510(k) Report

510(k) · Cleared

Recalled · Class I · 2024

2008T BlueStar Hemodailysis Machine

K222952510(k) TraditionalCleared Oct 27, 202230-day reviewClass II21 CFR 876.5860Recalls 12 (8 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Fresenius Medical Care Renal Therapies Group, LLC · KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System

Cleared on Oct 27, 2022 as substantially equivalent to K173972 (Fresenius Medical Care Renal Therapies Group, LLC, 2018) after a 30-day review. Its family tree reaches back 11 generations to 1994. 3 of the 23 devices in its family, counting itself, have a Class I recall. 3 later devices descend from it. It has been recalled 12 times, including 8 Class I recalls.

Decided
Oct 27, 2022
Review
30 days
Class
Class II
Product code
KDI
Recalls
12 (8 Class I)
Predicates
1

Its longest line of descent runs back 11 generations to K921456 (1994). 19 ancestors in all, the oldest K873155 (1987). 3 later devices descend from it, 1 citing it directly. 3 of the 23 devices in this family, counting this one, have a Class I recall.

Other ancestors, by generationDescendants, by generation 4 generations back: 1 other ancestor5 generations back: 2 other ancestors6 generations back: 2 other ancestors8 generations back: 2 other ancestors9 generations back: 1 other ancestor 1 generation forward: 1 descendant2 generations forward: 1 descendant3 generations forward: 1 descendant K173972 · 2008T BlueStar Hemodialysis Machine · Fresenius Medical Care Renal Therapies Group, LLC · 2018 · Class I recallK153449 · 2008K2 Hemodialysis Machine · Fresenius Medical Care Renal Therapies Group, LLC · 2016 · recalledK150708 · 2008K Hemodialysis Machine, 2008K@home Hemodialysis Machine, 2008T Hemodialysis Machine · Fresenius Medical Care Renal Therapies Group, LLC · 2015 · Class I recallK124035 · 2008K@Home Hemodialysis Machine · Fresenius Medical Care North America · 2013K121341 · Fresenius 2008T Hemodialysis Machine with Bibag System · Fresenius Medical Care North America · 2012 · recalledK120017 · Fresenius 2008 Hemodiaysis Machine with Bibag System · Fresenius Medical Care North America · 2012 · recalledK101715 · Fresenius 2008T Hemodialysis Machine Model 190573 · Fresenius Medical Care North America, Design Cente · 2010 · recalledK093902 · Modification to:Fresenius 2008T Hemodialysis Machine, Model 190573 · Fresenius Medical Care North America · 2010 · recalledK080964 · Fresenius Model 2008T Hemodialysis Machine · Fresenius Medical Care North America · 2009 · recalledK994267 · Fresenius 2008K · Fresenius Medical Care · 2000 · recalledK921456 · Fresenius 3008/FDS-08 Dialysate Delivery System · Fresenius USA, Inc. · 1994 · recalled−10−50
  1. Root · 11 generations backK921456 Fresenius 3008/FDS-08 Dialysate Delivery SystemFresenius USA, Inc. · 1994 · Recalled
  2. 7 generations between
  3. 3 generations backK150708 2008K Hemodialysis Machine, 2008K@home Hemodialysis Machine, 2008T…Fresenius Medical Care Renal Therapies Group, LLC · 2015 · Recalled (Class I)
  4. 2 generations backK153449 2008K2 Hemodialysis MachineFresenius Medical Care Renal Therapies Group, LLC · 2016 · Recalled
  5. 1 generation backK173972 2008T BlueStar Hemodialysis MachineFresenius Medical Care Renal Therapies Group, LLC · 2018 · Recalled (Class I)
  6. This deviceK222952 2008T BlueStar Hemodailysis MachineFresenius Medical Care Renal Therapies Group, LLC · 2022
  7. Descendant · cites it directly · 2 of its ownK231534 5008X Hemodialysis SystemFresenius Medical Care · 2024 · Recalled
  8. Descendant · 2 generations on · 1 of its ownK243505 5008X Hemodialysis SystemFresenius Medical Care Renal Therapies Group, LLC · 2025 · Recalled
  9. Descendant · 3 generations on · 0 of its ownK251851 DIASAFEplusUS (F00013010)Fresenius Medical Care Renal Therapies Group, LLC · 2025
Show all 11 generations
  1. Root · 11 generations backK921456 Fresenius 3008/FDS-08 Dialysate Delivery SystemFresenius USA, Inc. · 1994 · Recalled
  2. 10 generations backK994267 Fresenius 2008KFresenius Medical Care · 2000 · Recalled
  3. 9 generations backK080964 Fresenius Model 2008T Hemodialysis MachineFresenius Medical Care North America · 2009 · Recalled
  4. 8 generations backK093902 Modification to:Fresenius 2008T Hemodialysis Machine, Model 190573Fresenius Medical Care North America · 2010 · Recalled
  5. 7 generations backK101715 Fresenius 2008T Hemodialysis Machine Model 190573Fresenius Medical Care North America, Design Cente · 2010 · Recalled
  6. 6 generations backK120017 Fresenius 2008 Hemodiaysis Machine with Bibag SystemFresenius Medical Care North America · 2012 · Recalled
  7. 5 generations backK121341 Fresenius 2008T Hemodialysis Machine with Bibag SystemFresenius Medical Care North America · 2012 · Recalled
  8. 4 generations backK124035 2008K@Home Hemodialysis MachineFresenius Medical Care North America · 2013
  9. 3 generations backK150708 2008K Hemodialysis Machine, 2008K@home Hemodialysis Machine, 2008T…Fresenius Medical Care Renal Therapies Group, LLC · 2015 · Recalled (Class I)
  10. 2 generations backK153449 2008K2 Hemodialysis MachineFresenius Medical Care Renal Therapies Group, LLC · 2016 · Recalled
  11. 1 generation backK173972 2008T BlueStar Hemodialysis MachineFresenius Medical Care Renal Therapies Group, LLC · 2018 · Recalled (Class I)
  12. This deviceK222952 2008T BlueStar Hemodailysis MachineFresenius Medical Care Renal Therapies Group, LLC · 2022

