Liebel-Flarsheim Direct Digital Imaging System

K121838510(k) TraditionalCleared Jul 11, 201219-day reviewClass II21 CFR 892.1680Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Liebel-Flarsheim · MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)

Cleared on Jul 11, 2012 as substantially equivalent to K943581 (Mallinckrodt Group, Inc., 1995) and 2 other predicates after a 19-day review. Its family tree reaches back 9 generations to 1980. It has been recalled 2 times.

Decided
Jul 11, 2012
Review
19 days
Class
Class II
Product code
MQB
Recalls
2
Predicates
3

Its longest line of descent runs back 9 generations to K801623 (1980). 72 ancestors in all. No later device cites it as a predicate yet.

Other ancestors, by year cleared 2011: 1 other ancestor1995: 13 other ancestors2010: 1 other ancestor2001: 2 other ancestors2002: 5 other ancestors1999: 4 other ancestors2000: 3 other ancestors1998: 3 other ancestors1997: 5 other ancestors1996: 5 other ancestors1991: 5 other ancestors2004: 2 other ancestors1994: 3 other ancestors1993: 3 other ancestors1992: 2 other ancestors1989: 1 other ancestor1986: 2 other ancestors1988: 2 other ancestors1981: 1 other ancestorK801623 · Cavitron Ultrasonic Surg. Aspir. Cusa · Cavitron Corp. · 1980K962952 · Olympus Ultrasonic Surgical System (Usu) · Olympus Optical, Co. · 1996×2: 2 generations cleared close together, K972114 (1998), K981993 (1998) K994231 · Siemens Integrated Operating System (Sios) · Siemens Medical Corp. · 2000×2: 2 generations cleared close together, K052202 (2006), K061054 (2006) K062623 · Axiom Luminos DRF · Siemens Medical Solutions USA, Inc. · 2007 · recalledK101491 · Uroskop Omnia · Siemens Medical Solutions USA, Inc. · 2010 · recalled1980198519901995200020052010
  1. Root · 9 generations backK801623 Cavitron Ultrasonic Surg. Aspir. CusaCavitron Corp. · 1980
  2. 5 generations between
  3. 3 generations backK061054 Axiom Aristos FX PlusSiemens Medical Solutions USA, Inc. · 2006 · Recalled
  4. 2 generations backK062623 Axiom Luminos DRFSiemens Medical Solutions USA, Inc. · 2007 · Recalled
  5. 1 generation backK101491 Uroskop OmniaSiemens Medical Solutions USA, Inc. · 2010 · Recalled
  6. This deviceK121838 Liebel-Flarsheim Direct Digital Imaging SystemLiebel-Flarsheim · 2012
Show all 9 generations
  1. Root · 9 generations backK801623 Cavitron Ultrasonic Surg. Aspir. CusaCavitron Corp. · 1980
  2. 8 generations backK962952 Olympus Ultrasonic Surgical System (Usu)Olympus Optical, Co. · 1996
  3. 7 generations backK972114 Olympus Sonosurg SystemThe Olympus Optical Co. · 1998
  4. 6 generations backK981993 Olympus Integrated Endosurgery System EndoalphaThe Olympus Optical Co. · 1998
  5. 5 generations backK994231 Siemens Integrated Operating System (Sios)Siemens Medical Corp. · 2000
  6. 4 generations backK052202 Axiom Artis - Modular Angiographic SystemSiemens Medical Solutions USA, Inc. · 2006 · Recalled
  7. 3 generations backK061054 Axiom Aristos FX PlusSiemens Medical Solutions USA, Inc. · 2006 · Recalled
  8. 2 generations backK062623 Axiom Luminos DRFSiemens Medical Solutions USA, Inc. · 2007 · Recalled
  9. 1 generation backK101491 Uroskop OmniaSiemens Medical Solutions USA, Inc. · 2010 · Recalled
  10. This deviceK121838 Liebel-Flarsheim Direct Digital Imaging SystemLiebel-Flarsheim · 2012
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−9K801623Cavitron Ultrasonic Surg. Aspir. CusaCavitron Corp.1980—
−8K962952Olympus Ultrasonic Surgical System (Usu)Olympus Optical, Co.1996—
−7K972114Olympus Sonosurg SystemThe Olympus Optical Co.1998—
−6K981993Olympus Integrated Endosurgery System EndoalphaThe Olympus Optical Co.1998—
−5K994231Siemens Integrated Operating System (Sios)Siemens Medical Corp.2000—
−4K052202Axiom Artis - Modular Angiographic SystemSiemens Medical Solutions USA, Inc.200629
−3K061054Axiom Aristos FX PlusSiemens Medical Solutions USA, Inc.20063
−2K062623Axiom Luminos DRFSiemens Medical Solutions USA, Inc.200720
−1K101491Uroskop OmniaSiemens Medical Solutions USA, Inc.201012
0 (this device)K121838Liebel-Flarsheim Direct Digital Imaging SystemLiebel-Flarsheim2012—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198110
198620
198820
198910
199150
199220
199330
199430
1995130
199650
199750
199830
199940
200030
200120
200250
200420
201010
201110
FIG. 1 — Lineage of K121838, 1980–2012Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MQB? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MQB device

