510(k) · Cleared
ClearedIntellivue Patient Monitor
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Philips Medizinsysteme Boeblingen GmbH, Cardiac AN · MHX Monitor, Physiological, Patient (With Arrhythmia Detection Or Alarms)
Cleared on Jul 18, 2012 as substantially equivalent to K110474 (Philips Medizinsysteme Boeblingen GmbH, Cardiac AN, 2011) and 40 other predicates after a 163-day review. Its family tree reaches back 31 generations to 1994. 1 of the 272 devices in its family, counting itself, has a Class I recall. 46 later devices descend from it.
- Decided
- Jul 18, 2012
- Review
- 163 days
- Class
- Class II
- Product code
- MHX
- Recalls
- None
- Predicates
- 41
Its longest line of descent runs back 31 generations to K941811 (1994). 225 ancestors in all, the oldest K844187 (1985). 46 later devices descend from it, 9 citing it directly. 1 of the 272 devices in this family, counting this one, has a Class I recall.
- Root · 31 generations backK941811 HP M1175A, M1176A, Component Monitoring SystemHewlett-Packard Co. · 1994
- 27 generations between
- 3 generations backK100939 Intellivue Patient Monitor Model: MP2, MP5, MP5T, MP20, MP30, MP40…Philips Medizinsysteme Boeblingen GmbH, Cardiac AN · 2010
- 2 generations backK101449 Intellivue Patient MonitorPhilips Medizinsysteme Boeblingen GmbH, Cardiac AN · 2010
- 1 generation backK110622 Intellivue Patient MonitorPhilips Medizinsysteme Boeblingen GmbH, Cardiac AN · 2011 · Recalled
- This deviceK120366 Intellivue Patient MonitorPhilips Medizinsysteme Boeblingen GmbH, Cardiac AN · 2012
- Descendant · cites it directly · 41 of its ownK122439 Intellivue Patient MonitorPhilips Medizin Systeme Boeblingen GmbH · 2012 · Recalled
- Descendant · cites it directly · 37 of its ownK131872 Intellivue Patient Monitor MX500, Intellivue Patient Monitor MX550Philips Medizinsysteme Boeblingen GmbH, Cardiac AN · 2013 · Recalled
- Descendant · cites it directly · 37 of its ownK130849 Intellivue Patient Monitor MX400; Intellivue Patient Monitor MX450Philips Medizinsysteme Boeblingen GmbH, Cardiac AN · 2013 · Recalled
- Descendant · cites it directly · 36 of its ownK141015 Intellivue Patient Monitor MX400, Intellivue Patient Monitor MX450…Philips Medizinsysteme Boeblingen GmbH, Cardiac AN · 2014 · Recalled
- Descendant · cites it directly · 34 of its ownK150975 Philips IntelliVue MP2/MP5/MPSC Patient Monitors, Philips IntelliVue…Philips Medizin Systeme Boeblingen GmbH · 2015 · Recalled
Show all 31 generations
- Root · 31 generations backK941811 HP M1175A, M1176A, Component Monitoring SystemHewlett-Packard Co. · 1994
- 30 generations backK960934 HP M1205A Omnicare Component Monitor System Model 24Hewlett-Packard Co. · 1996
- 29 generations backK981376 Hewlett-Packard M1205A Viridia 24 CTHewlett-Packard Co. · 1999
- 28 generations backK990125 HP Viridia Component Monitoring System, Model Rev. KHewlett-Packard GmbH · 1999
- 27 generations backK990972 HP Viridia Component Monitoring System, HP Viridia CMS 24/26, HP…Hewlett-Packard GmbH · 1999
- 26 generations backK013199 Philips Component Compact Monitor,Model M1275B, Philips Compenent…Philips Medical Systems, Inc. · 2001
- 25 generations backK020531 Philips Multi-Parameter Module, Model M3001A; Philips Component…Philips Medical Systems · 2002
- 24 generations backK021778 Philips Component Compact Monitor, Release a.04, Model #M1275BPhilips Medical Systems · 2002
- 23 generations backK023871 M3046A (M2/M3/M4) Compact Portable Patient Monitor and MP60 & MP70…Philips Medical Systems · 2002
- 22 generations backK030038 Philips MP60, MP70 and MP90 Intellivue Patient Monitors with Portal…Philips Medical Systems, Inc. · 2003
- 21 generations backK031481 The Philips Intellivue MP60, MP70, and MP90 Patient Monitors…Philips Medizin Systeme Boblingen GmbH · 2003
- 20 generations backK032858 The Philips Intellivue MP40, MP50, MP60, MP70,and MP90 Patient…Philips Medizin Systeme Boblingen GmbH · 2003
- 19 generations backK040304 Philips Intellivue Patient Monitor, Models MP40 & MP50Philips Medizin Systeme Boeblingen GmbH · 2004
- 18 generations backK041235 Philips MP20, MP30, MP40, MP50. MP60, MP70 and MP90 Intellivue…Philips Medizin Systeme Boeblingen GmbH · 2004
- 17 generations backK042845 The Philips MP20, MP30, MP40, MP50, MP60, MP70, and MP90 Intellivue…Philips Medizin Systeme Boblingen GmbH · 2004
- 16 generations backK050141 The Philips MP20, MP30, MP40, MP50, MP60, MP70 and MP90 Intellivue…Philips Medizin Systeme Boeblingen GmbH · 2005
- 15 generations backK050762 Philips Intellivue Patient Monitor, Models MP20, MP30, MP40, MP50…Philips Medizin Systeme Boeblingen GmbH · 2005
