510(k) Report

510(k) · Cleared

Recalled · Class I · 2026

VOCSN Unified Respiratory System

K162877510(k) TraditionalCleared Apr 7, 2017175-day reviewClass II21 CFR 868.5895Recalls 16 (14 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ventec Life Systems, Inc. · CBK Ventilator, Continuous, Facility Use

Cleared on Apr 7, 2017 as substantially equivalent to K111610 (Respironics, Inc., 2011) and 4 other predicates after a 175-day review. Its family tree reaches back 10 generations to 1991. 12 of the 95 devices in its family, counting itself, have a Class I recall. 6 later devices descend from it. It has been recalled 16 times, including 14 Class I recalls.

Decided
Apr 7, 2017
Review
175 days
Class
Class II
Product code
CBK
Recalls
16 (14 Class I)
Predicates
5

Its longest line of descent runs back 10 generations to K902859 (1991). 88 ancestors in all, the oldest K841279 (1984). 6 later devices descend from it, 1 citing it directly. 12 of the 95 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2014: 1 other ancestor2012: 2 other ancestors2007: 2 other ancestors, 1 with a Class I recall12010: 2 other ancestors, 2 with a Class I recall22009: 2 other ancestors, 1 with a Class I recall12008: 1 other ancestor2006: 1 other ancestor2003: 4 other ancestors, 1 with a Class I recall12002: 4 other ancestors2000: 5 other ancestors1999: 5 other ancestors1998: 11 other ancestors, 2 with a Class I recall22005: 1 other ancestor2001: 5 other ancestors1996: 5 other ancestors2004: 2 other ancestors1994: 4 other ancestors1993: 1 other ancestor1991: 4 other ancestors1990: 4 other ancestors1985: 1 other ancestor1984: 2 other ancestors1997: 3 other ancestors1995: 3 other ancestors1992: 2 other ancestors1988: 1 other ancestor2019: 1 descendant cleared, 1 with a Class I recall12025: 2 descendants cleared2021: 1 descendant cleared, 1 with a Class I recall12024: 2 descendants clearedReference device: K060647 · LTV 1200 Ventilator · Pulmonetic Systems, Inc. · 2006Reference device: K110398 · Respironics Disposable Heated Wire Circuits · Respironics, Inc. · 2011Reference device: K973797 · Breathing Circuit Bacterial Filter · King Systems Corp. · 1998 · recalledK902859 · Servo Ventilator 300 · Siemens Life Support Systems · 1991 · recalledK960010 · Servo Ventilator 300 Computer Interface · Siemens Elema AB · 1996K970839 · Servo Ventilator 300A · Siemens Elema AB · 1997 · recalledK992788 · Bear Cub 750 PSV · Bear Medical Systems, Inc. · 1999K013642 · AVEA · Bird Products Corp. · 2002 · Class I recallK032451 · Bird Tbird Vela Ventilator · Bird Products Corp. · 2004 · recalled×3: 3 generations cleared close together, K052554 (2005), K053270 (2006), K061627 (2006) K111610 · Trilogy Series Ventilator with Oximetry · Respironics, Inc. · 2011 · Class I recall19801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 4 other ancestors, 1 with a Class I recall12 generations back: 12 other ancestors, 3 with a Class I recall33 generations back: 13 other ancestors, 1 with a Class I recall14 generations back: 21 other ancestors, 1 with a Class I recall15 generations back: 22 other