Coughassist T70 Device

K121955510(k) TraditionalCleared Dec 14, 2012164-day reviewClass II21 CFR 868.5905

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Respironics, Inc. · NHJ Device, Positive Pressure Breathing, Intermittent

Cleared on Dec 14, 2012 as substantially equivalent to K002598 (J. H. Emerson Co., 2000) and 1 other predicate after a 164-day review. Its family tree reaches back 2 generations to 2000. 3 of the 17 devices in its family, counting itself, have a Class I recall. 14 later devices descend from it.

Decided
Dec 14, 2012
Review
164 days
Class
Class II
Product code
NHJ
Recalls
None
Predicates
2

Its longest line of descent runs back 2 generations to K002598 (2000). 14 later devices descend from it, 6 citing it directly. 3 of the 17 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2012: 1 descendant cleared2014: 2 descendants cleared2025: 3 descendants cleared2020: 1 descendant cleared2017: 1 descendant cleared, 1 with a Class I recall12024: 3 descendants cleared2026: 1 descendant cleared2019: 1 descendant cleared, 1 with a Class I recall12021: 1 descendant cleared, 1 with a Class I recall1K002598 · Emerson Cough Assist Model CA-3000 W/Automatic Timing and Model CM-3000 W/Manual Timing · J. H. Emerson Co. · 2000K072292 · Negavent Plus Pegaso Cough, Model: DA-3 Plus · Dima Italia Srl · 2008200020052010201520202025
Descendants, by generation 1 generation forward: 6 descendants, 1 with a Class I recall12 generations forward: 3 descendants, 1 with a Class I recall13 generations forward: 4 descendants, 1 with a Class I recall14 generations forward: 1 descendant K072292 · Negavent Plus Pegaso Cough, Model: DA-3 Plus · Dima Italia Srl · 2008K002598 · Emerson Cough Assist Model CA-3000 W/Automatic Timing and Model CM-3000 W/Manual Timing · J. H. Emerson Co. · 20000
  1. Root · 2 generations backK002598 Emerson Cough Assist Model CA-3000 W/Automatic Timing and Model…J. H. Emerson Co. · 2000
  2. 1 generation backK072292 Negavent Plus Pegaso Cough, Model: DA-3 PlusDima Italia Srl · 2008
  3. This deviceK121955 Coughassist T70 DeviceRespironics, Inc. · 2012
  4. Descendant · cites it directly · 6 of its ownK162877 VOCSN Unified Respiratory SystemVentec Life Systems, Inc. · 2017 · Recalled (Class I)
  5. Descendant · 2 generations on · 5 of its ownK181166 Trilogy EvoRespironics, Inc. · 2019 · Recalled (Class I)
  6. Descendant · cites it directly · 4 of its ownK122111 Respironics Simplyclear DeviceRespironics, Inc. · 2012
  7. Descendant · 3 generations on · 2 of its ownK193586 Vivo 45 LSBreas Medical AB · 2021 · Recalled (Class I)
  8. Descendant · cites it directly · 1 of its ownK140598 Pegaso Cough, Pegaso a-Cough, Pegaso a-Cough PercDima Italia Srl · 2014

Showing 5 of 14 descendants. See all descendants

Descendants cleared per year

2012: 12014: 22017: 12019: 12020: 12021: 12024: 32025: 32026: 1
2012high 3 (2024) · low 0 (2013)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 14 · See all
  • K162877VOCSN Unified Respiratory SystemVentec Life Systems, Inc. · 2017 · cites it directly6descendants
  • K181166Trilogy EvoRespironics, Inc. · 2019 · 2 generations on5descendants
  • K122111Respironics Simplyclear DeviceRespironics, Inc. · 2012 · cites it directly4descendants
  • K193586Vivo 45 LSBreas Medical AB · 2021 · 3 generations on2descendants
  • K140598Pegaso Cough, Pegaso a-Cough, Pegaso a-Cough PercDima Italia Srl · 2014 · cites it directly1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K002598Emerson Cough Assist Model CA-3000 W/Automatic Timing and Model CM-3000…J. H. Emerson Co.2000—
−1K072292Negavent Plus Pegaso Cough, Model: DA-3 PlusDima Italia Srl2008—
0 (this device)K121955Coughassist T70 DeviceRespironics, Inc.2012—
+1K162877VOCSN Unified Respiratory SystemVentec Life Systems, Inc.2017Class I
+2K181166Trilogy EvoRespironics, Inc.2019Class I
+1K122111Respironics Simplyclear DeviceRespironics, Inc.2012—
+3K193586Vivo 45 LSBreas Medical AB2021Class I
+1K140598Pegaso Cough, Pegaso a-Cough, Pegaso a-Cough PercDima Italia Srl2014—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201210
201420
201711
201911
202010
202111
202430
202530
202610
FIG. 1 — Lineage of K121955, 2000–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new NHJ device