- 14 generations backK051106 The Philips MP20, MP30, MP40, MP50, MP60, MP70, MP80 and MP90…Philips Medizin Systeme Boeblingen GmbH · 2005
- 13 generations backK052801 Intellivue Patient Monitors with Release C.02, Models MP20, MP30…Philips Medizin Systeme Boeblingen GmbH · 2005
- 12 generations backK053522 Philips MP40, MP50, MP60, MP70, MP80 and MP90 Intellivue Patient…Philips Medizin Systeme Boeblingen GmbH · 2006
- 11 generations backK060221 The Phillips, Model MP20,MP30,MP40,MP50,MP60,MP70,MP80,and MP90…Philips Medizin Systeme Boeblingen GmbH · 2006
- 10 generations backK060541 The Philips MP20,MP30,MP40,MP50,MP60,MP70,MP80 and MP90 Intellivue…Philips Medizin Systeme Boblingen GmbH · 2006 · Recalled
- 9 generations backK061052 The Philips Intellivue Patient Monitors, Models…Philips Medical Systems North America, Inc. · 2006
- 8 generations backK061610 Philips Model MP20, MP30, MP40, MP50, MP60, MP70, MP80 and MP90…Philips Medizin Systeme Boeblingen GmbH · 2006
- 7 generations backK063315 Intellivue Patient Monitors, Models MP20, MP30, MP40, MP50, MP60…Philips Medizin Systeme Boeblingen GmbH · 2006
- 6 generations backK071426 MP2, X2, MP5, MP20, MP30, MP40, MP50, MP60, MP70, MP80, and MP90…Philips Medizinsysteme Boeblingen GmbH, Cardiac AN · 2007
- 5 generations backK083517 Intellivue Patient Monitor, Models MP2, MP5, MP20, MP30, MP40, MP50…Philips Medizinsysteme Boeblingen GmbH, Cardiac AN · 2009 · Recalled
- 4 generations backK091927 Intellivue Patient Monitors ,Models MP40, MP50, MP60, MP70, MP80, MP90Philips Medizinsysteme Boeblingen GmbH, Cardiac AN · 2009
- 3 generations backK100939 Intellivue Patient Monitor Model: MP2, MP5, MP5T, MP20, MP30, MP40…Philips Medizinsysteme Boeblingen GmbH, Cardiac AN · 2010
- 2 generations backK101449 Intellivue Patient MonitorPhilips Medizinsysteme Boeblingen GmbH, Cardiac AN · 2010
- 1 generation backK110622 Intellivue Patient MonitorPhilips Medizinsysteme Boeblingen GmbH, Cardiac AN · 2011 · Recalled
- This deviceK120366 Intellivue Patient MonitorPhilips Medizinsysteme Boeblingen GmbH, Cardiac AN · 2012
Showing 5 of 46 descendants. See all descendants
Descendants cleared per year
Class I recalls in this family
- K970209 Capnocheck Plus6 generations back · 1997
Most influential descendants
Showing 5 of 46 · See all- K122439Intellivue Patient MonitorPhilips Medizin Systeme Boeblingen GmbH · 2012 · cites it directly41descendants
- K131872Intellivue Patient Monitor MX500, Intellivue Patient Monitor MX550Philips Medizinsysteme Boeblingen GmbH, Cardiac AN · 2013 · cites it directly37descendants
- K130849Intellivue Patient Monitor MX400; Intellivue Patient Monitor MX450Philips Medizinsysteme Boeblingen GmbH, Cardiac AN · 2013 · cites it directly37descendants
- K141015Intellivue Patient Monitor MX400, Intellivue Patient Monitor MX450, Intellivue Patient…Philips Medizinsysteme Boeblingen GmbH, Cardiac AN · 2014 · cites it directly36descendants
- K150975Philips IntelliVue MP2/MP5/MPSC Patient Monitors, Philips IntelliVue X1 and X2…Philips Medizin Systeme Boeblingen GmbH · 2015 · cites it directly34descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −31 | K941811 | HP M1175A, M1176A, Component Monitoring System | Hewlett-Packard Co. | 1994 | — |
| −30 | K960934 | HP M1205A Omnicare Component Monitor System Model 24 | Hewlett-Packard Co. | 1996 | — |
| −29 | K981376 | Hewlett-Packard M1205A Viridia 24 CT | Hewlett-Packard Co. | 1999 | — |
| −28 | K990125 | HP Viridia Component Monitoring System, Model Rev. K | Hewlett-Packard GmbH | 1999 | — |
| −27 | K990972 | HP Viridia Component Monitoring System, HP Viridia CMS 24/26, HP Viridia… | Hewlett-Packard GmbH | 1999 | — |
| −26 | K013199 | Philips Component Compact Monitor,Model M1275B, Philips Compenent Rack, Model… | Philips Medical Systems, Inc. | 2001 | — |
| −25 | K020531 | Philips Multi-Parameter Module, Model M3001A; Philips Component Compact… | Philips Medical Systems | 2002 | — |
| −24 | K021778 | Philips Component Compact Monitor, Release a.04, Model #M1275B | Philips Medical Systems | 2002 | — |
| −23 | K023871 | M3046A (M2/M3/M4) Compact Portable Patient Monitor and MP60 & MP70 Intellivue… | Philips Medical Systems | 2002 | — |
| −22 | K030038 | Philips MP60, MP70 and MP90 Intellivue Patient Monitors with Portal Technology… | Philips Medical Systems, Inc. | 2003 | — |
| −21 | K031481 | The Philips Intellivue MP60, MP70, and MP90 Patient Monitors, Release a.20 | Philips Medizin Systeme Boblingen GmbH | 2003 | — |