Class I recalls in this family

Most influential descendants

Showing 3 of 3 · See all
  • K2315345008X Hemodialysis SystemFresenius Medical Care · 2024 · cites it directly2descendants
  • K2435055008X Hemodialysis SystemFresenius Medical Care Renal Therapies Group, LLC · 2025 · 2 generations on1descendant
  • K251851DIASAFEplusUS (F00013010)Fresenius Medical Care Renal Therapies Group, LLC · 2025 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−11K921456Fresenius 3008/FDS-08 Dialysate Delivery SystemFresenius USA, Inc.19943
−10K994267Fresenius 2008KFresenius Medical Care20005
−9K080964Fresenius Model 2008T Hemodialysis MachineFresenius Medical Care North America20094
−8K093902Modification to:Fresenius 2008T Hemodialysis Machine, Model 190573Fresenius Medical Care North America20102
−7K101715Fresenius 2008T Hemodialysis Machine Model 190573Fresenius Medical Care North America, Design Cente20101
−6K120017Fresenius 2008 Hemodiaysis Machine with Bibag SystemFresenius Medical Care North America20123
−5K121341Fresenius 2008T Hemodialysis Machine with Bibag SystemFresenius Medical Care North America20125
−4K1240352008K@Home Hemodialysis MachineFresenius Medical Care North America2013—
−3K1507082008K Hemodialysis Machine, 2008K@home Hemodialysis Machine, 2008T…Fresenius Medical Care Renal Therapies Group, LLC2015Class I
−2K1534492008K2 Hemodialysis MachineFresenius Medical Care Renal Therapies Group, LLC20162
−1K1739722008T BlueStar Hemodialysis MachineFresenius Medical Care Renal Therapies Group, LLC2018Class I
0 (this device)K2229522008T BlueStar Hemodailysis MachineFresenius Medical Care Renal Therapies Group, LLC2022—
+1K2315345008X Hemodialysis SystemFresenius Medical Care20241
+2K2435055008X Hemodialysis SystemFresenius Medical Care Renal Therapies Group, LLC20256
+3K251851DIASAFEplusUS (F00013010)Fresenius Medical Care Renal Therapies Group, LLC2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198710
199410
199710
200210
201110
201210
201320

