510(k) Report

510(k) · Cleared

Recalled · Class I · 2024

Pneupac Parapac Plus

K123957510(k) TraditionalCleared Sep 16, 2013269-day reviewClass II21 CFR 868.5925Recalls 7 (6 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Smiths Medical International, Ltd. · BTL Ventilator, Emergency, Powered (Resuscitator)

Cleared on Sep 16, 2013 as substantially equivalent to K020899 (Pneupac, Ltd., 2002) and 3 other predicates after a 269-day review. Its family tree reaches back 3 generations to 1980. 1 of the 12 devices in its family, counting itself, has a Class I recall. It has been recalled 7 times, including 6 Class I recalls.

Decided
Sep 16, 2013
Review
269 days
Class
Class II
Product code
BTL
Recalls
7 (6 Class I)
Predicates
4

Its longest line of descent runs back 3 generations to K801883 (1980). 11 ancestors in all. No later device cites it as a predicate yet. 1 of the 12 devices in this family, counting this one, has a Class I recall.

Other ancestors, by year cleared 2008: 1 other ancestor2005: 1 other ancestor2002: 1 other ancestor1996: 1 other ancestor1991: 1 other ancestor1983: 1 other ancestor1987: 1 other ancestor1981: 1 other ancestorK801883 · Peep Face Mask · Vital Signs, Inc. · 1980K982283 · Whisperflow Oxygen Flow Generator - Adjustable Model WF 8500 Whisperflow Oxygen Flow Generator - Fixed Model WF 8530 WH · Caradyne, Ltd. · 1999K013884 · Boussignac C.P.a.P. Device · Vygon SA · 200319801985199019952000200520102015
Other ancestors, by generation 1 generation back: 3 other ancestors2 generations back: 3 other ancestors3 generations back: 2 other ancestors K013884 · Boussignac C.P.a.P. Device · Vygon SA · 2003K982283 · Whisperflow Oxygen Flow Generator - Adjustable Model WF 8500 Whisperflow Oxygen Flow Generator - Fixed Model WF 8530 WH · Caradyne, Ltd. · 1999K801883 · Peep Face Mask · Vital Signs, Inc. · 19800
  1. Root · 3 generations backK801883 Peep Face MaskVital Signs, Inc. · 1980
  2. 2 generations backK982283 Whisperflow Oxygen Flow Generator - Adjustable Model WF 8500…Caradyne, Ltd. · 1999
  3. 1 generation backK013884 Boussignac C.P.a.P. DeviceVygon SA · 2003
  4. This deviceK123957 Pneupac Parapac PlusSmiths Medical International, Ltd. · 2013
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K801883Peep Face MaskVital Signs, Inc.1980—
−2K982283Whisperflow Oxygen Flow Generator - Adjustable Model WF 8500 Whisperflow…Caradyne, Ltd.1999—
−1K013884Boussignac C.P.a.P. DeviceVygon SA2003—
0 (this device)K123957Pneupac Parapac PlusSmiths Medical International, Ltd.2013—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198110
198310
198710
199110
199610
200210
200510
200810
FIG. 1 — Lineage of K123957, 1980–2013Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new BTL device

  1. #1 K242769 VentStar Resus heated (N) (MP17030); VentStar…score 11.2 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K020899Parapac 200 Medic, Model P200, & Parapac 200D Medic, Model P200D
Summary page 2
…vices for Pneupae® paraPAC plusTM Model 300/3 10 510(k) Ventilators Smiths Medical Pneupac® Parapac® P200D Medic Ventilator K020899 0 Two Carevent System ATV+ KO051469 Predicate Devices for Pneupac® para…
Pneupac, Ltd.2002—
*K051469Carevent Atv+ and Carevent Mri
Summary page 3
…essure (CPAP) control to provide CPAP therapy for spontaneously breathing patients which is also one of the features of the predicate device, the 0 Two Carevent Systems (K05 1469). This is achieved by connecting a Pneupac® paraPAd plusT l Disposable CPAP Circuit to the Pneupac® paraPAC plUSTM Mode! 310 Vent…
O-Two Medical Technologies, Inc.2005—
*K013884Boussignac C.P.a.P. Device
Summary page 2
…CPAP Circuit _____ Vygon Boussignac CPAP Device K013884 Pulmodyrie CHF Flow Generation System K080256 Device Description: The Pneupac® par…
Vygon SA2003—
*K080256Pulmodyne CHF Flow Generator System
Summary page 2
…Vygon Boussignac CPAP Device K013884 Pulmodyrie CHF Flow Generation System K080256 Device Description: The Pneupac® paraPAC plusT"l Model 300/3 10 Ventilators are gas powered portable medical devices intended…
Pulmodyne2008—