| −20 | K032858 | The Philips Intellivue MP40, MP50, MP60, MP70,and MP90 Patient Monitors… | Philips Medizin Systeme Boblingen GmbH | 2003 | — |
| −19 | K040304 | Philips Intellivue Patient Monitor, Models MP40 & MP50 | Philips Medizin Systeme Boeblingen GmbH | 2004 | — |
| −18 | K041235 | Philips MP20, MP30, MP40, MP50. MP60, MP70 and MP90 Intellivue Patient Monitors | Philips Medizin Systeme Boeblingen GmbH | 2004 | — |
| −17 | K042845 | The Philips MP20, MP30, MP40, MP50, MP60, MP70, and MP90 Intellivue Patient… | Philips Medizin Systeme Boblingen GmbH | 2004 | — |
| −16 | K050141 | The Philips MP20, MP30, MP40, MP50, MP60, MP70 and MP90 Intellivue Patient… | Philips Medizin Systeme Boeblingen GmbH | 2005 | — |
| −15 | K050762 | Philips Intellivue Patient Monitor, Models MP20, MP30, MP40, MP50, MP60, MP70… | Philips Medizin Systeme Boeblingen GmbH | 2005 | — |
| −14 | K051106 | The Philips MP20, MP30, MP40, MP50, MP60, MP70, MP80 and MP90 Intellivue… | Philips Medizin Systeme Boeblingen GmbH | 2005 | — |
| −13 | K052801 | Intellivue Patient Monitors with Release C.02, Models MP20, MP30, MP40, MP50… | Philips Medizin Systeme Boeblingen GmbH | 2005 | — |
| −12 | K053522 | Philips MP40, MP50, MP60, MP70, MP80 and MP90 Intellivue Patient Monitors… | Philips Medizin Systeme Boeblingen GmbH | 2006 | — |
| −11 | K060221 | The Phillips, Model MP20,MP30,MP40,MP50,MP60,MP70,MP80,and MP90 Intellivue… | Philips Medizin Systeme Boeblingen GmbH | 2006 | — |
| −10 | K060541 | The Philips MP20,MP30,MP40,MP50,MP60,MP70,MP80 and MP90 Intellivue Atient… | Philips Medizin Systeme Boblingen GmbH | 2006 | 2 |
| −9 | K061052 | The Philips Intellivue Patient Monitors, Models… | Philips Medical Systems North America, Inc. | 2006 | — |
| −8 | K061610 | Philips Model MP20, MP30, MP40, MP50, MP60, MP70, MP80 and MP90 Intellivue… | Philips Medizin Systeme Boeblingen GmbH | 2006 | — |
| −7 | K063315 | Intellivue Patient Monitors, Models MP20, MP30, MP40, MP50, MP60, MP70, MP80… | Philips Medizin Systeme Boeblingen GmbH | 2006 | — |
| −6 | K071426 | MP2, X2, MP5, MP20, MP30, MP40, MP50, MP60, MP70, MP80, and MP90 Intellivue… | Philips Medizinsysteme Boeblingen GmbH, Cardiac AN | 2007 | — |
| −5 | K083517 | Intellivue Patient Monitor, Models MP2, MP5, MP20, MP30, MP40, MP50, MO60… | Philips Medizinsysteme Boeblingen GmbH, Cardiac AN | 2009 | 2 |
| −4 | K091927 | Intellivue Patient Monitors ,Models MP40, MP50, MP60, MP70, MP80, MP90 | Philips Medizinsysteme Boeblingen GmbH, Cardiac AN | 2009 | — |
| −3 | K100939 | Intellivue Patient Monitor Model: MP2, MP5, MP5T, MP20, MP30, MP40, MP50… | Philips Medizinsysteme Boeblingen GmbH, Cardiac AN | 2010 | — |
| −2 | K101449 | Intellivue Patient Monitor | Philips Medizinsysteme Boeblingen GmbH, Cardiac AN | 2010 | — |
| −1 | K110622 | Intellivue Patient Monitor | Philips Medizinsysteme Boeblingen GmbH, Cardiac AN | 2011 | 2 |
| 0 (this device) | K120366 | Intellivue Patient Monitor | Philips Medizinsysteme Boeblingen GmbH, Cardiac AN | 2012 | — |
| +1 | K122439 | Intellivue Patient Monitor | Philips Medizin Systeme Boeblingen GmbH | 2012 | 4 |
| +1 | K131872 | Intellivue Patient Monitor MX500, Intellivue Patient Monitor MX550 | Philips Medizinsysteme Boeblingen GmbH, Cardiac AN | 2013 | 5 |
| +1 | K130849 | Intellivue Patient Monitor MX400; Intellivue Patient Monitor MX450 | Philips Medizinsysteme Boeblingen GmbH, Cardiac AN | 2013 | 8 |
| +1 | K141015 | Intellivue Patient Monitor MX400, Intellivue Patient Monitor MX450, Intellivue… | Philips Medizinsysteme Boeblingen GmbH, Cardiac AN | 2014 | 1 |
| +1 | K150975 | Philips IntelliVue MP2/MP5/MPSC Patient Monitors, Philips IntelliVue X1 and X2… | Philips Medizin Systeme Boeblingen GmbH | 2015 | 1 |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1985 | 1 | 0 |
| 1986 | 1 | 0 |
| 1987 | 1 | 0 |
| 1989 | 3 | 0 |
| 1990 | 7 | 0 |
| 1991 | 8 | 0 |
| 1992 | 7 | 0 |
| 1993 | 6 | 0 |
| 1994 | 2 | 0 |
| 1995 | 9 | 0 |
| 1996 | 10 | 0 |
| 1997 | 9 | 1 |
| 1998 | 17 | 0 |
| 1999 | 12 | 0 |
| 2000 | 20 | 0 |
| 2001 | 9 | 0 |
| 2002 | 16 | 0 |
| 2003 | 13 | 0 |
| 2004 | 16 | 0 |
| 2005 | 6 | 0 |
| 2006 | 7 | 0 |
| 2007 | 4 | 0 |
| 2008 | 3 | 0 |
| 2009 | 3 | 0 |
| 2010 | 3 | 0 |
| 2011 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2012 | 1 | 0 |
| 2013 | 4 | 0 |
| 2014 | 2 | 0 |
| 2015 | 2 | 0 |
| 2016 | 2 | 0 |
| 2017 | 3 | 0 |
| 2018 | 5 | 0 |
| 2019 | 4 | 0 |
| 2020 | 2 | 0 |
| 2021 | 1 | 0 |
| 2022 | 3 | 0 |
| 2023 | 6 | 0 |
| 2024 | 1 | 0 |
| 2025 | 6 | 0 |
| 2026 | 4 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new MHX device