  1. #1 K201528 Yushan X-Ray Flat Panel Detector with DROCscore 21.5 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K943581Hydradjust IV Urological Table
Summary page 1
…n: 21 CFR 892.1650 Device Class: Class 11 Product Code: MQB Predicate Devices: Hydra Vision Urological X-Ray System (K943581I) UROSKOP Omnia (K101491) Nexus DRF Digital X-ray Imaging System (K103416) Device Description: The Liebel-Flarsheimn Direct Dig…
Mallinckrodt Group, Inc.19953
*K101491Uroskop Omnia
Summary page 1
…1650 Device Class: Class 11 Product Code: MQB Predicate Devices: Hydra Vision Urological X-Ray System (K943581I) UROSKOP Omnia (K101491) Nexus DRF Digital X-ray Imaging System (K103416) Device Description: The Liebel-Flarsheimn Direct Digital Imaging System is a…
Siemens Medical Solutions USA, Inc.201012
*K103416Nexus DRF Digital X-Ray Imaging System
Summary page 1
…Product Code: MQB Predicate Devices: Hydra Vision Urological X-Ray System (K943581I) UROSKOP Omnia (K101491) Nexus DRF Digital X-ray Imaging System (K103416) Device Description: The Liebel-Flarsheimn Direct Digital Imaging System is a radiographic and fluoroscopy examination' table w…
Infimed, Inc.2011—

Also cleared against K943581: 2 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 11, 2021Class IIZ-1992-2021Radiation Control for Health and Safety ActFailure to comply with the performance standard for fluoroscopy systems. Per 21 CFR 1020.32(k)(6), the displayed AKR and cumulative air kerma shall not deviate from the actual values by more than ¿ 35 percent. The dosage displayed on the…
Feb 27, 2017Class IIZ-2834-2018Radiation Control for Health and Safety ActSoftware issue. The difference between the display and dosimeter readings In the Child/Pediatric automatic exposure mode is in the range of 52-65%. The display in all other modes deviated by less than 35% from the dosimeter readings.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 32 years and passes through 6 companies; a new generation was cleared every 2.0 years on median.
  • 25 of the 72 ancestors have a recall of any class.
  • 2 other devices also cite its predicate K943581.

How does its category compare?

All MQB devices

MQB Solid State X-Ray Imager (Flat Panel/Digital Imager): FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.

FDA decisions per year · MQB

8in 2025

2007: 132008: 82009: 152010: 112011: 242012: 222013: 212014: 322015: 212016: 222017: 312018: 162019: 222020: 192021: 212022: 312023: 102024: 82025: 82026 (partial year): 3
2007high 32 (2014) · low 8 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MQB

127in 2025

2007: 22 days median2008: 15 days median2009: 57 days median2010: 116 days median2011: 102 days median2012: 80 days median2013: 105 days median2014: 102 days median2015: 81 days median2016: 92 days median2017: 30 days median2018: 30 days median2019: 30 days median2020: 28 days median2021: 48 days median2022: 48 days median2023: 30 days median2024: 187 days median2025: 127 days median2026 (partial year): 245 days median
2007high 245 (2026) · low 15 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MQB

1in 2025

2007: 152008: 362009: 192010: 142011: 102012: 82013: 62017: 12018: 42019: 12020: 42021: 12025: 12026 (partial year): 1
2007high 36 (2008) · low 0 (2014)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MQB

141in 2025

2007: 252008: 302009: 342010: 392011: 1012012: 2672013: 1592014: 292015: 212016: 262017: 202018: 842019: 452020: 92021: 262022: 132023: 512024: 912025: 141
2007high 267 (2012) · low 9 (2020)2025

MAUDE reports naming product code MQB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200713221525
20088153630
200915571934
2010111161439
20112410210101
201222808267
2013211056159
201432102029
20152181021
20162292026
20173130120
20181630484
20192230145
2020192849
20212148126
20223148013
20231030051
20248187091
202581271141
202632451—

What is it indicated for?

The Liebel-Flarsheim Direct Digital Imaging System is a solid state detector X-ray system that facilitates digital radiologic and/or fluoroscopic procedures requiring a beam of diagnostic quality radiation and a flat imaging table primarily for urological applications such as functional x-ray diagnostics, endourology and minimal invasive urology/surgery. The system inky be used for urological, gastroenterological, and gynecological treatment, planning and diagnostic procedures. It is intended to replace fluoroscopic images formerly obtained through image intensifier technology. Not intended for mammography applications.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K121838: Liebel-Flarsheim Direct Digital Imaging System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K121838

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026