510(k) Report

510(k) · Cleared

Recalled · Class II · 2017

Fresenius 2008T Hemodialysis Machine

K120505510(k) SpecialCleared Mar 6, 201214-day reviewClass II21 CFR 876.5860Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Fresenius Medical Care North America, Design Cente · KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System

Cleared on Mar 6, 2012 as substantially equivalent to K113427 (Fresenius Medical Care North America, 2011) after a 14-day review. Its family tree reaches back 6 generations to 1994. 1 of the 24 devices in its family, counting itself, has a Class I recall. 17 later devices descend from it. It has been recalled 4 times.

Decided
Mar 6, 2012
Review
14 days
Class
Class II
Product code
KDI
Recalls
4
Predicates
1

Its longest line of descent runs back 6 generations to K921456 (1994). 17 later devices descend from it, 1 citing it directly. 1 of the 24 devices in this family, counting this one, has a Class I recall.

Descendants, by year cleared 2014: 1 descendant cleared2020: 2 descendants cleared2015: 1 descendant cleared2016: 1 descendant cleared2021: 2 descendants cleared2019: 1 descendant cleared2022: 2 descendants cleared2024: 4 descendants cleared2023: 2 descendants cleared, 1 with a Class I recall12026: 1 descendant clearedK921456 · Fresenius 3008/FDS-08 Dialysate Delivery System · Fresenius USA, Inc. · 1994 · recalledK994267 · Fresenius 2008K · Fresenius Medical Care · 2000 · recalledK080964 · Fresenius Model 2008T Hemodialysis Machine · Fresenius Medical Care North America · 2009 · recalledK093902 · Modification to:Fresenius 2008T Hemodialysis Machine, Model 190573 · Fresenius Medical Care North America · 2010 · recalled×2: 2 generations cleared close together, K111639 (2011), K113427 (2011) 1990200020102020
  1. Root · 6 generations backK921456 Fresenius 3008/FDS-08 Dialysate Delivery SystemFresenius USA, Inc. · 1994 · Recalled
  2. 5 generations backK994267 Fresenius 2008KFresenius Medical Care · 2000 · Recalled
  3. 4 generations backK080964 Fresenius Model 2008T Hemodialysis MachineFresenius Medical Care North America · 2009 · Recalled
  4. 3 generations backK093902 Modification to:Fresenius 2008T Hemodialysis Machine, Model 190573Fresenius Medical Care North America · 2010 · Recalled
  5. 2 generations backK111639 Fresenius 2008T Hemodialysis MachineFresenius Medical Care · 2011 · Recalled
  6. 1 generation backK113427 Fresenius 2008T Hemodialysis MachineFresenius Medical Care North America · 2011 · Recalled
  7. This deviceK120505 Fresenius 2008T Hemodialysis MachineFresenius Medical Care North America, Design Cente · 2012
  8. Descendant · cites it directly · 16 of its ownK140866 Tablo ConsoleHome Dialysis Plus · 2014
  9. Descendant · 2 generations on · 11 of its ownK150880 Tablo Console with Wireless CapabilityOutset Medical · 2015
  10. Descendant · 3 generations on · 10 of its ownK160881 Tablo ConsoleOutset Medical, Inc. · 2016
  11. Descendant · 4 generations on · 7 of its ownK200741 Tablo Hemodialysis SystemOutset Medical, Inc. · 2020 · Recalled
  12. Descendant · 5 generations on · 6 of its ownK211370 Tablo Hemodialysis SystemOutset Medical, Inc. · 2022 · Recalled

Showing 5 of 17 descendants. See all descendants

Descendants cleared per year

2014: 12015: 12016: 12019: 12020: 22021: 22022: 22023: 22024: 42026: 1
2014high 4 (2024) · low 0 (2017)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 17 · See all
  • K140866Tablo ConsoleHome Dialysis Plus · 2014 · cites it directly16descendants
  • K150880Tablo Console with Wireless CapabilityOutset Medical · 2015 · 2 generations on11descendants
  • K160881Tablo ConsoleOutset Medical, Inc. · 2016 · 3 generations on10descendants
  • K200741Tablo Hemodialysis SystemOutset Medical, Inc. · 2020 · 4 generations on7descendants
  • K211370Tablo Hemodialysis SystemOutset Medical, Inc. · 2022 · 5 generations on6descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K921456Fresenius 3008/FDS-08 Dialysate Delivery SystemFresenius USA, Inc.19943
−5K994267Fresenius 2008KFresenius Medical Care20005
−4K080964Fresenius Model 2008T Hemodialysis MachineFresenius Medical Care North America20094
−3K093902Modification to:Fresenius 2008T Hemodialysis Machine, Model 190573Fresenius Medical Care North America20102
−2K111639Fresenius 2008T Hemodialysis MachineFresenius Medical Care20111
−1K113427Fresenius 2008T Hemodialysis MachineFresenius Medical Care North America20114
0 (this device)K120505Fresenius 2008T Hemodialysis MachineFresenius Medical Care North America, Design Cente2012—
+1K140866Tablo ConsoleHome Dialysis Plus2014—
+2K150880Tablo Console with Wireless CapabilityOutset Medical2015—
+3K160881Tablo ConsoleOutset Medical, Inc.2016—
+4K200741Tablo Hemodialysis SystemOutset Medical, Inc.20201
+5K211370Tablo Hemodialysis SystemOutset Medical, Inc.20221