- #1 K251702 IntelliVue Patient Monitor MX750; IntelliVue…score 23.0 · 40 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (41)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K110474 | Philips Intellivue TCG10Summary page 2…ly cleared Philips IntelliVue Patient Monitors NP2, X2, MPS, MP5T, MPSSC, MP2O, NP30, MP4O, MPSO, MPGO, MP7O, MPSO, MP9O, MXEOO, MX700 and MX800 marketed pursuant to: K110474, K110622, K102562, p46tr#… | 2011 |
| * | K110622 | Intellivue Patient MonitorSummary page 2…ed Philips IntelliVue Patient Monitors NP2, X2, MPS, MP5T, MPSSC, MP2O, NP30, MP4O, MPSO, MPGO, MP7O, MPSO, MP9O, MXEOO, MX700 and MX800 marketed pursuant to: K110474, K110622, K102562, p46tr#… | 2011 |
| * | K102562 | Intellivue Patient Monitor, Models MP2, X2, MP5, MP5T, MP20, MP30, MP40, MP50, MP60…Summary page 2…s IntelliVue Patient Monitors NP2, X2, MPS, MP5T, MPSSC, MP2O, NP30, MP4O, MPSO, MPGO, MP7O, MPSO, MP9O, MXEOO, MX700 and MX800 marketed pursuant to: K110474, K110622, K102562, p46tr#… | 2010 |
| * | K101449 | Intellivue Patient MonitorSummary page 3…p46tr# KilZo)3Ct PHILIPS K(101449, K(100939, K(093268, K(091927, K(091395, K(090360, K(083517, 1(093228, K082633, K(082583, K(081793, K(072070, K(071426, K(063725,… | 2010 |
| * | K100939 | Intellivue Patient Monitor Model: MP2, MP5, MP5T, MP20, MP30, MP40, MP50, MP60, MP70…Summary page 3…p46tr# KilZo)3Ct PHILIPS K(101449, K(100939, K(093268, K(091927, K(091395, K(090360, K(083517, 1(093228, K082633, K(082583, K(081793, K(072070, K(071426, K(063725, K(063315,… | 2010 |
| * | K093268 | Intellivue Patient MonitorSummary page 3…p46tr# KilZo)3Ct PHILIPS K(101449, K(100939, K(093268, K(091927, K(091395, K(090360, K(083517, 1(093228, K082633, K(082583, K(081793, K(072070, K(071426, K(063725, K(063315, K(062392,… | 2010 |
| * | K091927 | Intellivue Patient Monitors ,Models MP40, MP50, MP60, MP70, MP80, MP90Summary page 3…p46tr# KilZo)3Ct PHILIPS K(101449, K(100939, K(093268, K(091927, K(091395, K(090360, K(083517, 1(093228, K082633, K(082583, K(081793, K(072070, K(071426, K(063725, K(063315, K(062392, K(062283,… | 2009 |
| * | K091395 | MP5 Intellivue Patient MonitorSummary page 3…# KilZo)3Ct PHILIPS K(101449, K(100939, K(093268, K(091927, K(091395, K(090360, K(083517, 1(093228, K082633, K(082583, K(081793, K(072070, K(071426, K(063725, K(063315, K(062392, K(062283, K061610,… | 2009 |
| * | K090360 | Philips M3001A Multi Measurement ServerSummary page 3…e 3 === KilZo)3Ct PHILIPS K(101449, K(100939, K(093268, K(091927, K(091395, K(090360, K(083517, 1(093228, K082633, K(082583, K(081793, K(072070, K(071426, K(063725, K(063315, K(062392, K(062283, K061610, 1(061052,… | 2009 |
| * | K083517 | Intellivue Patient Monitor, Models MP2, MP5, MP20, MP30, MP40, MP50, MO60, MP7-, MP80…Summary page 3…KilZo)3Ct PHILIPS K(101449, K(100939, K(093268, K(091927, K(091395, K(090360, K(083517, 1(093228, K082633, K(082583, K(081793, K(072070, K(071426, K(063725, K(063315, K(062392, K(062283, K061610, 1(061052, K(060541,… | 2009 |
| * | K083228 | Philips Intellivue MP5 Patient Monitor Our reading of the summary gave K093228, FDA’s number for Smith & Nephew Twinfix Ultra PK Suture Anchor (Smith & Nephew, Inc.). Corrected to the number one digit away whose FDA name carries the summary’s words for the device, “Philips IntelliVue Patient Monitors”. | 2008 |
| * | K082633 | Intellivue Patient Monitor, Intellivue XDS, Models MP2, MP5, MP20, MP30, MP40, MP50…Summary page 3…KilZo)3Ct PHILIPS K(101449, K(100939, K(093268, K(091927, K(091395, K(090360, K(083517, 1(093228, K082633, K(082583, K(081793, K(072070, K(071426, K(063725, K(063315, K(062392, K(062283, K061610, 1(061052, K(060541, K(060221, K(053522,… | 2008 |
| * | K082583 | Intellivue Patient Monitor, Intellibridge Module, Models MP40, MP50, MP60, MP70, MP80…Summary page 3…KilZo)3Ct PHILIPS K(101449, K(100939, K(093268, K(091927, K(091395, K(090360, K(083517, 1(093228, K082633, K(082583, K(081793, K(072070, K(071426, K(063725, K(063315, K(062392, K(062283, K061610, 1(061052, K(060541, K(060221, K(053522, K(052961,… | 2009 |
| * | K081793 | Philips MP5 and MP5T Intellivue Patient Monitors and TRX4841A Intellivue Telemetry…Summary page 3…KilZo)3Ct PHILIPS K(101449, K(100939, K(093268, K(091927, K(091395, K(090360, K(083517, 1(093228, K082633, K(082583, K(081793, K(072070, K(071426, K(063725, K(063315, K(062392, K(062283, K061610, 1(061052, K(060541, K(060221, K(053522, K(052961, K(052801,… | 2008 |
