Fresenius 2008K Home Hemodialysis Machine with Wireless Wetness Detector (Wetalert)

K121421510(k) SpecialCleared Jan 17, 2013248-day reviewClass II21 CFR 876.5860Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Fresenius Medical Care North America, Design Cente · ONW Hemodialysis System For Home Use

Cleared on Jan 17, 2013 as substantially equivalent to K070049 (Fresenius Medical Care North America, 2011) after a 248-day review. Its family tree reaches back 3 generations to 1994. 3 of the 16 devices in its family, counting itself, have a Class I recall. 12 later devices descend from it. It has been recalled 2 times.

Decided
Jan 17, 2013
Review
248 days
Class
Class II
Product code
ONW
Recalls
2
Predicates
1

Its longest line of descent runs back 3 generations to K921456 (1994). 12 later devices descend from it, 1 citing it directly. 3 of the 16 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2013: 1 descendant cleared2015: 1 descendant cleared, 1 with a Class I recall12014: 1 descendant cleared2016: 1 descendant cleared2018: 1 descendant cleared, 1 with a Class I recall12022: 1 descendant cleared, 1 with a Class I recall12023: 1 descendant cleared2024: 1 descendant cleared2025: 4 descendants clearedK921456 · Fresenius 3008/FDS-08 Dialysate Delivery System · Fresenius USA, Inc. · 1994 · recalledK994267 · Fresenius 2008K · Fresenius Medical Care · 2000 · recalledK070049 · Fresenius 2008K @Home with Wireless Wetness Detector · Fresenius Medical Care North America · 2011 · recalled1990200020102020
Descendants, by generation 1 generation forward: 1 descendant2 generations forward: 2 descendants, 1 with a Class I recall13 generations forward: 1 descendant4 generations forward: 1 descendant, 1 with a Class I recall15 generations forward: 2 descendants, 1 with a Class I recall16 generations forward: 4 descendants7 generations forward: 1 descendant K070049 · Fresenius 2008K @Home with Wireless Wetness Detector · Fresenius Medical Care North America · 2011 · recalledK994267 · Fresenius 2008K · Fresenius Medical Care · 2000 · recalledK921456 · Fresenius 3008/FDS-08 Dialysate Delivery System · Fresenius USA, Inc. · 1994 · recalled0+5
  1. Root · 3 generations backK921456 Fresenius 3008/FDS-08 Dialysate Delivery SystemFresenius USA, Inc. · 1994 · Recalled
  2. 2 generations backK994267 Fresenius 2008KFresenius Medical Care · 2000 · Recalled
  3. 1 generation backK070049 Fresenius 2008K @Home with Wireless Wetness DetectorFresenius Medical Care North America · 2011 · Recalled
  4. This deviceK121421 Fresenius 2008K Home Hemodialysis Machine with Wireless Wetness…Fresenius Medical Care North America, Design Cente · 2013
  5. Descendant · cites it directly · 11 of its ownK124035 2008K@Home Hemodialysis MachineFresenius Medical Care North America · 2013
  6. Descendant · 2 generations on · 9 of its ownK150708 2008K Hemodialysis Machine, 2008K@home Hemodialysis Machine, 2008T…Fresenius Medical Care Renal Therapies Group, LLC · 2015 · Recalled (Class I)
  7. Descendant · 3 generations on · 8 of its ownK153449 2008K2 Hemodialysis MachineFresenius Medical Care Renal Therapies Group, LLC · 2016 · Recalled
  8. Descendant · 4 generations on · 7 of its ownK173972 2008T BlueStar Hemodialysis MachineFresenius Medical Care Renal Therapies Group, LLC · 2018 · Recalled (Class I)
  9. Descendant · 5 generations on · 4 of its ownK231125 2008T BlueStar Hemodialysis MachineFresenius Medical Care Renal Therapies Group, LLC · 2023 · Recalled