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201410
201510
201610
201910
202020
202120
202220
202321
202440
202610
FIG. 1 — Lineage of K120505, 1994–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KDI device

  1. #1 K243874 FX CorAL 60 (F00012969); FX CorAL 80…score 18.2 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K113427Fresenius 2008T Hemodialysis Machine
Summary page 2
…Machine Special 510(k) Notification Device Description The Fresenius 2008T Hemodialysis Machine (K113427) is indicated for acute and chronic dialysis therapy. It is designed to provide hemodialysis treatment by controlling and monitor…
Fresenius Medical Care North America20114

What descends from it? (17)

See all 17

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K140866Tablo ConsoleHome Dialysis Plus201416

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 5, 2017Class IIZ-0829-2018Software Design ChangeWhile reviewing documentation for the next software release of the 2008T, an R&D technician identified that setting the UF goal to '0' introduces a discrepancy between the UF rate displayed and the actual UF pump rate. The software…
Mar 10, 2016Class IIZ-1325-2016Device DesignAdded Warning Statement: Fresenius 2008 Series Hemodialysis Machines False Blood Leak Alarm When Dialyzing Patients Treated with Hydroxocobalamin (or any form of Vitamin B-12)
Feb 20, 2015Class IIZ-1550-2015Software designThe Acetate value for GranuFlo on the Select Concentrate screen should be 8mEq/l but it is displayed as 4mEq/l. The correct value is shown on the dialysate screen.
May 24, 2013Class IIZ-1693-2013Nonconforming Material/ComponentIncorrect rubber nitrile O-rings distributed instead of the correct material EPDM. (Ethylene Propylene Dine Monomer (M-class) rubber)

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K233335Tablo® Hemodialysis SystemOutset Medical, Inc. · 2023 · 7 generations on · ◆ in the 2023 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 18 years and passes through 4 companies; a new generation was cleared every 1.4 years on median.
  • 6 of the 6 ancestors have a recall of any class.
  • Of its descendants, K140866 (Home Dialysis Plus, 2014) has gone furthest, with 16 descendants of its own.

How does its category compare?

All KDI devices

KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.

FDA decisions per year · KDI

9in 2025

2007: 112008: 62009: 42010: 92011: 52012: 62013: 112014: 82015: 52016: 52017: 82018: 22019: 42020: 92021: 22022: 72023: 72024: 72025: 92026 (partial year): 6
2007high 11 (2007) · low 2 (2018)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KDI

199in 2025

2007: 91 days median2008: 148 days median2009: 236 days median2010: 186 days median2011: 85 days median2012: 153 days median2013: 247 days median2014: 164 days median2015: 69 days median2016: 230 days median2017: 188 days median2019: 263 days median2020: 266 days median2022: 259 days median2023: 157 days median2024: 235 days median2025: 199 days median2026 (partial year): 154 days median
2007high 266 (2020) · low 69 (2015)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KDI

2in 2025

2007: 42009: 32010: 12011: 42012: 12013: 22014: 32015: 52016: 42017: 12018: 12019: 62021: 22022: 172023: 8, 8 Class I2024: 4, 2 Class I2025: 22026 (partial year): 2
2007high 17 (2022) · low 0 (2008)2026

Class I (10 since 2007) Class II/III not classified

Adverse event reports per year · KDI

2,786in 2025

2007: 9162008: 1,0422009: 5222010: 8232011: 2,2702012: 1,7652013: 2,3602014: 6,1542015: 4,1822016: 7662017: 9632018: 1,7162019: 1,5872020: 1,6332021: 1,9442022: 2,1802023: 2,6922024: 2,4412025: 2,786
2007high 6,154 (2014) · low 522 (2009)2025

MAUDE reports naming product code KDI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200711914916
2008614801,042
200942363522
201091861823
201158542,270
2012615311,765
20131124722,360
2014816436,154
201556954,182
201652304766
201781881963
20182—11,716
2019426361,587
2020926601,633
20212—21,944
20227259172,180
202371578 (8)2,692
202472354 (2)2,441
2025919922,786
202661542—

What is it indicated for?

Fresenius 2008T Hemodialysis Machine is indicated for acute and chronic dialysis therapy.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K120505: Fresenius 2008T Hemodialysis Machine.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K120505

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026