| * | K072070 | Philips MP2 and X2 Intellivue Patient MonitorsSummary page 3…Zo)3Ct PHILIPS K(101449, K(100939, K(093268, K(091927, K(091395, K(090360, K(083517, 1(093228, K082633, K(082583, K(081793, K(072070, K(071426, K(063725, K(063315, K(062392, K(062283, K061610, 1(061052, K(060541, K(060221, K(053522, K(052961, K(052801, K(051106,… | 2007 |
| * | K071426 | MP2, X2, MP5, MP20, MP30, MP40, MP50, MP60, MP70, MP80, and MP90 Intellivue Patient…Summary page 3…PHILIPS K(101449, K(100939, K(093268, K(091927, K(091395, K(090360, K(083517, 1(093228, K082633, K(082583, K(081793, K(072070, K(071426, K(063725, K(063315, K(062392, K(062283, K061610, 1(061052, K(060541, K(060221, K(053522, K(052961, K(052801, K(051106, K(050762,… | 2007 |
| * | K063725 | MP5 Intellivue Patient MonitorSummary page 3…PHILIPS K(101449, K(100939, K(093268, K(091927, K(091395, K(090360, K(083517, 1(093228, K082633, K(082583, K(081793, K(072070, K(071426, K(063725, K(063315, K(062392, K(062283, K061610, 1(061052, K(060541, K(060221, K(053522, K(052961, K(052801, K(051106, K(050762, K(050141,… | 2007 |
| * | K063315 | Intellivue Patient Monitors, Models MP20, MP30, MP40, MP50, MP60, MP70, MP80 and MP90Summary page 3…PHILIPS K(101449, K(100939, K(093268, K(091927, K(091395, K(090360, K(083517, 1(093228, K082633, K(082583, K(081793, K(072070, K(071426, K(063725, K(063315, K(062392, K(062283, K061610, 1(061052, K(060541, K(060221, K(053522, K(052961, K(052801, K(051106, K(050762, K(050141, K(042845,… | 2006 |
| * | K062392 | Philips MP5 Intellivue Patient MonitorSummary page 3…PHILIPS K(101449, K(100939, K(093268, K(091927, K(091395, K(090360, K(083517, 1(093228, K082633, K(082583, K(081793, K(072070, K(071426, K(063725, K(063315, K(062392, K(062283, K061610, 1(061052, K(060541, K(060221, K(053522, K(052961, K(052801, K(051106, K(050762, K(050141, K(042845, 1(041235,… | 2006 |
| * | K062283 | The Philips MP20, MP30, MP40, MP50, MP60, MP70, MP80, MP90 Intellivue Patient MonitorsSummary page 3…PHILIPS K(101449, K(100939, K(093268, K(091927, K(091395, K(090360, K(083517, 1(093228, K082633, K(082583, K(081793, K(072070, K(071426, K(063725, K(063315, K(062392, K(062283, K061610, 1(061052, K(060541, K(060221, K(053522, K(052961, K(052801, K(051106, K(050762, K(050141, K(042845, 1(041235, 1(040917,… | 2006 |
| * | K061610 | Philips Model MP20, MP30, MP40, MP50, MP60, MP70, MP80 and MP90 Intellivue Patient…Summary page 3…K(101449, K(100939, K(093268, K(091927, K(091395, K(090360, K(083517, 1(093228, K082633, K(082583, K(081793, K(072070, K(071426, K(063725, K(063315, K(062392, K(062283, K061610, 1(061052, K(060541, K(060221, K(053522, K(052961, K(052801, K(051106, K(050762, K(050141, K(042845, 1(041235, 1(040917, K(040304… | 2006 |
| * | K061052 | The Philips Intellivue Patient Monitors, Models MP20,MP30,MP40,MP50,MP60,MP70,MP80,MP90Summary page 3…K(100939, K(093268, K(091927, K(091395, K(090360, K(083517, 1(093228, K082633, K(082583, K(081793, K(072070, K(071426, K(063725, K(063315, K(062392, K(062283, K061610, 1(061052, K(060541, K(060221, K(053522, K(052961, K(052801, K(051106, K(050762, K(050141, K(042845, 1(041235, 1(040917, K(040304, K(033513… | 2006 |
| * | K060541 | The Philips MP20,MP30,MP40,MP50,MP60,MP70,MP80 and MP90 Intellivue Atient MonitorsSummary page 3…K(093268, K(091927, K(091395, K(090360, K(083517, 1(093228, K082633, K(082583, K(081793, K(072070, K(071426, K(063725, K(063315, K(062392, K(062283, K061610, 1(061052, K(060541, K(060221, K(053522, K(052961, K(052801, K(051106, K(050762, K(050141, K(042845, 1(041235, 1(040917, K(040304, K(033513, K(033444… | 2006 |
| * | K060221 | The Phillips, Model MP20,MP30,MP40,MP50,MP60,MP70,MP80,and MP90 Intellivue Patient…Summary page 3…K(091927, K(091395, K(090360, K(083517, 1(093228, K082633, K(082583, K(081793, K(072070, K(071426, K(063725, K(063315, K(062392, K(062283, K061610, 1(061052, K(060541, K(060221, K(053522, K(052961, K(052801, K(051106, K(050762, K(050141, K(042845, 1(041235, 1(040917, K(040304, K(033513, K(033444, K(032858… | 2006 |
| * | K053522 | Philips MP40, MP50, MP60, MP70, MP80 and MP90 Intellivue Patient Monitors, Release D.00Summary page 3…K(091395, K(090360, K(083517, 1(093228, K082633, K(082583, K(081793, K(072070, K(071426, K(063725, K(063315, K(062392, K(062283, K061610, 1(061052, K(060541, K(060221, K(053522, K(052961, K(052801, K(051106, K(050762, K(050141, K(042845, 1(041235, 1(040917, K(040304, K(033513, K(033444, K(032858, K(031481… | 2006 |
| * | K052961 | The Philips MP20, MP30, MP40, MP50, MP60, MP70, and MP90 Intellivue Patient Monitors…Summary page 3…K(090360, K(083517, 1(093228, K082633, K(082583, K(081793, K(072070, K(071426, K(063725, K(063315, K(062392, K(062283, K061610, 1(061052, K(060541, K(060221, K(053522, K(052961, K(052801, K(051106, K(050762, K(050141, K(042845, 1(041235, 1(040917, K(040304, K(033513, K(033444, K(032858, K(031481, K(030038… | 2005 |