ancestors, 1 with a Class I recall16 generations back: 6 other ancestors 1 generation forward: 1 descendant, 1 with a Class I recall12 generations forward: 4 descendants, 1 with a Class I recall13 generations forward: 1 descendant K111610 · Trilogy Series Ventilator with Oximetry · Respironics, Inc. · 2011 · Class I recallK061627 · Modification to I Vent 201 · Versamed Medical Systems, Inc. · 2006K053270 · IVENT 201 · Versamed Medical Systems, Inc. · 2006K052554 · Modification to: Ivent 201 Portable Ventilator · Versamed Medical Systems, Inc. · 2005K032451 · Bird Tbird Vela Ventilator · Bird Products Corp. · 2004 · recalledK013642 · AVEA · Bird Products Corp. · 2002 · Class I recallK992788 · Bear Cub 750 PSV · Bear Medical Systems, Inc. · 1999K970839 · Servo Ventilator 300A · Siemens Elema AB · 1997 · recalledK960010 · Servo Ventilator 300 Computer Interface · Siemens Elema AB · 1996K902859 · Servo Ventilator 300 · Siemens Life Support Systems · 1991 · recalled−10−50
  1. Root · 10 generations backK902859 Servo Ventilator 300Siemens Life Support Systems · 1991 · Recalled
  2. 6 generations between
  3. 3 generations backK053270 IVENT 201Versamed Medical Systems, Inc. · 2006
  4. 2 generations backK061627 Modification to I Vent 201Versamed Medical Systems, Inc. · 2006
  5. 1 generation backK111610 Trilogy Series Ventilator with OximetryRespironics, Inc. · 2011 · Recalled (Class I)
  6. This deviceK162877 VOCSN Unified Respiratory SystemVentec Life Systems, Inc. · 2017
  7. Descendant · cites it directly · 5 of its ownK181166 Trilogy EvoRespironics, Inc. · 2019 · Recalled (Class I)
  8. Descendant · 2 generations on · 2 of its ownK193586 Vivo 45 LSBreas Medical AB · 2021 · Recalled (Class I)
  9. Descendant · 2 generations on · 0 of its ownK232365 Vivo 45 LSOconnell Regulatory Consultants, Inc. · 2025
  10. Descendant · 3 generations on · 0 of its ownK233452 Vivo 45 LSBreas Medical AB · 2025 · Recalled
  11. Descendant · 2 generations on · 0 of its ownK241135 RESPOND (R) VentilatorCorvent Medical, Inc. · 2024
Show all 10 generations
  1. Root · 10 generations backK902859 Servo Ventilator 300Siemens Life Support Systems · 1991 · Recalled
  2. 9 generations backK960010 Servo Ventilator 300 Computer InterfaceSiemens Elema AB · 1996
  3. 8 generations backK970839 Servo Ventilator 300ASiemens Elema AB · 1997 · Recalled
  4. 7 generations backK992788 Bear Cub 750 PSVBear Medical Systems, Inc. · 1999
  5. 6 generations backK013642 AVEABird Products Corp. · 2002 · Recalled (Class I)
  6. 5 generations backK032451 Bird Tbird Vela VentilatorBird Products Corp. · 2004 · Recalled
  7. 4 generations backK052554 Modification to: Ivent 201 Portable VentilatorVersamed Medical Systems, Inc. · 2005
  8. 3 generations backK053270 IVENT 201Versamed Medical Systems, Inc. · 2006
  9. 2 generations backK061627 Modification to I Vent 201Versamed Medical Systems, Inc. · 2006
  10. 1 generation backK111610 Trilogy Series Ventilator with OximetryRespironics, Inc. · 2011 · Recalled (Class I)
  11. This deviceK162877 VOCSN Unified Respiratory SystemVentec Life Systems, Inc. · 2017