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202410
202520
FIG. 1 — Lineage of K222952, 1994–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KDI device

  1. #1 K243874 FX CorAL 60 (F00012969); FX CorAL 80…score 18.2 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K1739722008T BlueStar Hemodialysis Machine
Summary page 4
…he predicate device for the 2008T BlueStar Hemodialysis Machine (hereinafter referred to as the “2008T Machine”) is the 2008T BlueStar Hemodialysis Machine cleared under K173972. 5.4. Device Description 5.4.1. Device Identification The 2008T Machines with updated hydraulics will be implemented…
Fresenius Medical Care Renal Therapies Group, LLC2018Class I

Also cleared against K173972: 1 other device →

What descends from it? (3)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K2315345008X Hemodialysis SystemFresenius Medical Care20242

Has anything in its family been recalled? (3 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 17, 2024Class IIZ-1867-2024Device DesignAn increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
Apr 17, 2024Class IIZ-1869-2024Device DesignAn increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
Apr 17, 2024Class IIZ-1870-2024Device DesignAn increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
Apr 17, 2024Class IIZ-1868-2024Device DesignAn increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
Sep 6, 2023Class IZ-0028-2024Device DesignPotential PCBA leaching from tubing of hemodialysis machines.
Sep 6, 2023Class IZ-0031-2024Device DesignPotential PCBA leaching from tubing of hemodialysis machines.
Sep 6, 2023Class IZ-0035-2024Device DesignPotential PCBA leaching from tubing of hemodialysis machines.
Sep 6, 2023Class IZ-0029-2024Device DesignPotential PCBA leaching from tubing of hemodialysis machines.
Sep 6, 2023Class IZ-0030-2024Device DesignPotential PCBA leaching from tubing of hemodialysis machines.
Sep 6, 2023Class IZ-0032-2024Device DesignPotential PCBA leaching from tubing of hemodialysis machines.
Sep 6, 2023Class IZ-0033-2024Device DesignPotential PCBA leaching from tubing of hemodialysis machines.
Sep 6, 2023Class IZ-0034-2024Device DesignPotential PCBA leaching from tubing of hemodialysis machines.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K1739722008T BlueStar Hemodialysis MachineFresenius Medical Care Renal Therapies Group, LLC · 2018 · Direct predicate · ◆ on the line of descent in FIG. 18 Class I recalls
  • K1507082008K Hemodialysis Machine, 2008K@home Hemodialysis Machine, 2008T Hemodialysis MachineFresenius Medical Care Renal Therapies Group, LLC · 2015 · 3 generations back · ◆ on the line of descent in FIG. 18 Class I recalls

What does this lineage say?

  • The line of descent spans 28 years and passes through 5 companies; a new generation was cleared every 1.4 years on median.
  • 13 of the 19 ancestors have a recall of any class; 2 have a Class I recall.
  • Of its descendants, K231534 (Fresenius Medical Care, 2024) has gone furthest, with 2 descendants of its own.
  • 1 other device also cites its predicate K173972.

How does its category compare?

All KDI devices

KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2022, this device’s year, is highlighted.