| * | K052801 | Intellivue Patient Monitors with Release C.02, Models MP20, MP30, MP40, MP50, MP60…Summary page 3…K(083517, 1(093228, K082633, K(082583, K(081793, K(072070, K(071426, K(063725, K(063315, K(062392, K(062283, K061610, 1(061052, K(060541, K(060221, K(053522, K(052961, K(052801, K(051106, K(050762, K(050141, K(042845, 1(041235, 1(040917, K(040304, K(033513, K(033444, K(032858, K(031481, K(030038, 1(023871… | 2005 |
| * | K051106 | The Philips MP20, MP30, MP40, MP50, MP60, MP70, MP80 and MP90 Intellivue Patient MonitorsSummary page 3…1(093228, K082633, K(082583, K(081793, K(072070, K(071426, K(063725, K(063315, K(062392, K(062283, K061610, 1(061052, K(060541, K(060221, K(053522, K(052961, K(052801, K(051106, K(050762, K(050141, K(042845, 1(041235, 1(040917, K(040304, K(033513, K(033444, K(032858, K(031481, K(030038, 1(023871. The new… | 2005 |
| * | K050762 | Philips Intellivue Patient Monitor, Models MP20, MP30, MP40, MP50, MP60, MP70 and MP90Summary page 3…K082633, K(082583, K(081793, K(072070, K(071426, K(063725, K(063315, K(062392, K(062283, K061610, 1(061052, K(060541, K(060221, K(053522, K(052961, K(052801, K(051106, K(050762, K(050141, K(042845, 1(041235, 1(040917, K(040304, K(033513, K(033444, K(032858, K(031481, K(030038, 1(023871. The new SpO2 intel… | 2005 |
| * | K050141 | The Philips MP20, MP30, MP40, MP50, MP60, MP70 and MP90 Intellivue Patient MonitorsSummary page 3…(082583, K(081793, K(072070, K(071426, K(063725, K(063315, K(062392, K(062283, K061610, 1(061052, K(060541, K(060221, K(053522, K(052961, K(052801, K(051106, K(050762, K(050141, K(042845, 1(041235, 1(040917, K(040304, K(033513, K(033444, K(032858, K(031481, K(030038, 1(023871. The new SpO2 intelligent ala… | 2005 |
| * | K042845 | The Philips MP20, MP30, MP40, MP50, MP60, MP70, and MP90 Intellivue Patient MonitorsSummary page 3…(081793, K(072070, K(071426, K(063725, K(063315, K(062392, K(062283, K061610, 1(061052, K(060541, K(060221, K(053522, K(052961, K(052801, K(051106, K(050762, K(050141, K(042845, 1(041235, 1(040917, K(040304, K(033513, K(033444, K(032858, K(031481, K(030038, 1(023871. The new SpO2 intelligent alarm delay f… | 2004 |
| * | K041235 | Philips MP20, MP30, MP40, MP50. MP60, MP70 and MP90 Intellivue Patient MonitorsSummary page 3…(072070, K(071426, K(063725, K(063315, K(062392, K(062283, K061610, 1(061052, K(060541, K(060221, K(053522, K(052961, K(052801, K(051106, K(050762, K(050141, K(042845, 1(041235, 1(040917, K(040304, K(033513, K(033444, K(032858, K(031481, K(030038, 1(023871. The new SpO2 intelligent alarm delay feature of… | 2004 |
| * | K040917 | Philips Anesthesia Gas Monitor, Model M1026BSummary page 3…(071426, K(063725, K(063315, K(062392, K(062283, K061610, 1(061052, K(060541, K(060221, K(053522, K(052961, K(052801, K(051106, K(050762, K(050141, K(042845, 1(041235, 1(040917, K(040304, K(033513, K(033444, K(032858, K(031481, K(030038, 1(023871. The new SpO2 intelligent alarm delay feature of the Philip… | 2004 |
| * | K040304 | Philips Intellivue Patient Monitor, Models MP40 & MP50Summary page 3…(063725, K(063315, K(062392, K(062283, K061610, 1(061052, K(060541, K(060221, K(053522, K(052961, K(052801, K(051106, K(050762, K(050141, K(042845, 1(041235, 1(040917, K(040304, K(033513, K(033444, K(032858, K(031481, K(030038, 1(023871. The new SpO2 intelligent alarm delay feature of the Philips Intelliv… | 2004 |
| * | K033513 | Easi Ecg AlgorithmSummary page 3…(063315, K(062392, K(062283, K061610, 1(061052, K(060541, K(060221, K(053522, K(052961, K(052801, K(051106, K(050762, K(050141, K(042845, 1(041235, 1(040917, K(040304, K(033513, K(033444, K(032858, K(031481, K(030038, 1(023871. The new SpO2 intelligent alarm delay feature of the Philips Intellivue Patient… | 2004 |
| * | K033444 | Philips Medical Systems, M3012A- Hemodynamic Extension to the Multi-Measurement ServerSummary page 3…(062392, K(062283, K061610, 1(061052, K(060541, K(060221, K(053522, K(052961, K(052801, K(051106, K(050762, K(050141, K(042845, 1(041235, 1(040917, K(040304, K(033513, K(033444, K(032858, K(031481, K(030038, 1(023871. The new SpO2 intelligent alarm delay feature of the Philips Intellivue Patient Monitors… | 2003 |
| * | K032858 | The Philips Intellivue MP40, MP50, MP60, MP70,and MP90 Patient Monitors, Release B.O.Summary page 3…(062283, K061610, 1(061052, K(060541, K(060221, K(053522, K(052961, K(052801, K(051106, K(050762, K(050141, K(042845, 1(041235, 1(040917, K(040304, K(033513, K(033444, K(032858, K(031481, K(030038, 1(023871. The new SpO2 intelligent alarm delay feature of the Philips Intellivue Patient Monitors called 'Sm… | 2003 |
