510(k) Report

510(k) · Cleared

Cleared

Aesculap Sterilcontainer System

K053389510(k) TraditionalCleared Feb 15, 200672-day reviewClass II21 CFR 880.6850

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Aesculap, Inc. · FRG Wrap, Sterilization

Cleared on Feb 15, 2006 as substantially equivalent to K022978 (Case Medical, Inc., 2003) and 2 other predicates after a 72-day review. Its family tree reaches back 2 generations to 1987. 12 later devices descend from it.

Decided
Feb 15, 2006
Review
72 days
Class
Class II
Product code
FRG
Recalls
None
Predicates
3

Its longest line of descent runs back 2 generations to K871202 (1987). 6 ancestors in all, the oldest K792558 (1979). 12 later devices descend from it, 4 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1979: 1 other ancestor1997: 3 other ancestors2014: 1 descendant cleared2006: 1 descendant cleared2012: 1 descendant cleared2008: 1 descendant cleared2017: 1 descendant cleared2020: 2 descendants cleared2024: 1 descendant cleared2022: 1 descendant cleared2025: 3 descendants clearedK871202 · Flashpak Container System · Performance Instruments Corp. · 1987K022978 · Steritite Container with Flashtite Valve Plate (S) & Meditray Products · Case Medical, Inc. · 200319801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 3 other ancestors 1 generation forward: 4 descendants2 generations forward: 4 descendants3 generations forward: 1 descendant4 generations forward: 3 descendants K022978 · Steritite Container with Flashtite Valve Plate (S) & Meditray Products · Case Medical, Inc. · 2003K871202 · Flashpak Container System · Performance Instruments Corp. · 19870
  1. Root · 2 generations backK871202 Flashpak Container SystemPerformance Instruments Corp. · 1987
  2. 1 generation backK022978 Steritite Container with Flashtite Valve Plate (S) & Meditray ProductsCase Medical, Inc. · 2003
  3. This deviceK053389 Aesculap Sterilcontainer SystemAesculap, Inc. · 2006
  4. Descendant · cites it directly · 9 of its ownK073168 Aesculap Primeline Container LTD.Aesculap, Inc. · 2008
  5. Descendant · cites it directly · 7 of its ownK112671 Sterilcontainer SystemAesculap, Ltd. · 2012
  6. Descendant · 2 generations on · 5 of its ownK192056 JJ Series Container SystemAesculap, Inc. · 2020
  7. Descendant · 3 generations on · 4 of its ownK214041 Aesculap AICON ContainerAesculap, Inc. · 2022
  8. Descendant · cites it directly · 1 of its ownK131407 Bahadir Sterilization TraysBahadir USA Corp. · 2014

Showing 5 of 12 descendants. See all descendants

Descendants cleared per year

2006: 12008: 12012: 12014: 12017: 12020: 22022: 12024: 12025: 3
2006high 3 (2025) · low 0 (2007)2026

Most influential descendants

Showing 5 of 12 · See all
  • K073168Aesculap Primeline Container LTD.Aesculap, Inc. · 2008 · cites it directly9descendants
  • K112671Sterilcontainer SystemAesculap, Ltd. · 2012 · cites it directly7descendants
  • K192056JJ Series Container SystemAesculap, Inc. · 2020 · 2 generations on5descendants
  • K214041Aesculap AICON ContainerAesculap, Inc. · 2022 · 3 generations on4descendants
  • K131407Bahadir Sterilization TraysBahadir USA Corp. · 2014 · cites it directly1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K871202Flashpak Container SystemPerformance Instruments Corp.1987—
−1K022978Steritite Container with Flashtite Valve Plate (S) & Meditray ProductsCase Medical, Inc.2003—
0 (this device)K053389Aesculap Sterilcontainer SystemAesculap, Inc.2006—
+1K073168Aesculap Primeline Container LTD.Aesculap, Inc.2008—
+1K112671Sterilcontainer SystemAesculap, Ltd.2012—
+2K192056JJ Series Container SystemAesculap, Inc.2020—
+3K214041Aesculap AICON ContainerAesculap, Inc.2022—
+1K131407Bahadir Sterilization TraysBahadir USA Corp.2014—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
197910
199730

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200610
200810
201210
201410
201710
202020
202210
202410
202530
FIG. 1 — Lineage of K053389, 1987–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FRG? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRG device

