510(k) · Cleared
Recalled · Class II · 2017Hydradjust IV Urological Table
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Mallinckrodt Group, Inc. · IXR Table, Radiographic, Tilting
Cleared on Jan 4, 1995 after a 166-day review. FDA lists a 510(k) statement for it but has not posted the file, so its predicates are not known. 50 later devices descend from it. It has been recalled 3 times.
- Decided
- Jan 4, 1995
- Review
- 166 days
- Class
- Class II
- Product code
- IXR
- Recalls
- 3
- Predicates
- 0
50 later devices descend from it, 3 citing it directly.
- This deviceK943581 Hydradjust IV Urological TableMallinckrodt Group, Inc. · 1995
- Descendant · cites it directly · 47 of its ownK011311 UrologicPck Electronic Industry and Trade Co, Ltd. · 2001
- Descendant · 2 generations on · 45 of its ownK042236 G-SCANEsaote, S.p.A. · 2004
- Descendant · 3 generations on · 43 of its ownK060956 G-Scan-Model 9800020000, E-Scan Model 9800010000 and E-Scan…Esaote, S.p.A. · 2006
- Descendant · 3 generations on · 34 of its ownK063207 S-Scan Mri SystemEsaote, S.p.A. · 2006
- Descendant · 4 generations on · 32 of its ownK061429 Dynamic Mri Software Option for C-Scan, E-Scan XQ and E-Scan OperaEsaote, S.p.A. · 2006
Showing 5 of 50 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 50 · See all- K011311UrologicPck Electronic Industry and Trade Co, Ltd. · 2001 · cites it directly47descendants
- K042236G-SCANEsaote, S.p.A. · 2004 · 2 generations on45descendants
- K060956G-Scan-Model 9800020000, E-Scan Model 9800010000 and E-Scan…Esaote, S.p.A. · 2006 · 3 generations on43descendants
- K063207S-Scan Mri SystemEsaote, S.p.A. · 2006 · 3 generations on34descendants
- K061429Dynamic Mri Software Option for C-Scan, E-Scan XQ and E-Scan OperaEsaote, S.p.A. · 2006 · 4 generations on32descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| 0 (this device) | K943581 | Hydradjust IV Urological Table | Mallinckrodt Group, Inc. | 1995 | — |
| +1 | K011311 | Urologic | Pck Electronic Industry and Trade Co, Ltd. | 2001 | — |
| +2 | K042236 | G-SCAN | Esaote, S.p.A. | 2004 | — |
| +3 | K060956 | G-Scan-Model 9800020000, E-Scan Model 9800010000 and E-Scan… | Esaote, S.p.A. | 2006 | — |
| +3 | K063207 | S-Scan Mri System | Esaote, S.p.A. | 2006 | — |
| +4 | K061429 | Dynamic Mri Software Option for C-Scan, E-Scan XQ and E-Scan Opera | Esaote, S.p.A. | 2006 | — |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1999 | 1 | 0 |
| 2001 | 1 | 0 |
| 2004 | 1 | 0 |
| 2005 | 2 | 0 |
| 2006 | 3 | 0 |
| 2008 | 4 | 0 |
| 2009 | 2 | 0 |
| 2010 | 2 | 0 |
| 2011 | 4 | 0 |
| 2012 | 8 | 0 |
| 2013 | 3 | 0 |
| 2015 | 4 | 0 |
| 2016 | 3 | 0 |
| 2017 | 1 | 0 |
| 2019 | 4 | 0 |
| 2021 | 1 | 0 |
| 2022 | 1 | 0 |
| 2024 | 2 | 0 |
| 2026 | 3 | 0 |
Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in IXR? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
- Candidate predicates for a device like this, rankedin the report
- Recall history across each candidate’s familyin the report
What was it cleared against?
FDA lists a 510(k) statement for it but has not posted the file, so its predicates are not known.
What descends from it? (50)
See all 503 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Feb 27, 2017 | Class II | Z-2833-2018 | Radiation Control for Health and Safety Act | Software issue. The difference between the display and dosimeter readings In the Child/Pediatric automatic exposure mode is in the range of 52-65%. The display in all other modes deviated by less than 35% from the dosimeter readings. |
| Apr 25, 2013 | Class II | Z-1753-2013 | Employee error | A complaint was received in which a hospital physicist indicated the dosage was too high on the HUT system. During the service call the systems Maximum Fluoro Dose was found to be 30 R/min which is three times the Maximum Fluoro Dose… |
| Jun 14, 2010 | Not classified | Z-1904-2010 | Process control | Covidien received complaints of the Sedecal Generator failing and producing smoke at customer sites. The failure modes are located in different parts of the generator and will lead to the generator becoming inoperative. |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- Of its descendants, K011311 (Pck Electronic Industry and Trade Co, Ltd., 2001) has gone furthest, with 47 descendants of its own.
How does its category compare?
All IXR devicesIXR Table, Radiographic, Tilting: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · IXR
FDA has decided no IXR device since 2007.
Median 510(k) review, days · IXR
Too few IXR 510(k)s in any year since 2007 to give a median.
Recalls per year · IXR
1in 2025
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · IXR
1in 2024
MAUDE reports naming product code IXR, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 0 | — | 1 | 4 |
| 2008 | 0 | — | 4 | 8 |
| 2009 | 0 | — | 0 | 1 |
| 2010 | 0 | — | 5 | 12 |
| 2011 | 0 | — | 0 | 10 |
| 2012 | 0 | — | 0 | 164 |
| 2013 | 0 | — | 0 | 196 |
| 2014 | 0 | — | 0 | 120 |
| 2015 | 0 | — | 0 | 82 |
| 2016 | 0 | — | 1 | 58 |
| 2017 | 0 | — | 1 | 25 |
| 2018 | 0 | — | 0 | 17 |
| 2019 | 0 | — | 1 | 8 |
| 2020 | 0 | — | 0 | 12 |
| 2021 | 0 | — | 0 | 5 |
| 2022 | 0 | — | 0 | 2 |
| 2023 | 0 | — | 0 | 1 |
| 2024 | 0 | — | 0 | 1 |
| 2025 | 0 | — | 1 | 0 |
| 2026 | 0 | — | 0 | — |
https://510kreport.com/device/K943581 · printed Nov 6, 2026
Cite this page
510(k) Report. “K943581: Hydradjust IV Urological Table.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K943581
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026