| * | K031481 | The Philips Intellivue MP60, MP70, and MP90 Patient Monitors, Release a.20Summary page 3…061610, 1(061052, K(060541, K(060221, K(053522, K(052961, K(052801, K(051106, K(050762, K(050141, K(042845, 1(041235, 1(040917, K(040304, K(033513, K(033444, K(032858, K(031481, K(030038, 1(023871. The new SpO2 intelligent alarm delay feature of the Philips Intellivue Patient Monitors called 'Smart Alarm… | 2003 |
| * | K030038 | Philips MP60, MP70 and MP90 Intellivue Patient Monitors with Portal Technology and M3185…Summary page 3…061052, K(060541, K(060221, K(053522, K(052961, K(052801, K(051106, K(050762, K(050141, K(042845, 1(041235, 1(040917, K(040304, K(033513, K(033444, K(032858, K(031481, K(030038, 1(023871. The new SpO2 intelligent alarm delay feature of the Philips Intellivue Patient Monitors called 'Smart Alarm Delay' is… | 2003 |
| * | K023871 | M3046A (M2/M3/M4) Compact Portable Patient Monitor and MP60 & MP70 Intellivue Patient…Summary page 3…060541, K(060221, K(053522, K(052961, K(052801, K(051106, K(050762, K(050141, K(042845, 1(041235, 1(040917, K(040304, K(033513, K(033444, K(032858, K(031481, K(030038, 1(023871. The new SpO2 intelligent alarm delay feature of the Philips Intellivue Patient Monitors called 'Smart Alarm Delay' is substantia… | 2002 |
| * | K060576 | Nellcor Oximax N-600X Pulse OximeterSummary page 3…nitors called 'Smart Alarm Delay' is substantially equivalent to the 'SatSeconds' alarm management technique of the Nelloor Pulse Oximeter N-GO0x, marketed pursuant to K(060576. 4. Description of the device The Philips IntelliVue Patient Monitors family comprises the multiparameter patient monitor serie… | 2006 |
What descends from it? (46)
See all 469 later devices cite it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K122439 | Intellivue Patient Monitor | 2012 | 41 |
| K131872 | Intellivue Patient Monitor MX500, Intellivue Patient Monitor MX550 | 2013 | 37 |
| K130849 | Intellivue Patient Monitor MX400; Intellivue Patient Monitor MX450 | 2013 | 37 |
| K141015 | Intellivue Patient Monitor MX400, Intellivue Patient Monitor MX450, Intellivue Patient… | 2014 | 36 |
| K150975 | Philips IntelliVue MP2/MP5/MPSC Patient Monitors, Philips IntelliVue X1 and X2… | 2015 | 34 |
| K151681 | IntelliVue Patient Monitor | 2015 | 32 |
| K133973 | Tempus Pro with Enhanced Features | 2014 | 3 |
| K131913 | Intellivue CL SPO2 Pod and CL NBP Pod, Philips Intellivue Patient Monitors MP5, MP5T… | 2013 | 0 |
| K131829 | Intellivue Patient Monitor | 2013 | 0 |
Has anything in its family been recalled? (1 Class I)
No FDA recall names this device.
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K970209Capnocheck PlusBci Intl., Inc. · 1997 · 6 generations back · ◆ in the 1997 ancestors bar of FIG. 11 Class I recall
What does this lineage say?
- The line of descent spans 18 years and passes through 8 companies; a new generation was cleared every 4 months on median.
- 35 of the 225 ancestors have a recall of any class; 1 has a Class I recall.
- Of its descendants, K122439 (Philips Medizin Systeme Boeblingen GmbH, 2012) has gone furthest, with 41 descendants of its own.
- 7 other devices also cite its predicate K110474.
How does its category compare?
All MHX devicesMHX Monitor, Physiological, Patient (With Arrhythmia Detection Or Alarms): FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.
FDA decisions per year · MHX
8in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · MHX
192in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · MHX
21so far in 2026, the most since 2007
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · MHX
3,761in 2025
MAUDE reports naming product code MHX, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 15 | 96 | 3 | 131 |
| 2008 | 16 | 58 | 2 | 232 |
| 2009 | 13 | 57 | 7 | 382 |
| 2010 | 20 | 90 | 13 | 955 |
| 2011 | 12 | 81 | 16 | 949 |
| 2012 | 16 | 88 | 11 | 676 |
| 2013 | 21 | 120 | 14 | 1,997 |
| 2014 | 13 | 248 | 6 | 1,609 |
| 2015 | 14 | 203 | 5 | 1,284 |
| 2016 | 13 | 149 | 7 | 1,105 |
| 2017 | 14 | 134 | 6 | 1,115 |
| 2018 | 11 | 203 | 12 | 2,157 |
| 2019 | 7 | 243 | 6 | 1,462 |
| 2020 | 11 | 196 | 0 | 1,714 |
| 2021 | 5 | 195 | 0 | 2,423 |
| 2022 | 15 | 189 | 6 | 1,470 |
| 2023 | 8 | 229 | 3 | 2,254 |
| 2024 | 5 | 181 | 3 | 2,814 |
| 2025 | 8 | 192 | 15 | 3,761 |
| 2026 | 14 | 205 | 21 | — |
What is it indicated for?