  1. #1 K234050 HALYARD® ONE-STEP® Sterilization Wrap…score 15.8 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K022978Steritite Container with Flashtite Valve Plate (S) & Meditray Products
Summary page 1
…Aesculap,, Inc. believes that the SterilcontainerSystem for Flash Sterilization is substantially equivalent to: Case Medical Steritite Container with Flashtite Valve (K022978) Riley Medical Flashpak Container System (K871202) Aesculap Sterilcontainer System (K792558) DEVICE DESCRIPTION The Flash compat…
Case Medical, Inc.2003—
*K871202Flashpak Container System
Summary page 1
…rSystem for Flash Sterilization is substantially equivalent to: Case Medical Steritite Container with Flashtite Valve (K022978) Riley Medical Flashpak Container System (K871202) Aesculap Sterilcontainer System (K792558) DEVICE DESCRIPTION The Flash compatible Aesculap Sterilcontainer is designed as a con…
Performance Instruments Corp.1987—
*K792558Container System
Summary page 1
…tially equivalent to: Case Medical Steritite Container with Flashtite Valve (K022978) Riley Medical Flashpak Container System (K871202) Aesculap Sterilcontainer System (K792558) DEVICE DESCRIPTION The Flash compatible Aesculap Sterilcontainer is designed as a container system that will allow for steriliz…
Aesculap Instruments Corp.1979—

Also cleared against K022978: 3 other devices →

What descends from it? (12)

See all 12

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K073168Aesculap Primeline Container LTD.Aesculap, Inc.20089
K112671Sterilcontainer SystemAesculap, Ltd.20127
K131407Bahadir Sterilization TraysBahadir USA Corp.20141
K062510Aesculap Sterilcontainer SystemAesculap20060

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 19 years and passes through 3 companies; a new generation was cleared every 15.9 years on median.
  • 1 of the 6 ancestors has a recall of any class.
  • Of its descendants, K073168 (Aesculap, Inc., 2008) has gone furthest, with 9 descendants of its own.
  • 3 other devices also cite its predicate K022978.

How does its category compare?

All FRG devices

FRG Wrap, Sterilization: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · FRG

9in 2025

2007: 32008: 22009: 42010: 42011: 62012: 102013: 62014: 82015: 122016: 92017: 92018: 52019: 92020: 52021: 42022: 92023: 92024: 112025: 92026 (partial year): 4
2007high 12 (2015) · low 2 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRG

151in 2025

2007: 182 days median2009: 171 days median2010: 213 days median2011: 110 days median2012: 133 days median2013: 146 days median2014: 177 days median2015: 196 days median2016: 131 days median2017: 260 days median2018: 150 days median2019: 258 days median2020: 122 days median2021: 170 days median2022: 217 days median2023: 230 days median2024: 182 days median2025: 151 days median2026 (partial year): 247 days median
2007high 260 (2017) · low 110 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRG

2in 2022

2013: 22017: 292020: 142022: 2
2007high 29 (2017) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FRG

18in 2025

2007: 12008: 22009: 22011: 122012: 42013: 52014: 132015: 92016: 202017: 72018: 222019: 102020: 212021: 92022: 42023: 132024: 212025: 18
2007high 22 (2018) · low 0 (2010)2025

MAUDE reports naming product code FRG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007318201
20082—02
2009417102
2010421300
20116110012
20121013304
2013614625
20148177013
20151219609
20169131020
20179260297
20185150022
20199258010
202051221421
2021417009
2022921724
20239230013
202411182021
20259151018
202642470—

What is it indicated for?

The Aesculap Sterilcontainer is a reusable sterilization container system (consisting of a solid bottom, a perforated lid w/ filter retention plates, and disposable paper filters) intended to be used to enclose another medical device that is to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed device and also maintain sterility of the enclosed device until used. This container has been validated with stainless…

Read the full statement (167 words)

The Aesculap Sterilcontainer is a reusable sterilization container system (consisting of a solid bottom, a perforated lid w/ filter retention plates, and disposable paper filters) intended to be used to enclose another medical device that is to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed device and also maintain sterility of the enclosed device until used. This container has been validated with stainless steel lumens, hinged, and knurled instruments (stainless steel lumens of greater than 3 mm inner diameter or less than 400 mm in length). This container system is compatible for use in pre-vacuum (steam) flash sterilization. The SterilContainer System for includes accessories such as baskets, trays, and racks.

The following containers (lids & bottoms) were validated for pre-vacuum Flash sterilization:

Full Size: Lid JK489/ Bottoms JK440, JK441, JK442, and JK444

Three-Quarter Size: Lid JK789/ Bottoms JK740, JK741, and JK742

Half Size: Lid JK389/ Bottoms JK339, JK340, JK341, JK342, and JK344

Mini Containers: Lid JK174/ Bottoms JK187 and JK188

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K053389: Aesculap Sterilcontainer System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K053389

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026