Showing 5 of 12 descendants. See all descendants

Descendants cleared per year

2013: 12014: 12015: 12016: 12018: 12022: 12023: 12024: 12025: 4
2013high 4 (2025) · low 0 (2017)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 12 · See all
  • K1240352008K@Home Hemodialysis MachineFresenius Medical Care North America · 2013 · cites it directly11descendants
  • K1507082008K Hemodialysis Machine, 2008K@home Hemodialysis Machine, 2008T Hemodialysis MachineFresenius Medical Care Renal Therapies Group, LLC · 2015 · 2 generations on9descendants
  • K1534492008K2 Hemodialysis MachineFresenius Medical Care Renal Therapies Group, LLC · 2016 · 3 generations on8descendants
  • K1739722008T BlueStar Hemodialysis MachineFresenius Medical Care Renal Therapies Group, LLC · 2018 · 4 generations on7descendants
  • K2311252008T BlueStar Hemodialysis MachineFresenius Medical Care Renal Therapies Group, LLC · 2023 · 5 generations on4descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K921456Fresenius 3008/FDS-08 Dialysate Delivery SystemFresenius USA, Inc.19943
−2K994267Fresenius 2008KFresenius Medical Care20005
−1K070049Fresenius 2008K @Home with Wireless Wetness DetectorFresenius Medical Care North America20117
0 (this device)K121421Fresenius 2008K Home Hemodialysis Machine with Wireless Wetness Detector…Fresenius Medical Care North America, Design Cente2013—
+1K1240352008K@Home Hemodialysis MachineFresenius Medical Care North America2013—
+2K1507082008K Hemodialysis Machine, 2008K@home Hemodialysis Machine, 2008T…Fresenius Medical Care Renal Therapies Group, LLC2015Class I
+3K1534492008K2 Hemodialysis MachineFresenius Medical Care Renal Therapies Group, LLC20162
+4K1739722008T BlueStar Hemodialysis MachineFresenius Medical Care Renal Therapies Group, LLC2018Class I
+5K2311252008T BlueStar Hemodialysis MachineFresenius Medical Care Renal Therapies Group, LLC20234

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201310
201410
201511
201610
201811
202211
202310
202410
202540
FIG. 1 — Lineage of K121421, 1994–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new ONW device

  1. #1 K242269 SC+ Hemodialysis Device (SC-14269); Dialysate…score 11.0 · 10 generations · no Class I recall
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K070049Fresenius 2008K @Home with Wireless Wetness Detector
Summary page 1
…1per §876.5860 Legally Marketed Predicate Device (unmodified devices): The Fresenius 2008K@home Hemodialysis Machine with Wireless Wetness Detector (WetAlert) system (K070049) originally 03 February 2011. Device Description: The 2008K@home Hemodialysis Machine with Wireless Wetness Detector is intended…
Fresenius Medical Care North America20117

What descends from it? (12)

See all 12

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K1240352008K@Home Hemodialysis MachineFresenius Medical Care North America201311

Has anything in its family been recalled? (3 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 20, 2015Class IIZ-1549-2015Software designThe Acetate value for GranuFlo on the Select Concentrate screen should be 8mEq/l but it is displayed as 4mEq/l. The correct value is shown on the dialysate screen.
Dec 16, 2014Class IIZ-1015-2015Under Investigation by firmUnqualified power assemblies shipped to customers.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K1507082008K Hemodialysis Machine, 2008K@home Hemodialysis Machine, 2008T Hemodialysis MachineFresenius Medical Care Renal Therapies Group, LLC · 2015 · 2 generations on · ◆ in the 2015 descendants bar of FIG. 18 Class I recalls
  • K1739722008T BlueStar Hemodialysis MachineFresenius Medical Care Renal Therapies Group, LLC · 2018 · 4 generations on · ◆ in the 2018 descendants bar of FIG. 18 Class I recalls
  • K2229522008T BlueStar Hemodailysis MachineFresenius Medical Care Renal Therapies Group, LLC · 2022 · 5 generations on · ◆ in the 2022 descendants bar of FIG. 18 Class I recalls

What does this lineage say?

  • The line of descent spans 19 years and passes through 4 companies; a new generation was cleared every 5.9 years on median.
  • 3 of the 3 ancestors have a recall of any class.
  • Of its descendants, K124035 (Fresenius Medical Care North America, 2013) has gone furthest, with 11 descendants of its own.

How does its category compare?

All ONW devices

ONW Hemodialysis System For Home Use: FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.

FDA decisions per year · ONW

1in 2024

2011: 12013: 22024: 1
2007high 2 (2013) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · ONW

Too few ONW 510(k)s in any year since 2007 to give a median.

Recalls per year · ONW

3in 2013

2011: 12013: 3
2007high 3 (2013) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · ONW

1in 2025

2012: 12013: 82014: 352015: 282016: 192017: 112018: 122019: 32020: 62021: 72022: 52023: 12025: 1
2007high 35 (2014) · low 0 (2007)2025

MAUDE reports naming product code ONW, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20100—00
20111—10
20120—01
20132—38
20140—035
20150—028
20160—019
20170—011
20180—012
20190—03
20200—06
20210—07
20220—05
20230—01
20241—00
20250—01
20260—0—

What is it indicated for?

The Fresenius 2008K@home is indicated for acute and chronic dialysis therapy in an acute or chronic facility. The 2008K@home is also indicated for hemodialysis in the home and must be observed by a trained and qualified person as prescribed by their physician.

The Wireless Wetness Detector is indicated for use with the Fresenius 2008K@home hemodialysis machine and is an optional accessory to aid in the detection of blood and water leaks during hemodialysis. Home hemodialysis using the detector must be observed by a trained and qualified person as prescribed by their physician.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K121421: Fresenius 2008K Home Hemodialysis Machine with Wireless Wetness Detector (Wetalert).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K121421

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026