Showing 5 of 6 descendants. See all descendants

Descendants cleared per year

2019: 12021: 12024: 22025: 2
2019high 2 (2024) · low 0 (2020)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 6 · See all
  • K181166Trilogy EvoRespironics, Inc. · 2019 · cites it directly5descendants
  • K193586Vivo 45 LSBreas Medical AB · 2021 · 2 generations on2descendants
  • K232365Vivo 45 LSOconnell Regulatory Consultants, Inc. · 2025 · 2 generations on0descendants
  • K233452Vivo 45 LSBreas Medical AB · 2025 · 3 generations on0descendants
  • K241135RESPOND (R) VentilatorCorvent Medical, Inc. · 2024 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−10K902859Servo Ventilator 300Siemens Life Support Systems19913
−9K960010Servo Ventilator 300 Computer InterfaceSiemens Elema AB1996—
−8K970839Servo Ventilator 300ASiemens Elema AB19974
−7K992788Bear Cub 750 PSVBear Medical Systems, Inc.1999—
−6K013642AVEABird Products Corp.2002Class I
−5K032451Bird Tbird Vela VentilatorBird Products Corp.20041
−4K052554Modification to: Ivent 201 Portable VentilatorVersamed Medical Systems, Inc.2005—
−3K053270IVENT 201Versamed Medical Systems, Inc.2006—
−2K061627Modification to I Vent 201Versamed Medical Systems, Inc.2006—
−1K111610Trilogy Series Ventilator with OximetryRespironics, Inc.2011Class I
0 (this device)K162877VOCSN Unified Respiratory SystemVentec Life Systems, Inc.2017—
+1K181166Trilogy EvoRespironics, Inc.2019Class I
+2K193586Vivo 45 LSBreas Medical AB2021Class I
+2K232365Vivo 45 LSOconnell Regulatory Consultants, Inc.2025—
+3K233452Vivo 45 LSBreas Medical AB20251
+2K241135RESPOND (R) VentilatorCorvent Medical, Inc.2024—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198420
198510
198810
199040
199140
199220
199310
199440
199530
199650
199730
1998112
199950
200050
200150
200240
200341
200420
200510
200610
200721
200810
200921
201022
201220
201410

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201911
202111
202420
202520
FIG. 1 — Lineage of K162877, 1991–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

12 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CBK device

  1. #1 K252663 Inbentus Versatilescore 16.0 · 17 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (5)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K111610Trilogy Series Ventilator with Oximetry
Summary page 5
…nal Premarket Notification Primary Predicate Device Mechanical ventilator component Information: Respironics Trilogy Series Ventilator (K111610) Reference Predicate Device Mechanical ventilator component Information: CareFusion Palmtop Ventil…
Respironics, Inc.2011Class I
*K070594Palmtop Ventilator, Models PTV-8, PTV-10
Summary page 5
…es Ventilator (K111610) Reference Predicate Device Mechanical ventilator component Information: CareFusion Palmtop Ventilator PTV-8 & -10 (K070594) CareFusion LTV 1200 Ventilator (K060647) Oxygen concentra…
Pulmonetic Systems, Inc.2007Class I
*K120785Omni 3 Oxygen System
Summary page 5
…usion LTV 1200 Ventilator (K060647) Oxygen concentrator component Omni 3 (eQuinox) Oxygen System (K120785) Cough Assist component Respironics Cough Assist T70 (K12…
Caire, Inc.2012—
*K121955Coughassist T70 Device
Summary page 5
…Omni 3 (eQuinox) Oxygen System (K120785) Cough Assist component Respironics Cough Assist T70 (K121955) Suction Pump component Precision Medical Easy Go Vac PM…
Respironics, Inc.2012—
*K140179Easy GO Vac
Summary page 5
…espironics Cough Assist T70 (K121955) Suction Pump component Precision Medical Easy Go Vac PM66 (K140179) Heated Patient Circuits Respironics (Philips) Disposable…
Precision Medical, Inc.2014—
†K060647LTV 1200 VentilatorPulmonetic Systems, Inc.2006—
†K110398Respironics Disposable Heated Wire CircuitsRespironics, Inc.2011—
†K973797Breathing Circuit Bacterial FilterKing Systems Corp.19981

Also cleared against K111610: 4 other devices →

What descends from it? (6)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K181166Trilogy Evo Class IRespironics, Inc.20195