Models MP2 and X2: The monitor is indicated for use by healthcare professionals whenever there is a need for monitoring the physiological parameters of patients. The monitor is intended to be used for monitoring and recording of, and to generate alarms, for, multiple physiological parameters of adults, pediatrics, and neonates. The monitor is intended for use by trained healthcare professionals in a hospital environment. The monitor is also intended for…
Read the full statement (1,075 words)
Models MP2 and X2:
The monitor is indicated for use by healthcare professionals whenever there is a need for monitoring the physiological parameters of patients.
The monitor is intended to be used for monitoring and recording of, and to generate alarms, for, multiple physiological parameters of adults, pediatrics, and neonates. The monitor is intended for use by trained healthcare professionals in a hospital environment.
The monitor is also intended for use during patient transport inside and outside of the hospital environment.
The monitor is only for use on one patient at a time. It is not intended for home use. Not a therapeutic device. The monitor is for prescription use only.
The ECG measurement is intended to be used for diagnostic recording of rhythm and detailed morphology of complex cardiac complexes (according to AANI EC 11).
ST segment monitoring is intended for use with adult patients only and is not clinically validated for use with neonatal and pediatric patients.
Models MPS, MP5T, and MP5SC:
The monitors are indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients.
The monitors are intended to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults, pediatrics, and neonates. The monitors are intended for use by trained healthcare professionals in a hospital environment.
The MP5, MP5SC and MPST monitors are also intended for use during patient transport inside the hospital environment; only the MAPS monitor is for use during patient transport outside of the hospital environment. The MPS, MP5SC and MPST when used with the TRx4841A/TRx4S5lA IntelliVue Telemetry System Transceiver or with the IntelliVue Cableless Measurement Devices, are intended for use in a hospital environment and during patient transport inside the hospital environment.
The monitors are only for use on one patient at a time. They are not intended for home use. Not therapeutic devices. The monitors are for prescription use only.
The ECG measurement is intended to be used for diagnostic recording of rhythm and detailed morphology of complex cardiac complexes (according to AAMI EC 11).
ST segment monitoring is intended for use with adult patients only and is not clinically validated for use with neonatal and pediatric patients.
The Predictive Temperature unit is intended for use with adult and pediatric patients in a hospital environment.
The SSC Sepsis Protocol, in the ProtocolWatch clinical decision support tool, is intended for use with adult patients only.
The derived measurement Pulse Pressure Variation (PPV) is intended for use with sedated patients receiving controlled mechanical ventilation and mainly free from cardiac arrhythmia. The PPV measurement has been validated only for adult patients.
Models MP2O, MP3O, MP4O, MP5O, MP6O, MP7O, MP8O, and MP9O:
The monitors are indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients.
The monitors are intended to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults, pediatrics, and neonates. The monitors are intended for use by trained healthcare professionals in a hospital environment.
The MP20/MP3O/MP4O/MPSO monitors are additionally intended for use in transport situations within hospital environments.
The monitors are only for use on one patient at a time. They are not intended for home use. Not therapeutic devices. The monitors are for prescription use only.
The ECG measurement is intended to be used for diagnostic recording of rhythm and detailed morphology of complex cardiac complexes (according to AAl EC 11).
ST segment monitoring is intended for use with adult patients only and is not clinically validated for use with neonatal and pediatric patients.
The derived measurement Pulse Pressure Variation (PPV) is intended for use with sedated patients receiving controlled mechanical ventilation and mainly free from cardiac arrhythmia. The PPV measurement has been validated only for adult patients.
The transcutaneous gas measurement (tcGas) is restricted to neonatal patients only.
HIS is intended for use under the direct supervision of a licensed health care practitioner or by personnel trained in its proper use. It is intended for use on adult and pediatric patients within a hospital or medical facility providing patient care to monitor the state of the brain by data acquisition of EEG signals. The BIS may be used as an aid in monitoring the effects of certain anesthetic agents.
Use of HIS monitoring to help guide anesthetic administration may be associated with the reduction of the incidence of awareness with recall in adults during general anesthesia and sedation.
The SSC Sepsis Protocol, in the ProtocolWatch clinical decision support tool, is intended for use with adult patients only.
Models MX600, NX700, and Ma800:
The monitors are indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients.
The monitors are intended to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults, pediatrics, and neonates. The monitors are intended for use by trained healthcare professionals in a hospital environment.
The monitors are only for use on one patient at a time. They are not intended for home use. Not therapeutic devices. The monitors are for prescription use only.
The EGG measurement is intended to be used for diagnostic recording of rhythm and detailed morphology of complex cardiac complexes (according to AAl EC 11).
ST segment monitoring is intended for use with adult patients only and is not clinically validated for use with neonatal and pediatric patients.
The transcutaneous gas measurement (tcGas) is restricted to neonatal patients only.
HIS is intended for use under the direct supervision of a licensed health care practitioner or by personnel trained in its proper use. It is intended for use on adult and pediatric patients within a hospital or medical facility providing patient care to monitor the state of the brain by data acquisition of EEG signals. The HIS may be used as an aid in monitoring the effects of certain anesthetic agents.
Use of HIS monitoring to help guide anesthetic administration may be associated with the reduction of the incidence of awareness with recall in adults during general anesthesia and sedation.
The SSC Sepsis Protocol, in the ProtocolWatch clinical decision support tool, is intended for use with adult patients only.
The derived measurement Pulse Pressure Variation (PPV) is intended for use with sedated patients receiving controlled mechanical ventilation and mainly free from cardiac arrhythmia. The PPV measurement has been validated only for adult patients.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K120366 · printed Nov 3, 2026
Cite this page
510(k) Report. “K120366: Intellivue Patient Monitor.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K120366
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026