FDA decisions per year · KDI

9in 2025

2007: 112008: 62009: 42010: 92011: 52012: 62013: 112014: 82015: 52016: 52017: 82018: 22019: 42020: 92021: 22022: 72023: 72024: 72025: 92026 (partial year): 6
2007high 11 (2007) · low 2 (2018)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KDI

199in 2025

2007: 91 days median2008: 148 days median2009: 236 days median2010: 186 days median2011: 85 days median2012: 153 days median2013: 247 days median2014: 164 days median2015: 69 days median2016: 230 days median2017: 188 days median2019: 263 days median2020: 266 days median2022: 259 days median2023: 157 days median2024: 235 days median2025: 199 days median2026 (partial year): 154 days median
2007high 266 (2020) · low 69 (2015)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KDI

2in 2025

2007: 42009: 32010: 12011: 42012: 12013: 22014: 32015: 52016: 42017: 12018: 12019: 62021: 22022: 172023: 8, 8 Class I2024: 4, 2 Class I2025: 22026 (partial year): 2
2007high 17 (2022) · low 0 (2008)2026

Class I (10 since 2007) Class II/III not classified

Adverse event reports per year · KDI

2,786in 2025

2007: 9162008: 1,0422009: 5222010: 8232011: 2,2702012: 1,7652013: 2,3602014: 6,1542015: 4,1822016: 7662017: 9632018: 1,7162019: 1,5872020: 1,6332021: 1,9442022: 2,1802023: 2,6922024: 2,4412025: 2,786
2007high 6,154 (2014) · low 522 (2009)2025

MAUDE reports naming product code KDI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200711914916
2008614801,042
200942363522
201091861823
201158542,270
2012615311,765
20131124722,360
2014816436,154
201556954,182
201652304766
201781881963
20182—11,716
2019426361,587
2020926601,633
20212—21,944
20227259172,180
202371578 (8)2,692
202472354 (2)2,441
2025919922,786
202661542—

What is it indicated for?

The 2008T BlueStar Hemodialysis Machine is indicated for acute and chronic dialysis therapy in a healthcare facility. Additional therapy options for patients receiving hemodialysis include: Isolated Ultrafiltration, Sustained Low Efficiency Dialysis (SLED), and low volume hemodialysis (patients weighing > 20 kg and < 40 kg). This machine accommodates the use of both low flux and high flux dialyzers. The SLED therapy option is not to be used for patients…

Read the full statement (280 words)

The 2008T BlueStar Hemodialysis Machine is indicated for acute and chronic dialysis therapy in a healthcare facility.

Additional therapy options for patients receiving hemodialysis include: Isolated Ultrafiltration, Sustained Low Efficiency Dialysis (SLED), and low volume hemodialysis (patients weighing > 20 kg and < 40 kg). This machine accommodates the use of both low flux and high flux dialyzers. The SLED therapy option is not to be used for patients weighing < 40 kg. The 2008T BlueStar Hemodialysis Machine is not to be used for plasma replacement therapies, for patients weighing less than 20 kg, or for renal therapies using substitution fluid.

bibag System (Optional):

The bibag system is used with three stream proportioning Hemodialysis Machines equipped with the bibag module such as the 2008T BlueStar Hemodialysis Machine and intended for use in bicarbonate hemodialysis for acute and chronic renal failure. The bibag is intended for extracorporeal bicarbonate hemodialysis according to a physician’s prescription.

Crit-Line Clip Monitor (CLiC) (Optional):

The Crit-Line Clip Monitor is used with the 2008T BlueStar Hemodialysis Machine to non-invasively measure hematocrit, oxygen saturation and percent change in blood volume. The CLiC device measures hematocrit, percent change in blood volume and oxygen saturation in real time for application in the treatment of dialysis patients with the intended purpose of providing a more effective treatment for both the dialysis patient and the clinician. Based on the data that the monitor provides, the clinician/nurse, under physician direction, intervenes (i.e., increases or decreases the rate at which fluid is removed from the blood) in order to remove the maximum amount of fluid from the dialysis patient without the patient experiencing the common complications of dialysis which include nausea, cramping and vomiting

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K222952: 2008T BlueStar Hemodailysis Machine.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K222952

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026