Has anything in its family been recalled? (12 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 17, 2026Class IZ-0129-2027Under Investigation by firmVentilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
Sep 17, 2026Class IZ-0130-2027Under Investigation by firmVentilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
Sep 17, 2026Class IZ-0127-2027Under Investigation by firmVentilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
Sep 17, 2026Class IZ-0124-2027Under Investigation by firmVentilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
Sep 17, 2026Class IZ-0131-2027Under Investigation by firmVentilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
Sep 17, 2026Class IZ-0123-2027Under Investigation by firmVentilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
Sep 17, 2026Class IZ-0128-2027Under Investigation by firmVentilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
Sep 17, 2026Class IZ-0125-2027Under Investigation by firmVentilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
Sep 17, 2026Class IZ-0126-2027Under Investigation by firmVentilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
Sep 17, 2026Class IZ-0122-2027Under Investigation by firmVentilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
Sep 17, 2026Class IZ-0132-2027Under Investigation by firmVentilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
May 8, 2026Class IIZ-2525-2026Vendor change controlRespiratory system intended to provide continuous/intermittent ventilator support, may have a damaged back-up battery supporting backup alarm function, which may result in the backup alarm not activating during loss of primary power and…
Mar 25, 2026Class IZ-1942-2026Process change controlRespiratory system intended to provide continuous/intermittent ventilator support, may not have been evaluated under all intended high-pressure conditions during production testing, so an oxygen leak condition could occur that may lead to…
Feb 3, 2025Class IZ-1196-2025Component change controlMulti-Function Ventilators were serviced using incorrect parts which have the potential to cause unexpected shutdown, or when using an active circuit could result in inaccurate tidal volume monitor, not triggering on patient efforts, less…
Jan 9, 2024Class IZ-1070-2024Process controlThe bonded spiral wrap may detach before or during ventilation due to a manufacturing issue. Detachment of the spiral wrap can compromise structural integrity, functionality and/ or performance resulting in occlusions or leaks of the…
May 10, 2021Class IIZ-1902-2021Device DesignDue to a component failure, ventilators may unexpectedly shut down or have a loss of blower function while in patient use. The shutdown potentially prevents the ventilator from providing ventilation therapy.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K111610Trilogy Series Ventilator with OximetryRespironics, Inc. · 2011 · Direct predicate · ◆ on the line of descent in FIG. 12 Class I recalls
  • K070594Palmtop Ventilator, Models PTV-8, PTV-10Pulmonetic Systems, Inc. · 2007 · Direct predicate · ◆ in the 2007 ancestors bar of FIG. 12 Class I recalls
  • K093905Trilogy 202 VentilatorRespironics, Inc. · 2010 · 2 generations back · ◆ in the 2010 ancestors bar of FIG. 12 Class I recalls
  • K093416Trilogy 200 VentilatorRespironics, Inc. · 2010 · 2 generations back · ◆ in the 2010 ancestors bar of FIG. 14 Class I recalls
  • K083526Trilogy 100 VentilatorRespironics Inc., Sleep & Home Respiratory Group · 2009 · 2 generations back · ◆ in the 2009 ancestors bar of FIG. 15 Class I recalls
  • K013642AVEABird Products Corp. · 2002 · 3 generations back · ◆ on the line of descent in FIG. 113 Class I recalls
  • K981072Esprit VentilatorRespironics, Inc. · 1998 · 3 generations back · ◆ in the 1998 ancestors bar of FIG. 11 Class I recall
  • K970460Nellcor Puritan Bennett, 840 Ventilator System (Model 840)Puritan Bennett Corp. · 1998 · 4 generations back · ◆ in the 1998 ancestors bar of FIG. 11 Class I recall
  • K023366Datex-Ohmeda 7900 Ventilator EnhancementsDatex-Ohmeda, Inc. · 2003 · 5 generations back · ◆ in the 2003 ancestors bar of FIG. 11 Class I recall
  • K181166Trilogy EvoRespironics, Inc. · 2019 · Cites it directly · ◆ in the 2019 descendants bar of FIG. 131 Class I recalls
  • K193586Vivo 45 LSBreas Medical AB · 2021 · 2 generations on · ◆ in the 2021 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 26 years and passes through 7 companies; a new generation was cleared every 1.9 years on median.
  • 31 of the 88 ancestors have a recall of any class; 9 have a Class I recall.
  • Of its descendants, K181166 (Respironics, Inc., 2019) has gone furthest, with 5 descendants of its own.
  • 4 other devices also cite its predicate K111610.