Also cleared against K020899: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 31, 2024Class IZ-3090-2024OtherSmiths Medical became aware of an issue related to a potential for the patient outlet connector to loosen/detach from the paraPAC Plus P300 and P310 ventilators impacting the active ventilation function.
May 31, 2024Class IZ-3091-2024OtherSmiths Medical became aware of an issue related to a potential for the patient outlet connector to loosen/detach from the paraPAC Plus P300 and P310 ventilators impacting the active ventilation function.
May 31, 2024Class IZ-2974-2024Under Investigation by firmThere is a potential for inadvertent tidal volume knob movement from the original setting when set at high (1000-1500 mL) and low (70-150 mL) settings.
May 31, 2024Class IZ-2973-2024Under Investigation by firmThere is a potential for inadvertent tidal volume knob movement from the original setting when set at high (1000-1500 mL) and low (70-150 mL) settings.
Feb 7, 2024Class IZ-1274-2024Under Investigation by firmWhen a paraPAC plus ventilator is switched to the operating mode of VENTILATE, the ventilator may intermittently provide continuous positive gas flow instead of the intended cycling like a human breath. This non-cycling and continuous…
Feb 7, 2024Class IZ-1275-2024Under Investigation by firmWhen a paraPAC plus ventilator is switched to the operating mode of VENTILATE, the ventilator may intermittently provide continuous positive gas flow instead of the intended cycling like a human breath. This non-cycling and continuous…
Jul 15, 2021Class IIIZ-0163-2022Process controlTwo devices were labeled with the same serial number

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 33 years and passes through 4 companies; a new generation was cleared every 10.7 years on median.
  • None of the 11 ancestors has a recall.
  • 1 other device also cites its predicate K020899.

How does its category compare?

All BTL devices

BTL Ventilator, Emergency, Powered (Resuscitator): FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.

FDA decisions per year · BTL

1in 2025

2007: 52008: 42009: 52010: 32011: 22013: 12014: 72015: 12017: 22018: 12019: 12020: 12021: 12023: 12025: 1
2007high 7 (2014) · low 0 (2012)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · BTL

203in 2014

2007: 122 days median2008: 198 days median2009: 173 days median2010: 140 days median2014: 203 days median
2007high 203 (2014) · low 122 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · BTL

2in 2024

2007: 12011: 12012: 2, 2 Class I2013: 3, 1 Class I2015: 3, 2 Class I2019: 3, 3 Class I2021: 12024: 2, 2 Class I
2007high 3 (2013) · low 0 (2008)2026

Class I (10 since 2007) Class II/III not classified

Adverse event reports per year · BTL

165in 2025

2007: 52008: 372009: 692010: 962011: 752012: 1152013: 1032014: 702015: 792016: 802017: 1012018: 1972019: 3412020: 4542021: 5062022: 7552023: 7312024: 6672025: 165
2007high 755 (2022) · low 5 (2007)2025

MAUDE reports naming product code BTL, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007512215
20084198037
20095173069
20103140096
20112—175
20120—2 (2)115
20131—3 (1)103
20147203070
20151—3 (2)79
20160—080
20172—0101
20181—0197
20191—3 (3)341
20201—0454
20211—1506
20220—0755
20231—0731
20240—2 (2)667
20251—0165
20260—0—

What is it indicated for?

The Pneupace paraPAC plusTm Model 300/310 Ventilators are gas-powered emergency and transport portable ventilators that are primarily intended for use in transport applications in vehicles including fixed and rotary winged aircraft. They are suitable for emergency use at the accident scene, intra and inter-hospital transport and within medical facilities including medical imaging systems to 3 Teslas. They should only be used under the constant…

Read the full statement (99 words)

The Pneupace paraPAC plusTm Model 300/310 Ventilators are gas-powered emergency and transport portable ventilators that are primarily intended for use in transport applications in vehicles including fixed and rotary winged aircraft. They are suitable for emergency use at the accident scene, intra and inter-hospital transport and within medical facilities including medical imaging systems to 3 Teslas. They should only be used under the constant supervision or'trained healthcare professionals. The devices are intended to provide ventilatory support to adults, children and infants (above approx. 10kg).

The devices also provide free flow oxygen therapy and CPAP therapy for spontaneously breathing patients.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K123957: Pneupac Parapac Plus.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K123957

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026