  1. #1 K260604 BiWaze Clear Systemscore 14.0 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K002598Emerson Cough Assist Model CA-3000 W/Automatic Timing and Model CM-3000 W/Manual Timing
Summary page 2
…standards). Predicate Device Name(s) and Emerson CoughAssist, Model CA-3000 - 510(k) numbers: K002598 Dima Italia Negavent DA-3 Plus Pegaso - ________________________K072292…
J. H. Emerson Co.2000—
*K072292Negavent Plus Pegaso Cough, Model: DA-3 Plus
Summary page 2
…k) numbers: K002598 Dima Italia Negavent DA-3 Plus Pegaso - ________________________K072292 Reason for submission: Device modifications Description of Device The Philips Respironics CoughAssist…
Dima Italia Srl2008—

Also cleared against K002598: 6 other devices →

What descends from it? (14)

See all 14

6 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K162877VOCSN Unified Respiratory System Class IVentec Life Systems, Inc.20176
K122111Respironics Simplyclear DeviceRespironics, Inc.20124
K140598Pegaso Cough, Pegaso a-Cough, Pegaso a-Cough PercDima Italia Srl20141
K242438ClearoBreas Medical AB20250
K191912BiWaze CoughAbmrc, LLC20200
K140605Minipegaso Cough, Minipegaso a-Cough, Minipegaso a-Coug PercDima Italia Srl20140

Has anything in its family been recalled? (3 Class I)

No FDA recall names this device.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K162877VOCSN Unified Respiratory SystemVentec Life Systems, Inc. · 2017 · Cites it directly · ◆ in the 2017 descendants bar of FIG. 114 Class I recalls
  • K181166Trilogy EvoRespironics, Inc. · 2019 · 2 generations on · ◆ in the 2019 descendants bar of FIG. 131 Class I recalls
  • K193586Vivo 45 LSBreas Medical AB · 2021 · 3 generations on · ◆ in the 2021 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 12 years and passes through 3 companies; a new generation was cleared every 7.1 years on median.
  • None of the 2 ancestors has a recall.
  • Of its descendants, K162877 (Ventec Life Systems, Inc., 2017) has gone furthest, with 6 descendants of its own.
  • 6 other devices also cite its predicate K002598.

How does its category compare?

All NHJ devices

NHJ Device, Positive Pressure Breathing, Intermittent: FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.

FDA decisions per year · NHJ

1in 2025

2008: 12012: 22013: 22014: 32016: 12020: 32022: 12024: 12025: 12026 (partial year): 2
2007high 3 (2014) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · NHJ

190in 2020

2014: 207 days median2020: 190 days median
2007high 207 (2014) · low 190 (2020)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · NHJ

11so far in 2026, the most since 2007

2024: 1, 1 Class I2026 (partial year): 11, 4 Class I
2007high 11 (2026) · low 0 (2007)2026

Class I (5 since 2007) Class II/III not classified

Adverse event reports per year · NHJ

35in 2025

2007: 162008: 72009: 12010: 22012: 22013: 22014: 12015: 22016: 22017: 62018: 22019: 12020: 32021: 52022: 62023: 172024: 402025: 35
2007high 40 (2024) · low 0 (2011)2025

MAUDE reports naming product code NHJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—016
20081—07
20090—01
20100—02
20110—00
20122—02
20132—02
2014320701
20150—02
20161—02
20170—06
20180—02
20190—01
2020319003
20210—05
20221—06
20230—017
20241—1 (1)40
20251—035
20262—11 (4)—

What is it indicated for?

The Philips Respironics CoughAssist T70 is intended for use on adult or pediatric patients unable to cough or clear secretions effectively. It may be used either with a facemask or mouthpiece, or with an adapter to a patient's endotracheal or tracheostomy tube. The device is intended to be used in the hospital, institutional environment or in the home.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K121955: Coughassist T70 Device.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K121955

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.