How does its category compare?

All CBK devices

CBK Ventilator, Continuous, Facility Use: FDA decisions, review times, recalls and adverse event reports over 20 years. 2017, this device’s year, is highlighted.

FDA decisions per year · CBK

4in 2025

2007: 112008: 182009: 122010: 142011: 132012: 172013: 52014: 52015: 72016: 42017: 62018: 42019: 82020: 52021: 62023: 52024: 92025: 42026 (partial year): 3
2007high 18 (2008) · low 0 (2022)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CBK

163in 2025

2007: 142 days median2008: 115 days median2009: 122 days median2010: 150 days median2011: 192 days median2012: 207 days median2013: 277 days median2014: 439 days median2015: 266 days median2016: 323 days median2017: 280 days median2018: 183 days median2019: 435 days median2020: 176 days median2021: 267 days median2023: 309 days median2024: 203 days median2025: 163 days median2026 (partial year): 235 days median
2007high 439 (2014) · low 115 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CBK

14in 2024

2007: 172008: 232009: 82010: 92011: 122012: 2, 1 Class I2013: 5, 2 Class I2014: 5, 4 Class I2015: 6, 4 Class I2016: 4, 1 Class I2017: 12018: 3, 3 Class I2019: 14, 13 Class I2020: 42021: 21, 19 Class I2022: 8, 4 Class I2023: 4, 3 Class I2024: 14, 14 Class I2026 (partial year): 15, 13 Class I
2007high 23 (2008) · low 0 (2025)2026

Class I (81 since 2007) Class II/III not classified

Adverse event reports per year · CBK

13,991in 2025

2007: 1,5312008: 1,9332009: 1,7662010: 1,7992011: 1,8772012: 5,5092013: 7,9212014: 7,6002015: 11,8762016: 16,0562017: 14,4512018: 9,8492019: 9,1002020: 11,0962021: 13,3222022: 15,9742023: 24,9282024: 20,3362025: 13,991
2007high 24,928 (2023) · low 1,531 (2007)2025

MAUDE reports naming product code CBK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200711142171,531
200818115231,933
20091212281,766
20101415091,799
201113192121,877
2012172072 (1)5,509
201352775 (2)7,921
201454395 (4)7,600
201572666 (4)11,876
201643234 (1)16,056
20176280114,451
201841833 (3)9,849
2019843514 (13)9,100
20205176411,096
2021626721 (19)13,322
20220—8 (4)15,974
202353094 (3)24,928
2024920314 (14)20,336
20254163013,991
2026323515 (13)—

What is it indicated for?

The VOCSN Unified Respiratory System is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. It may be used in invasive and non-invasive applications. The VOCSN is intended for pediatric through adult patients weighing at least 5 kg. It is intended for use in home, hospital, institutional and transport settings, including portable applications. The integral oxygen concentrator…

Read the full statement (106 words)

The VOCSN Unified Respiratory System is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. It may be used in invasive and non-invasive applications. The VOCSN is intended for pediatric through adult patients weighing at least 5 kg. It is intended for use in home, hospital, institutional and transport settings, including portable applications.

The integral oxygen concentrator is intended for the administration of supplemental oxygen. The integral suction pump is intended for airway fluid removal and oral/ pharyngeal hygiene. The integral cough assist option is intended for patients who are additionally unable to cough or clear secretions effectively.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K162877: VOCSN Unified Respiratory System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K162877

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026