025: 151 days median2026 (partial year): 247 days median
2007high 260 (2017) · low 110 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRG

2in 2022

2013: 22017: 292020: 142022: 2
2007high 29 (2017) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FRG

18in 2025

2007: 12008: 22009: 22011: 122012: 42013: 52014: 132015: 92016: 202017: 72018: 222019: 102020: 212021: 92022: 42023: 132024: 212025: 18
2007high 22 (2018) · low 0 (2010)2025

MAUDE reports naming product code FRG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007318201
20082—02
2009417102
2010421300
20116110012
20121013304
2013614625
20148177013
20151219609
20169131020
20179260297
20185150022
20199258010
202051221421
2021417009
2022921724
20239230013
202411182021
20259151018
202642470—

What is it indicated for?

FlashPak® is a reusable rigid container system to be used during flash sterilization by hospitals and healthcare facilities. It is intended to enable flash sterilization of the enclosed devices and prevent recontamination during immediate transport to the point of use. The container is compatible with gravity-displacement flash sterilization using a 10 minute cycle at 132° C and with pre-vacuum flash sterilization using a 4 minute cycle at 132° C.…

Read the full statement (110 words)

FlashPak® is a reusable rigid container system to be used during flash sterilization by hospitals and healthcare facilities. It is intended to enable flash sterilization of the enclosed devices and prevent recontamination during immediate transport to the point of use. The container is compatible with gravity-displacement flash sterilization using a 10 minute cycle at 132° C and with pre-vacuum flash sterilization using a 4 minute cycle at 132° C. FlashPak is recommended for surface sterilization of stainless steel instruments and for lumens with the following limits: gravity-displacement (5.5mm inner diameter or larger and up to 184mm in length), pre-vacuum (1mm inner diameter or larger and up to 203mm in length).

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K113776: Flashpak.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K113776

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026